Rev k rapid digestors fume removal systems
Technical Reference Summary
Covered Models & Document Scope
This manual covers the Labconco Rapid Digestor-25 (catalog 2301200), Rapid Digestor-4 (catalog 2308000), and the optional 25-Place Fume Removal System (2350025) and 4-Place Fume Removal System (2354000). The 25-place digestor is a 208-240V AC unit with 25 ceramic heaters; the 4-place digestor is a 115V AC unit with 4 heaters. The 25-place fume removal system includes a water jet eductor, stand, and drip pan; the 4-place system does not include these items.
Section-by-Section Technical Breakdown
Electrical and physical specifications for the 25-place and 4-place Rapid Digestors, including voltage, current, fusing, dimensions, weights, sample capacity, temperature range, and heat-up times.
- 25-Place: 208-240V AC single phase, 16 A max at 230V, 25 A fuses, 50/60 Hz.
- 4-Place: 115V AC, 6 A max, 8 A fusing, 50/60 Hz.
- Both models: ambient to 450°C, sample size up to 5 g, 250 ml straight or constricted tubes.
- Heat-up to 410°C: approximately 1 hour (25-place) or 45 minutes (4-place).
Instructions for unpacking and installing the digestor and control unit, including electrical connections, placement under a fume removal system, and assembly of the optional fume removal system.
- Locate digestor under a suitable acid-resistant fume removal system; control unit must be outside the hood away from corrosive vapors.
- Connect power cord from control unit to digestor and the small gray sensor cord to the 2-prong receptacle.
- 25-place requires 208-240V AC 3-wire installation fused at 25 A; 4-place requires 115V AC.
- Fume removal system: assemble glass manifold, secure with stainless steel retainer clips, connect to water aspirator with high-temperature acid-resistant hose.
General Kjeldahl digestion procedure for nitrate-free samples and operation of the optional fume removal system, including reagent addition, temperature settings, digestion time, and cooling/dilution steps.
- Set temperature to 410°C and allow approximately one hour for the 25-place unit to reach temperature.
- Add 9 g K2SO4, 0.42 g HgO, and 15 ml H2SO4 to samples in digestion tubes.
- Digest at 410°C for 45 minutes for complete nitrogen recovery; reduce temperature or sample size if foaming is uncontrollable.
- After digestion, cool tubes 5-8 minutes and dilute with approximately 100 ml water before a salt cake forms.
- Fume removal: adjust water flow to maximum for first 5-7 minutes, then reduce to the point where all fumes are removed.
Cleaning and maintenance procedures for the digestor and fume removal system, including ceramic heater cores, glass manifold, and drip pan.
- Clean surfaces with detergent and water when unit is cool and disconnected from power.
- Keep ceramic heater cores clear of foreign material for free tube insertion.
- Disassemble and wash glass manifold with detergent and water; inspect each bulb for obstructed vent holes.
- Rinse drip pan with water after each use.
Table of symptoms, probable causes, and corrective measures for common electrical and heating faults.
- Main pilot light does not illuminate: check ON-OFF switch, main power fuse/circuit breaker, 25 A fuse (8 A on 4-place), and switch continuity.
- No heat with pilot light illuminated: check sensor continuity across pins 1 & 2 of female sensor receptacle.
- Indicator lamp not illuminated: check bulb, resistor, and ceramic heater continuity.
- Uneven heat (not applicable to 4-place): check strip heater continuity.
Replacement procedures for main power switch, strip heater, indicator lamp, resistor, ceramic heater, and sensor.
- Always disconnect main line cord before servicing.
- Main power switch: remove #10-24 screws from control unit sides, replace switch, rewire per wiring diagram.
- Strip heater (25-place only): remove six screws from bottom of digestor, tilt heater containment pan, replace heater.
- Ceramic heater: thread wires through hole in insulator board; do not bend wires or allow folding between heater casing and insulator board.
- Sensor: disconnect 2-pin connector (left side of 25-place, rear of 4-place), replace heater-sensor assembly.
Catalog numbers and descriptions for replacement parts for the 25-place and 4-place digestors, controllers, and fume removal systems.
- 25-Place Digestor parts: heater assembly 2303400, resistor 1324900, strip heater 1345900, heater assembly 2302000, tubes 2303025/2304025, flask rack 2305100.
- 25-Place Controller parts: pilot light 1270700, switch 1324000, control assembly 2311900.
- 4-Place Digestor parts: heater assembly 2308400, tubes 2303005/2304005, flask rack 2309700.
- 4-Place Controller parts: pilot light 1276900, switch 1325700, control assembly 2312400.
- Fume Removal System 2350025 parts: water jet eductor 1357500, manifold 2351300, glass pipe 2353200, tubing 1540000, clip 2351700.
- Fume Removal System 2354000 parts: tubing 1622600, glass pipe 2308500.
Wiring diagrams for the 25-place and 4-place controllers and digestors, plus detailed drawings for the water eductor assembly and installation methods.
- Wiring diagram for 25-place controller (p.29).
- Wiring diagram for 4-place controller (p.30).
- Wiring diagram for 25-place digestor (p.31).
- Wiring diagram for 4-place digestor (p.32).
- Water eductor assembly and installation methods (p.33-34).
