Technical Reference Summary

Covered Models & Document Scope

This manual (Part #3939201, Rev. C) covers Labconco Purifier® Non-Ventilated PCR Enclosures and Purifier Tissue Culture Enclosures, models 3951400, 3951420, 3952400, 3952420. The PCR variant includes a digital UV light timer; the Tissue Culture variant has a built-in UV light for secondary surface decontamination. Both are non-ventilated, circulation-free enclosures with a switched interior outlet.

Section-by-Section Technical Breakdown

pp. 8-9
Prerequisites: Support, Location, Electrical and Space Requirements

Before installation the site must be level and of solid construction, with a dedicated electrical source nearby. The enclosure should be placed away from high-traffic areas, HVAC diffusers, fans, radiators, doorways and windows to avoid air disturbances. Standard duplex receptacles are required; the enclosure provides a switched auxiliary outlet rated 115V/8A or 230V/4A. Space requirements are given in Appendix B.

  • Support structure: base cabinet plus chemically resistant work surface; mobile stand or cart allowable.
  • Avoid high-traffic areas, overhead/wall HVAC diffusers, fans, radiators, lab equipment producing air currents, and doorways/windows that may be opened.
  • Do not position the unit so the main disconnect device is difficult to operate.
  • Auxiliary switched outlet rated 115V, 8 amps or 230V, 4 amps.
  • Enclosure dimensions are in Appendix B.
pp. 10-13
Installation on Supporting Structure, Work Surface and Electrical Connection

The enclosure weighs 100 lbs (45 kg) and must be moved with a floor jack or mechanical lift; do not tilt onto a hand truck or lift by the front lower support. The work surface must be at least as wide and deep as the enclosure, smooth, rigid, non-porous and chemically resistant, with a dished recessed area. The enclosure front must align within 1/8" of the work surface front; four mounting screws secure it. Electrical connection requires a nearby grounded receptacle and a properly rated detachable power cord; building wiring must be by a licensed electrician.

  • Enclosure weight: 100 lbs (45 kg).
  • Do not lift by the front lower support; do not move by tilting onto a hand truck.
  • Work surface must be at least as wide and deep as the enclosure and have a dished recessed area for spills.
  • Align enclosure front within 1/8" of work surface front.
  • Insert four mounting screws with minimum 1/8" clearance under the head, slide rear and lower front bracket flanges under screw heads, then tighten.
  • Seal enclosure to work surface with a bead of silicone sealant.
  • Use only an adequately rated detachable power cord; building outlet wiring by licensed electrician per local codes.
pp. 17
Using Your Enclosure

Operation covers planning, start-up, work techniques, unloading and shutdown. Clean and surface disinfect the interior before use, use UV as required by protocol, turn on the fluorescent light, and raise the sash only for loading. Keep all materials and operations inside the work area; cover open trays and containers before removal. At shutdown, turn off the fluorescent light, install the UV sash closure panel, clean and disinfect, then turn on UV per protocol.

  • Only raise the sash for loading.
  • Wear long-sleeved lab coat, rubber gloves, protective eyewear and mask if appropriate.
  • Keep all materials inside and perform all operations inside the work area.
  • Cover open trays, weigh vessels and containers before removing from the enclosure.
  • Shutdown: fluorescent light off, UV sash closure panel installed, clean/disinfect, then UV on as determined by protocol.
pp. 18-19
Routine Maintenance and UV/Fluorescent Lamp Replacement

Maintenance is scheduled weekly (clean interior), monthly (clean exterior with glass cleaner, plus weekly activities) and annually (replace UV lamp, verify UV panel safety switch, plus monthly activities). UV lamp replacement requires disconnecting power, rotating out the old lamp, installing the new one, and testing with the UV closure panel installed. Fluorescent lamp replacement requires removing two Phillips screws on top of the light assembly, lifting and rotating the assembly while protecting the UV closure panel sensor wire, then replacing the lamp. Lamps contain mercury and must be disposed of as hazardous waste.

  • Weekly: clean enclosure interior appropriate for the application.
  • Monthly: clean exterior surfaces, particularly front, with cloth and glass cleaner; perform weekly activities.
  • Annually: replace UV lamp, verify UV panel safety switch operation, perform monthly activities.
  • UV lamp replacement: disconnect power, rotate/remove old lamp, reinstall new lamp, test with UV closure panel installed.
  • Fluorescent lamp replacement: remove two Phillips machine screws on top of light, lift/rotate assembly upside down, protect UV closure panel sensor wire, replace lamp and reassemble.
  • Lamps contain mercury; dispose as hazardous waste per local laws; see www.lamprecycle.org.
pp. 20-21
Accessories: Work Surfaces, Storage Cabinets and Stands

Optional dished epoxy work surfaces are available in black or gray, 4-foot wide, 48" x 26.66" x 1". Storage cabinets and base stands/mobile stands are listed with catalog numbers by width and door configuration. Telescoping base stands are available with fixed feet or 5" casters, 48" wide x 29.25" deep, adjustable height 27.5" to 33.5".

