Rev g protector work stations user manual

Technical Reference Summary

Covered Models & Document Scope

Covers Labconco Protector Work Stations: 3930000 (115V) and 3930020 (230V) integral-blower models with carbon filters exhausting to the room; 3930001 (115V) and 3930021 (230V) remote-blower/house-exhaust models for outside exhaust, also usable with FilterMate Portable Exhauster and carbon/HEPA filters. Accessories covered include Guardian 500 Airflow Monitor kits 3944700/3944701, Guardian Digital 1000 kits 3908800/3908801, work surfaces 3908402 and 6942003 (with sink), base stand 3746702, transition adapters 3912400/3912401/3912403, and exhaust cover plate 3913100.

Section-by-Section Technical Breakdown

pp. 9-10
Exhaust and Blower Requirements

Defines exhaust volume, enclosure static pressure loss and noise for the 4' Protector Work Station at face velocities of 60, 80 and 100 fpm. Blower selection must combine these values with total system static pressure loss for outside exhaust. A dedicated blower or house exhaust is mandatory, and a house exhaust system requires an adjustable damper.

  • 4' enclosure at 60 fpm: 150 CFM, 0.06 in. static pressure loss, 57-59 dBA.
  • 4' enclosure at 80 fpm: 200 CFM, 0.10 in. static pressure loss, 62-64 dBA.
  • 4' enclosure at 100 fpm: 250 CFM, 0.16 in. static pressure loss, 69-71 dBA.
  • Enclosure must be connected to either a dedicated blower or a house exhaust system.
  • If connected directly to a house exhaust system, an adjustable damper (or valve) must be installed to control airflow, especially when one house system serves multiple enclosures.
pp. 10, 15
Electrical Requirements

Specifies power connection for integral and remote blower configurations. The remote blower may be fed from the switched outlet on the back of the upper light housing, limited to 6 amps. Wiring must be performed by a licensed electrician per local codes and NEC-2002.

  • Standard duplex electrical receptacles should be nearby for the enclosure and other equipment.
  • Remote blower models include iris pass-throughs for electrical cords through the back of the enclosure.
  • Remote blower may be connected to the blower switch via an outlet on the back of the upper light housing.
  • Recommended maximum amperage for the remote blower circuit is 6 amps.
  • The maximum circuit load for the blower is 6 amps.
  • Grounding connection shall not be made to the terminal box cover of the remote blower.
  • Building supply for the remote blower should include overload protection (switch or circuit breaker) marked as the disconnecting device.
pp. 12-13
Installing the Enclosure on a Supporting Structure and Work Surface

Describes work surface requirements and the six-step mounting procedure. The work surface must be smooth, rigid, non-porous, chemical-resistant and have a dished recess for spills. The enclosure front must align within 0.3 inch of the work surface front, and mounting screws require at least 1/8 inch clearance under the head.

  • Supporting structure must safely support the combined weight of enclosure and related equipment.
  • Work surface should be at least as wide as the enclosure.
  • Front of enclosure aligned within 0.3" of the front of the work surface.
  • Insert supplied mounting screws in four holes with minimum 1/8" clearance under the head.
  • On 6942003 work surface with sink, fasteners are omitted and the enclosure is secured with silicone sealant.
  • Sink plumbing: supply valves 1/2" OD slip joint by 1/4" NPT female; drain 1-1/2" NPT female; optional polypropylene P-trap P/N 1432600.
pp. 15-17
Connecting to the Exhaust System

Covers exhaust connection for remote-exhaust models. The enclosure accepts a 2" x 10" nominal transition adapter; Labconco offers top or bottom adapters for 5" hose or 6" duct. Ductwork weight must be supported independently of the enclosure superstructure and the system installed by a qualified HVAC contractor.

  • Exhaust connection accepts a 2" x 10" (5.1 cm x 25.4 cm) nominal transition adapter.
  • Transition adapters available for top or bottom exhaust and for 5" dia. hose or 6" dia. duct.
  • Exhaust ductwork weight must be supported independently of the enclosure superstructure.
  • Exhaust system should be installed by a qualified HVAC contractor.
  • Duct material should match the enclosure, procedures and chemical applications.
pp. 18
Installation of Carbon Filters for Room Exhaust

Details carbon filter installation behind the perforated baffle for integral-blower models. Two filters must be installed with the gasket side on the downstream exhaust side. When combining formaldehyde and organic vapor filters, the formaldehyde filter must be downstream (on top) of the organic vapor filter.

