CentriVap Pro Centrifugal Concentrators - User's Manual
Technical Reference Summary
Covered Models & Document Scope
Covers Labconco CentriVap Pro centrifugal vacuum concentrators: standard benchtop (85101xx, 85111xx, 85121xx), benchtop with Heat Boost (85102xx, 85112xx, 85122xx), DNA (85301xx, 85311xx), and refrigerated (85201xx, 85211xx, 85221xx). 115 V 60 Hz and 230 V 50/60 Hz variants; 230 V 50 Hz variants for refrigerated models. Optional Heat Boost, CentriZap strobe light, Out of Balance Detect, and vacuum pump oil reminder.
Section-by-Section Technical Breakdown
Defines the dedicated outlet, earthing, and current draw for each model family. Standard and Heat Boost models draw 12 A at 115 V (7.4 A unit + 4.6 A max pump) and 8 A at 230 V (4.6 A + 2.3 A max pump). DNA and refrigerated models follow the same current limits.
- Dedicated outlet with appropriate circuit breaker close to the product; must provide protective earthing ground.
- 115 V models: 12 A total (7.4 A + 4.6 A max pump).
- 230 V models: 8 A total (4.6 A + 2.3 A max pump).
- Refrigerated 230 V 50 Hz variants: 8 A (4.6 A + 2.3 A max pump).
Specifies fume-hood exhaust and user-supplied vacuum pump performance and electrical compatibility. Aqueous samples need 82 L/min free air flow and <2 mBar ultimate pressure; volatile samples can use a diaphragm pump at 42 L/min and 200 mBar. Pump inlet must accept 0.50 ID hose and current draw must not exceed 8 A (115 V) or 4.5 A (230 V).
- Locate within a fume hood if hazardous or flammable solvents are used; vent pump exhaust to fume hood.
- Aqueous samples: pump ≥82 L/min free air flow, <2 mBar ultimate pressure.
- Volatile samples: diaphragm pump 42 L/min, 200 mBar.
- Pump inlet fitting for 0.50 ID hose.
- 115 V pump: NEMA 5-15P plug, ≤8 A; 230 V pump: reverse IEC 320 plug, ≤4.5 A.
- Wetted parts should be PTFE or PTFE-coated for aggressive samples.
Requires a cold trap or equivalent when using a non-corrosion-resistant mechanical pump, and recommends an acid-neutralizing secondary trap when evaporating acids. Notes that stainless steel and aluminum components can be degraded by acids; contact Labconco before evaporating acids.
- Cold Trap or equivalent mandatory with non-corrosion-resistant mechanical pump.
- Acid neutralizing secondary trap advisable for acids.
- Stainless steel and aluminum components can be degraded by acids.
- Moisture, solvent, and radioisotope traps available as accessories.
Covers unpacking, component verification, emergency lid access, hose connections, electrical connection, rotor installation, ground wire, and levelling. The refrigeration system draws air in on the right and exhausts on the left, requiring at least 3 inches clearance to adjacent walls. The middle levelling foot must contact the surface after levelling.
- Minimum 3 inches between sides of Cold Trap or Refrigerated CentriVap and adjacent walls.
- Emergency lid access: insert small screwdriver into round hole on right side near top behind control panel; remove vacuum hose first if chamber under vacuum.
- Rotor must be flat (not conical) on bottom and must not touch chamber when placed on shaft.
- Ground wire from Cold Trap stainless-steel elbow to solvent catch pot grounding clip when draining.
- Level front-to-back, side-to-side, corner-to-corner; ensure middle levelling foot contacts surface.
Provides material compatibility ratings (A/B/C/D) for chamber, bearings, lid gasket, rotor shaft, valve, O-rings, fittings, rotor, rotor hub, and tubing against acids, bases, and solvents. Ratings C and D indicate moderate and severe degradation respectively. Includes maintenance guidance for rotary vane and diaphragm pumps and color-change indicators for secondary traps.
- C = moderate degradation/questionable use; D = severe degradation, infrequent use recommended, immediate thorough cleaning required.
- Standard chamber: epoxy-coated aluminum; Acid Resistant chamber: PTFE-coated aluminum.
- Standard rotor shaft: stainless steel; Acid Resistant rotor shaft: Hastelloy.
- Rotary vane pump oil must be changed if cloudy, shows particles, or is discolored.
- Acid trap changes off-white to purple when used up; moisture trap changes blue to pink; solvent trap molecular sieve does not change color.
Describes the 300 W heater and optional 300 W Heat Boost (benchtop only), timed vacuum pump outlet that delays pump start until rotor reaches speed, lid safety switch and latch, and touch-screen control of vacuum, temperature, speed, and time. The Cold Trap protects the pump; glass trap insert recommended for corrosive applications.
- 300 W heater; optional 300 W Heat Boost on Benchtop Concentrators only.
- Timed outlet delays vacuum pump start until rotor reaches operating speed.
- Safety switch prevents start with lid open; latch prevents opening while rotor turns.
