Rev g users manual logicvue
Technical Reference Summary
Covered Models & Document Scope
Logic Vue Class II Enclosure, catalog numbers 3300400/-10 and 3300401/-11 (4-foot models) and 3300600/-10 and 3300601/-11 (6-foot models); -10/-11 suffixes denote 115 V and 230 V variants respectively. Optional accessories covered: Canopy Connection Kit, Air-Tight Damper, Large Tube Pass-Thru Kit, Service Fixture, Airflow Sensor, Access Door, Computer Mount.
Section-by-Section Technical Breakdown
The enclosure must be installed on a level, solid floor away from doors, fans, ventilation registers, chemical fume hoods and other air-handling devices, with room windows kept closed. Minimum clearances are specified above, to the right, front and rear for airflow, electrical access and hatch operation.
- Minimum 6 in (150 mm) clearance above the product for proper airflow.
- Minimum 24 in (610 mm) clearance on the right side for access to the electrical box.
- Minimum 36 in (915 mm) front and rear clearance to open Front and Rear Hatches and not restrict exhaust airflow; rear clearance may be reduced to 24 in (610 mm) if the Rear Hatch need not be opened.
- If exhausting back into the laboratory, allow at least 24 in (610 mm) between the exhaust opening and any large vertical surface or wall.
Each model requires a dedicated, hard-wired single-phase supply with a circuit breaker marked as the main disconnect, located no more than 10 ft (3 m) from the product. A power cord must not be used; wiring must be performed by a licensed electrician.
- 3300400/-10: 6 A typical operating current, 115 V, 60 Hz, 30 A circuit, 10 AWG (6 mm²) service wire.
- 3300401/-11: 3 A typical operating current, 230 V, 50/60 Hz, 20 A circuit, 12 AWG (4 mm²) service wire.
- 3300600/-10: 8 A typical operating current, 115 V, 60 Hz, 30 A circuit, 10 AWG (6 mm²) service wire.
- 3300601/-11: 4 A typical operating current, 230 V, 50/60 Hz, 20 A circuit, 12 AWG (4 mm²) service wire.
- Service wire must be rated at 80 °C minimum; dedicated circuit must provide a protective earthing ground.
- Supply conductors L1, L2 and earth ground are landed on DIN-rail terminals inside the electrical box; wiring diagram is on the inside of the electrical box cover.
Utility service lines must be ¼ in O.D. brass, copper or stainless steel with an accessible shut-off valve. Accessory service valves are rated for 40 PSI (275 kPa); higher line pressure requires a regulator.
- Tubing must be ¼ in O.D. soft metal, fully deburred, routed through a slot in the lower end panel without contacting electrical wires.
- Tube inserts approximately ¾ in (19 mm) into the 90° fitting; tighten nut hand-tight plus at least ¾ turn with a 7/16 in wrench.
- Close the service valve in the enclosure, slowly open the line shut-off valve, and leak-test all fittings.
- Flammable gases or solvents and high-pressure gases should be avoided in the enclosure.
When the optional Canopy Connection Kit is installed, the enclosure may be connected to a dedicated building exhaust system. Duct sizing, straight-run lengths, balancing damper and exhaust volumes are specified.
- Clearance must accommodate a 12-inch diameter duct; avoid an elbow directly above the exhaust connection.
- Provide a straight length of 10 duct diameters between the enclosure connection and any elbow, and between subsequent elbows.
- Canopy inlet relief valve draws a maximum of 200 CFM (340 m³/h); drawing additional room air can cause unstable operation.
- An air-tight balancing damper is highly recommended at the canopy duct collar for fine adjustment and isolation.
- Manifolded exhaust should be avoided; if used, a fast-acting variable air volume valve (e.g. venturi) is required.
- Recommended duct vacuum is 0.21 in H₂O (52 Pa); vacuums below 0.05 in H₂O (12 Pa) or above 0.5 in H₂O (125 Pa) may cause erratic operation and an alarm.
Installation covers crate removal, moving the heavy enclosure, installing the four levelling feet with seismic restraint, and reassembly for narrow/low door clearance. The levelling feet are mandatory for full load rating and seismic compliance.
- Two persons minimum for unpacking; lifting equipment must be certified to 2000 lbs.
- Four levelling feet must be installed to level the work area, allow full load capacity, and comply with seismic restraint regulations.
- With shipping casters installed, work surface load rating is reduced to 165 lbs (75 kg) per 2 linear feet (60 cm).
- Raise one end approximately ¼ in (6 mm) with the scissor jack; chock the casters on the end not being raised.
- Plenum/blower lift bolts (4) are turned counterclockwise in a pattern of no more than 10 revolutions each to lower the assembly evenly.
A qualified certifier must certify the enclosure before initial use and at least annually or after relocation/service. Tests include downflow velocity profile, inflow volume, airflow patterns, HEPA filter leak, and optional cleanliness, lighting and noise tests.