  • Unthread knobs on the perforated baffle in the upper corners and slowly lower the baffle.
  • All carbon filters install with the gasket side on the downstream exhaust side.
  • Two carbon filters must be installed behind the perforated baffle for capacity and air mixing.
  • Formaldehyde filter must be placed downstream of the organic vapor filter (on top when baffle is tilted down).
  • Organic Vapor filter 3924200, 10.5 lbs; Formaldehyde filter 3924201, 13.3 lbs; Ammonia/Amines filter 3924202, 15.2 lbs.
  • Carbon filters provide no particulate protection, only odor control for low-level concentrations below OSHA TWA.
pp. 19-22
Connecting the Electrical Supply Source to the Enclosure

Gives the power connection procedure for 115V and 230V models. Integral blower power cord connects from the motorized impeller outlet to the back of the upper light control panel. At 50 Hz, integral blower maximum airflow is reduced by 17%. 230V models ship without a plug.

  • 115V: connect supplied power cord to IEC inlet on back of enclosure; connect integral blower cord from motorized impeller outlet to upper light control panel.
  • At 50 Hz operation, integral blower maximum airflow is reduced by 17%.
  • 230V models ship without a plug; install appropriate plug per local codes.
  • Do not use any detachable power cord not adequately rated for the unit.
  • Remote blower wires connect to the switched outlet on the back of the upper light control panel, switched by the blower switch.
pp. 22
Setting the Face Velocity with the Speed Control Adjustment

Explains how to set face velocity on integral-blower models using the speed control behind the upper light control panel. Face velocity must be adjusted between 60 and 100 fpm; turning the control counterclockwise increases velocity and clockwise decreases it. Lowest appropriate velocity gives quietest operation.

  • Face velocity should be adjusted from 60 to 100 fpm.
  • Speed control is located behind the upper light control panel.
  • Turn speed control counterclockwise to increase face velocity, clockwise to decrease.
  • Face velocity measured with an anemometer per the Industrial Ventilation Manual.
  • Working at the lowest face velocity appropriate for the application gives the quietest operation.
pp. 22-23
Validating the Protector Work Station Enclosure

Defines the face velocity validation procedure and ASHRAE 110-1995 test results. Readings are taken in two rows across the sash opening, 6 inches from the ends and every 12 inches, with eight readings averaged for the 4' enclosure. All enclosures achieve an as-manufactured rating below 0.05 ppm at 4 lpm.

  • Two rows of readings: first row 3" down from upper sash foil, second row 3" up from work surface.
  • Readings 6" from ends and evenly spaced every 12"; eight readings averaged for 4' enclosure.
  • Tested per ASHRAE 110-1995; as-manufactured rating AM <0.05 ppm at 4 lpm.
  • Face velocity profiles and smoke testing should be performed frequently per the organization's quality system.
pp. 24-29
High Performance Features

Describes the aerodynamic design elements that provide containment: Clean-Sweep air foil, upper sash foil, upper dilution air supply, zoned rear perforated baffle, side-entry air foils, 20-degree slope, safety glass sash, utility ports, and the built-in motorized impeller with speed control. Remote blower models provide 23" working depth versus 17.19" for integral blower models.

  • Upper dilution air supply introduces 5 to 7 percent of required air volume.
  • Remote blower model working depth 23"; integral blower model working depth 17.19".
  • Sash is 1/4 inch tempered safety glass with spring-loaded latch; must be down for normal operation.
  • Ergonomic slope of 20 degrees.
  • Utility ports with iris on remote exhaust models; ships with solid and iris plugs.
  • Guardian 500 and Guardian 1000 Digital airflow monitors are available options on all models.
pp. 35-58
Routine Maintenance Schedule

Provides the periodic maintenance schedule for the enclosure, including filter replacement determination and component replacement procedures for the fluorescent light, motorized impeller and speed control.

  • Determination of when to replace carbon filters and how to replace them.
  • Calculating carbon filter life.
  • Fluorescent light replacement.
  • Motorized impeller replacement.
  • Speed control replacement.
pp. 37-43
Guardian 500 Airflow Monitor

Covers component identification, installation, calibration and alarm activation for the Guardian 500 Airflow Monitor kits 3944700 (115V) and 3944701. The monitor provides audio/visual alarm for low airflow conditions.

  • Kit numbers 3944700 (115V) and 3944701.
  • Includes component identification, installation procedure, calibration procedure and alarm activation.
  • Audio/visual alarm alerts user to low airflow conditions.
pp. 44-51
Guardian Digital 1000 Airflow Monitor

Covers operation, installation and calibration of the Guardian Digital 1000 kits 3908800 (115V) and 3908801. Includes an alternate calibration procedure for constant volume conditions. The monitor displays face velocity, provides RS232 output, night setback mode and auxiliary relay ports.

  • Kit numbers 3908800 (115V) and 3908801.
  • Displays face velocity value.
  • Provides RS232 output, night setback mode and several auxiliary relay ports.
  • Alternate calibration procedure for constant volume conditions.
pp. 66-68
Troubleshooting

Provides a problem/cause/corrective action table covering containment, blower operation, lighting and airflow monitor malfunctions. If suggested corrective actions do not solve the problem, contact Labconco.

