Rev f precise ca basic glove box

Technical Reference Summary

Covered Models & Document Scope

Covers Labconco Precise® Controlled Atmosphere Glove Box and Precise® Basic Glove Box models. Ordering codes: Precise CA models 5220100 (115V/60Hz), 5220120 (230V/50Hz), 5220121 (230V/60Hz), 5220130 (100V/50Hz), 5220131 (100V/60Hz); Precise Basic models 5220000 (115V/60Hz), 5220020 (230V/50Hz), 5220021 (230V/60Hz), 5220030 (100V/50Hz), 5220031 (100V/60Hz). Precise CA includes six valves, pressure relief valve and two chamber pressure gauges; Precise Basic omits these for user customization.

Section-by-Section Technical Breakdown

pp. 6-9
Prerequisites: Support, Location, Exhaust, Electrical, Plumbing and Space Requirements

Defines site preparation: rigid bench/stand (floor or wall mounted, not both), typical standing height 35–40 in (889–1016 mm), work surface depth 29–30 in (737–762 mm). The pressure relief valve activates above +6 in or below -6 in water gauge; exhaust options include unfiltered 4 in duct, HEPA and/or carbon filtered vent kits with 4 in thimble connection. Electrical ratings and plumbing valve assignments are specified.

  • Bench/stand rigidly mounted to floor or wall, not both; typical standing height 35–40 in (889–1016 mm); Labconco stands 33–40 in (838–1016 mm).
  • Work surface depth 29–30 in (737–762 mm); height 35–40 in for standing operation.
  • Pressure relief valve activates above +6 in WG or below -6 in WG.
  • Exhaust options: 4 in nominal duct to outside; HEPA and/or carbon filtered kits with 4 in thimble duct connection; remote blowers have 6 in inlet requiring 4x6 in transition.
  • Electrical: entire glove box 115V/8A, 230V/4A, or 100V/8A; interior duplex 115V/5A, 230V/2A, or 100V/5A.
  • Gas inlet lines set at 10–30 psi via gas supply pressure regulator.
  • Main chamber has four valves (two for gas inlet/vacuum outlet, two for recirculating drying train); transfer chamber has two valves.
  • All valves use 3/8 in (9.5 mm) compression fittings or 3/8 in hose barbs.
pp. 10-12
Unpacking, Lifting and Installing the Glove Box

Each Precise Glove Box weighs 185 lbs (83 kg). Move with a flat low dolly under the shipping skid or a floor jack; do not lift by the upper sheet metal façade or front panel. The work surface must support the combined weight and be as wide as the glove box; front of glove box aligned with front of work surface.

  • Weight: 185 lbs (83 kg) per model.
  • Do not lift by upper sheet metal façade and front panel; damage to liner can occur.
  • Work surface must be as wide as the entire glove box and support combined weight.
  • Height 35–40 in (889–1016 mm) appropriate for standing operation.
pp. 12-16
Connecting the Exhaust to the Pressure Relief Valve

The pressure relief valve exhausts back into the room unless an optional vent kit is installed. Vent kits: Ventilation Kit #5242700 (unfiltered to outside), HEPA Filtered Pressure Relief Vent Kit #5238500, Carbon Filtered Pressure Relief Vent Kit #5241200; HEPA and carbon kits may be combined. For Precise Basic, adding pressure relief valve #5230600 requires drilling a 2.06 in (52.3 mm) hole in the raised boss (4.00 in dia x 1.00 in high) and exposing 2.00–18 UNS threads.

  • Ventilation Kit #5242700 for unfiltered exhaust to outside.
  • HEPA Filtered Pressure Relief Vent Kit #5238500.
  • Carbon Filtered Pressure Relief Vent Kit #5241200.
  • HEPA and carbon filtered kits may be used together.
  • Pressure relief valve #5230600 for Precise Basic; drill 2.06 in (52.3 mm) hole in 4.00 in (101.6 mm) dia x 1.00 in (25.4 mm) high raised boss.
  • Exhaust ductwork weight must be supported independently of glove box superstructure.
pp. 17
Installation of Gloves to the Glove Ports

Gloves are installed with thumbs up and right/left orientation by stretching the beaded cuff into the glove port groove nearest the window. A separate 8 in diameter O-ring is installed over the gloves into the outer groove and secured with stainless steel band clamps.

