CentriVap Centrifugal Concentrators & Cold Trap (7539800 Rev L)
Technical Reference Summary
Covered Models & Document Scope
This manual covers Labconco CentriVap Centrifugal Concentrators and CentriVap Cold Traps. Concentrator catalog numbers 7810010–7810041 (standard, heat boost, glass lid, acid resistant options; 115V/60Hz and 230V/50Hz/60Hz). Cold Trap catalog numbers 7811020–7811040 (-50°C), 7460020–7460040 (-85°C), 7385020–7385040 (-105°C); 115V/60Hz and 230V/50Hz/60Hz. Concentrator Systems 7982010–7984037 (Aqueous, Acid, Solvent) and Gel Dryers 7475010–7475012. 230V 50Hz models comply with CE regulations.
Section-by-Section Technical Breakdown
The Concentrator requires a dedicated grounded outlet: 15 A circuit breaker/fuse for 115V 60Hz models and 8 A for 230V 50/60Hz models. The Cold Trap also requires a dedicated outlet: 15 A for 115V -50°C and -85°C models, 20 A for 115V -105°C models, and 8 A for 230V models. Power cords may be replaced with an approved cord of suitable style if the supplied cord does not match the receptacle.
- Concentrator 115V (60 Hz): 15 A circuit breaker or fuse.
- Concentrator 230V (50/60 Hz): 8 A circuit breaker or fuse.
- Cold Trap 115V (-50°C & -85°C): 15 A at 115V (60 Hz).
- Cold Trap 115V (-105°C): 20 A at 115V (60 Hz).
- Cold Trap 230V: 8 A at 230V (50/60 Hz).
- Dedicated grounded electrical outlet required for each unit.
The Concentrator must be placed on a stable, flat, level surface. The system should be located within a fume hood if hazardous or flammable solvents are used, and the vacuum pump should always be vented to a fume hood. No person or hazardous material should be within 12 inches of the operating Concentrator.
- Stable, flat, level surface required.
- Fume hood required if hazardous or flammable solvents are used.
- Vacuum pump exhaust must be vented to a fume hood.
- No person or hazardous material within 12 inches during operation.
- Accessory secondary trap reduces solvent exhaust but does not replace fume hood venting.
The user must provide a vacuum pump. For aqueous samples, a pump with 82 L/min free air flow and <2 mBar ultimate pressure is adequate; more volatile samples can use a diaphragm pump with 42 L/min free airflow and 200 mBar vacuum. The pump inlet must accept 0.50 ID hose, and pumps for 115V models need a NEMA 5-15P plug while 230V models need a reverse IEC 320 plug to connect to the Concentrator outlet.
- Aqueous samples: 82 L/min free air flow, <2 mBar ultimate pressure.
- Volatile samples: diaphragm pump, 42 L/min free airflow, 200 mBar vacuum.
- Pump inlet fitting suitable for 0.50 ID hose.
- 115V pump: 115V, 15 A NEMA 5-15P plug.
- 230V pump: reverse IEC 320 plug.
- Wetted parts should be PTFE or PTFE-coated for aggressive samples.
- Explosion-proof or solvent-suitable pump recommended for flammable solvents.
When using a non-corrosion-resistant mechanical pump, the CentriVap Cold Trap or equivalent must be used in the vacuum line to protect the pump from solvent vapors. For acids, a soda lime acid secondary trap is advisable. Several CentriVap components are stainless steel or aluminum and can be degraded by acids; contact Labconco before evaporating acids.
- Cold Trap or equivalent required with non-corrosion-resistant mechanical pumps.
- Soda lime acid secondary trap recommended when evaporating acids.
- Stainless steel and aluminum components can be degraded by acids.
- Moisture, solvent, and radioisotope traps are available as accessories.
