Labconco puricare brochure 4 57
Technical Reference Summary
Covered Models & Document Scope
PuriCare Open Access Stations (4' and 5' nominal widths; catalog numbers 3820001, 3820011, 3820023, 3820024, 3820000, 3820010, 3820033, 3820034) and PuriCare Procedure Stations (3', 4', 5', 6' nominal widths; ordering code 3/4/5/6 + electrical configuration). Accessories and base stands are also covered.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Nominal width | 4' p.8 ↗ | Open Access Station |
| Nominal width | 5' p.8 ↗ | Open Access Station |
| Exterior width A | 48.4" (123.0 cm) p.8 ↗ | Open Access Station, 4' |
| Exterior width A | 58.4" (148.3 cm) p.8 ↗ | Open Access Station, 5' |
| Interior width B | 44.4" (123.0 cm) p.8 ↗ | Open Access Station, 4' |
| Interior width B | 48.5" (123.2 cm) p.8 ↗ | Open Access Station, 5' |
| Electrical requirements | 115 volts, 60 Hz, 12 amps p.8 ↗ | Open Access Station, 3820001/3820000 |
| Electrical requirements | 100 volts, 50/60 Hz, 12 amps p.8 ↗ | Open Access Station, 3820011/3820010 |
| Electrical requirements | 230 volts, 50/60 Hz, 6 amps p.8 ↗ | Open Access Station, 3820024/3820034 |
| Plug type | 115 volts, 15 amps p.8 ↗ | Open Access Station, 3820001/3820000 |
| Plug type | 100 volts, 15 amps p.8 ↗ | Open Access Station, 3820011/3820010 |
| Plug type | Schuko p.8 ↗ | Open Access Station, 3820023/3820033 |
| Plug type | British (UK) p.8 ↗ | Open Access Station, 3820024/3820034 |
| Nominal width | 3' p.11 ↗ | Procedure Station |
| Nominal width | 4' p.11 ↗ | Procedure Station |
| Nominal width | 5' p.11 ↗ | Procedure Station |
| Nominal width | 6' p.11 ↗ | Procedure Station |
| Energy requirements | 12A p.11 ↗ | Procedure Station, 3', 100V or 115V |
| Energy requirements | 6A p.11 ↗ | Procedure Station, 3', 230V |
| Energy requirements | 16A p.11 ↗ | Procedure Station, 4'/5'/6', 100V or 115V |
| Energy requirements | 8A p.11 ↗ | Procedure Station, 4'/5'/6', 230V |
| Watts | 195 W p.11 ↗ | Procedure Station, 3' |
| Watts | 240 W p.11 ↗ | Procedure Station, 4' |
| Watts | 340 W p.11 ↗ | Procedure Station, 5' |
| Watts | 490 W p.11 ↗ | Procedure Station, 6' |
Operating Limits
- HEPA filter efficiency: 99.99+% at 0.3 microns p.10 ↗
- Sash opening height: 12" p.10 ↗
- Sash opening height (Open Access): 14" (36 cm) p.6 ↗
- Work surface liquid capacity: 1 liter p.6 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| field 1 p.11 ↗ | 3 / 4 / 5 / 6 | Nominal width of cabinet: 3 = 3', 4 = 4', 5 = 5', 6 = 6' |
| field 2 p.11 ↗ | 115V, 15A | Electrical configuration and plug: 115V, 15A for North America |
| field 2 p.11 ↗ | 115V, 20A | Electrical configuration and plug: 115V, 20A for North America |
| field 2 p.11 ↗ | 230V | Electrical configuration and plug: 230V for North America |
| field 2 p.11 ↗ | 230V Schuko | Electrical configuration and plug: 230V Schuko |
| field 2 p.11 ↗ | 230V China/Australia | Electrical configuration and plug: 230V for China/Australia |
| field 2 p.11 ↗ | 230V Great Britain (UK) | Electrical configuration and plug: 230V for Great Britain (UK) |
| field 2 p.11 ↗ | 230V India | Electrical configuration and plug: 230V for India |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| 3820001 p.8 ↗ | 3820001 | Open Access Station, 4', 115 volts, 60 Hz, 12 amps | ||
| 3820011 p.8 ↗ | 3820011 | Open Access Station, 4', 100 volts, 50/60 Hz, 12 amps | ||
| 3820023 p.8 ↗ | 3820023 | Open Access Station, 4', Schuko | ||
| 3820024 p.8 ↗ | 3820024 | Open Access Station, 4', 230 volts, 50/60 Hz, 6 amps, British (UK) | ||
| 3820000 p.8 ↗ | 3820000 | Open Access Station, 5', 115 volts, 60 Hz, 12 amps | ||
| 3820010 p.8 ↗ | 3820010 | Open Access Station, 5', 100 volts, 50/60 Hz, 12 amps | ||
| 3820033 p.8 ↗ | 3820033 | Open Access Station, 5', Schuko | ||
| 3820034 p.8 ↗ | 3820034 | Open Access Station, 5', 230 volts, 50/60 Hz, 6 amps, British (UK) |