Unistat® TR 7 Temperature Control Unit — Operation Manual
Technical Reference Summary
Covered Models & Document Scope
Operation manual for the Unistat® TR series temperature control units (models listed on page 5 of the manual). Covers variants with/without lifting eyes, with/without casters, with/without leveling feet, transportation lock Types A/B/C, water-cooled units, and controller variants TR400, TR401, TR402 (HT-Circulator/Controller split units) plus optional Com.G@te® interface. Document version V2.6.0en/07.03.22//17.12.
Section-by-Section Technical Breakdown
Describes in-plant transport, lifting with or without lifting eyes, mounting/removing leveling feet, positioning with casters, and the three transportation lock types. Transport must be upright only; casters are not suitable for transport and are loaded symmetrically with 25% of total mass. Lifting eyes are designed only for a load without inclination (0°).
- Only transport the temperature control unit in an upright position; horizontal transport damages the compressor.
- Only transport an emptied temperature control unit to avoid overflowing thermal fluid.
- Casters (if present) are not suitable for transport; they are symmetrically loaded with 25% of the total mass.
- Lifting eyes are only designed for a load without inclination (0°); always use adequately dimensioned load handling attachments.
- Crane or industrial truck lifting force must be equal to or greater than the weight of the temperature control unit (see datasheet).
- Leveling feet can only be mounted while the unit is lifted; do not lower until all leveling feet are mounted.
- Parking brakes and/or leveling feet must be activated/unscrewed before putting the unit into operation.
Specifies ambient conditions, EMC-specific notes and installation conditions for the temperature control unit. Also covers recommended temperature control and cooling water hoses, wrench sizes and torques, and water-cooled units.
- Ambient conditions and EMC-specific notes must be observed.
- Recommended temperature control and cooling water hoses are specified.
- Wrench sizes and torques are given for connections.
- Special notes apply to temperature control units with water cooling.
Covers unscrewing/activating leveling feet, opening/closing the ball valve on >Drain< [8], checking the knurled screw on >Emptying of residues< [10], connecting the hose to >Overflow< [12], inert gas (nitrogen) connections, connecting the controller (TR401, TR402 only), and connecting the functional earth.
- Open/close ball valve on >Drain< [8].
- Check the knurled screw from the >Emptying of residues< [10].
- Connect the hose to the >Overflow< [12].
- Connections for inert gas (nitrogen) are provided.
- Connecting the temperature control unit with the controller applies only to TR401 and TR402.
- Connect the functional earth.
Describes connecting an externally closed application and connecting to the power supply via socket with protective earth (PE) or via hard wiring, including converting the power supply connection.
- Connection using socket with protective earth (PE).
- Connection via hard wiring is possible.
- The power supply connection can be converted.
Describes general and other functions of the temperature control unit, information on thermal fluids, planning notes for tests, the Pilot ONE® controller functional overview, clock/event function, rechargeable accumulator, programmable event functions, touch screen operation, display and control instruments, and function examples such as displaying software version, Start & Stop, copying settings to a USB flash drive, and restoring factory settings.
- Pilot ONE® controller functional overview.
- Clock/event function with rechargeable accumulator and programmable event function ('Alarm clock event', 'Program event').
- Operation via touch screen and touchbuttons with categories, sub-categories and dialogs.
- Function examples: display software version, Start & Stop, save/load settings to/from USB flash drive, restore factory settings with or without overtemperature protection.
Covers turning the unit on/off, setting overtemperature protection (OT limit: heating, OT expansion vessel, Process Safety, Display OT values), testing overtemperature protection, adjusting the Delta T limiter, selecting internal or process temperature control, monitoring Pt100 sensors, auto/expert control parameter configuration, setpoint thresholds and setpoint setting, and filling, venting, degassing and draining.
- Overtemperature protection threshold can only be set by entering a code randomly generated by Pilot ONE®.
- Process Safety mode requires the compressor to be set to Permanently ON ([System Settings] > [Power/ECO Settings] > [Compressor ON/OFF/AUTO] > [Permanently ON]).
- Delta T limiter can be adjusted and changed.
- Temperature control can be selected as internal or process temperature.
- Pt100 temperature sensors are monitored.
- Auto/expert mode with 'Find parameters', 'Control Dynamics', 'Fluid Properties', 'Display parameters', manual configuration and reset parameters.
- Filling, venting, degassing and draining procedures for externally closed applications.
Describes automatic operation, starting and ending the temperature control process, and temperature control via a created temperature control program including starting and ending/cancelling the program.
- Starting and ending the temperature control process.
- Starting and ending/cancelling a temperature control program.
Covers interfaces at the Pilot ONE® controller (Ethernet RJ45 10/100 Mbps, USB 2.0 host and device), interfaces at the Unistat® TR400 (Pt100 process controller sensor jack, service interface, POKO floating contact alarm, ECS standby jack), interfaces at TR401/TR402 (HT-Circulator and Controller backs), optional Com.G@te® interfaces (LEVEL, POKO, AIF Reg-E-Prog, ECS, RS232/RS485), and firmware update.
- 10/100 Mbps Ethernet for RJ45 network sockets.
- USB-2.0 interface, host and device.
- Pt100 process controller sensor connection jack.
- Service interface.
- Connector POKO (floating contact) alarm.
- Jack ECS (External Control Signal) standby.
- Optional Com.G@te®: Jack LEVEL, POKO, AIF Reg-E-Prog, ECS, RS232/RS485 serial.