The Concentrator, Cold Trap, and vacuum pump may be arranged in either order. The Cold Trap draws air in through the right side and exhausts through the left side, so at least 3 inches must be allowed between the Cold Trap sides and adjacent walls. Hoses connect the Concentrator outlet to a Cold Trap cover barb, then from the other barb to the pump inlet, or through an optional Secondary Trap.
- Minimum 3 inches clearance between Cold Trap sides and walls.
- Cold Trap airflow: intake right side, exhaust left side.
- Hose from Concentrator rear tube to Cold Trap cover barb fitting.
- Second hose from remaining Cold Trap barb to pump inlet.
- If Secondary Trap used: Cold Trap cover to Secondary Trap 'out', then Secondary Trap remaining connector to pump inlet.
- Secure all hoses with supplied clamps.
Plug the Concentrator and Cold Trap power cords into suitable receptacles. The vacuum pump power cord plugs into the receptacle on the back of the Concentrator, and the pump's own switch (if present) must be turned ON; the Concentrator then controls the pump.
- Concentrator power cord to suitable receptacle.
- Cold Trap power cord to suitable receptacle.
- Vacuum pump power cord to receptacle on back of Concentrator.
- Turn pump switch ON if equipped; pump is controlled by Concentrator.
Place the rotor on the Concentrator shaft and rotate slightly to engage the drive pin with the rotor hub slots. The top of the shaft should align with the top of the rotor hub. Do not use a damaged rotor.
- Engage drive pin in shaft with slots in rotor hub.
- Top of shaft aligned with top of rotor hub.
- Do not use a rotor showing any signs of damage.
Attach one end of the included ground wire to the stainless steel elbow on the Cold Trap drain tubing. When draining the Cold Trap, attach the other end (grounding clip) to the solvent catch pot to eliminate electrostatic spark ignition risk.
- Ground wire to stainless steel elbow on Cold Trap drain tubing.
- Grounding clip to solvent catch pot during draining.
- Prevents electrostatic spark ignition.
The manual provides a material compatibility table rating components (acrylic lid, EPDM gasket, stainless steel, Viton O-rings, polypropylene, anodized aluminum, PVC tubing, etc.) against acids, bases, and solvents with C (moderate degradation/questionable use) and D (severe degradation/infrequent use, immediate cleaning) ratings. It also gives maintenance guidance for rotary vane pumps, cold trap draining, and secondary trap saturation indicators.
- C = moderate degradation, questionable use; D = severe degradation, infrequent use recommended, immediate thorough cleaning required.
- Acrylic lid rated D for many solvents; glass lid option available.
- Stainless steel cold trap chamber can be attacked by some compounds; glass trap insert option available.
- Drain Cold Trap immediately after ice melts and flush with water.
- Do not chip ice off Cold Trap walls.
- Do not start a rotary vane pump when Cold Trap contains liquid.
- Rotary vane pump oil must be changed if cloudy, particulate, or discolored.
- Acid trap changes off-white to purple when spent; moisture trap blue to pink; solvent trap molecular sieve does not change color.
Proper balance of heat, vacuum, and centrifugal force is required; a trial run without sample is recommended for new protocols. Sample tubes should normally be filled no more than approximately half full and must fit the rotor without being loose.
- Trial run without sample recommended for unfamiliar protocols.
- Sample tubes filled no more than approximately half full.
- Tubes must be compatible with rotor and not loose.
- Rotors available for various tube sizes.
For smooth operation, samples must be loaded symmetrically to distribute weight evenly in the rotor, especially when running less than a full load.
- Load samples symmetrically.
- Distribute weight evenly in the rotor.
- Critical when running less than a full load.
Set temperature as high as possible without damaging or bumping the sample. Two timers are available: RUN TIME turns off the entire concentrator (rotation, heater, vacuum pump, and bleeds vacuum), while HEATER TIME turns off only the heater before the run ends. The CentriVap can alarm and turn off after a preset period.
- Temperature set as high as possible without sample damage or bumping.
- RUN TIME stops rotation, heater, vacuum pump, and bleeds vacuum.
