Rev c capture portable fuming systems user manual

Technical Reference Summary

Covered Models & Document Scope

Covers the Labconco CApture Portable Fuming Systems, models 3420000 (115 VAC, 60 Hz, 1 Amp) and 3420020, 3420022, 3420024 (230 VAC, 50/60 Hz, 0.5 Amp). Optional accessories include the Exhaust Connection Kit (#3425400).

Section-by-Section Technical Breakdown

pp. 6-7
Prerequisites

Lists required supplies (CA glue, aluminum heating pans, Rain-X or Rain Block treatment) and gives electrical and exhaust requirements. Electrical data are tabulated per model, and notes cover circuit breaker ratings, power cord length, and inverter operation.

  • CA glue capacity of filter is approximately 2 oz (57 g); do not use thickening or gap-filling formulations.
  • Aluminum heating pans: 10 shipped; bag of 250 available as Labconco part 34252002.
  • Exhaust Connection Kit optional; do not add additional duct beyond that supplied.
  • Model 3420000: 115 VAC, 60 Hz, 1 Amp; models 34200-20, -22, -24: 230 VAC, 50/60 Hz, 0.5 Amp.
  • Circuit breaker: 10 A for 115 V models, 5 A for 230 V models; power cords 3 m (10 ft).
  • Inverter operation: 12 VDC to AC, output rating at least 100 W, protected against current overload.
pp. 8-10
Theory of Operation

Explains CA polymerization, controlled heating to generate fumes, the clear polycarbonate chamber, the two-stage filter system, and the blower. Includes temperature limits and filter efficiency data.

  • CA boiling point approximately 120°F (50°C); decomposition products above 250°F (121°C); hydrogen cyanide can be produced above 400°F (204°C).
  • Controller setpoint range: 120°F (50°C) to 350°F (177°C); redundant safety switch limits heater to 350°F (177°C).
  • Clear chamber must be treated with Rain-X or Rain Block; protection lasts up to 20 uses if undisturbed.
  • First filter stage: polyester loft particulate filter, MERV 11 (65-79% of 1.0-3.0 µm particles; >85% of 3.0-10 µm particles).
  • Second filter stage: granular activated carbon (GAC) adsorbs gaseous CA.
  • Filter life: minimum 100 fuming cycles (about 2 oz or 57 ml of CA).
pp. 13-22
Using the System

Covers unpacking, transporting, component identification, handling hints for CA, setup, heater adjustment, fuming procedure, and humidity adjustment. Provides step-by-step operating instructions and key parameter values.

  • System weight approximately 22 lbs (10 kg); transport only in transport mode with heater/filter module nested and transport lid locked.
  • Setup: release clamps, remove foam insert, lower leveling foot, treat chamber, select hooks/shelves, seal chamber, plug in and switch main power ON.
  • Heater adjustment: push control switch to Heat; setpoint range 120-350°F (50-177°C); temperature stabilizes within 3-5 minutes at setpoint ±15°F.
  • Fuming: use 4-6 drops of CA (never exceed 8 drops); typical development 3-5 minutes with 4 drops at 350°F (177°C).
  • After fuming, switch to Vent and let chamber vent at least 3-5 minutes before opening.
  • Humidity: optimum 70-90% RH, best at approximately 80%; one drop of water raises RH by approximately 2%.
pp. 23-25
Maintaining the System

Provides a routine maintenance schedule and service operations including filter replacement and circuit breaker reset. Specifies cleaning intervals and treatment reapplication.

  • Weekly: clean stainless steel face plate with soap/detergent or LabSolutions wipes (part #1570000); reapply Rain-X or Rain Block if chamber interior cleaned.
  • Monthly: clean exterior surfaces with damp cloth or LabSolutions wipes.
  • After every 20 development cycles: clean clear chamber with mild detergent or glass cleaner, then treat with Rain-X or Rain Block.
  • After every 100 development cycles: replace filter cartridge.
  • Filter replacement: unplug, separate module, loosen captive thumbscrew, remove panel, pull cartridge out of seal and flexible connector, install replacement, ensure seal not deformed.
  • Circuit breaker reset: press switch on rear of steel module to ON (I) position.
pp. 26
Troubleshooting

Presents a troubleshooting table with problems, causes, and corrective actions for common system failures. Covers power, print development, fume generation, heater, fan, and controller errors.

