Rev g protector controlled atmosphere glove boxes
Technical Reference Summary
Covered Models & Document Scope
Covers Labconco Protector Controlled Atmosphere Glove Boxes, base catalog models 50600 (manual pressure control) and 50800 (auto pressure control), each in fiberglass or stainless steel liner, single or double width, and combination variants (auto pressure control with blower/filter conversion). Configured catalog numbers are formed by appending a two-digit liner width code to the base model: 10/12 for single/double manual and auto, 60/62 for combination single/double. Electrical variants: North America 100-115V 50/60 Hz, British (UK) 230V 50/60 Hz, Schuko 230V 50/60 Hz, China 230V 50/60 Hz, Australia 230V 50/60 Hz, and North America 230V 60 Hz only.
Section-by-Section Technical Breakdown
Site preparation requires a level, solid support rigidly mounted to floor or wall (not both), with a typical standing height of 35-40 in (889-1016 mm). The glove box rests on a 29-30 in (737-762 mm) deep work surface. Vibration-sensitive work benefits from corner locations and isolation.
- Bench/stand rigidly mounted to floor or wall, but not both.
- Standing height 35-40 in (889-1016 mm); Labconco adjustable stands 33-40 in (838-1016 mm).
- Work surface depth 29-30 in (737-762 mm).
- Avoid vibration-producing equipment on the bench; building corners have less vibration.
- Marble slab with dampening pads is an effective low-cost vibration isolator.
The pressure relief bubbler protects the main chamber from over/under pressurization and glass breakage, activating above +5 in w.g. or below -5 in w.g. It must be filled with 0.87 specific gravity oil to 6 in (152 mm) measured from the bottom of the stainless steel bubbler. Exhaust may be routed outside or through the optional HEPA Filter Exhaust Kit 5441403.
- Bubbler activates at pressures/vacuums above +5 in w.g. (1245 Pa) or below -5 in w.g. (-1245 Pa).
- Fill with 0.87 specific gravity oil to 6 in (152 mm) from bubbler bottom.
- Exhaust pipe is 3/4 in diameter; 3/4 in I.D. neoprene hose kit available.
- HEPA Filter Exhaust Kit 5441403 for filtered room exhaust.
- Exhaust ductwork must be supported independently of the glove box superstructure.
Single-width units require two line cords; double-width units require three. Each line cord is rated 100-115V 12A or 230V 12A. Interior duplex receptacles are rated 10A maximum; the exterior vacuum pump outlet is rated 100-115V 12A or 230V 6A (8A for 100-115V models, 4.5A for 230V models per Chapter 3).
- Single-width: two line cords; double-width: three line cords.
- Line cords each rated 100-115V 12A or 230V 12A.
- Interior duplex: 115V/230V/100V, 10A maximum.
- Exterior vacuum pump outlet: 100-115V 12A or 230V 6A; rated 8A (100-115V) or 4.5A (230V).
- 230V models: one cord at 12A, other at 6A for vacuum pump outlet.
- Foot switch (3-position connector) supplied on auto pressure control and combination models.
The glove box has two manual valves on the main chamber and two on the transfer chamber for inert gas inlet and vacuum outlet. Inert gas supply pressure must never exceed 10-14 psi (69-96 kPa) via a supply regulator. Solenoid valves use 3/8 in compression fittings or 3/8 in hose barbs; rear purifier connections are 3/4 in compression.
- Inert gas line pressure must never exceed 10-14 psi (69-96 kPa).
- Two valves on main chamber, two on transfer chamber; transfer lines may be tied to main lines.
- Solenoid valves: 3/8 in (9.5 mm) compression fittings or 3/8 in hose barbs.
- Rear purifier/drying train connections: two 3/4 in compression fittings.
- Primary tubing: 3/8 in O.D. x 3/16 in I.D. with tube support insert.
- Gas and Vacuum Tubing Connection Kits 5245100 or 5245200.
The glove box should remain in its shipping container until placed at the final location. Weights are 350 lb/157 kg (fiberglass), 500 lb/225 kg (stainless), and 1000 lb/450 kg (stainless doublewide). The work surface must support the combined weight and be as wide as the glove box.