- HEATER TIME turns off heater only; rotation and vacuum continue until RUN TIME expires.
- Audible alarm and automatic shutoff at end of set time.
Some models have a secondary Heat Boost heater on the upper sidewall of the chamber. It is controlled by the microprocessor to maintain the set point temperature at the bottom of the chamber and prevent excessive sample overheating.
- Secondary heater located on sidewall near top of chamber.
- Microprocessor maintains bottom-of-chamber set point.
- Prevents excessive overheating of samples.
The control panel includes an LCD display, program buttons, Run/Stop, Preheat, Increase/Decrease, Set Point Select, and an optional Heat Boost button. The display shows set point and actual parameters.
- LCD shows set point and actual measured conditions.
- Program buttons initiate stored runs.
- Run/Stop starts or stops a run.
- Preheat turns on heater before loading samples.
- Set Point Select cycles through parameters.
- Heat Boost button activates auxiliary heater if equipped.
Preheat is initiated with the PREHEAT button and set point adjusted with increase/decrease. Nine programs can be stored; programs 1–3 start with a single button press. Temperature set point ranges from OFF to 99°C, with HI = 100°C. Run Time and Heater Time range 1–999 minutes or ON for continuous operation. Timers remember stop time and resume; programs 1–3 reset timers on restart, other programs reset by holding start for five seconds.
- Nine programs storable; PROG 1/2/3 start with one button press.
- Temperature set point: OFF to 99°C; HI = 100°C.
- Run Time and Heater Time: 1 to 999 minutes, or ON for continuous.
- STOP remembers timer value; RUN resumes countdown.
- PROG 1/2/3 resets timers to original set point and starts new run.
- For other programs, hold start button five seconds to reset timers.
Operating sequence: turn on Cold Trap, turn on Concentrator, select program or set parameters, preheat if desired, enable Heat Boost if equipped, load vials (no more than half full) into rotor, close lid, and press RUN. The latch locks the lid, rotor starts, vacuum break valve closes, and vacuum pump starts after rotor reaches operating speed. Set points can be altered during a run.
- Cold Trap ON: amber indicator; green when at operating temperature.
- Lid safety switch prevents starting with lid open.
- Latch locks lid closed during run.
- Vacuum pump starts after rotor reaches operating speed.
- Set points can be changed during run via SELECT and INCREASE/DECREASE.
- Alarm sounds and all functions cease at end of set time.
- Allow rotor to stop before opening lid.
Cold Trap temperature may rise as solvent load increases and the green indicator may turn off; it reilluminates as load decreases. On power-up the CentriVap returns to the prior RUN or STOP mode and attempts to continue the run if it was running. Pressing STOP halts all functions; after rotor stops the lid may be opened, and RUN resumes at same set points with timer continuing from stop point.
- Cold Trap green indicator may turn off under heavy solvent load and return as load decreases.
- Power-up restores previous RUN/STOP mode.
- If power lost during RUN, unit attempts to resume programmed set points on power-up.
- STOP halts all functions; lid can be opened after rotor stops.
- RUN after STOP resumes same set points and timer countdown.
Maintenance is minimal. Disconnect power before servicing and decontaminate if hazardous materials were used. As-needed tasks include cleaning spills and surfaces, cleaning lid and gasket with mild non-abrasive soap, checking vacuum pump oil level (add if less than 1 inch/25.4 mm above MIN), changing cloudy/particulate/discolored oil, neutralizing acids immediately after runs, and emptying the Cold Trap before each run.
- Disconnect power before servicing.
- Decontaminate before servicing if hazardous materials used.
- Clean lid and gasket with soft cloth and mild non-abrasive soap.
- Pump oil level between MIN and MAX; add if less than 1 inch (25.4 mm) above MIN.
- Change pump oil if cloudy, particulate, or discolored.
- Neutralize and clean immediately after acid use.
- Empty Cold Trap completely before next run.