  • Contaminants outside enclosure: improper technique, baffle/exhaust restriction, external airflow disruption, improper face velocity.
  • Blower won't operate: power, circuit breaker/GFI, wiring, switch, motorized impeller.
  • Low face velocity: sash not closed, clogged HEPA on FilterMate, loaded carbon filter.
  • Lights do not work: breaker, lamp installation/wiring, lamp, switch, ballast.
  • Airflow monitor malfunction: power, cables, alarm set point, blocked sensor airway, alarm silencing behavior.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
4' Protector Work Station at 60 fpm requires 150 CFM with 0.06 in. enclosure static pressure loss and produces 57-59 dBA. p.9 ↗ Use these values with total system static pressure loss to size a remote blower for low-flow operation. 4' Protector Work Station
4' Protector Work Station at 80 fpm requires 200 CFM with 0.10 in. static pressure loss and produces 62-64 dBA. p.9 ↗ Intermediate airflow setting for higher containment demand; noise increases with face velocity. 4' Protector Work Station
4' Protector Work Station at 100 fpm requires 250 CFM with 0.16 in. static pressure loss and produces 69-71 dBA. p.9 ↗ Maximum recommended face velocity; highest exhaust volume and noise of the three settings. 4' Protector Work Station
Face velocity must be adjusted from 60 to 100 fpm using the speed control behind the upper light control panel; counterclockwise increases velocity. p.22 ↗ Set the lowest velocity appropriate for the application to minimize noise while maintaining containment. Integral blower models 3930000/3930020
At 50 Hz operation, integral blower maximum airflow is reduced by 17%. p.19 ↗ Account for reduced airflow when sizing and validating 50 Hz installations. Integral blower models
Maximum circuit load for the blower is 6 amps; recommended maximum amperage for the remote blower circuit is 6 amps. p.10 ↗ Do not exceed 6 A on the switched blower outlet; provide overload protection marked as disconnecting device. All models with remote blower connection
Enclosure front must be aligned within 0.3" of the front of the work surface; mounting screws require minimum 1/8" clearance under the head. p.12 ↗ Ensures proper seating and airflow at the lower air foil; misalignment can affect containment. All models
Two carbon filters must be installed behind the perforated baffle with gasket side on the downstream exhaust side; formaldehyde filter must be downstream of organic vapor filter. p.18 ↗ Correct filter count and order are required for capacity, air mixing and proper adsorption. Integral blower models 3930000/3930020
Carbon filter part numbers and weights: Organic Vapor 3924200 (10.5 lbs), Formaldehyde 3924201 (13.3 lbs), Ammonia/Amines 3924202 (15.2 lbs). p.18 ↗ Select filter type based on chemical class; concentrations must not exceed IDLH in work area and TWA in exhaust. Integral blower models
Exhaust connection accepts a 2" x 10" (5.1 cm x 25.4 cm) nominal transition adapter; adapters available for top or bottom exhaust and 5" hose or 6" duct. p.15 ↗ Order correct transition adapter for the chosen duct routing and diameter. Remote exhaust models 3930001/3930021
Remote blower model provides 23" working depth; integral blower model provides 17.19" working depth. p.24 ↗ Choose remote blower model when deeper interior workspace is required. 3930001/3930021 vs 3930000/3930020
Upper dilution air supply introduces 5 to 7 percent of the required air volume. p.24 ↗ Part of the total exhaust volume; account for it in airflow balance and blower sizing. All models
Enclosures tested per ASHRAE 110-1995 achieve as-manufactured rating AM <0.05 ppm at 4 lpm. p.23 ↗ Baseline containment performance; field certification should be arranged separately. All models
Face velocity validation: two rows of readings, 6" from ends and every 12", first row 3" down from upper sash foil, second row 3" up from work surface; eight readings averaged for 4' enclosure. p.22 ↗ Standardized method for verifying average face velocity within 60-100 fpm. 4' Protector Work Station
Sash is 1/4 inch tempered safety glass and must remain in the down position during normal operation. p.25 ↗ Operating with sash open compromises containment. All models
General acid use and perchloric acid use are prohibited in this enclosure; high-level radioisotope materials are prohibited. p.26 ↗ Select a different enclosure type for acid or radioisotope applications. All models
Enclosure weight is either 85 or 105 lbs (37 or 46 kg) depending on model. p.11 ↗ Verify supporting structure and lifting equipment capacity; do not lift by the front air foil. All models
Sink work surface 6942003 plumbing: supply valves 1/2" OD slip joint by 1/4" NPT female; drain 1-1/2" NPT female; optional P-trap P/N 1432600. p.13 ↗ Order correct plumbing fittings and P-trap for histology/pathology work surface. Work surface 6942003

