Technical Reference Summary

Covered Models & Document Scope

This manual covers the Labconco Rapid Digestor-25 (catalog 2301200), Rapid Digestor-4 (catalog 2308000), and the optional 25-Place Fume Removal System (2350025) and 4-Place Fume Removal System (2354000). The 25-place digestor is a 208-240V AC unit with 25 ceramic heaters; the 4-place digestor is a 115V AC unit with 4 heaters. The 25-place fume removal system includes a water jet eductor, stand, and drip pan; the 4-place system does not include these items.

Section-by-Section Technical Breakdown

pp. 9-10
Specifications

Electrical and physical specifications for the 25-place and 4-place Rapid Digestors, including voltage, current, fusing, dimensions, weights, sample capacity, temperature range, and heat-up times.

  • 25-Place: 208-240V AC single phase, 16 A max at 230V, 25 A fuses, 50/60 Hz.
  • 4-Place: 115V AC, 6 A max, 8 A fusing, 50/60 Hz.
  • Both models: ambient to 450°C, sample size up to 5 g, 250 ml straight or constricted tubes.
  • Heat-up to 410°C: approximately 1 hour (25-place) or 45 minutes (4-place).
pp. 11-12
Installation

Instructions for unpacking and installing the digestor and control unit, including electrical connections, placement under a fume removal system, and assembly of the optional fume removal system.

  • Locate digestor under a suitable acid-resistant fume removal system; control unit must be outside the hood away from corrosive vapors.
  • Connect power cord from control unit to digestor and the small gray sensor cord to the 2-prong receptacle.
  • 25-place requires 208-240V AC 3-wire installation fused at 25 A; 4-place requires 115V AC.
  • Fume removal system: assemble glass manifold, secure with stainless steel retainer clips, connect to water aspirator with high-temperature acid-resistant hose.
pp. 13-15
Normal Operation

General Kjeldahl digestion procedure for nitrate-free samples and operation of the optional fume removal system, including reagent addition, temperature settings, digestion time, and cooling/dilution steps.

  • Set temperature to 410°C and allow approximately one hour for the 25-place unit to reach temperature.
  • Add 9 g K2SO4, 0.42 g HgO, and 15 ml H2SO4 to samples in digestion tubes.
  • Digest at 410°C for 45 minutes for complete nitrogen recovery; reduce temperature or sample size if foaming is uncontrollable.
  • After digestion, cool tubes 5-8 minutes and dilute with approximately 100 ml water before a salt cake forms.
  • Fume removal: adjust water flow to maximum for first 5-7 minutes, then reduce to the point where all fumes are removed.
pp. 16
Routine Maintenance

Cleaning and maintenance procedures for the digestor and fume removal system, including ceramic heater cores, glass manifold, and drip pan.

  • Clean surfaces with detergent and water when unit is cool and disconnected from power.
  • Keep ceramic heater cores clear of foreign material for free tube insertion.
  • Disassemble and wash glass manifold with detergent and water; inspect each bulb for obstructed vent holes.
  • Rinse drip pan with water after each use.
pp. 17-18
Troubleshooting

Table of symptoms, probable causes, and corrective measures for common electrical and heating faults.

  • Main pilot light does not illuminate: check ON-OFF switch, main power fuse/circuit breaker, 25 A fuse (8 A on 4-place), and switch continuity.
  • No heat with pilot light illuminated: check sensor continuity across pins 1 & 2 of female sensor receptacle.
  • Indicator lamp not illuminated: check bulb, resistor, and ceramic heater continuity.
  • Uneven heat (not applicable to 4-place): check strip heater continuity.
pp. 19-21
Service

Replacement procedures for main power switch, strip heater, indicator lamp, resistor, ceramic heater, and sensor.

  • Always disconnect main line cord before servicing.
  • Main power switch: remove #10-24 screws from control unit sides, replace switch, rewire per wiring diagram.
  • Strip heater (25-place only): remove six screws from bottom of digestor, tilt heater containment pan, replace heater.
  • Ceramic heater: thread wires through hole in insulator board; do not bend wires or allow folding between heater casing and insulator board.
  • Sensor: disconnect 2-pin connector (left side of 25-place, rear of 4-place), replace heater-sensor assembly.
pp. 23-28
Replacement Parts

Catalog numbers and descriptions for replacement parts for the 25-place and 4-place digestors, controllers, and fume removal systems.

  • 25-Place Digestor parts: heater assembly 2303400, resistor 1324900, strip heater 1345900, heater assembly 2302000, tubes 2303025/2304025, flask rack 2305100.
  • 25-Place Controller parts: pilot light 1270700, switch 1324000, control assembly 2311900.
  • 4-Place Digestor parts: heater assembly 2308400, tubes 2303005/2304005, flask rack 2309700.
  • 4-Place Controller parts: pilot light 1276900, switch 1325700, control assembly 2312400.
  • Fume Removal System 2350025 parts: water jet eductor 1357500, manifold 2351300, glass pipe 2353200, tubing 1540000, clip 2351700.
  • Fume Removal System 2354000 parts: tubing 1622600, glass pipe 2308500.
pp. 29-36
Wiring Diagrams & Detailed Drawings

Wiring diagrams for the 25-place and 4-place controllers and digestors, plus detailed drawings for the water eductor assembly and installation methods.