  • Blower and controller won't turn on: check outlet and circuit breaker.
  • Prints not developing properly: insufficient humidity; add 1-2 drops of water in a heating pan.
  • Little or no fumes: heater temperature too low or pan bottom dented/bent.
  • Heater temperature not changing: possible over-temperature switch failure or defective controller.
  • Fan does not run: fan or 12 V power supply failure.
  • Controller displays 'Er' or 'oo' or '-': controller error.
pp. 27
Components

Lists replacement parts and accessories with part numbers and quantities. Includes clear chamber, heating pans, flexible connector, filter cartridge, shelf, hook, and foam storage insert.

  • Clear chamber: part 3421901.
  • Heating pans, package of 250: part 3425202.
  • Flexible connector: part 3424500.
  • Filter cartridge: part 3421801.
  • Shelf: part 3422203; Hook: part 3422202.
  • Foam storage insert: part 3425500.
pp. 28
Dimensions

Provides the interior dimensions of the clear chamber.

  • Clear chamber interior: 16" (406 mm) H x 8" (203 mm) W x 8" (203 mm) D.
pp. 29
Specifications

Lists electrical data and environmental operating conditions for the system.

  • Electrical: 3420000 115 VAC, 60 Hz, 1 Amp, 75 W; 34200-20, -22, -24 230 VAC, 50/60 Hz, 0.5 Amp, 75 W.
  • Indoor use only; maximum altitude 6562 ft (2000 m).
  • Ambient temperature range: 41°F to 104°F (5°C to 40°C).
  • Maximum relative humidity: 80% up to 88°F (31°C), decreasing linearly to 50% at 104°F (40°C).
  • Main supply voltage fluctuations not to exceed ±10% of nominal.
  • Transient overvoltages per Installation Category II: up to 1500 V for 115 V models, 2500 V for 230 V models.
  • Pollution degree 2 environment.
pp. 30
Accessories

Lists optional and replacement accessories with part numbers, descriptions, and weights.

  • Clear Chamber (#3421901): replacement clear chamber with latches, 2 lb (0.9 kg).
  • Heating Pans (#3425202): package of 250 aluminum disposable pans, 1 lb (0.5 kg).
  • Filter Cartridge (#3421801): replacement particulate/GAC filter, lasts 100 cycles, 5 lbs (2.3 kg).
  • Exhaust Connection Kit (#3425400): allows remote exhaust or fume hood connection; includes 10 ft (1.2 m) extended, 4" (100 mm) diameter collapsible duct, 1 lb (0.5 kg).
  • Replacement Hook (#3422202): for hanging evidence, 1 lb (0.5 kg).
  • Replacement Shelf (#3422203): for supporting small evidence, 1 lb (0.5 kg).
  • Perforated Shelf Cover (#3426101): stainless steel cover fits over Shelf #3426103, 1 lb (0.5 kg).

