Rev f purifier vertical clean benches
Technical Reference Summary
Covered Models & Document Scope
Labconco Purifier® Vertical Clean Benches in 2', 3', 4' and 6' widths, catalog numbers 3970200/3970201/3970203/3970204, 3970220/3970221/3970223/3970224, 3970300/3970301/3970303/3970304, 3970320/3970321/3970323/3970324, 3970400/3970401/3970403/3970404, 3970420/3970421/3970423/3970424, 3660600/3660601/3660603/3660604, 3660620/3660621/3660623/3660624. Variants include 115 V and 230 V electrical configurations, models with or without internal UV light and UV sash closure, and models with or without the optional Guardian™ Airflow Monitor.
Section-by-Section Technical Breakdown
Defines ISO Class 5 cleanliness as equivalent to Federal Standard 209E Class 100 and gives the maximum permitted particle concentrations per ISO 14644-1. The table lists concentration limits for particle sizes from 0.1 µm to 5 µm for ISO Classes 1 through 9.
- ISO Class 5 is equivalent to Class 100 (100 particles 0.5 µm or larger per cubic foot).
- ISO Class 5 limit is 3 520 particles/m³ for particles ≥0.5 µm.
- ISO Class 5 limit is 100 000 particles/m³ for particles ≥0.1 µm and 23 700 particles/m³ for ≥0.2 µm.
- ISO Class 5 limit is 10 200 particles/m³ for ≥0.3 µm, 832 particles/m³ for ≥1 µm and 29 particles/m³ for ≥5 µm.
Specifies that the supporting structure must be level and of solid construction, typically a base cabinet or base stand with a chemically-resistant work surface. The work surface must be at least as wide and deep as the enclosure and able to support the combined weight of the clean bench and related equipment.
- Supporting structure usually consists of a base cabinet or base stand and chemically-resistant work surface.
- Area must be level and of solid construction.
- Work surface should be at least as wide and deep as the enclosure.
- Front of the clean bench should be aligned within 0.30" (8 mm) of the front of the work surface.
Recommends placing the clean bench away from high-traffic areas, HVAC diffusers, fans, radiators, doorways and windows that may be opened, to avoid disrupting airflow patterns. The main disconnect device must remain easily operable.
- Avoid high-traffic areas where walking might cause air disturbance.
- Avoid overhead or wall HVAC diffusers, fans, radiators or other equipment producing air currents.
- Avoid doorways or windows that may be opened.
- Do not position the unit so that it is difficult to operate the main disconnect device.
Provides downflow velocity, out-flow exhaust volume, noise pressure and average outflow face velocity for each enclosure width at 30, 40, 50 and 60 fpm settings. The enclosure is factory set to deliver 40–55 fpm downflow velocity measured 6 inches below the upper diffuser per IEST-RP-CC002.
- 2' model: 30 fpm gives 120 CFM, 53–57 dB(A), 85 fpm outflow face velocity; 60 fpm gives 250 CFM, 62–63 dB(A), 170 fpm.
- 3' model: 30 fpm gives 180 CFM, 58–60 dB(A), 85 fpm; 60 fpm gives 370 CFM, 66–70 dB(A), 170 fpm.
- 4' model: 30 fpm gives 240 CFM, 58–62 dB(A), 85 fpm; 60 fpm gives 500 CFM, 66–70 dB(A), 170 fpm.
- 6' model: 30 fpm gives 360 CFM, 61–65 dB(A), 85 fpm; 60 fpm gives 740 CFM, 69–73 dB(A), 170 fpm.
Requires standard duplex electrical receptacles nearby for connecting the clean bench or other equipment. Enclosures include iris pass-throughs for routing electrical cords through the back without leaving a large hole.
- Standard duplex electrical receptacles should be nearby.
- Iris pass-throughs allow electrical cords through the back of the enclosure.
- 115 V models connect via supplied power cord to IEC inlet on back of unit.