  • Wiring diagram for 25-place controller (p.29).
  • Wiring diagram for 4-place controller (p.30).
  • Wiring diagram for 25-place digestor (p.31).
  • Wiring diagram for 4-place digestor (p.32).
  • Water eductor assembly and installation methods (p.33-34).

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
25-Place Rapid Digestor operates on 208-240V AC single phase, 50/60 Hz, draws 16 A max at 230V, and requires 25 A fusing. p.9 ↗ Electrical supply must be sized for 25 A and 208-240V; use 3-wire installation. Rapid Digestor-25 (2301200)
4-Place Rapid Digestor operates on 115V AC, 50/60 Hz, draws 6 A max, and requires 8 A fusing. p.10 ↗ Standard 115V laboratory outlet with 8 A protection is sufficient. Rapid Digestor-4 (2308000)
Temperature range is ambient to 450°C; heat-up time to 410°C is approximately one hour for the 25-place and 45 minutes for the 4-place. p.9 ↗ Allow sufficient warm-up time before starting digestion; set point dial can be set up to 450°C. Both models
Maximum sample size is 5 g for solid materials or approximately 25 ml for liquid samples. p.5 ↗ Do not exceed these limits to avoid foaming and incomplete digestion. Both models
Digestion tubes are 250 ml borosilicate glass, either straight or constricted with volumetric mark, 11-15/16"L x 1.65 OD. p.9 ↗ Use only specified tubes for proper fit in ceramic heaters and rack. Both models
General Kjeldahl digestion for feed/grain: add 9 g K2SO4, 0.42 g HgO, and 15 ml H2SO4; digest at 410°C for 45 minutes. p.13 ↗ This is the standard reagent recipe and digestion time for complete nitrogen recovery. Both models
After digestion, cool tubes 5-8 minutes and dilute with approximately 100 ml water before a salt cake forms. p.14 ↗ Timing is critical: dilution must occur before caking but after the digest is cool enough to handle the exothermic reaction. Both models
Fume removal system water jet eductor requires 2-3 gallons/minute water flow and 1/2" male connection (1/2" x 3/8" nipple provided). p.7 ↗ Ensure water supply and drain can handle 2-3 GPM; drain must be capable of this flow. Fume Removal System 2350025
During initial digestion stages, adjust water flow to maximum for 5-7 minutes, then reduce to the point where all fumes are removed. p.15 ↗ Excessive water flow throughout digestion can cause acid removal, sample caking, and nitrogen loss. Fume Removal System 2350025
Exhaust manifold requires a tube in each corner and the center of the digestor for support; a full rack is not necessary. p.15 ↗ Minimum tube placement pattern must be followed to support the manifold. Fume Removal System 2350025
Control unit must be located outside the hood away from vapors; digestor must be placed in an acid-resistant venting system. p.11 ↗ Corrosive vapors will damage the control unit if placed inside the hood. Both models
The 25-place digestor uses a 25 A fuse on the control unit; the 4-place uses an 8 A fuse. p.17 ↗ Use correct fuse rating for troubleshooting and replacement. Both models
Sensor continuity is checked across pins 1 & 2 of the female sensor receptacle; open circuit indicates defective sensor. p.17 ↗ Use an ohmmeter to diagnose no-heat conditions with pilot light illuminated. Both models
Ceramic heater replacement: thread wires through hole in insulator board; do not bend wires or allow folding between heater casing and insulator board. p.20 ↗ Improper wire routing greatly reduces heater life. Both models
Perchloric acid must not be used with this equipment due to risk of rapid oxidation and unstable residue. p.14 ↗ Use alternative digestion methods for perchloric acid; equipment is not compatible. Both models
Fume removal system 2354000 does not include water jet eductor, stand, or drip pan; a non-metallic water aspirator such as Nalgene 6140 is suitable. p.12 ↗ For 4-place system, provide separate aspirator and drain arrangement. Fume Removal System 2354000

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Main pilot light does not illuminate p.17 ↗ ON-OFF switch in off position Turn switch to ON position
Main pilot light does not illuminate p.17 ↗ Main power fuse blown or circuit breaker tripped Replace fuse or reset circuit breaker
Main pilot light does not illuminate p.17 ↗ 25 Amp fuse on control unit blown (8 Amp on Rapid Digestor 4-Place) Check continuity of fuse with ohmmeter. Replace fuse if blown. If fuse continues to blow, locate cause.
Main pilot light does not illuminate p.17 ↗ ON-OFF switch defective Replace switch. Refer to Service Section for replacement procedure.
No heat applied to unit. Main pilot light illuminated p.17 ↗ Sensor defective Check continuity of sensor by placing ohmmeter across pins 1 & 2 of female sensor receptacle. If resistance measurements show sensor to be open, replace sensor by following Service Section on Sensor Replacement.
Indicator lamp not illuminated p.17 ↗ Bulb defective Check continuity of bulb with ohmmeter and replace if defective following Service Section-Indicator Lamp Replacement.
Indicator lamp not illuminated p.17 ↗ Defective resistor Check continuity of resistor for the row of heaters behind indicator lamp not illuminated using ohmmeter.
Indicator lamp not illuminated p.17 ↗ Defective ceramic heater Check continuity of each heater contained in the row behind indicator lamp that is not illuminated. Replace defective heater following Ceramic Heater Replacement section.
Uneven heat applied to unit (not applicable to Rapid Digestor 4-Place) p.18 ↗ Strip heater defective Check continuity of each strip heater and replace defective heater following Strip Heater Replacement section.