- Fiberglass model weight: 350 lb/157 kg.
- Stainless model weight: 500 lb/225 kg.
- Stainless doublewide weight: 1000 lb/450 kg.
- Do not lift by upper sheet metal façade or front panel.
- Work surface height 35-40 in (889-1016 mm); front of glove box aligned with front of work surface.
Gloves are installed with thumbs up and right/left orientation by stretching the beaded cuff into the glove port groove nearest the window. A separate 8 in diameter O-ring is installed over the gloves into the outer groove and secured with stainless steel band clamps.
- Thumbs up, right/left orientation.
- Beaded glove cuff stretched into groove nearest the window.
- 8 in diameter O-ring over gloves in outer groove.
- Stainless steel band clamps secure the O-ring.
Each glove box is helium leak tested at the factory to no leak greater than 1 x 10^-6 cc/sec at +5 in helium. Oxygen permeation is <0.16 ppm/min (stainless) and <0.30 ppm/min (fiberglass) per ISO 10648-2, meeting Class 1 (<1.67 ppm/min). Pressure leak decay tests use -1000 Pa for acceptance and -250 Pa for operational testing.
- Factory helium leak test: no leak >1 x 10^-6 cc/sec at +5 in helium.
- Stainless permeation <0.16 ppm/min; fiberglass <0.30 ppm/min.
- ISO 10648-2 Class 1 acceptance level: <1.67 ppm/min (<5 x 10^-4 hr^-1).
- Acceptance pressure leak decay test at -1000 Pa (-4 in w.g.); operational at -250 Pa (-1 in w.g.).
- 0.1°C temperature fall equals -34 Pa (-0.14 in w.g.) pressure change.
- 0.01 in Hg barometric rise equals -34 Pa (-0.14 in w.g.) pressure change.
Fiberglass models maintain oxygen to 0.4 ppm and moisture to 2 ppm; stainless models to 0.2 ppm and 1 ppm. Manual models use analog gauges (-5 to +5 in w.g. main, 0 to -29.5 in Hg transfer); auto models use digital LCD and automate up to 499 main-chamber and 199 transfer-chamber cycles. All models have four manual valves, pressure relief bubbler, 3/8 in compression fittings, and interior duplex receptacles.
- Fiberglass attainable purity: O2 0.4 ppm, H2O 2 ppm (installed purifier).
- Stainless attainable purity: O2 0.2 ppm, H2O 1 ppm (installed purifier).
- Manual gauges: main -5 to +5 in w.g.; transfer 0 to -29.5 in Hg.
- Auto pressure controller: main -5 to +5 in w.g.; transfer 0 to -29.5 in Hg; up to 499/199 cycles.
- Viewing window 36 in w x 27.6 in high, 3/8 in laminated safety glass.
- Gloves: neoprene, 0.015 in thick, 30 in long, size 9 3/4; doublewide has two pair.
- Glove ports: two 8 in ID epoxy-coated aluminum, 17 in apart; doublewide has four.
- Lighting: two 25 W fluorescent (four on doublewide); fiberglass >120 fc, stainless >80 fc.
- Doublewide main chamber volume 34 cu ft.
Start-up involves turning on the vacuum pump, opening the inert gas tank valve, and turning on the light; the AtmosPure Re-Gen Gas Purifier can start once atmosphere is below 5000 ppm. Transfer chamber evacuation/fill typically takes 1-5 cycles to reach below 1% O2/H2O; main chamber normally requires 75-100 cycles. Four operating modes cover main chamber pressure control, main chamber purge/fill, transfer chamber purge/fill, and setup menu.
- Start purifier once inert atmosphere below 5000 ppm is established.
- Transfer chamber: 1-5 evacuation/fill cycles to reduce O2/H2O below 1%.
- Main chamber: normally 75-100 evacuation/fill cycles to reduce below 1%.
- Porous materials may require longer sustained vacuum purges.
- Ensure inner transfer door is closed before operating transfer chamber gas/vacuum valves.
- Operation Mode 1: Glove Box Pressure Control; Mode 2: Glove Box Evacuation (Purge)/Fill; Mode 3: Transfer Chamber Evacuation (Purge) and Fill; Mode 4: Setup Menu.