- 230 V models are shipped without a plug; install plug per local codes.
Describes leveling the base cabinet and work surface, securing the work surface, and mounting the clean bench using four supplied screws with at least 1/8" clearance under the screw heads. The work surface should be smooth, rigid, non-porous, chemically resistant and have a dished recessed area for spills.
- Level the base cabinet or stand and the work surface.
- Secure work surface to base cabinet or stand with structural adhesive or silicone sealant.
- Insert four mounting screws allowing minimum 1/8" clearance under the head.
- Slide rear flange and front air foil flanges under screw heads and tighten four screws.
The HEPA filter is shipped installed with the gasket on the downstream side and is leak checked at Labconco. A second leak check is recommended before use and at least annually thereafter. All seams downstream of the HEPA filter are jacketed by positive pressure.
- HEPA filter is shipped installed with gasket on downstream side.
- Second leak check recommended before using the enclosure and at least annually thereafter.
- All seams downstream of the HEPA filter are jacketed by positive pressure.
- HEPA filter part numbers: 2' = 3707900 (qty 1), 3' = 3707901 (qty 1), 4' = 3707902 (qty 1), 6' = 3707901 (qty 2).
115 V models use the supplied power cord connected to the IEC inlet on the back of the clean bench; at 50 Hz operation maximum airflow is reduced by 17%. 230 V models are shipped without a plug and require an appropriate plug installed per local codes.
- 115 V models: connect supplied power cord to IEC electrical supply plug on back.
- At 50 Hz operation, blower maximum airflow is reduced by 17%.
- 230 V models: shipped without a plug; install appropriate plug per local codes.
- Do not use any detachable power cord not adequately rated for the unit.
The speed control behind the front panel adjusts downflow velocity, which should be 30–60 fpm for ISO Class 5 conditions. The unit is factory set at 40–55 fpm; measurements are taken 6 inches below the interior upper diffuser using an anemometer.
- Downflow velocity should be 30–60 fpm for Class 100 (ISO Class 5) conditions.
- Factory setting is 40–55 fpm downflow velocity.
- Measure downflow velocity 6" below the interior upper diffuser per IEST-RP-CC002.
- Turn speed control counterclockwise to increase and clockwise to decrease downflow velocity.
Defines the downflow test grid for each model width, with readings taken 6 inches below the top of the interior upper diffuser. A certified technician should complete validation; the clean benches are tested at Labconco before shipment.
- 2' model: 9 test points (3 points in 3 rows); back row 6" from rear; rows 6.57" apart; side columns 6" from sidewalls; center column 5.32" from first/last column.
- 3' model: 12 test points (4 points in 3 rows); two inner columns spaced 7.54" from first and last column.
- 4' model: 15 test points (5 points in 3 rows); three inner columns spaced 8.66" from first and last column.
- 6' model: 21 test points (7 points in 3 rows); five inner columns spaced 9.77" from each other and from first/last column.
After the clean bench is set in place and wired, a bead of silicone sealant is recommended to seal it to the work surface and prevent materials from collecting under the walls.
- Seal after the clean bench is set in place and wired.
- Use a bead of silicone sealant between clean bench and work surface.
- Sealing prevents materials from collecting under the walls.
Describes the components of the clean bench including aerodynamic lower air foil, 10-degree ergonomic slope, 25-inch internal depth, safety glass sash with spring-loaded latch, utility ports with iris, optional Guardian™ Airflow Monitor, inherently safe impeller, speed control, fluorescent lamp, filter pressure gauge, HEPA filter, UV lamp and closure panel, upper diffuser screen and prefilter.
- Ergonomic slope of 10 degrees provides maximum visibility and reduces glare.
- Internal depth of 25 inches.
- Safety glass sash with spring-loaded latch must be down for normal operation.
- Prefilter (roughing) should be replaced quarterly.
- HEPA filter rated to remove 99.99% of particles 0.3 micron in size.
