Technical Reference Summary

Covered Models & Document Scope

This manual covers Labconco Fume Adsorbers in 30" and 60" widths with 115V or 230V electrical configurations, and three carbon filter types: Organic Vapor (models 3955200/3955220 for 30", 3955400/3955420 for 60"), Formaldehyde (3955201/3955221 for 30", 3955401/3955421 for 60"), and Ammonia (3955202/3955222 for 30", 3955402/3955422 for 60"). Units ship with organic vapor carbon filter(s); 30" units have one filter, 60" units have two.

Section-by-Section Technical Breakdown

pp. 5-6
Site Prerequisites

Before installation, the site must provide adequate support, electrical service, and space, and must avoid air disturbances. The supporting structure is typically a laboratory work surface. Electrical receptacles should be nearby, and the nameplate on the rear of the enclosure gives rated electrical service.

  • Supporting structure usually consists of a work surface in the laboratory.
  • Avoid high-traffic areas, HVAC diffusers, fans, radiators, and doorways/windows that may be opened.
  • Do not position the unit so that the main disconnect device is difficult to operate.
  • Standard duplex electrical receptacles should be nearby; see nameplate for rated electrical service.
pp. 8-9
Installation and Setup

The enclosure must be placed on a chemical-resistant work surface that can safely support its weight. Carbon filters are installed with the gasket side on the downstream exhaust side. Electrical connection uses an IEC inlet; 115V models ship with a cord, 230V models ship without a plug. Face velocity must be validated after installation.

  • 30" wide Fume Adsorber weighs 62 lbs; 60" wide Fume Adsorber-2 weighs 125 lbs.
  • Work surface should be at least as wide as the enclosure.
  • Remove four (or eight) Phillips screws on the perforated baffle to access filters.
  • All carbon filters are installed with the gasket side on the downstream exhaust side.
  • 115V models: connect supplied power cord to IEC inlet. 230V models: install appropriate plug per local codes.
  • Average face velocity should be 35 fpm; do not use enclosure below 25 fpm.
pp. 10-11
Performance Features

The Fume Adsorber uses a built-in motorized impeller to draw air through a carbon filter and exhaust filtered air back into the room. Airflow capacity is 100 CFM for 30" units and 200 CFM for 60" units. A perforated baffle provides uniform airflow and access to the carbon filter.

  • Built-in motorized impeller provides up to 35 fpm average face velocity.
  • Airflow capacity: 100 CFM on 30" wide unit, 200 CFM on 60" wide unit.
  • Perforated baffle provides uniform removal of noxious fumes and must be removed to access the carbon filter.
  • Blower switch provides ON/OFF control.
  • Carbon filter removes small amounts of fumes and chemicals.
pp. 14-15, 13
Routine Operation

Daily work procedures include starting the blower, checking baffle slots, wearing personal protective equipment, and working as far back in the enclosure as possible. After work, allow the blower to run for two to three minutes to purge airborne contaminants before shutdown.

  • Turn on exhaust blower with ON/OFF switch.
  • Check baffle air slots for obstructions.
  • Keep all materials inside the enclosure and perform contaminated operations as far to the rear as possible.
  • After work, allow enclosure to operate for two to three minutes undisturbed to purge airborne contaminants.
  • Leave blower on for at least one minute after work; if a chemical is spilled, do not switch off blower until all traces are removed.
pp. 2, 12, 16-18
Chemical Compatibility and Limits

The enclosure is intended only for chemicals with a TWA of 100 ppm or greater, unless an impregnated specialized filter is used. Acids, perchloric acid, and high-level radioisotopes are prohibited. Inlet concentration must never exceed IDLH, and the exhaust must not exceed the TWA.

  • Only chemicals with TWA of 100 ppm or greater are suitable, unless using impregnated filters for formaldehyde, formalin, ammonia, or amines.
  • General acid use and perchloric acid use are prohibited.
  • High-level radioisotope materials are prohibited.
  • Inlet concentration must never exceed IDLH; exhaust must not exceed TWA.
  • Formaldehyde filters can adsorb up to 30% organic vapors; ammonia filters can adsorb up to 30% organic vapors.
  • Carbon filters do not provide particulate protection.
pp. 19-22
Maintenance and Filter Replacement

Routine maintenance includes weekly cleaning and airflow checks, and monthly face velocity measurement. Carbon filters must be replaced when breakthrough is indicated by odor, time, detector tube, or analytical instrumentation. Filter life can be estimated using Labconco's modeling program, conservatively at 10% of published Paramount values.

