Unistat P640w Technical Data Sheet

Contents

  1. Overview p. 1
    Product introduction and general description of the Unistat P640w unit
  2. Spec pp. 1-2
    Technical specifications and performance data according to DIN 12876
  3. p. 2 table, 17 rows
  4. p. 1 table, 29 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **918x963x1771 mm
Power supply requirement400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat P640w (Order-No.: 5020.0101.01)

CharacteristicValue
Operating temperature range-60...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power24 kW
Cooling power with Thermooil at 200°C32 kW
Cooling power with Thermooil at 100°C32 kW
Cooling power with Thermooil at 20°C35 kW
Cooling power with Ethanol at 0°C35 kW
Cooling power with Ethanol at -20°C30 kW
Cooling power with Ethanol at -40°C14 kW
Cooling power with Ethanol at -60°C3,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Circulation pump: MK pump max. delivery201 l/min
Circulation pump: MK pump max. delivery pressure5,3 bar
Delivery at 1,0 bar179 l/min
Delivery at 2,0 bar154 l/min
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min

Значения приведены дословно; строки охлаждения сведены по параметру (теплоноситель/температура) в одну колонку характеристик.

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase 50 Hz supply, so the electrical infrastructure must be verified before installation.

Determines the mains connection, cabling and protection needed at the installation site.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

The cooling water supply must be kept between 5 °C and 20 °C inlet temperature for the stated performance.

Cooling water temperature directly affects achievable cooling power and must be planned for.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

Cooling water pipes must be insulated if the water temperature can drop below the condensation point.

Prevents condensation and corrosion damage in the cooling water circuit.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

The cooling water circuit requires a differential pressure between 1 and 4 bar and must not exceed 6 bar.

Defines the required cooling water supply pressure and protects the unit from overpressure.

Source
“min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar max. cooling water pressure 6 bar”
Installation

Three-phase devices drawing less than 63 A are supplied with a cable but without a plug.

The customer must provide and fit the appropriate plug/connection on site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The pump connection is M38x1.5 male and must mate with ball-socket fittings per DIN 3863 and sleeve nuts per DIN 3870.

Ensures correct selection of hoses and fittings for the external circuit.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Installation

The unit needs sufficient space around its 918x963x1771 mm footprint and must be installed in an ambient range of 5–40 °C.

Space and ambient conditions affect access, ventilation and reliable operation.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Operation

The pump delivers 201 l/min max and 5.3 bar max, dropping to 42 l/min at 5.0 bar, so high-pressure-drop loops must be checked against the curve.

Confirms whether the pump can overcome the pressure drop of the external application.

Source
“max. delivery 201 l/min max. delivery pressure 5,3 bar”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s.

Limits the choice of heat transfer fluid, especially at low temperatures.

Source
“max. permissible kin. viscosity 100 mm²/s”
Operation

The unit has a minimum filling capacity of 8.9 l and a 19.9 l expansion tank.

Determines the required heat transfer fluid volume and expansion allowance.

Source
“min. filling capacity 8,9 l Filling capacity expansion tank 19,9 l”
Safety

The refrigerant R-452A has a GWP of 2141 and is classified A1/H280, which carries regulatory and handling obligations.

Affects refrigerant handling, leak checks and environmental compliance.

Source
“R-452A (A1, H280) Global Warming Potential (GWP) 2141”
Safety

The unit is safety class III/FL with adjustable overtemperature protection per DIN 12876.

Relevant for risk assessment and protection of temperature-sensitive external applications.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Compliance

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Must be checked against the site power quality to avoid damage or malfunction.

Source
“-5% voltage and +2% frequency -> not allowed!”
Compliance

The equipment is EMC class A, group 1 per EN55011, which is intended for industrial environments.

Class A equipment may cause interference in residential areas and constrains installation location.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”