Selection
The unit has no active refrigeration, so reaching 65 °C requires adequate heat loss from the application or suitable water cooling.
Prevents specifying the T330 for cooling duties below ambient without an external heat sink or water cooling.
Source
“These devices do not have an active refrigeration system.”
Installation
For air-cooled operation the application heat loss must be between 0.1 kW and 2.8 kW, and ambient must be well below 60 °C to reach 65 °C.
Defines the permissible process load window and ambient limit for air-cooled installations.
Source
“The heat loss must be between 0.1kW and 2.8kW, depending on the unit.”
Installation
Three-phase devices drawing less than 63 A are supplied with cable but without a plug, so the customer must provide the plug.
Affects electrical installation planning and bill of materials for the 400 V 3~ supply.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance
Voltage and frequency tolerances must be applied simultaneously: -5% voltage with +2% frequency is not allowed.
Prevents commissioning on a supply that violates EN60034-1 combined tolerance limits.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
Leakage current exceeds 3.5 mA, which requires appropriate protective earthing and RCD selection.
Impacts electrical safety design and choice of residual current protective devices.
Source
“Attention: leakage current > 3,5mA”
Compliance
The unit is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.
Determines EMC environment suitability and possible need for filtering or separation.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The unit requires 400 V 3~ 50 Hz supply and has IP20 protection, so it must be installed in a dry indoor location.
Defines the site power infrastructure and environmental protection requirements.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
Ambient temperature must stay between 5 °C and 40 °C for safe operation.
Sets the allowable room conditions for installation and ventilation planning.
Source
“min. ambient temperature 5 °C”
Selection
The pump delivers up to 96 l/min at 3.5 bar with M30x1.5 male connection, and the fluid viscosity must not exceed 50 mm²/s.
Ensures the circulation loop and thermofluid choice match pump capability and connection size.
Source
“max. permissible kin. viscosity 50 mm²/s”
Operation
Minimum filling capacity is 3 l and the expansion tank holds 24 l, which sets the minimum fluid charge and expansion allowance.
Guides thermofluid quantity and expansion volume planning for the loop.
Source
“min. filling capacity 3 l”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.
Confirms built-in safety limiting for unattended operation and compliance with DIN 12876.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Installation
Space requirements around the unit must be respected per the operating conditions.
Prevents overheating and ensures service access during installation.
Source
“Please respect space requirements.”