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Contaminants outside of enclosure. p.70 ↗ Improper user techniques for the enclosure. See "Certifying the Enclosure" Chapter 3 and "Safety Precautions" Chapter 4 sections in the manual. (Ref. Appendix D). Chapter 8
Contaminants outside of enclosure. p.70 ↗ Restriction of the baffle air slots or blockage of the exhaust outlet. Remove obstruction to ensure that all air slots and the exhaust outlet are unobstructed. Chapter 8
Contaminants outside of enclosure. p.70 ↗ External factors are disrupting the enclosure airflow patterns or acting as a source of contamination. See "Location Requirements" Chapter 2, "Certifying the Enclosure" Chapter 3, and "Safety Precautions" Chapter 4 sections of this manual. (Ref. Appendix D). Chapter 8
Contaminants outside of enclosure. p.70 ↗ Enclosure has improper face velocity. Have enclosure certified and check exhaust system. Check carbon filters for loading. Check accessory FilterMate HEPA filters for loading. Adjust enclosure speed control. Enclosure should have an average face velocity of 60-100 fpm. Chapter 8
Blower won't operate. p.70 ↗ Unit not plugged into outlet. Plug the enclosure into appropriate electrical service. Chapter 8
Blower won't operate. p.70 ↗ Circuit breaker(s) or Ground Fault Interrupter. Reset circuit breaker. Chapter 8
Blower won't operate. p.70 ↗ Blower wiring is disconnected. Inspect blower wiring. Chapter 8
Blower won't operate. p.70 ↗ Blower switch is defective. Replace switch. Chapter 8
Blower won't operate. p.70 ↗ Motorized impeller or blower is defective. Replace motorized impeller or blower. See Chapter 6. Chapter 8
Low face velocity or poor containment of contaminants. p.71 ↗ Enclosure sash not closed. Close sash to the lowest position. Chapter 8
Low face velocity or poor containment of contaminants. p.71 ↗ HEPA filter clogged on FilterMate. Replace HEPA filter or increase speed. Chapter 8
Low face velocity or poor containment of contaminants. p.71 ↗ Carbon filter loaded with chemicals. Replace carbon filter. Chapter 8
Blower and lights won't operate. p.71 ↗ Unit not plugged into outlet. Plug enclosure into appropriate electrical service. Chapter 8
Lights do not work. p.71 ↗ Circuit breaker(s) tripped. Reset or replace circuit breaker. Chapter 8
Lights do not work. p.71 ↗ Lamp not installed properly. Inspect lamp installation. Chapter 8
Lights do not work. p.71 ↗ Lamp wiring disconnected. Inspect lamp wiring. Chapter 8
Lights do not work. p.71 ↗ Defective lamp. Replace lamp. Chapter 8
Lights do not work. p.71 ↗ Light switch is defective. Replace light switch. Chapter 8
Lights do not work. p.71 ↗ Defective electronic ballast. Replace ballast. Chapter 8
Airflow monitor malfunction. No power. No lights. No display. p.71 ↗ Power cable to circuit board is disconnected and needs to be connected. Sensor cable needs to be connected. Power supply is not plugged into proper voltage; plug in power supply. Verify that all airflow monitor interface cables are connected. Check circuit breaker on enclosure. Connect power cable to circuit board, connect sensor cable, plug power supply into proper voltage, verify all interface cables connected, check circuit breaker on enclosure. Chapter 8
No audible alarm. p.71 ↗ Alarm has been temporarily silenced using "SILENCE ALARM" or "enter" buttons. Restore audible alarm (refer to monitor manual). Chapter 8
Wrong alarm set point. p.71 ↗ Airflow monitor was not properly adjusted. Repeat calibration steps outlined in this manual in Chapter 6 or in the supplied Guardian 1000 Digital Monitor Manual. Chapter 8
Constant audible alarm. p.71 ↗ Airflow or calibration of airflow monitor. Check airflow and calibration of airflow monitor. See Chapter 6. Chapter 8
Continuous alarm. p.72 ↗ Face velocity of the enclosure may have changed; HEPA filter on accessory FilterMate may have become loaded. Check face velocity; if incorrect, adjust speed control to increase face velocity. If face velocity is correct, calibrate the airflow monitor as outlined in Chapter 6. Chapter 8
Monitor alarms; air way to airflow monitor sensor is blocked by insects, dust or debris. p.72 ↗ Airway to airflow monitor sensor blocked by insects, dust or debris. Lightly clean the airway with clean air. Be careful not to touch sensitive electrical components. Chapter 8
Audible disable will not stay operational. p.72 ↗ An alarm condition must be continuously present before the audible alarm can be silenced. If flow conditions fluctuate near the alarm set point, the airflow monitor will automatically reset it. Bring the enclosure airflow into proper operating parameters or adjust the alarm set point lower. Chapter 8