- If ice has melted, glass trap insert must be emptied before restarting Cold Trap.
The manual covers installation of a glass lid, secondary chemical trap, vacuum gauge, CentriZap strobe light, and glass trap in the Cold Trap. These accessories extend chemical resistance, monitoring, and sample visualization capabilities.
- Glass lid installation for chemical resistance.
- Secondary chemical trap installation in vacuum line.
- Vacuum gauge installation.
- CentriZap strobe light installation.
- Glass trap insert installation in Cold Trap.
The Troubleshooting chapter lists display error codes sensed by internal self-check systems and other potential problems with causes and corrective actions. Display codes include Heat Sensor, Close Lid, Latch Fail, Mem Fail P1, and MOTOR ERROR. Other problems include unit not operating, excessive vibration, sample odor, reduced evaporation rate, no/poor vacuum, low condensate recovery, Cold Trap not cooling, frequent oil changes, and unit starting and shutting off.
- Display error codes: Heat Sensor, Close Lid, Latch Fail, Mem Fail P1, MOTOR ERROR.
- Disconnect power before corrective action.
- Other problems table covers operation, vibration, odor, evaporation, vacuum, Cold Trap, oil, and start/stop issues.
- Contact Labconco if corrective actions do not solve the problem.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Concentrator 115V (60 Hz) models require a 15 A circuit breaker or fuse; 230V (50/60 Hz) models require 8 A. p.8 ↗ | Size dedicated branch circuits accordingly for each Concentrator. | All Concentrator models |
| Cold Trap 115V (-50°C & -85°C) models require 15 A at 115V (60 Hz); 115V (-105°C) models require 20 A at 115V (60 Hz); 230V models require 8 A at 230V (50/60 Hz). p.8 ↗ | Provide a dedicated outlet with the correct breaker rating for the selected Cold Trap temperature class. | Cold Trap models 7811020–7811040, 7460020–7460040, 7385020–7385040 |
| Aqueous samples: vacuum pump with 82 L/min free air flow and <2 mBar ultimate pressure is adequate. More volatile samples: diaphragm pump with 42 L/min free airflow and 200 mBar vacuum. p.9 ↗ | Select pump performance based on sample volatility; the Concentrator does not include a pump. | All Concentrator models |
| Vacuum pump inlet fitting must be suitable for 0.50 ID hose. 115V pumps need a 115V, 15 A NEMA 5-15P plug; 230V pumps need a reverse IEC 320 plug. p.9 ↗ | Pump must be compatible with the Concentrator's rear receptacle for automatic control. | All Concentrator models |
| Minimum 3 inches clearance required between Cold Trap sides and adjacent wall surfaces. p.14 ↗ | Maintain airflow through the Cold Trap cabinet (intake right, exhaust left) to avoid performance loss. | All Cold Trap models |
| Heater may be programmed up to 100°C; heater element may normally run at 110°C; thermal fuse limits maximum to 141°C and auto-resets below 120°C. p.18 ↗ | Solvents must be compatible with these temperatures; do not evaporate solvents with autoignition below 180°C. | All Concentrator models |
| Do not evaporate solvents classified as NEC Group A, B, or C; only non-flammable or Group D solvents with autoignition temperature ≥180°C. p.18 ↗ | Verify solvent classification before use; the CentriVap is not explosion-proof. | All Concentrator models |
| Temperature set point range: OFF to 99°C; display 'HI' = 100°C. Run Time and Heater Time range 1–999 minutes or 'ON' for continuous. p.23 ↗ | Program parameters within these limits; use ON for continuous operation without end-of-run alarm. | All Concentrator models |