  • Glove ports are 8 in nominal size with double grooves.
  • Install beaded glove cuff into groove nearest window.
  • Install separate 8 in diameter O-ring over gloves into outer groove.
  • Secure O-ring with stainless steel band clamps.
pp. 17
Connecting the Electrical Supply Source

115V/60Hz, 100V/50Hz and 100V/60Hz models connect the supplied power cord to the IEC inlet on the back; these models are rated 8 A total with interior duplex maximum 5 A. 230V/50Hz and 230V/60Hz models ship without a plug; install appropriate plug per local codes; rated 4 A total with interior duplex 2 A.

  • 115V/100V models: 8 A total; interior duplex max 5 A.
  • 230V models: 4 A total; interior duplex 2 A.
  • 230V models shipped without attached plug; install per local codes.
  • Power cord connects to IEC inlet on back of glove box.
pp. 18-24
Connecting Plumbing Lines for Gas, Vacuum and Drying Train

Primary connection method uses 3/8 in O.D. x 3/16 in I.D. tubing with a tube support insert inside soft tubing; alternative uses tube connector and hose clamp. Accessory 0.2 micron filters can be installed in vacuum and inert gas lines. Tubing kits: 5245100 (manual valve), 5245200 (automatic pressure controller), 5242501 (oxygen & moisture removal), 5242500 (dual moisture/solvent removal).

  • Primary method: 3/8 in O.D. x 3/16 in I.D. tubing with tube support insert.
  • Alternative method: tube connector and hose clamp (hose clamp not supplied).
  • 0.2 micron filter for vacuum line and inert gas supply line.
  • Tubing Connection Kit 5245100 for manual valve operation.
  • Tubing Connection Kit 5245200 with Automatic Pressure Controller.
  • Tubing Connection Kit 5242501 for oxygen & moisture removal.
  • Tubing Connection Kit 5242500 for dual moisture/solvent removal.
pp. 24-27
Validating the Glove Box: Leak Testing per ISO 10648-2

Each glove box is helium leak tested at factory to no leaks greater than 1 x 10^-4 cc/sec at +5 in helium. The Precise CA meets ISO 10648-2 Class 1 (hourly leak rate ≤ 5 x 10^-4 hr^-1). Oxygen method and pressure change method formulas are given. Acceptance test at -1000 Pa (-4 in WG); operational test at -250 Pa (-1 in WG).

  • Factory helium leak test: no leaks > 1 x 10^-4 cc/sec at +5 in helium.
  • ISO 10648-2 Class 1: hourly leak rate ≤ 5 x 10^-4 hr^-1.
  • Oxygen method formula: Tf = 300*(O2f - O2i)/(t*10^6).
  • Pressure change method formula: Tf = (60/t)*((PnT1/P1Tn)-1).
  • Acceptance test negative pressure: -1000 Pa (-4 in WG); operational test: -250 Pa (-1 in WG).
  • 0.1°C fall = -34 Pa (-0.14 in WG) equivalent pressure change; 0.01 in Hg rise = -34 Pa (-0.14 in WG).
pp. 28-30
Performance Features

Describes construction and features: one-piece molded polyethylene shell, six manual valves (four main, two transfer), transfer chamber vacuum gauge 0–29 in Hg, main chamber pressure gauge -5 to +5 in WG, 8 in glove ports, 0.015 in thick 30 in long neoprene gloves size 9¾, 1/4 in laminated safety glass window with 26.4 x 12.5 in opening, fluorescent lamp 30–40 foot candles, phenolic work surface, interior switched duplex, pressure relief valve set at ±6 in WG with operational maximum ±5 in WG.