- 230 V 50 Hz models comply with CE regulations.
- Glass Trap insert recommended for corrosive applications.
Describes the VaporWorks touch-screen interface, initial power-up timing, home screen, pre-conditioning, CentriZap, method creation, step parameters, End Point Assist, Smart Dry, data logging, settings, calibration, network/Wi-Fi, security, diagnostics, and software updates. Temperature can be set 30°C to 100°C (-10°C for refrigerated), rotor speed 500–1800 rpm, vacuum 0–300 mBar.
- Initial power-up: display off 10–15 s, animated start-up, ~90 s to fully load.
- Temperature range 30°C to 100°C; -10°C for Refrigerated model; can be set off.
- Rotor speed 500–1800 rpm or off; off only allowed when setting up a method (desiccator use).
- Vacuum 0–300 mBar; can be set to minimum or off.
- End Point Assist requires cold trap, ≥36 mL total liquid, ≥45°C, and ≤45 mBar; only one step per method.
- Smart Dry pulses vacuum valve to flush filtered air after last step; cannot be combined with EPA.
- Data log default rate 1 minute; up to nine parameters; file naming LOG_YYYY-MM-DD_XX.
- Calibration tolerances: temperature ±3°C, vacuum ±10 mbar.
Covers planning, glassware selection, loading, setting guidelines, Heat Boost operation, and operational notes. Emphasizes balancing rotors, using lowest speed that prevents boiling, and not filling tubes, plates, or flasks beyond stated volume.
- Do not fill tubes, plates, or flasks more than the stated volume.
- Adjust rotor speed to rotor weight to minimize glassware breakage.
- Run at lowest speed that prevents samples from boiling.
- Heat Boost available on Benchtop and Acid Resistant models; can be enabled/disabled in Configuration.
Provides maintenance safety precautions and a recommended maintenance schedule. Maintenance reminders default to 2000 hours for inlet filter and condenser cleaning and 500 hours for vacuum pump oil. Lifetime hours for motor, compressor, and vacuum pump are viewable in Settings > Maintenance.
- Inlet filter change reminder default: 2000 hours.
- Condenser cleaning reminder default: 2000 hours (Refrigerated only).
- Vacuum pump oil reminder default: 500 hours (Standard or Refrigerated only).
- Lifetime hours for motor, compressor, and vacuum pump viewable in Settings > Maintenance.
Lists rotors, traps and vacuum pumps, secondary chemical trap installation, vacuum gauge installation, and CentriZap strobe light installation. Accessories include moisture, solvent, and radioisotope traps; glass trap insert for corrosive applications.
- Rotors available; previous-generation rotors usable if flat-bottomed and not touching chamber.
- Secondary chemical trap installation described.
- Vacuum gauge installation described.
- CentriZap strobe light installation described.
- Moisture, solvent, and radioisotope traps available.
Lists common symptoms and alert codes with causes and remedies: excessive vibration, sample odor, reduced evaporation rate, EE memory error, no/poor vacuum, cold trap condensate recovery, frequent pump oil changes, TA high temperature, TS temperature sensor, VS vacuum sensor, LF latch fail, LO lid open, ME motor error, and TC/TH temperature not reached.
- Alert codes: EE, TA, TS, VS, LF, LO, ME, TC, TH.
- Symptoms include excessive vibration, sample odor in lab, reduced evaporation rate, no or poor vacuum.
- Cold trap condensate recovery less than normal and frequent vacuum pump oil changes are covered.
- Each alert has cause and remedy in the troubleshooting section.
Appendix C provides power data, sound pressure data, and environmental conditions for the CentriVap Pro. Power data matches the electrical requirements table; sound pressure and environmental conditions are listed.
- Power data per model family as in Section 2.
- Sound pressure data provided.