| Nine programs can be stored; programs 1, 2, and 3 start with a single button press and reset timers to original set point. p.23 ↗ | Store frequently used protocols in programs 1–3 for one-button startup. | All Concentrator models |
| Sample tubes should normally be filled no more than approximately half full. p.20 ↗ | Prevents sample loss from bumping/foaming during centrifugation under vacuum. | All Concentrator models |
| Vacuum pump oil level should be between MIN and MAX; add oil if less than 1 inch (25.4 mm) above MIN. Change oil if cloudy, particulate, or discolored. p.26 ↗ | Regular oil inspection extends rotary vane pump life and prevents contamination. | Systems using rotary vane vacuum pumps |
| Do not store flammable or hazardous solvents within 12 inches (300 mm) of the CentriVap; no person or hazardous material within 12 inches during operation. p.9 ↗ | Maintain a 12-inch exclusion zone around the operating Concentrator. | All Concentrator models |
| Concentrator weighs over 42 lbs (19 kg); Cold Trap weighs over 70 lbs (33 kg). Use mechanical lift or at least two persons for manual lifting. p.10 ↗ | Plan installation with adequate lifting equipment or personnel. | All Concentrator and Cold Trap models |
| Acid trap changes from off-white to purple when spent; moisture trap changes from blue to pink when spent; solvent trap molecular sieve does not change color. p.18 ↗ | Monitor secondary trap color for acid and moisture traps; solvent trap requires extra care to determine saturation. | Systems with optional secondary traps |
| If ice has melted in the glass trap insert, it must be emptied before the Cold Trap is started again. p.26 ↗ | Prevents liquid carryover into the vacuum pump and oil contamination. | Cold Traps with glass trap insert |
| Emergency lid access without power: insert a small screwdriver into the small round hole on the left side of the case near the top behind the control panel to unlock the lid latch. p.14 ↗ | Allows sample recovery during power failure without defeating the latch during operation. | All Concentrator models |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Heat Sensor p.35 ↗ | Sensor failure | Replace sensor assembly. | Chapter 7 Troubleshooting |
| Heat Sensor p.35 ↗ | Connection failure | Repair connection. | Chapter 7 Troubleshooting |
| Close Lid p.35 ↗ | Lid open | Close lid. | Chapter 7 Troubleshooting |
| Latch Fail p.35 ↗ | Solenoid failure | Check component. | Chapter 7 Troubleshooting |
| Latch Fail p.35 ↗ | Switch or sensor failure | Check connections. | Chapter 7 Troubleshooting |
| Mem Fail P1 p.35 ↗ | Memory failure | Push program button #1. | Chapter 7 Troubleshooting |
| Mem Fail P1 p.35 ↗ | New memory IC chip | Push program button #1. | Chapter 7 Troubleshooting |
| Mem Fail P1 p.35 ↗ | Bad memory IC chip | Call Labconco – Replace IC chip or control PCB. | Chapter 7 Troubleshooting |
| MOTOR ERROR p.35 ↗ | Defective motor | Replace motor. | Chapter 7 Troubleshooting |
| MOTOR ERROR p.35 ↗ | Hall effect sensor failure | Replace sensor. | Chapter 7 Troubleshooting |
| MOTOR ERROR p.35 ↗ | Wire failure | Replace wire. | Chapter 7 Troubleshooting |
| Unit will not operate p.36 ↗ | Unit not connected to electrical power | Connect unit to proper electrical receptacle. | Chapter 7 Troubleshooting |
| Unit will not operate p.36 ↗ | Circuit breaker blown | Correct electrical problem and reset circuit breaker by pressing button. | Chapter 7 Troubleshooting |
| Unit will not operate p.36 ↗ | Lid open | Close lid. | Chapter 7 Troubleshooting |
| Excessive vibration p.36 ↗ | Sample tubes not located symmetrically in rotor | Reposition sample tubes. | Chapter 7 Troubleshooting |