- Optional Guardian™ Airflow Monitor available on all models.
Covers planning, start-up, loading, work techniques, final purging and unloading. The blower must operate unobstructed for 1 minute before work and 2–3 minutes after work to purge airborne contaminants.
- Turn on blower and light; turn off UV light and remove UV closure panel if equipped.
- Allow blower to operate unobstructed for 1 minute before work.
- After loading, wait one minute to purge airborne contaminants.
- After work, allow 2–3 minutes undisturbed operation before shutting down blower.
- Wipe interior surfaces with mild household detergent or disinfectant; do not use abrasive cleaners, bleach or solvents.
Suitable applications are based on IEST guidelines and include procedures traditionally performed on an open bench where an ISO Class 5 environment is required. The clean bench provides product protection only and must not be used with biohazardous or hazardous solids.
- Use only for operations requiring sample protection from environmental contamination.
- Provides product protection only; no personnel protection.
- Biohazardous materials or other hazardous solids cannot be used.
- HEPA filters are only effective for entrapment of particulate matter.
Defines weekly, monthly and annual maintenance tasks including cleaning, airflow checks, downflow velocity measurement, UV lamp replacement, certification and prefilter replacement.
- Weekly: wipe interior with disinfectant or cleaner; clean exterior with damp cloth; check airflow using visible smoke.
- Monthly: determine actual downflow velocity 6" below upper diffuser; increase speed control or change HEPA filter if velocity drops below recommended.
- Annually: replace UV lamps if equipped; verify UV closure panel switch; have enclosure validated by qualified certification technician; change prefilter.
HEPA filters typically last two to five years depending on room air cleanliness and operating time. Replace when downflow velocity drops below 30–60 fpm with speed control at full speed, or if the filter fails the HEPA Filter Leak Test.
- Typical HEPA filter life is two to five years.
- Replace when downflow velocity drops below recommended 30–60 fpm and speed control is at full speed.
- Replace if HEPA filter fails the HEPA Filter Leak Test.
Only a qualified certifier should service the HEPA filter, and the enclosure must be certified after replacement. The procedure involves unplugging, removing prefilter and front panel, loosening filter clamp bolts with a 9/16" deep socket, and installing the new filter with correct orientation.
- Unplug enclosure and remove prefilter on top.
- Remove front panel by loosening two screws.
- Use 9/16" deep socket to loosen filter clamp bolts on top.
- Cover new HEPA filter gasket with light coating of silicone grease if desired.
- Enclosure MUST be certified after HEPA filter replacement.
Describes the procedure for leak testing the HEPA filter installation, referenced from Chapter 6. A second leak check is recommended before use and at least annually.
- Leak test recommended before using enclosure and at least annually thereafter.
- Consult Safety Officer and Chapter 6 for the HEPA Filter Leak Test procedure.
- All seams downstream of HEPA filter are jacketed by positive pressure.
Covers calibration and operation of the optional Guardian™ Airflow Monitor, including alarm set point adjustment and silencing. The monitor provides audio/visual alarm for low airflow conditions.
- Guardian™ Airflow Monitor continuously monitors airflow.
- Audio/visual alarm alerts user to low airflow conditions.
- Alarm can be temporarily silenced using SILENCE ALARM.
- Calibration steps outlined in Chapter 6.
A qualified certification technician should test the clean bench before initial use. The clean bench should be validated annually and whenever relocated.
- Qualified certification technician should test before initial use.
- Validate annually and whenever relocated.
- Labconco performed extensive performance testing to validate clean benches for product protection under ISO Class 5 conditions.
Covers replacement of the fluorescent lamp located above the work area behind the control panel. The lamp ballast is located behind the control panel.
- Fluorescent lamp located above work area, out of contact with clean air.
- Safety glass window beneath lamp distributes light evenly.
- Lamp ballast located behind control panel.