  • Weekly: wipe interior/exterior with mild soap; operate exhaust and note airflow using visible smoke.
  • Monthly: measure face velocity; average should be 35±10 fpm using calibrated thermal anemometer.
  • Test carbon filters with gas detector tubes at intervals of 20% of estimated filter life; for formaldehyde or carcinogens, every 10%.
  • Replace carbon filters when outlet concentration approaches inlet concentration or odor/TWA is exceeded.
  • Filter replacement: remove rear perforated baffle, remove upper support bracket, loosen lower bracket, replace filter with gasket on downstream side.
  • For 60" wide units, there are two carbon filters.
pp. 30
Electrical and Environmental Specifications

Electrical specifications are 2 Amps at 115V, 50/60 Hz, and 1 Amp at 230V, 50/60 Hz. Environmental conditions include indoor use only, maximum altitude 6562 feet (2000 meters), ambient temperature 41° to 104°F (5° to 40°C), and maximum relative humidity 80% for temperatures up to 88°F (31°C), decreasing linearly to 50% at higher temperatures.

  • 115V models: 2 Amps, 50/60 Hz.
  • 230V models: 1 Amp, 50/60 Hz.
  • Indoor use only.
  • Maximum altitude: 6562 feet (2000 meters).
  • Ambient temperature range: 41° to 104°F (5° to 40°C).
  • Maximum relative humidity: 80% for temperatures up to 88°F (31°C), decreasing linearly to 50% relative humidity at higher temperatures.
pp. 26-28
Replacement Parts

Appendix A lists commonly requested replacement parts including canopy hoods, perforated baffles, filter clamps, motorized impellers, switches, circuit breakers, power cords, and carbon filters. Part numbers are provided for 30" and 60" widths and for 115V and 230V variants.

