Educational Fume Hoods & Enclosures Catalog

Technical Reference Summary

Covered Models & Document Scope

Protector ClassMate Laboratory Hoods: vertical-rising sash models and combination horizontal-sliding/vertical-rising sash models in 4', 5' and 6' nominal widths; electrical variants 100-115V 50/60 Hz and 208-230V 50/60 Hz; ordering code 1 6 0 0 1 with width, electrical/service-fixture digits.

Technical Data

Parameter Value Condition
Nominal width 4' ft p.6 ↗ 4' Hood
Nominal width 5' ft p.6 ↗ 5' Hood
Nominal width 6' ft p.6 ↗ 6' Hood
Overall width (A) 48.0" (121.9 cm) in (cm) p.6 ↗ 4' Hood
Overall width (A) 60.0" (152.4 cm) in (cm) p.6 ↗ 5' Hood
Overall width (A) 72.0" (182.9 cm) in (cm) p.6 ↗ 6' Hood
Dimension B 43.5" (110.5 cm) in (cm) p.6 ↗ 4' Hood
Dimension B 55.5" (141.0 cm) in (cm) p.6 ↗ 5' Hood
Dimension B 67.5" (171.5 cm) in (cm) p.6 ↗ 6' Hood
Dimension C 20.5" (52.2 cm) in (cm) p.6 ↗ 4' Hood
Dimension C 26.5" (67.3 cm) in (cm) p.6 ↗ 5' Hood
Dimension C 32.5" (82.6 cm) in (cm) p.6 ↗ 6' Hood
Dimension D 24.0" (61.0 cm) in (cm) p.6 ↗ 4' Hood
Dimension D 30.0" (76.2 cm) in (cm) p.6 ↗ 5' Hood
Dimension D 36.0" (91.4 cm) in (cm) p.6 ↗ 6' Hood
Sash stop height 18.0" (45.7 cm) in (cm) p.6 ↗
Work surface depth 1.25" in p.3 ↗ Spill trough depth
Electrical requirements 100-115V, 50/60 Hz, 10A V, Hz, A p.8 ↗ code 00
Electrical requirements 208-230V, 50/60 Hz, 5A V, Hz, A p.8 ↗ code 20
Airflow 785 CFM p.8 ↗ 4' Hood, vertical-rising sash, 100 fpm, s.p. 0.44"
Airflow 995 CFM p.8 ↗ 5' Hood, vertical-rising sash, 100 fpm, s.p. 0.54"
Airflow 1225 CFM p.8 ↗ 6' Hood, vertical-rising sash, 100 fpm, s.p. 0.72"
Airflow 405 CFM p.8 ↗ 4' Hood, combination sash, 100 fpm, s.p. 0.12"
Airflow 520 CFM p.8 ↗ 5' Hood, combination sash, 100 fpm, s.p. 0.15"
Airflow 635 CFM p.8 ↗ 6' Hood, combination sash, 100 fpm, s.p. 0.19"

Operating Limits

  • Electrical supply, 100-115V variant: 100-115V, 50/60 Hz, 10A p.8 ↗
  • Electrical supply, 208-230V variant: 208-230V, 50/60 Hz, 5A p.8 ↗
  • Sash stop height: 18.0" (45.7 cm) p.6 ↗

Type Code Decoder

Position Token Meaning
field WIDTH p.8 ↗ 4 = 4' (122 cm); 6 = 6' (183 cm) Nominal width of fume hood
field ELECTRICAL REQUIREMENTS p.8 ↗ 00 = 100-115V, 50/60 Hz, 10A; 20 = 208-230V, 50/60 Hz, 5A Electrical requirements
field SERVICE FIXTURES & GFCI p.8 ↗ — = No Electrical; 02 = Two Service Fixtures & GFCI Service fixtures and GFCI option