- Downflow test: probe 12 in (30.5 cm) above work surface, 6.0 in (152 mm) test-point spacing, 28 points for 4-ft models and 44 points for 6-ft models; average 40, 45 or 50 fpm (±3 fpm).
- Inflow volume: 278-306 cfm (472-520 m³/h) for 4-ft models; 411-452 cfm (698-768 m³/h) for 6-ft models; average inflow velocity 100-110 fpm.
- HEPA leak test: scan at no more than 1 in from filter media and no more than 2 in/s traverse rate; photometer warmed up at least 5 minutes; generator air pressure 23 psi.
- Certifier password: [LIGHT] [UP] [TIMER] [TIMER] [OK/MUTE]; zero-point calibration password: [LIGHT] [UP] [TIMER] [DOWN] [OK/MUTE].
The MyLogic operating system provides the home screen, keypad, menus, timers, airflow alert, alarms, security lock, USB/BMS data output and automatic operation options (NightSmart/SmartStart).
- Alarms include Power Loss, Sash, Blower Failure and Running Exhaust alarms.
- Airflow Alert can be reset from the Tools submenu.
- BMS data output via USB cable with configurable output rate and defined data string format.
- Security lock and units-of-measure settings are accessible from the Settings submenu.
Configuration is restricted to certifiers/installers and covers verification of installed accessories and blower speeds via the Info Screen. Calibration covers blower speed adjustment, filter life gauge setting, airflow sensor calibration and zero-point calibration.
- Info Screen shows current configuration and blower speeds; it is not password protected.
- Supply blower speed is adjusted to set average downflow; exhaust blower speed is adjusted to set inflow volume.
- Airflow sensor replacement requires zero-point calibration using the separate password.
- Filter Life Gauge is set during calibration.
Maintenance covers safety precautions, a recommended maintenance schedule, circuit breaker reset, drip tray removal, pre-filter replacement, LED lamp replacement and storage.
- Resetting a circuit breaker is a user service operation.
- Pre-filter and LED lamps are replaceable maintenance items.
- Drip tray is removable for cleaning and for exhaust HEPA filter leak testing.
- Decontamination is required before maintenance in contaminated areas, before HEPA filter changes, before certification entry, before relocation, before program changes, and after gross spills.
Advanced procedures cover external dress panel removal, sash sensor adjustment/replacement, supply and exhaust HEPA filter replacement, blower/motor replacement, display board replacement, preparation for gaseous sterilization, service fixture installation and wiring diagrams.
- Supply and exhaust HEPA filters are replaced separately.
- Supply and exhaust blower/motor assemblies are replaced separately.
- Wiring diagrams are provided for 100-120 V and 208-240 V configurations.
- Gaseous sterilization preparation procedure is included.
Accessories include Canopy Connection, Large Tube Pass-Thru Kit, Service Fixture, Airflow Sensor, Access Door and Computer Mount.
- Canopy Connection Kit enables connection to a remote building exhaust system.
- Airflow Sensor is field-installable and requires zero-point calibration after installation.
- Service Fixture can be field installed at any time; the kit contains installation instructions.
- Large Tube Pass-Thru Kit, Access Door and Computer Mount are available options.
Appendices provide the parts list, overall dimensions, power data, motor specifications, environmental conditions and a quick chart reference.
- Power data and motor specifications are listed in Appendix C.
- Environmental conditions for operation are specified in Appendix C.
- Overall product dimensions are in Appendix B.
- Quick chart reference in Appendix D includes HEPA leak test data such as theoretical aerosol concentration and Laskin nozzle count.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Minimum clearance above the product is 6 in (150 mm); right side 24 in (610 mm); front and rear 36 in (915 mm), reducible to 24 in (610 mm) at the rear if the Rear Hatch is not opened. p.12 ↗ | Site planning must reserve these clearances to maintain airflow, electrical access and hatch operation. | whole series |
| Catalog 3300400/-10 requires 115 V, 60 Hz, 30 A single-phase circuit with 10 AWG (6 mm²) wire rated 80 °C minimum; typical operating current 6 A. p.12 ↗ | Electrical supply and breaker must be sized to the 30 A circuit, not the 6 A operating current. | 3300400, 3300401, 3300600, 3300601 variants as tabulated |
| Catalog 3300401/-11 requires 230 V, 50/60 Hz, 20 A single-phase circuit with 12 AWG (4 mm²) wire; typical operating current 3 A. p.12 ↗ | 230 V variants use a smaller service wire and 20 A circuit than 115 V variants. | 3300401, -11; 3300601, -11 |
| The dedicated circuit breaker must be marked as the main disconnect and located no more than 10 ft (3 m) from the product. p.12 ↗ | Installation layout must place the junction box/breaker within 10 ft and label it as the disconnecting device. | whole series |