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| 25-Place Rapid Digestor operates on 208-240V AC single phase, 50/60 Hz, draws 16 A max at 230V, and requires 25 A fusing. p.9 ↗ | Electrical supply must be sized for 25 A and 208-240V; use 3-wire installation. | Rapid Digestor-25 (2301200) |
| 4-Place Rapid Digestor operates on 115V AC, 50/60 Hz, draws 6 A max, and requires 8 A fusing. p.10 ↗ | Standard 115V laboratory outlet with 8 A protection is sufficient. | Rapid Digestor-4 (2308000) |
| Temperature range is ambient to 450°C; heat-up time to 410°C is approximately one hour for the 25-place and 45 minutes for the 4-place. p.9 ↗ | Allow sufficient warm-up time before starting digestion; set point dial can be set up to 450°C. | Both models |
| Maximum sample size is 5 g for solid materials or approximately 25 ml for liquid samples. p.5 ↗ | Do not exceed these limits to avoid foaming and incomplete digestion. | Both models |
| Digestion tubes are 250 ml borosilicate glass, either straight or constricted with volumetric mark, 11-15/16"L x 1.65 OD. p.9 ↗ | Use only specified tubes for proper fit in ceramic heaters and rack. | Both models |
| General Kjeldahl digestion for feed/grain: add 9 g K2SO4, 0.42 g HgO, and 15 ml H2SO4; digest at 410°C for 45 minutes. p.13 ↗ | This is the standard reagent recipe and digestion time for complete nitrogen recovery. | Both models |
| After digestion, cool tubes 5-8 minutes and dilute with approximately 100 ml water before a salt cake forms. p.14 ↗ | Timing is critical: dilution must occur before caking but after the digest is cool enough to handle the exothermic reaction. | Both models |
| Fume removal system water jet eductor requires 2-3 gallons/minute water flow and 1/2" male connection (1/2" x 3/8" nipple provided). p.7 ↗ | Ensure water supply and drain can handle 2-3 GPM; drain must be capable of this flow. | Fume Removal System 2350025 |
| During initial digestion stages, adjust water flow to maximum for 5-7 minutes, then reduce to the point where all fumes are removed. p.15 ↗ | Excessive water flow throughout digestion can cause acid removal, sample caking, and nitrogen loss. | Fume Removal System 2350025 |
| Exhaust manifold requires a tube in each corner and the center of the digestor for support; a full rack is not necessary. p.15 ↗ | Minimum tube placement pattern must be followed to support the manifold. | Fume Removal System 2350025 |
| Control unit must be located outside the hood away from vapors; digestor must be placed in an acid-resistant venting system. p.11 ↗ | Corrosive vapors will damage the control unit if placed inside the hood. | Both models |
| The 25-place digestor uses a 25 A fuse on the control unit; the 4-place uses an 8 A fuse. p.17 ↗ | Use correct fuse rating for troubleshooting and replacement. | Both models |
| Sensor continuity is checked across pins 1 & 2 of the female sensor receptacle; open circuit indicates defective sensor. p.17 ↗ | Use an ohmmeter to diagnose no-heat conditions with pilot light illuminated. | Both models |
| Ceramic heater replacement: thread wires through hole in insulator board; do not bend wires or allow folding between heater casing and insulator board. p.20 ↗ | Improper wire routing greatly reduces heater life. | Both models |
| Perchloric acid must not be used with this equipment due to risk of rapid oxidation and unstable residue. p.14 ↗ | Use alternative digestion methods for perchloric acid; equipment is not compatible. | Both models |
| Fume removal system 2354000 does not include water jet eductor, stand, or drip pan; a non-metallic water aspirator such as Nalgene 6140 is suitable. p.12 ↗ | For 4-place system, provide separate aspirator and drain arrangement. | Fume Removal System 2354000 |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Main pilot light does not illuminate p.17 ↗ | ON-OFF switch in off position | Turn switch to ON position | |
| Main pilot light does not illuminate p.17 ↗ | Main power fuse blown or circuit breaker tripped | Replace fuse or reset circuit breaker | |
| Main pilot light does not illuminate p.17 ↗ | 25 Amp fuse on control unit blown (8 Amp on Rapid Digestor 4-Place) | Check continuity of fuse with ohmmeter. Replace fuse if blown. If fuse continues to blow, locate cause. | |
| Main pilot light does not illuminate p.17 ↗ | ON-OFF switch defective | Replace switch. Refer to Service Section for replacement procedure. | |
| No heat applied to unit. Main pilot light illuminated p.17 ↗ | Sensor defective | Check continuity of sensor by placing ohmmeter across pins 1 & 2 of female sensor receptacle. If resistance measurements show sensor to be open, replace sensor by following Service Section on Sensor Replacement. | |
| Indicator lamp not illuminated p.17 ↗ | Bulb defective | Check continuity of bulb with ohmmeter and replace if defective following Service Section-Indicator Lamp Replacement. | |
| Indicator lamp not illuminated p.17 ↗ | Defective resistor | Check continuity of resistor for the row of heaters behind indicator lamp not illuminated using ohmmeter. | |
| Indicator lamp not illuminated p.17 ↗ | Defective ceramic heater | Check continuity of each heater contained in the row behind indicator lamp that is not illuminated. Replace defective heater following Ceramic Heater Replacement section. | |
| Uneven heat applied to unit (not applicable to Rapid Digestor 4-Place) p.18 ↗ | Strip heater defective | Check continuity of each strip heater and replace defective heater following Strip Heater Replacement section. |