Chapter 6 provides a routine maintenance schedule, initial certification and annual recertification, fluorescent light replacement, window replacement, solenoid valve replacement, and auto pressure controller circuit board replacement.
- Initial certification by qualified technician before first use.
- Recertification at least annually after installation.
- Fluorescent light replacement procedure.
- Window replacement procedure.
- Solenoid valve replacement procedure.
- Auto pressure controller circuit board replacement procedure.
Chapter 8 provides a problem/cause/corrective action table covering lights, electrical receptacles, auto pressure controller solenoid valves, and auto pressure controller operation.
- Lights will not operate: check outlet, circuit breaker, lamp installation/wiring, lamp, light switch, ballast.
- No power to electrical receptacle: check plug, circuit breaker, receptacle switch, receptacle.
- Auto Pressure Controller Solenoid Valves will not operate: check plumbing orientation (VAC out to pump, GAS in from supply) or replace solenoid valve.
- Auto Pressure Controller will not operate: check power, circuit breaker, LCD display/circuit board.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Inert gas supply pressure must never exceed 10-14 psi (69-96 kPa) through a gas supply pressure regulator. p.9 ↗ | Prevents overpressurization and glass breakage; install a regulator on the inert gas supply. | All Protector Controlled Atmosphere Glove Box models |
| Pressure relief bubbler activates at pressures above +5 in w.g. (1245 Pa) or vacuums below -5 in w.g. (-1245 Pa). p.8 ↗ | Defines the safe operating pressure envelope; exceeding it vents through the bubbler. | All models |
| Bubbler must be filled with 0.87 specific gravity oil to a level of 6 in (152 mm) measured vertically from the bottom of the stainless steel bubbler. p.8 ↗ | Correct oil level is required for proper pressure relief function. | All models |
| Single-width units require two line cords; double-width units require three line cords, each rated 100-115V 12A or 230V 12A. p.9 ↗ | Provide sufficient dedicated electrical circuits near the installation site. | Single-width vs double-width models |
| Interior electrical duplex is rated 10A maximum at 115V, 230V, or 100V; exterior vacuum pump outlet is rated 100-115V 12A or 230V 6A (8A for 100-115V, 4.5A for 230V). p.9 ↗ | Limit interior equipment load to 10A; select vacuum pump compatible with outlet rating. | All models; doublewide has additional interior duplex |
| Solenoid valves use 3/8 in (9.5 mm) compression fittings or 3/8 in (9.5 mm) hose barbs; rear purifier connections are two 3/4 in compression fittings. p.10 ↗ | Plumbing hardware must match these fitting sizes. | All models |
| Primary gas/vacuum tubing is 3/8 in O.D. x 3/16 in I.D.; a tube support insert must be placed inside soft tubing to prevent disconnection. p.17 ↗ | Use correct tubing and inserts; alternative is tube connector with hose clamp. | All models |
| Factory helium leak test: no individual leak greater than 1 x 10^-6 cc/sec at +5 inches helium. p.1 ↗ | Baseline integrity specification for acceptance and recertification. | All models |
| Oxygen permeation: stainless steel <0.16 ppm/min; fiberglass <0.30 ppm/min per ISO 10648-2 at -4 in w.g. p.1 ↗ | Both meet ISO 10648-2 Class 1 (<1.67 ppm/min); stainless offers lower permeation. | Stainless vs fiberglass liner |
| Attainable purity with installed purifier: fiberglass O2 0.4 ppm / H2O 2 ppm; stainless O2 0.2 ppm / H2O 1 ppm. p.2 ↗ | Select liner material based on required atmosphere purity. | Fiberglass vs stainless models |
| Pressure leak decay acceptance test at -1000 Pa (-4 in w.g.); operational test at -250 Pa (-1 in w.g.). p.23 ↗ | Use these setpoints for ISO 10648-2 pressure change method validation. | All models |
