Introduces the Unistat P520 temperature control system
Specp. 2
Technical data according to DIN 12876 standard
p. 2table, 12 rows
p. 1table, 30 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 2,0 bar
46 l/min
Delivery at 2,5 bar
32 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
min. filling capacity
5,1 l
Filling capacity expansion tank
17 l
Overall dimensions WxDxH **
1290x900x1400 mm
Power supply factory configured (3 Phase)
400V 3~ 50Hz
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/24
** Please respect space requirements. See operating conditions at www.huber-online.com
Unistat P520 Technical data according to DIN 12876
Parameter
Value
Order-No.
1072.0023.01
Operating temperature range
-55...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
6 kW
Cooling power with Thermooil at 250°C
6 kW
Cooling power with Thermooil at 200°C
6 kW
Cooling power with Thermooil at 100°C
6 kW
Cooling power with Ethanol at 0°C
6 kW
Cooling power with Ethanol at -20°C
4,2 kW
Cooling power with Ethanol at -40°C
1,5 kW
Cooling power with Ethanol at -50°C
0,65 kW
Refrigeration machine
air-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-1270 (A3, H220)
Global Warming Potential (GWP)
0
UN-number
UN 3358
Circulation pump
MK pump
max. delivery
82 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
75 l/min
Delivery at 1,0 bar
66 l/min
Delivery at 1,5 bar
57 l/min
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is a refrigerated heating circulator with an air-cooled condenser, so it rejects heat to ambient air and needs adequate room ventilation.
Air-cooled machines require sufficient ambient airflow and room heat load planning at selection and installation.
Source
“Refrigerated Heating Circulator with air-cooled cooling machine and optical level indicator.”
Selection
The P-models are specifically intended for high-pressure-drop circuits, so they suit flow-through chemistry or semiconductor applications.
Confirms the pump head is matched to restrictive external circuits rather than low-pressure baths.
Source
“Unistat „P“ Models: Circulating pumps with a high discharge pressure for applications with high pressure drops, e.g. in the Flow-Through chemistry or in the Semicon industry.”
Installation
The pump connection is M30x1.5 male, so mating hoses or fittings must match this thread.
Wrong thread on hoses or adapters prevents leak-free connection and commissioning.
Source
“Pump connection
M30x1,5 male”
Installation
The unit is factory-configured for 400 V three-phase 50 Hz supply, so site power must match before delivery.
Three-phase supply and correct voltage/frequency must be confirmed at the installation site.