Covers replacement of the optional UV lamp used for surface disinfection when the clean bench is not in use. The UV closure panel must be in place for the UV lamp to work.
- UV lamp found only on some models.
- UV light rays will not penetrate the plane of the sash and lower air foil.
- UV closure panel must be in place for UV lamp to work.
- Replace UV lamps annually if equipped.
Covers replacement of the motorized impeller/blower. The impeller has a negative pressure plenum surrounding the positive pressure impeller so that leaks are captured and refiltered.
- Inherently safe impeller has negative pressure plenum surrounding positive pressure impeller.
- If leak occurs, unfiltered air is captured and refiltered.
- Replace motorized impeller or blower if defective; see Chapter 6.
Covers replacement of the speed control located behind the front panel, used to adjust downflow velocity.
- Speed control located behind front panel.
- Speed control regulates speed of motorized impeller.
- Used by certifier to validate and adjust downflow velocity.
Covers acceptable modifications and accessories for the clean bench, including the optional Guardian™ Airflow Monitor, UV light and UV sash closure, and accessory work surfaces.
- Guardian™ Airflow Monitor available as option on all models.
- UV light and UV sash closure available as option.
- Accessory work surface is dished and contoured to contain spills.
- Iris plugs included with User's Manual for utility ports.
Provides a table of problems, probable causes and corrective actions for issues including improper ISO Class 5 conditions, blower failure, light failure, airflow monitor malfunction and alarm issues.
- ISO Class 5 air conditions not present: check user techniques, external airflow disruptions, downflow velocity.
- Blower won't operate: check power, circuit breaker, wiring, switch, motorized impeller.
- Light does not work: check lamp installation, wiring, lamp, switch, ballast, UV closure panel, magnetic switch, UV timer circuit board.
- Airflow monitor malfunction: check power, cables, fuses, alarm set point, calibration, airway blockage.
Appendix A contains labeled diagrams of all components and replacement parts for the clean benches, including HEPA filter part numbers by model size.
- HEPA filter part numbers: 2' = 3707900, 3' = 3707901, 4' = 3707902, 6' = 3707901 (qty 2).
- Labeled diagrams of all components.
- Replacement parts listed in Appendix A.
Appendix B contains comprehensive diagrams showing all dimensions for the clean benches.
- Comprehensive dimension diagrams for all models.
- Refer to Appendix B for complete Vertical Clean Bench dimensions.
Appendix C contains electrical requirements, environmental conditions, specifications and wiring diagrams for the clean benches.
- Electrical requirements for clean benches.
- Wiring diagrams included.
- Environmental conditions and specifications.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Downflow velocity should be 30–60 fpm for ISO Class 5 conditions; factory setting is 40–55 fpm. p.17 ↗ | Adjust speed control to maintain velocity within this range for proper ISO Class 5 operation. | All models |
| Downflow velocity measurements must be taken 6 inches below the interior upper diffuser per IEST-RP-CC002. p.17 ↗ | Use an anemometer at the specified distance for accurate velocity readings during validation and maintenance. | All models |
| HEPA filter is rated at least 99.99% efficient for 0.3-micron particles. p.5 ↗ | Ensures ISO Class 5 air conditions inside the enclosure for product protection. | All models |
| HEPA filter part numbers: 2' model = 3707900 (qty 1), 3' model = 3707901 (qty 1), 4' model = 3707902 (qty 1), 6' model = 3707901 (qty 2). p.16 ↗ | Order correct filter part number and quantity for replacement. | 2', 3', 4', 6' models |
| At 50 Hz operation, blower maximum airflow is reduced by 17%. p.16 ↗ | Account for reduced airflow when operating 115 V models at 50 Hz; may require speed control adjustment. | 115 V models |