  • Canopy Hood, Plastic 30": 6961000; 60": 6961001.
  • Baffle, 30" Perforated: 6961500; 60": 6961501.
  • Motorized Impeller, 115V: 6962000; 230V: 6962001.
  • Switch, Rocker-DPST: 1307000.
  • Circuit Breaker 2 Amp (115V): 1327205; 1 Amp (115V): 1327206.
  • Power Cord, 115V: 1334500; 230V: 1334100.
  • Carbon filters: Organic Vapor 6961300, Formaldehyde 6961301, Ammonia 6961302.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
30" wide Fume Adsorber weighs 62 lbs; 60" wide Fume Adsorber-2 weighs 125 lbs. p.8 ↗ Work surface must be able to safely support these weights; use caution when lifting or moving. 30" and 60" models
Average face velocity should be 35 fpm; do not use enclosure with average velocity less than 25 fpm. p.9 ↗ Validate face velocity after installation and monthly; consult safety officer if below 25 fpm. all models
Airflow capacity is 100 CFM on 30" wide unit and 200 CFM on 60" wide unit. p.11 ↗ Verify airflow matches model; insufficient airflow may indicate filter loading or blower issue. 30" and 60" models
115V models: maximum circuit load 2 amps; 230V models: maximum circuit load 1 amp. p.9 ↗ Ensure electrical supply and circuit breaker are rated accordingly. 115V and 230V models
Carbon filters must be installed with gasket side on the downstream exhaust side. p.8 ↗ Incorrect orientation may cause leakage and reduced adsorption efficiency. all carbon filter types
Formaldehyde filters can adsorb up to 30% organic vapors; ammonia filters can adsorb up to 30% organic vapors. p.2 ↗ These filters provide limited organic vapor capacity; not a substitute for organic vapor filters for high organic loads. Formaldehyde and Ammonia filter models
Only chemicals with a TWA of 100 ppm or greater are suitable, unless using impregnated specialized filters. p.1 ↗ Do not use chemicals with TWA below 100 ppm; consult Labconco for chemical assessment. all models
Inlet concentration must never exceed IDLH; exhaust concentration must not exceed TWA. p.8 ↗ Monitor both inlet and exhaust concentrations; use detector tubes or analytical instrumentation. all models
Test carbon filters with gas detector tubes at intervals of 20% of estimated filter life; for formaldehyde or carcinogens, every 10%. p.19 ↗ More frequent testing required for hazardous chemicals to prevent breakthrough. all models
There is a 5/16" diameter hole on the upper right side for inserting the detector tube. p.19 ↗ Use this port for sampling exhaust concentration without disassembly. all models
Filter life for Fume Adsorbers with integral impeller is conservatively calculated at 10% of published Paramount values. p.22 ↗ Use this factor when estimating filter life from chemical guide data. all models
For formaldehyde, use 10% of impregnated carbon weight; 7.5 lbs Formasorb carbon adsorbs 0.75 lbs (340 g) formaldehyde, equivalent to 851 ml of 37% formalin. p.23 ↗ Provides capacity limit for formaldehyde filter; replace when this capacity is reached. Formaldehyde filter 6961301
For ammonia, use 10% of impregnated carbon weight; 8.75 lbs Ammonasorb II carbon adsorbs 0.88 lbs (400 g) ammonia, equivalent to 800 ml of 50% ammonia solution. p.23 ↗ Provides capacity limit for ammonia filter; replace when this capacity is reached. Ammonia filter 6961302
Environmental conditions: indoor use only, maximum altitude 6562 feet (2000 meters), ambient temperature 41° to 104°F (5° to 40°C), maximum relative humidity 80% for temperatures up to 88°F (31°C), decreasing linearly to 50% at higher temperatures. p.30 ↗ Install only in environments meeting these conditions to ensure proper operation and safety. all models
Leave the blower on for at least one minute after work; if a chemical is spilled, do not switch off blower until all traces have been removed. p.13 ↗ Ensures purging of contaminants and prevents accumulation. all models
After work, allow enclosure to operate for two to three minutes undisturbed to purge airborne contaminants before shutting off blower. p.15 ↗ Recommended final purging procedure for routine daily work. all models
Face velocity profiles and smoke testing should be performed at least monthly. p.9 ↗ Regular validation ensures containment performance. all models
Carbon filters are installed behind upper and lower filter clamps and are removed by Phillips screws. p.8 ↗ Filter replacement requires removing perforated baffle and support brackets. all models

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Contaminants outside of enclosure. p.24 ↗ Improper user techniques for the enclosure. See Chapter 3 and Chapter 4 sections in the manual. (Ref. Appendix D). Chapter 7
Contaminants outside of enclosure. p.24 ↗ Restriction of the baffle air slots or blockage of the exhaust outlet. Remove obstruction to ensure that all air slots and the exhaust outlet are unobstructed. Chapter 7
Contaminants outside of enclosure. p.24 ↗ External factors are disrupting the enclosure airflow patterns or acting as a source of contamination. See “Location Requirements” Chapter 2, “Certifying the Enclosure” Chapter 3, and “Safety Precautions” Chapter 4 sections of this manual. (Ref. Appendix D). Chapter 7
Contaminants outside of enclosure. p.24 ↗ Enclosure has improper face velocity. Have the enclosure’s exhaust system checked. Check carbon filters for loading. Enclosure should have an average face velocity of 35 fpm. Chapter 7
Contaminants outside of enclosure. p.24 ↗ Filtered exhausted. Change and replace the carbon filter. Chapter 7
Blower won’t operate. p.25 ↗ Unit not plugged into outlet. Plug the enclosure into appropriate electrical service. Chapter 7
Blower won’t operate. p.25 ↗ Circuit breaker(s) or Ground Fault Interrupter. Reset circuit breaker. Chapter 7
Blower won’t operate. p.25 ↗ Blower wiring is disconnected. Inspect blower wiring and re-connect. Chapter 7
Blower won’t operate. p.25 ↗ Blower switch is defective. Replace switch. Chapter 7
Blower won’t operate. p.25 ↗ Motorized impeller or blower is defective. Replace motorized impeller or blower. Chapter 7