  • Main chamber pressure gauge range: -5 to +5 in water gauge.
  • Transfer chamber vacuum gauge range: 0 to 29 in Hg vacuum.
  • Pressure relief valve set to relieve above +6 or below -6 in WG; operational maximum ±5 in WG.
  • Gloves: neoprene, 0.015 in thick, 30 in long, size 9¾, one pair.
  • Window: 1/4 in laminated safety glass; opening 26.4 in wide x 12.5 in high (670 x 317 mm).
  • Fluorescent lamp: 30–40 foot candles on work surface.
  • Interior duplex: 115V/5A, 230V/2A, or 100V/5A; innermost outlet labeled AUX is switched.
pp. 33-35
Using Your Glove Box: Routine Daily Work Procedures

Covers planning, start-up, loading, purge/fill cycles, work techniques, transferring materials out, and shutdown. Transfer chamber typically requires 1–5 purge/fill cycles to reduce oxygen and moisture below 1%; main chamber 75–100 cycles to below 1% and additional 30–50 cycles to below 0.3% (3,000 ppm). Recirculating drying trains can maintain O2 <5 ppm and moisture <50 ppm.

  • Transfer chamber: 1–5 purge/fill cycles to reduce O2 and moisture below 1%.
  • Main chamber: 75–100 purge/fill cycles to below 1%; additional 30–50 cycles to below 0.3% (3,000 ppm).
  • Recirculating drying trains maintain O2 <5 ppm and moisture <50 ppm.
  • Maintain slight positive pressure to preserve atmosphere; slightly negative pressure when working with hazardous compounds.
  • After work, purge and fill main chamber approximately 5–10 times; transfer chamber 1–5 times.
  • Never apply high vacuum purge on sensitive electronic moisture and oxygen monitors.
pp. 36
Maintaining Your Glove Box

Routine maintenance schedule includes weekly cleaning and pressure decay charting, initial certification before first use, annual recertification, fluorescent light replacement, window replacement, pressure relief valve replacement, and service valve/pressure gauge replacement.

  • Weekly: wipe interior with disinfectant/cleaner; clean exterior; chart pressure decay.
  • Initial certification by qualified technician before first use.
  • Recertification at least annually after installation.
  • Fluorescent lamp replaceable from outside the glove box.
  • Window replacement and pressure relief valve replacement procedures included.
  • Service valve and pressure gauge replacement procedures included.
pp. 47
Electrical System Troubleshooting

Table of problems, causes and corrective actions for lights not operating and no power to duplex electrical. The glove box has only two electrical circuits.