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Model 3420000 requires 115 VAC, 60 Hz, 1 Amp; models 34200-20, -22, -24 require 230 VAC, 50/60 Hz, 0.5 Amp. p.7 ↗ Select model according to available mains voltage; use a circuit breaker rated 10 A for 115 V models or 5 A for 230 V models. whole series
Power cords supplied are 3 meters (10 feet) long. p.7 ↗ Locate the system within 3 m of a suitable outlet; use an extension cord only if adequately rated. whole series
The system can be operated on an inverter converting 12 VDC to AC, with output rating at least 100 W and current overload protection. p.7 ↗ Enables mobile operation from a vehicle 12 V supply; ensure inverter meets minimum rating. whole series
CA boiling point is approximately 120°F (50°C); decomposition products appear above 250°F (121°C); hydrogen cyanide can be produced above 400°F (204°C). p.9 ↗ Never exceed 350°F (177°C) setpoint; the redundant safety switch limits heater to 350°F (177°C). whole series
Controller setpoint range is 120°F (50°C) to 350°F (177°C). p.9 ↗ Program setpoint within this range; temperatures stabilize within 3-5 minutes at setpoint ±15°F. whole series
Clear chamber must be treated with Rain-X or Rain Block; protection lasts up to 20 uses if undisturbed. p.9 ↗ Reapply treatment after every 20 development cycles or whenever chamber interior is cleaned. whole series
First filter stage is MERV 11, removing 65-79% of 1.0-3.0 µm particles and >85% of 3.0-10 µm particles. p.9 ↗ Provides particulate filtration; replace filter cartridge after 100 development cycles or when odor is detected. whole series
Filter life is a minimum of 100 fuming cycles using about 2 oz (57 ml) of CA. p.9 ↗ Plan filter replacement after 100 cycles; part number 3421801. whole series
Never exceed 8 drops of CA per development cycle; typical procedure uses 4-6 drops. p.11 ↗ Exceeding 8 drops may overload filter and produce excessive fumes; use 4 drops for typical 3-5 minute development at 350°F (177°C). whole series
Optimum relative humidity for fingerprint development is 70-90%, best at approximately 80%; one drop of water raises RH by approximately 2%. p.20 ↗ Use the humidity chart to add water drops: e.g., to raise RH by 30%, add 15 drops. whole series
Typical development time is 3-5 minutes with 4 drops of CA at 350°F (177°C). p.21 ↗ Monitor development and switch to Vent when prints are developed as desired. whole series
After fuming, let chamber vent for at least 3-5 minutes before opening. p.22 ↗ Ensures CA fumes are removed before retrieving evidence. whole series
System weight is approximately 22 lbs (10 kg). p.11 ↗ Follow safe-lifting guidelines when transporting; use transport mode with heater/filter module nested and transport lid locked. whole series
Clear chamber interior dimensions: 16" (406 mm) H x 8" (203 mm) W x 8" (203 mm) D. p.28 ↗ Limits evidence size; ensure evidence fits within chamber. whole series
Environmental limits: indoor use, max altitude 6562 ft (2000 m), ambient 41-104°F (5-40°C), max RH 80% up to 88°F (31°C) decreasing to 50% at 104°F (40°C). p.29 ↗ Install in a location meeting these conditions; voltage fluctuations not to exceed ±10%. whole series
Exhaust Connection Kit (#3425400) includes 10 ft (1.2 m) extended, 4" (100 mm) diameter collapsible duct; do not add additional duct. p.6 ↗ Additional duct reduces exhaust flow and may cause incomplete venting. whole series
Heating pans should be discarded after the second use; do not use pans with dented or warped bottoms. p.16 ↗ Residue buildup affects heat transfer; damaged pans heat CA inefficiently. whole series
Store opened CA glue in a freezer or airtight container with silica gel to prevent setting. p.16 ↗ Extends usable life of CA by limiting moisture exposure. whole series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Blower and controller won’t turn on p.26 ↗ Unit not plugged into outlet Plug the cabinet into appropriate electrical service.
Blower and controller won’t turn on p.26 ↗ Circuit breaker tripped Reset circuit breaker.
Prints are not developing properly p.26 ↗ Insufficient humidity in the clear chamber Place 1-2 drops of water in a heating pan, and place on the heater.
Little or no fumes are being produced p.26 ↗ Heater temperature is too low. Increase heater temperature
Little or no fumes are being produced p.26 ↗ Bottom of heater pan is dented/bent Discard damaged pan, and use one with flat bottom.
Heater temperature not changing on controller (“out” indicator is on) p.26 ↗ Heater /Over temperature switch failure Contact Labconco Product Service Department.
Heater temperature not changing on controller (out indicator is off) p.26 ↗ Controller is defective Contact Labconco Product Service Department.
Fan does not run p.26 ↗ Fan or 12 volt power supply failure Contact Labconco Product Service Department.
Controller displays “Er” or “oo” or “-” p.26 ↗ Controller error Contact Labconco Product Service Department.