| 230 V models are shipped without a plug; install appropriate plug per local codes. p.16 ↗ | Requires field installation of plug matching local electrical specifications. | 230 V models |
| Each clean bench model weighs between 125 to 305 lbs (55 to 140 kg). p.12 ↗ | Ensure supporting structure can safely support combined weight of clean bench and related equipment. | All models |
| Front of clean bench should be aligned within 0.30" (8 mm) of the front of the work surface. p.13 ↗ | Proper alignment ensures stable installation and airflow. | All models |
| Mounting screws require minimum 1/8" clearance under the head for positioning the clean bench. p.14 ↗ | Allows sliding rear flange and front air foil flanges under screw heads during installation. | All models |
| Prefilter (roughing) should be replaced quarterly. p.24 ↗ | Regular prefilter replacement enhances HEPA filter life by trapping large particles. | All models |
| HEPA filters typically last two to five years with normal usage before requiring replacement. p.28 ↗ | Plan for HEPA filter replacement based on room air cleanliness and operating time. | All models |
| Replace HEPA filter when downflow velocity drops below 30–60 fpm with speed control at full speed, or if filter fails leak test. p.30 ↗ | Indicates filter loading; replacement restores proper airflow. | All models |
| Use a 9/16" deep socket to loosen filter clamp bolts when replacing HEPA filter. p.31 ↗ | Required tool for HEPA filter replacement procedure. | All models |
| Enclosure MUST be certified after HEPA filter replacement. p.31 ↗ | Only a qualified certifier should service the HEPA filter; certification ensures proper operation. | All models |
| Downflow test grid for 2' model: 9 test points (3 points in 3 rows); back row 6" from rear; rows 6.57" apart; side columns 6" from sidewalls; center column 5.32" from first/last column. p.17 ↗ | Use specified grid for accurate average downflow velocity validation. | 2' model |
| Downflow test grid for 3' model: 12 test points (4 points in 3 rows); two inner columns spaced 7.54" from first and last column. p.17 ↗ | Use specified grid for accurate average downflow velocity validation. | 3' model |
| Downflow test grid for 4' model: 15 test points (5 points in 3 rows); three inner columns spaced 8.66" from first and last column. p.17 ↗ | Use specified grid for accurate average downflow velocity validation. | 4' model |
| Downflow test grid for 6' model: 21 test points (7 points in 3 rows); five inner columns spaced 9.77" from each other and from first/last column. p.17 ↗ | Use specified grid for accurate average downflow velocity validation. | 6' model |
| Internal depth of 25 inches and ergonomic slope of 10 degrees. p.23 ↗ | Provides necessary workspace depth and reduces operator fatigue. | All models |
| Allow blower to operate unobstructed for 1 minute before work and 2–3 minutes after work to purge airborne contaminants. p.26 ↗ | Ensures ISO Class 5 conditions are established before use and contaminants are purged before shutdown. | All models |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| ISO Class 5 air conditions are not present inside the enclosure. p.46 ↗ | Improper user techniques for the enclosure. | See "Certifying the Enclosure" Chapter 3 and "Safety Precautions" Chapter 4 sections in the manual. (Ref. Appendix D). | Chapter 8 |
| ISO Class 5 air conditions are not present inside the enclosure. p.46 ↗ | External factors are disrupting the enclosure airflow patterns. | See "Location Requirements" Chapter 2, "Certifying the Enclosure" Chapter 3, and "Safety Precautions" Chapter 4 sections of this manual. (Ref. Appendix D). | Chapter 8 |
| ISO Class 5 air conditions are not present inside the enclosure. p.46 ↗ | Enclosure has improper downflow velocity. | Have enclosure certified and check exhaust system. Check HEPA filters for loading. Adjust enclosure speed control. Enclosure should have an average downflow velocity of 30-60 fpm. | Chapter 8 |
