Unistat P530w Technical Data

Contents

  1. Overview p. 1
    Product introduction and general description of the Unistat P530w unit
  2. Spec p. 2
    Technical data according to DIN 12876 including performance parameters
  3. p. 2 table, 25 rows
  4. p. 1 table, 29 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 2,0 bar154 l/min
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C830 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,2 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
Net weight439 kg
sound pressure level +/- 4 dB(A)62 dB(A)
Power supply requirement400V 3~ 50Hz
max. current31,8 A
Fuse3x32A
Pressure equipment categoryII
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S608826

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat P530w (Order-No.: 5002.0101.01)

CharacteristicValue
Operating temperature range-55...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Thermooil at 250°C21 kW
Cooling power with Thermooil at 200°C21 kW
Cooling power with Thermooil at 100°C21 kW
Cooling power with Thermooil at 20°C21 kW
Cooling power with Ethanol at 0°C16 kW
Cooling power with Ethanol at -20°C9 kW
Cooling power with Ethanol at -40°C3 kW
Cooling power with Ethanol at -50°C1 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
Refrigerant quantity1,35 kg
CO2 equivalent0,0 t
Circulation pump: MK pump max. delivery201 l/min
Circulation pump: MK pump max. delivery pressure5,3 bar
Circulation pump: MK pump Delivery at 1,0 bar179 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 32 A fuse per phase, so the electrical installation must be sized accordingly.

Determines the site power supply, cable cross-section and protection device needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

The cooling water circuit must be supplied at 1–4 bar differential pressure and no more than 6 bar, with 830 l/h consumption at 15 °C.

Sizing of the cooling water supply, pump and piping must meet these pressure and flow limits.

Source
“min. cooling water differential pressure 1 bar”
Installation

Cooling water intake must stay between 5 °C and 20 °C, and pipes must be insulated if water can drop below the condensation point.

Out-of-range cooling water reduces cooling power and can cause condensation and corrosion.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity drops sharply with ethanol at low temperatures (16 kW at 0 °C down to 1 kW at -50 °C), so low-temperature duty must be checked against the curve.

Selection for sub-zero processes must account for the reduced cooling power at the required temperature.

Source
“at -50°C 1 kW”
Safety

The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant, so the installation area must respect flammability requirements.

A3 classification affects room ventilation, ignition-source control and site safety assessment.

Source
“R-1270 (A3, H220)”
Safety

Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must be verified.

High leakage current influences the choice of residual current device and earthing arrangement.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.

The customer must procure and fit the mains plug, affecting installation planning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are ±5 % and ±2 % but must not be combined in the same direction, e.g. -5 % voltage with +2 % frequency is not allowed.

Supply quality at the site must be checked to avoid operating outside the permitted envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Operation

The pump delivers 201 l/min at zero head but only 42 l/min at 5 bar, so high-pressure-drop loops must be checked against the pump curve.

Ensures the circulation pump can meet the flow demand of the external circuit at its pressure drop.

Source
“max. delivery pressure 5,3 bar”
Installation

Pump connections are M38x1.5 male with bore shape Y (60°) per DIN 3863, so matching hoses and fittings must be ordered.

Determines the hose/fitting specification needed for connecting the external circuit.

Source
“Pump connection M38x1,5 male”
Operation

Minimum filling capacity is 8.2 l and the expansion tank holds 19.9 l, which sets the heat-transfer fluid charge and expansion allowance.

Needed to calculate fluid volume, expansion volume and cost of heat-transfer fluid.

Source
“min. filling capacity 8,2 l”
Installation

The unit is IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in a wet or hot environment.

Site conditions and enclosure placement must respect the IP rating and ambient limits.

Source
“Degree of Protection IP20”
Compliance

The equipment is EMC class A group 1, so it is intended for industrial environments and may cause interference in residential areas.

Affects site selection and EMC coordination with sensitive equipment nearby.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Water quality must be controlled to avoid corrosion and fouling of these materials.

Source
“Please use suitable cooling water.”