| Sample odor in lab p.36 ↗ | Vent hose exhausting into lab area | Redirect hose to fume hood. | Chapter 7 Troubleshooting |
| Evaporation rate is reduced p.36 ↗ | Heater inoperable | Contact Labconco. | Chapter 7 Troubleshooting |
| Evaporation rate is reduced p.36 ↗ | Vacuum pump failure | Check pump. | Chapter 7 Troubleshooting |
| Evaporation rate is reduced p.36 ↗ | Obstruction in hose | Remove obstruction or replace hose. | Chapter 7 Troubleshooting |
| Evaporation rate is reduced p.36 ↗ | Lack of adequate vacuum | See below. | Chapter 7 Troubleshooting |
| No vacuum/poor vacuum p.36 ↗ | Pump not on | Turn on pump. | Chapter 7 Troubleshooting |
| No vacuum/poor vacuum p.36 ↗ | Control valve open | Check control valve. | Chapter 7 Troubleshooting |
| No vacuum/poor vacuum p.36 ↗ | Leaks in lines or connectors or gasket | Locate and repair. | Chapter 7 Troubleshooting |
| No vacuum/poor vacuum p.36 ↗ | Foreign material on lid gasket | Clean gasket and lid. | Chapter 7 Troubleshooting |
| No vacuum/poor vacuum p.36 ↗ | Pump is not functioning properly | Check pump by locating vacuum gauge closer to pump and close off rest of system. Check pump oil for cloudiness or particles and change. | Chapter 7 Troubleshooting |
| No vacuum/poor vacuum p.36 ↗ | Pump is not functioning properly | If pump is faulty, seek authorized service or replace pump. | Chapter 7 Troubleshooting |
| No vacuum/poor vacuum p.36 ↗ | Cold Trap or Concentrator gasket is not sealing properly | Check gasket for cleanliness. Adjust gasket as needed. | Chapter 7 Troubleshooting |
| No vacuum/poor vacuum p.36 ↗ | Cold Trap lid not seated | Hold lid down until vacuum is initiated. | Chapter 7 Troubleshooting |
| No vacuum/poor vacuum (cont.) p.37 ↗ | Ice formed on Cold Trap lid sealing surface | Defrost and wipe dry. | Chapter 7 Troubleshooting |
| No vacuum/poor vacuum (cont.) p.37 ↗ | Secondary Trap Cannister not fitted properly | Tighten all connections to and from the Secondary Trap Cannister. | Chapter 7 Troubleshooting |
| No vacuum/poor vacuum (cont.) p.37 ↗ | Secondary Trap Insert is spent | Replace with new insert. | Chapter 7 Troubleshooting |
| No vacuum/poor vacuum (cont.) p.37 ↗ | New Secondary Trap has moisture in it | Run vacuum pump for 24 hours to remove moisture. | Chapter 7 Troubleshooting |
| Recovery of condensate in Cold Trap is less than normally expected p.37 ↗ | Cold Trap is not ON | Check to make sure switch is ON and condensing unit fan is moving air out of the rear of the cabinet. | Chapter 7 Troubleshooting |
| Cold Trap does not cool down p.37 ↗ | Cold Trap is not ON | Turn vacuum pump OFF and allow Cold Trap to cool for at least 30 min. to reach temperature. | Chapter 7 Troubleshooting |
| Cold Trap does not cool down p.37 ↗ | Cold Trap is not ON | -50°C temperature can be checked with a solvent thermometer or digital thermometer. | Chapter 7 Troubleshooting |
| Frequent oil change needed in pump p.37 ↗ | Secondary Trap insert is spent | Change insert often. | Chapter 7 Troubleshooting |
| Frequent oil change needed in pump p.37 ↗ | Cold Trap is not emptied after each run and dried | Empty the traps (glass or stainless steel) after each run and replace. | Chapter 7 Troubleshooting |
| Frequent oil change needed in pump p.37 ↗ | Vacuum too strong for chemical | Use a Secondary Trap insert and diaphragm pump. | Chapter 7 Troubleshooting |
| Unit starts and shuts off p.37 ↗ | Latch optical sensor improperly calibrated | Unplug power cord. Wait 10 seconds. Plug in power cord. | Chapter 7 Troubleshooting |