| Blower won’t operate. p.47 ↗ | Unit not plugged into outlet. | Plug the enclosure into appropriate electrical service. | Chapter 8 |
| Blower won’t operate. p.47 ↗ | Circuit breaker(s) or ground fault Interrupter. | Reset circuit breaker. | Chapter 8 |
| Blower won’t operate. p.47 ↗ | Blower wiring is disconnected. | Inspect blower wiring. | Chapter 8 |
| Blower won’t operate. p.47 ↗ | Blower switch is defective. | Replace switch. | Chapter 8 |
| Blower won’t operate. p.47 ↗ | Motorized impeller or blower is defective. | Replace motorized impeller or blower. See Chapter 6. | Chapter 8 |
| Poor internal ISO Class 5 air conditions p.47 ↗ | Enclosure sash not closed. | Close sash. | Chapter 8 |
| Poor internal ISO Class 5 air conditions p.47 ↗ | HEPA filter clogged. | Replace HEPA filter or increase speed. | Chapter 8 |
| Blower and lights won’t operate. p.47 ↗ | Unit not plugged into outlet. | Plug enclosure into appropriate electrical service. | Chapter 8 |
| Blower and lights won’t operate. p.47 ↗ | Circuit breaker(s) tripped. | Reset or replace circuit breaker. | Chapter 8 |
| Light does not work. p.47 ↗ | Lamp not installed properly. | Inspect lamp installation. | Chapter 8 |
| Light does not work. p.47 ↗ | Lamp wiring disconnected. | Inspect lamp wiring. | Chapter 8 |
| Light does not work. p.47 ↗ | Defective lamp. | Replace lamp. | Chapter 8 |
| Light does not work. p.47 ↗ | Light switch is defective. | Replace light switch. | Chapter 8 |
| Light does not work. p.47 ↗ | Defective electronic ballast. | Replace ballast. | Chapter 8 |
| Light does not work. p.47 ↗ | UV closure panel not installed. | Install UV closure panel (UV lights only). | Chapter 8 |
| Light does not work. p.47 ↗ | Magnetic switch for UV closure panel defective. | Replace switch (UV lights only). | Chapter 8 |
| Light does not work. p.47 ↗ | UV timer circuit board defective. | Replace circuit board (UV lights only). | Chapter 8 |
| Airflow monitor malfunction. No power. No lights. No display. p.47 ↗ | Power cable to circuit board is disconnected and needs to be connected. Sensor cable needs to be connected. Verify that all airflow monitor interface cables are connected. Check fuses on enclosure. | Connect power cable and sensor cable; verify all airflow monitor interface cables are connected; check fuses on enclosure. | Chapter 8 |
| No audible alarm. p.47 ↗ | Alarm has been temporarily silenced using “SILENCE ALARM.” | Press SILENCE ALARM again or wait for alarm to reset. | Chapter 8 |
| Airflow monitor malfunction (Continued) Wrong alarm set point. p.48 ↗ | Airflow monitor was not properly adjusted. | Repeat calibration steps outlined in this manual in Chapter 6. | Chapter 8 |
| Constant audible alarm. p.48 ↗ | Airflow monitor calibration or airflow issue. | Check airflow and calibration of airflow monitor. See Chapter 6. | Chapter 8 |
| Continuous alarm. p.48 ↗ | Downflow velocity of the enclosure may have changed; HEPA filter may have become loaded. | Check downflow velocity; if incorrect, adjust speed control to increase downflow velocity. If downflow velocity is correct, calibrate the airflow monitor outlined in Chapter 6. | Chapter 8 |
| Monitor alarms; air way to airflow monitor sensor is blocked by insects, dust or debris. p.48 ↗ | Airway to airflow monitor sensor blocked. | Lightly clean the airway with clean air. Be careful not to touch sensitive electrical components. | Chapter 8 |
| Audible disable will not stay operational. p.48 ↗ | An alarm condition must be continuously present before the audible alarm can be silenced. If flow conditions fluctuate near the alarm set point, the airflow monitor will automatically reset it. | Bring the enclosure airflow into proper operating parameters or adjust the alarm set point lower. | Chapter 8 |