- Environmental conditions provided.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| 115 V models draw 12 A total (7.4 A unit + 4.6 A max pump); 230 V models draw 8 A total (4.6 A + 2.3 A max pump). p.12 ↗ | Size dedicated circuit breaker and outlet for at least these currents; use protective earthing. | All 115 V and 230 V models |
| Vacuum pump current draw must not exceed 8 A for 115 VAC models and 4.5 A for 230 VAC models. p.13 ↗ | Select a pump compatible with the concentrator's internal receptacle and current limit. | All models |
| Aqueous samples require a pump with 82 L/min free air flow and <2 mBar ultimate pressure; volatile samples can use a diaphragm pump at 42 L/min and 200 mBar. p.13 ↗ | Match pump performance to sample volatility; lower ultimate pressure needed for aqueous. | All models except DNA (DNA includes pump) |
| Pump inlet fitting must be suitable for 0.50 ID hose. p.13 ↗ | Verify hose and fittings compatibility before installation. | All models |
| Minimum 3 inches clearance between sides of Cold Trap or Refrigerated CentriVap and adjacent walls. p.23 ↗ | Prevent airflow restriction that could degrade refrigeration performance. | Cold Trap and Refrigerated CentriVap Pro |
| Heater may be programmed to 100°C, element may run at 110°C, thermal fuse limits to 141°C and resets below 120°C. p.27 ↗ | Do not evaporate solvents with autoignition temperature below 180°C; only NEC Group D or non-flammable solvents. | All models |
| Temperature set range 30°C to 100°C (-10°C for Refrigerated model) or off. p.33 ↗ | Refrigerated model can cool below ambient; benchtop and DNA rely on passive cooling and cannot go below ~30°C ambient. | All models; -10°C only Refrigerated |
| Rotor speed adjustable 500 to 1800 rpm or off; off only allowed when setting up a method. p.33 ↗ | Use lowest speed that prevents boiling to minimize glassware breakage; off enables desiccator use. | All models |
| Vacuum level adjustable 0 to 300 mBar; can be set lower than pump can achieve. p.33 ↗ | Verify pump capability; higher vacuum may not be achievable depending on pump capacity. | All models |
| End Point Assist requires cold trap connected, total liquid volume ≥36 mL, temperature ≥45°C, and vacuum ≤45 mBar; only one step per method. p.36 ↗ | EPA may not function outside these conditions; Labconco tested only these ranges. | Benchtop, Acid Resistant, and Refrigerated models |
| Calibration tolerances: temperature ±3°C, vacuum ±10 mbar. p.44 ↗ | Use these tolerances when verifying sensor accuracy. | All models |
| Maintenance reminders default: 2000 hours for inlet filter and condenser cleaning; 500 hours for vacuum pump oil. p.42 ↗ | Plan maintenance intervals accordingly; can be modified in Settings > Maintenance. | All models; condenser cleaning Refrigerated only; pump oil Standard or Refrigerated only |
| Data log default rate 1 minute; selectable intervals 10 s, 30 s, 60 s, 5 min, 10 min. p.42 ↗ | Adjust logging rate for resolution vs storage; up to nine parameters collected. | All models |
| Initial power-up: display off 10–15 s, animated start-up, ~90 s to fully load. p.29 ↗ | Allow sufficient time before expecting full operation. | All models |
| Do not store flammable or hazardous solvents within 12 inches (300 mm) of the CentriVap Pro. p.27 ↗ | Maintain clearance for fire safety. | All models |
| Acid trap changes off-white to purple when used up; moisture trap changes blue to pink; solvent trap molecular sieve does not change color. p.26 ↗ | Monitor trap color; replace solvent trap on schedule since no color change occurs. | Models with secondary traps |
| Rotary vane pump oil must be changed if cloudy, shows particles, or is discolored; run pump with inlet blocked after run to purge contaminants. p.26 ↗ | Extends pump life; do not run at elevated vacuum level with inlet blocked or oil will be expelled. | Models using rotary vane pumps |
| Rotor must be flat (not conical) on bottom and must not touch chamber when placed on shaft. p.24 ↗ | Verify rotor compatibility; damaged rotors must not be used. | All models |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Excessive Vibration p.69 ↗ | Not specified in available text. | Not specified in available text. | Section 10 Troubleshooting |
| Sample Odor in Lab p.69 ↗ | Not specified in available text. | Not specified in available text. | Section 10 Troubleshooting |
| Evaporation Rate is Reduced p.70 ↗ | Not specified in available text. | Not specified in available text. | Section 10 Troubleshooting |
| EE -Memory Error Alert p.70 ↗ | Not specified in available text. | Not specified in available text. | Section 10 Troubleshooting |
| No Vacuum or Poor Vacuum p.71 ↗ | Not specified in available text. | Not specified in available text. | Section 10 Troubleshooting |
| Cold Trap Condensate Recovery Less than Normal p.73 ↗ | Not specified in available text. | Not specified in available text. | Section 10 Troubleshooting |
| Frequent Vacuum Pump Oil Changes Needed p.73 ↗ | Not specified in available text. | Not specified in available text. | Section 10 Troubleshooting |
| TA - High Temperature Alert p.74 ↗ | Not specified in available text. | Not specified in available text. | Section 10 Troubleshooting |
| TS -Temperature Sensor Alert p.74 ↗ | Not specified in available text. | Not specified in available text. | Section 10 Troubleshooting |
| VS -Vacuum Sensor Alert p.74 ↗ | Not specified in available text. | Not specified in available text. | Section 10 Troubleshooting |
| LF -LatchFail Alert p.75 ↗ | Not specified in available text. | Not specified in available text. | Section 10 Troubleshooting |
| LO -Lid Open Alert p.75 ↗ | Not specified in available text. | Not specified in available text. | Section 10 Troubleshooting |
| ME -Motor Error Alert p.75 ↗ | Not specified in available text. | Not specified in available text. | Section 10 Troubleshooting |
| TC and TH -Temperature Not Reached Alerts p.75 ↗ | Not specified in available text. | Not specified in available text. | Section 10 Troubleshooting |