| A 0.1°C temperature fall equals -34 Pa (-0.14 in w.g.) pressure change; a 0.01 in Hg barometric rise equals -34 Pa (-0.14 in w.g.). p.23 ↗ | Correct leak test results for temperature and barometric pressure changes. | All models |
| Auto pressure controller automates main chamber evacuation/fill up to 499 cycles and transfer chamber up to 199 cycles. p.10 ↗ | Cycle limits define controller capability for purging routines. | Auto pressure control and combination models |
| Manual gauges: main chamber -5 to +5 in w.g.; transfer chamber 0 to -29.5 in Hg. Auto controller regulates same ranges. p.27 ↗ | Operating pressure ranges for manual vs auto control. | Manual vs auto pressure control models |
| Transfer chamber typically requires 1-5 evacuation/fill cycles to reduce O2/H2O below 1%; main chamber normally 75-100 cycles. p.33 ↗ | Plan purge cycles accordingly; porous materials may need longer sustained vacuum. | All models |
| AtmosPure Re-Gen Gas Purifier can be started once an inert atmosphere below 5000 ppm is established. p.32 ↗ | Purge the glove box before starting the purifier. | Models with AtmosPure Re-Gen Gas Purifier |
| Glove box weights: fiberglass 350 lb/157 kg; stainless 500 lb/225 kg; stainless doublewide 1000 lb/450 kg. p.11 ↗ | Work surface and lifting equipment must support these weights. | Fiberglass, stainless, stainless doublewide |
| Glove ports: two 8 in ID epoxy-coated aluminum, double-grooved, mounted 17 in apart; doublewide has four ports. p.28 ↗ | Glove sizing and port spacing for operator reach. | Single-width vs doublewide |
| Viewing window: 36 in wide x 27.6 in high, 3/8 in thick laminated safety glass; doublewide has two viewing areas. p.27 ↗ | Window replacement must use 3/8 in laminated safety glass. | All models |
| HEPA Filter Exhaust Kit part number 5441403 for filtered bubbler exhaust to room; 3/4 in I.D. neoprene hose kit for outside exhaust. p.8 ↗ | Order correct exhaust accessory based on routing. | All models |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Lights will not operate. p.58 ↗ | Unit not plugged into outlet. | Plug unit into appropriate electrical service. | Chapter 8 |
| Lights will not operate. p.58 ↗ | Circuit breaker(s) tripped. | Reset or replace circuit breaker. | Chapter 8 |
| Lights will not operate. p.58 ↗ | Lamp not installed properly. | Inspect lamp installation. | Chapter 8 |
| Lights will not operate. p.58 ↗ | Lamp wiring disconnected. | Inspect lamp wiring. | Chapter 8 |
| Lights will not operate. p.58 ↗ | Defective lamp. | Replace lamp. | Chapter 8 |
| Lights will not operate. p.58 ↗ | Light switch is defective. | Replace light switch. | Chapter 8 |
| Lights will not operate. p.58 ↗ | Defective electronic ballast. | Replace ballast. | Chapter 8 |
| No power to electrical receptacle. p.58 ↗ | Unit not plugged in. | Plug unit into electrical service. | Chapter 8 |
| No power to electrical receptacle. p.58 ↗ | Circuit breaker(s) tripped. | Reset or replace circuit breaker. | Chapter 8 |
| No power to electrical receptacle. p.58 ↗ | Receptacle switch is defective. | Replace receptacle switch. | Chapter 8 |
| No power to electrical receptacle. p.58 ↗ | Receptacle is defective. | Replace receptacle. | Chapter 8 |
| Auto Pressure Controller Solenoid Valves will not operate. p.58 ↗ | Plumbing is installed incorrectly. | Check plumbing and insure VAC is "out" to vacuum pump and GAS is "in" from inert gas supply. | Chapter 8 |
| Auto Pressure Controller Solenoid Valves will not operate. p.58 ↗ | Solenoid Valve is defective. | Replace solenoid valve. | Chapter 8 |
| Auto Pressure Controller will not operate. p.58 ↗ | Unit not plugged into outlet. | Plug unit into appropriate electrical service. | Chapter 8 |
| Auto Pressure Controller will not operate. p.58 ↗ | Circuit breaker(s) tripped. | Reset or replace circuit breaker. | Chapter 8 |
| Auto Pressure Controller will not operate. p.58 ↗ | LCD Display is defective. | Replace circuit board or power supply circuit board. | Chapter 8 |