| Accessory service valves are rated for 40 PSI (275 kPa); higher supply pressure requires a pressure regulator. p.12 ↗ | Facility gas/utility supply must be regulated to ≤40 PSI before connection. | models with service fixtures |
| Canopy inlet relief valve draws a maximum of 200 CFM (340 m³/h); exhaust duct clearance must accommodate a 12-inch diameter duct. p.13 ↗ | Exhaust system must be sized for the enclosure exhaust volume plus up to 200 CFM room air, and duct routing must fit a 12-inch duct. | models with Canopy Connection Kit |
| Recommended duct vacuum is 0.21 in H₂O (52 Pa); below 0.05 in H₂O (12 Pa) or above 0.5 in H₂O (125 Pa) may cause erratic operation and an alarm. p.14 ↗ | Exhaust balancing must target 0.21 in H₂O and stay within the 0.05-0.5 in H₂O window. | canopy-connected models |
| 4-foot models: exhaust volume 1200-1320 cfm (2040-2240 m³/h) with concurrent balance value 1320-1452 cfm (2242-2467 m³/h). p.14 ↗ | Exhaust system must be capable of moving these volumes at the stated negative pressure. | 33004xx (4-foot) |
| 6-foot models: exhaust volume 1575-1775 cfm (2675-3015 m³/h) with concurrent balance value 1733-1953 cfm (2944-3318 m³/h). p.14 ↗ | 6-foot models require substantially higher exhaust capacity than 4-foot models. | 33006xx (6-foot) |
| Work surface load limit is 250 lbs (113 kg) per 2 linear feet (60 cm) with levelling feet installed; reduced to 165 lbs (75 kg) per 2 linear feet if shipping casters remain. p.20 ↗ | Levelling feet must be installed to achieve the full load rating and seismic restraint. | whole series |
| Downflow acceptance: average 40, 45 or 50 fpm (0.20, 0.23 or 0.25 m/s) ±3 fpm (0.015 m/s), measured 12 in (30.5 cm) above the work surface with 6.0 in (152 mm) test-point spacing; 28 points for 4-ft, 44 points for 6-ft models. p.50 ↗ | Certification must use the specified grid and acceptance band; supply blower speed is adjusted to meet it. | whole series |
| Inflow acceptance: 278-306 cfm (472-520 m³/h) for 4-ft models and 411-452 cfm (698-768 m³/h) for 6-ft models, average inflow velocity 100-110 fpm (0.51-0.56 m/s). p.53 ↗ | Exhaust blower speed is adjusted to bring inflow volume into this range. | whole series |
| HEPA leak scan: sampling nozzle no more than 1 in from filter media, traverse rate no more than 2 in/s, photometer warm-up at least 5 minutes, aerosol generator air pressure 23 psi, oil level within 1/8 in (3 mm) of the level line. p.55 ↗ | Certifier must follow these scan parameters for valid leak-test results. | whole series |
| Service tubing must be ¼ in O.D. soft metal, inserted approximately ¾ in (19 mm) into the fitting, nut tightened hand-tight plus at least ¾ turn with a 7/16 in wrench. p.47 ↗ | Plumbing connections must follow this insertion depth and torque procedure to avoid leaks. | models with service fixtures |
| Certifier password is [LIGHT] [UP] [TIMER] [TIMER] [OK/MUTE]; zero-point calibration password is [LIGHT] [UP] [TIMER] [DOWN] [OK/MUTE]. p.49 ↗ | Calibration and certification access requires these key sequences; airflow sensor replacement requires the zero-point password. | whole series |
| Blowers reach stable operation generally when the Please Wait notification is dismissed, about 90 seconds after start. p.51 ↗ | Certification tests and airflow readings should be taken only after this stabilization period. | whole series |
| If large equipment cannot have its outer panels removed for downflow, set average downflow no higher than 40 fpm (0.20 m/s). p.51 ↗ | Reduces high-velocity zones between equipment and hatch walls. | whole series |
| Enclosure must be recertified at least annually and whenever relocated or serviced; filter integrity and airflow performance must be verified before use. p.20 ↗ | Maintenance planning must include annual certification and post-service recertification. | whole series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Blower and Lights not working p.156 ↗ | Not stated in the extracted text of this manual. | Not stated in the extracted text of this manual. | Chapter 13 |
| Blower only will not start p.157 ↗ | Not stated in the extracted text of this manual. | Not stated in the extracted text of this manual. | Chapter 13 |
| Lights only will not illuminate p.158 ↗ | Not stated in the extracted text of this manual. | Not stated in the extracted text of this manual. | Chapter 13 |
| Airflow Alert activating p.159 ↗ | Not stated in the extracted text of this manual. | Not stated in the extracted text of this manual. | Chapter 13 |
| Filter Life Gauge not at 100% when new p.160 ↗ | Not stated in the extracted text of this manual. | Not stated in the extracted text of this manual. | Chapter 13 |
| Contamination in the work area p.161 ↗ | Not stated in the extracted text of this manual. | Not stated in the extracted text of this manual. | Chapter 13 |