- Firmware update procedure.
Covers messages from the temperature control unit, replacing the Pilot ONE®, maintenance including function check and visual inspection, replacing temperature control or coolant hoses, cleaning the hat-type strainer (dirt trap) on water-cooled units, emptying the drip tray, thermal fluid inspection/replacement and circuit cleaning, cleaning surfaces, inspecting the mechanical seal, plug contacts, and decontamination/repairs.
- Messages from the temperature control unit are described.
- Replacing the Pilot ONE® controller.
- Function check and visual inspection.
- Replacing temperature control hoses and coolant hoses.
- Clean hat-type strainer (dirt trap) on water-cooled temperature control units.
- Empty the drip tray.
- Thermal fluid inspection, replacement and circuit cleaning.
- Inspect the mechanical seal and plug contacts.
Covers safety instructions and basic principles for shutting down, switch-off, draining the temperature control unit, draining the cooling water, uninstalling an external application, transportation lock, deinstalling the collecting container, packing and shipping.
- Switch-off procedure.
- Draining the temperature control unit and cooling water.
- Uninstalling an external application.
- Transportation lock for shipping.
- Deinstalling the collecting container, packing and shipping.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Only transport the temperature control unit in an upright position; horizontal transport causes damage to the compressor. p.21 ↗ | Plan rigging and transport routes to keep the unit upright at all times. | whole series |
| Casters (if present) are not suitable for transport and are symmetrically loaded with 25% of the total mass of the temperature control unit. p.21 ↗ | Use an industrial truck or crane for transport; use casters only for positioning at the place of installation. | units with casters |
| Lifting eyes are only designed for a load without inclination (0°); the crane or industrial truck must have a lifting force equal to or greater than the weight of the temperature control unit. p.21 ↗ | Select lifting equipment rated for the unit weight and avoid angled lifts. | units with lifting eyes |
| Leveling feet can only be mounted while the temperature control unit is lifted; do not lower the unit until all leveling feet have been mounted. p.22 ↗ | Secure the unit against slipping/lowering before mounting leveling feet; never stand or lie under the unit. | units with leveling feet removed for shipping |
| The overtemperature protection threshold can only be set if a code randomly generated by Pilot ONE® is entered correctly. p.18 ↗ | Prevents accidental changes to the safety limit; document the code procedure for authorized personnel. | whole series |
| Process Safety mode requires the compressor to be set to Permanently ON via [System Settings] > [Power/ECO Settings] > [Compressor ON/OFF/AUTO] > [Permanently ON]. p.19 ↗ | Configure the compressor setting before relying on Process Safety to keep cooling active at the overtemperature trigger value. | whole series |
| Low level protection is achieved via a pressure sensor in the fluid circuit; air in the system (fill level too low, inadequately vented) prevents the pressure from reaching the specified value, interrupting temperature control and circulation. p.19 ↗ | Ensure proper filling and venting to avoid low-level trips. | whole series |
| Power interruption response can be set via Pilot ONE®: OFF/Standby (default), ON/Temperature control active, or Power Failure Auto. p.20 ↗ | Select the response appropriate to the process; Power Failure Auto resumes temperature control automatically after a power outage. | whole series |
| Alarm response can be configured to switch off temperature control or control temperature to a safe setpoint (2nd setpoint). p.20 ↗ | Define alarm strategy based on process risk. | whole series |
| The temperature control unit must NOT be installed or started up within an ATEX zone without an Ex px cabinet. p.15 ↗ | Verify area classification before installation; use Ex px cabinet where required and observe the ATEX annex. | whole series |
| Use with medical devices (e.g. in vitro diagnostic procedure) or for direct foodstuff temperature control is NOT permissible. p.16 ↗ | Confirm application suitability before selection. | whole series |
| Do not permanently operate the temperature control unit with very dynamic and high temperature changes recurring in short intervals; the guarantee expires if exposed to this type of application. p.16 ↗ | Avoid rapid cyclic thermal loading of the plate heat exchanger; consult Huber for such applications. | whole series |
| The responsible body must install a condensation water / thermal fluid drip tray below the temperature control unit. p.17 ↗ | Include a drip tray in the installation design; national legislation may prescribe it. | whole series |
| The >Mains isolator< [36] (if present) can be locked in the off position to facilitate safe system installation and maintenance. p.17 ↗ | Develop a lock-out/tag-out procedure for the energy source in accordance with local regulations (e.g. CFR 1910.147 for the US). | units with mains isolator |
| Emergency strategy: turn the >Mains isolator< [36] to '0'; for units with additional >Appliance switch< [37], also turn it to '0'; for units with >Emergency stop switch< [70], press it then turn the >Main switch< [36] to '0'. p.19 ↗ | Identify the switch combination from the wiring diagram and train operators on the emergency shutdown sequence. | whole series |
| Connecting the temperature control unit with the controller applies only to TR401 and TR402. p.32 ↗ | For TR401/TR402, follow the separate controller connection procedure. | TR401, TR402 |
| The Pilot ONE® controller provides 10/100 Mbps Ethernet RJ45 and USB-2.0 host/device interfaces. p.69 ↗ | Plan network and USB connectivity for data logging and firmware updates. | whole series |
| Optional Com.G@te® provides LEVEL, POKO (floating contact) alarm, AIF Reg-E-Prog, ECS standby and RS232/RS485 serial interfaces. p.75 ↗ | Select Com.G@te® option when serial or external control signal integration is required. | units with Com.G@te® option |