  • Lights won't operate: check plug, circuit breaker, lamp installation, lamp wiring, lamp, light switch, electronic ballast.
  • No power to duplex electrical: check plug, circuit breaker, duplex switch.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Pressure relief valve activates at pressures above +6 in water gauge or vacuums below -6 in water gauge. p.7 ↗ Exhaust or vent the relief valve to prevent room contamination; operational maximum is ±5 in WG. Precise CA (standard); Precise Basic with optional valve #5230600
Gas inlet lines should be set at 10–30 psi using a gas supply pressure regulator. p.8 ↗ Prevents overpressurization of the main chamber. Whole series
All valves utilize 3/8 in (9.5 mm) compression fittings or 3/8 in (9.5 mm) hose barbs. p.9 ↗ Plumbing connections must use compatible 3/8 in fittings or soft tubing with hose barbs. Whole series
Entire glove box electrical rating: 115V/8A, 230V/4A, or 100V/8A. Interior duplex: 115V/5A, 230V/2A, or 100V/5A. p.8 ↗ Size branch circuit and interior loads accordingly; do not exceed duplex rating. Whole series by voltage variant
Factory helium leak test: no leaks greater than 1 x 10^-4 cc/sec at positive pressure of 5 inches helium. p.5 ↗ Baseline integrity; field validation should confirm Class 1 leak rate ≤ 5 x 10^-4 hr^-1. Whole series
ISO 10648-2 Class 1 hourly leak rate ≤ 5 x 10^-4 hr^-1. p.25 ↗ Acceptance test at -1000 Pa (-4 in WG); operational test at -250 Pa (-1 in WG). Precise CA
Main chamber pressure gauge range: -5 to +5 inches water gauge. Transfer chamber vacuum gauge: 0 to 29 inches mercury vacuum. p.29 ↗ Operating pressure should remain within ±5 in WG; full vacuum requires steel reinforced high vacuum sleeve. Precise CA (gauges optional on Basic)
Gloves: neoprene, 0.015 in thick, 30 in long, size 9¾, one pair supplied. p.29 ↗ Replacement gloves must match port size and O-ring; 8 in nominal glove ports. Whole series
Window opening when frame removed: 26.4 in wide x 12.5 in high (670 mm x 317 mm). p.29 ↗ Large apparatus must fit within this opening for loading. Whole series
Transfer chamber typically requires 1–5 purge/fill cycles to reduce oxygen and moisture below 1%. p.34 ↗ Porous materials may require longer sustained vacuum purges. Whole series
Main chamber: 75–100 purge/fill cycles reduce moisture and oxygen below 1%; additional 30–50 cycles reduce to below 0.3% (3,000 ppm). p.29 ↗ Plan gas consumption and cycle time accordingly; Automatic Pressure Controller automates up to 199 cycles. Whole series
Recirculating drying trains can maintain oxygen levels below 5 ppm and moisture levels below 50 ppm. p.34 ↗ Required for sustained low-level atmosphere; use accessory monitors to verify. Whole series
Pressure relief valve #5230600 for Precise Basic requires drilling a 2.06 in (52.3 mm) hole in a 4.00 in (101.6 mm) dia x 1.00 in (25.4 mm) high raised boss; 2.00–18 UNS threads must be exposed. p.13 ↗ Field modification possible; vent kits attach to these threads. Precise Basic
Vent kits: #5242700 (unfiltered), #5238500 (HEPA filtered), #5241200 (carbon filtered); all include 4 in diameter thimble duct connection. p.13 ↗ Remote blowers have 6 in inlet; use 4x6 in duct transition. Whole series
Primary tubing connection: 3/8 in O.D. x 3/16 in I.D. with tube support insert inside soft tubing. p.18 ↗ Prevents tube disconnection; alternative uses tube connector and hose clamp. Whole series
0.1°C fall in temperature equals -34 Pa (-0.14 in WG) pressure change; 0.01 in Hg barometric rise equals -34 Pa (-0.14 in WG). p.26 ↗ Leak decay tests must correct for temperature and barometric pressure. Whole series
Each Precise Glove Box model weighs 185 lbs (83 kg). p.10 ↗ Work surface and lifting equipment must support this weight plus related equipment. Whole series
Work surface depth 29–30 in (737–762 mm); standing height 35–40 in (889–1016 mm). p.7 ↗ Bench must be sized accordingly; Labconco stands adjustable 33–40 in. Whole series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Lights won’t operate. p.47 ↗ Unit not plugged into outlet. Plug unit into appropriate electrical service. Chapter 8
Lights won’t operate. p.47 ↗ Circuit breaker(s) tripped. Reset or replace circuit breaker. Chapter 8
Lights won’t operate. p.47 ↗ Lamp not installed properly. Inspect lamp installation. Chapter 8
Lights won’t operate. p.47 ↗ Lamp wiring disconnected. Inspect lamp wiring. Chapter 8
Lights won’t operate. p.47 ↗ Defective lamp. Replace lamp. Chapter 8
Lights won’t operate. p.47 ↗ Light switch is defective. Replace light switch. Chapter 8
Lights won’t operate. p.47 ↗ Defective electronic ballast. Replace ballast. Chapter 8
No power to duplex electrical. p.47 ↗ Unit not plugged in. Plug unit into electrical service. Chapter 8
No power to duplex electrical. p.47 ↗ Circuit breaker(s) tripped. Reset or replace circuit breaker. Chapter 8
No power to duplex electrical. p.47 ↗ Duplex switch is defective. Replace duplex switch. Chapter 8