Unistat T330 HT Technical Data

Contents

  1. Overview p. 1
    Product name and general description of the Unistat T330 HT unit
  2. Spec p. 2
    Technical data according to DIN 12876 for the unit
  3. p. 1 table, 29 rows

Unistat T330 HT — Technical data according to DIN 12876

CharacteristicValue
Order-No.1004.0055.01
Operating temperature range65... 300 °C
Temperature stability at 70°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
External sensorPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power24 kW
Cooling power at 300°C18 kW
Cooling power at 200°C10 kW
Cooling power at 100°C3,5 kW
Circulation pump:MK pump
max. delivery96 l/min
max. delivery pressure3,5 bar
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
min. filling capacity7 l
Filling capacity expansion tank24 l
Overall dimensions WxDxH **540x749x1700 mm
Power supply (3 Phase)400V 3~ 50Hz
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.1.0/25

** Overall dimensions WxDxH — размеры указаны приблизительно и могут быть изменены.

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit covers a working temperature range of 65 to 300 °C, so it is only suitable for high-temperature closed-loop or open-bath applications, not for sub-ambient work.

Defines the process window and rules out low-temperature duties during equipment selection.

Source
“Operating temperature range 65... 300 °C”
Installation

The unit requires a 400 V three-phase 50 Hz supply, so the electrical infrastructure must be verified before ordering.

Determines the site power connection and whether a suitable three-phase feed is available.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and fit the site plug.

Affects installation planning and the electrical bill of materials at the site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The pump connection is an M30x1.5 male thread, so mating hoses or adapters must match this thread.

Prevents mismatch of process hoses and adapters during piping installation.

Source
“Pump connection M30x1,5 male”
Installation

The pump port bore follows DIN 3863 shape Y (60°) and requires ball-socket fittings with DIN 3870 sleeve nuts, so standard fittings may not seal correctly.

Ensures correct fitting selection and leak-free connections on site.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Installation

The enclosure is IP20, so the unit must be installed in a dry indoor location protected from water and dust ingress.

Constrains the installation environment and rules out washdown or outdoor siting.

Source
“Degree of Protection IP20”
Installation

Ambient temperature must stay between 5 °C and 40 °C, so the room climate must be checked before commissioning.

Out-of-range ambient conditions can derate performance or trip the unit.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The unit needs a minimum fill of 7 l and has a 24 l expansion tank, so the loop volume and fluid inventory must be planned accordingly.

Determines the required thermofluid quantity and expansion volume for the installation.

Source
“min. filling capacity 7 l Filling capacity expansion tank 24 l”
Safety

Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must account for this higher leakage.

Affects electrical safety design and the choice of residual-current protection.

Source
“Attention: leakage current > 3,5mA”
Safety

The unit is classified III / FL according to DIN 12876 and has adjustable overtemperature protection, so it can be applied in flammable-liquid duties only within that classification.

Governs whether the unit may be used with flammable fluids and what safety measures apply.

Source
“Safety classification III / FL”
Operation

The pump delivers up to 96 l/min at 3.5 bar, so the external circuit pressure drop must be checked against this curve.

Ensures the circulation pump can overcome the resistance of the connected process loop.

Source
“max. delivery 96 l/min max. delivery pressure 3,5 bar”
Operation

The maximum permissible kinematic viscosity is 50 mm²/s, so the thermofluid must be selected to stay within this limit at operating temperature.

Prevents pump overload and poor heat transfer from an overly viscous fluid.

Source
“max. permissible kin. viscosity 50 mm²/s”
Operation

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

Protects the compressor and electronics from an out-of-tolerance supply combination.

Source
“-5% voltage and +2% frequency -> not allowed!”
Compliance

The unit is EMC class A group 1 to EN55011, so it is intended for industrial environments and may cause interference in residential areas.

Determines EMC suitability of the installation site and possible filtering needs.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”