  • Work surface catalog #3908402 (black) and #3908405 (gray): 48" x 26.66" x 1".
  • Storage cabinets available in solvent and acid versions, widths 12" to 48", with dual doors, right/left hinge, self-closing and ADA variants.
  • Telescoping base stand #3746702 with fixed feet: 48" w x 29.25" d x 27.5" to 33.5" h.
  • Telescoping base stand #3746712 with 5" casters: 48" w x 29.25" d x 27.5" to 33.5" h.
pp. 23-25
Replacement Parts

Appendix A lists replacement parts with item numbers, quantities, part numbers and descriptions, including glass panels, sash assembly, wiper, latch components, air foil brackets, header, bushings, labels, digital UV timer PCB, switches, fluorescent lamp (FT25T8 25W), UV lamp (G30T8 30W), ballasts, power inlets and cords, outlets, circuit breakers, UV closure panels, magnetic switch and rubber bumpers.

  • UV lamp: part #1271300, G30T8, 30 Watt.
  • Fluorescent lamp: part #9721901, FT25T8, 25 Watt.
  • Digital UV timer PCB: part #3945300.
  • UV ballast 115V: #1233400; UV ballast 230V: #1233600.
  • Circuit breaker 10 Amp 115V: #1327201; circuit breaker 5 Amp 230V: #1327204 (qty 2).
  • UV closure panels: 2 ft #3560510, 3 ft #3560511, 4 ft #3560512.
  • Magnetic switch assembly: #1322800.
pp. 27
Specifications

Appendix C gives electrical specifications and environmental conditions for all Purifier Passive PCR and Tissue Culture Enclosures. Electrical: 10 A at 115V/60 Hz or 5 A at 230V/50/60 Hz; auxiliary outlet load 8 A at 115V and 4 A at 230V. Environmental: indoor use only, maximum altitude 6562 ft (2000 m), ambient 41° to 104°F (5° to 40°C), humidity 80% up to 88°F decreasing linearly to 50% at 104°F, supply voltage fluctuation ±10%, transient over-voltages per Installation Category II (1500V for 115V models, 2500V for 230V models), Pollution Degree 2.

  • Electrical: 10 Amps, 115V, 60 Hz; 5 Amps, 230V, 50/60 Hz.
  • Auxiliary outlet load: 8 Amps at 115V, 4 Amps at 230V.
  • Indoor use only; maximum altitude 6562 feet (2000 meters).
  • Ambient temperature range 41° to 104°F (5° to 40°C).
  • Maximum relative humidity 80% for temperatures up to 88°F (31°C), decreasing linearly to 50% at 104°F (40°C).
  • Main supply voltage fluctuations not to exceed ±10% of nominal.
  • Transient over-voltages per Installation Category II: up to 1500V for 115V models and 2500V for 230V models.
  • Pollution Degree 2 environment.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Enclosure weight is 100 lbs (45 kg). p.10 ↗ Use a mechanical lift truck or floor jack; if manual lifting is required, follow safe-lifting guidelines and do not lift by the front lower support. whole series
The front of the enclosure must be aligned within 1/8" of the front of the work surface. p.11 ↗ Critical for proper mounting and sealing; verify alignment before tightening the four mounting screws. whole series
Mounting screws require a minimum of 1/8" clearance under the head for positioning the enclosure. p.12 ↗ Allows the rear flange and lower front bracket flanges to slide under the screw heads before final tightening. whole series
Auxiliary switched outlet is rated for either 115V, 8 amps or 230V, 4 amps. p.9 ↗ Do not exceed these loads when connecting equipment to the interior switched outlet. whole series
Electrical specifications: 10 Amps, 115V, 60 Hz or 5 Amps, 230V, 50/60 Hz. p.27 ↗ Size the branch circuit and protective device accordingly; the unit draws up to 10 A at 115V or 5 A at 230V. whole series
UV light timer has eight settings: 5, 10, 15, 30, 60, 120, 240 minutes and infinitely ON. p.15 ↗ Select the required surface disinfection time on the control panel for PCR protocols. Purifier PCR Enclosure only
The UV sash closure panel must be in place for the UV light to operate. p.16 ↗ UV light will not activate without the closure panel; this is a safety interlock via magnetic switch. whole series
Internal depth is 23" and interior height is 22". p.15 ↗ Ensures procedures and auxiliary equipment fit inside the enclosure without extending outside. whole series
Ergonomic slope and front of 20°. p.15 ↗ Provides maximum visibility and comfort, reduces glare and operator fatigue. whole series
Environmental ambient temperature range: 41° to 104°F (5° to 40°C). p.27 ↗ Install only in indoor locations within this temperature range; operation outside may impair performance. whole series
Maximum relative humidity: 80% for temperatures up to 88°F (31°C), decreasing linearly to 50% at 104°F (40°C). p.27 ↗ Avoid condensing conditions; temporary conductivity from condensation is expected per Pollution Degree 2. whole series
Main supply voltage fluctuations must not exceed ±10% of nominal voltage. p.27 ↗ Verify supply stability; out-of-range fluctuations may affect operation and are excluded from warranty. whole series
Transient over-voltages per Installation Category II: up to 1500V for 115V models and 2500V for 230V models. p.27 ↗ Installation must be suitable for Category II over-voltage conditions per IEC 1010. whole series
Maximum altitude: 6562 feet (2000 meters). p.27 ↗ Do not install above this altitude; cooling and electrical clearance may be affected. whole series
UV lamp is G30T8, 30 Watt (part #1271300); fluorescent lamp is FT25T8, 25 Watt (part #9721901). p.24 ↗ Use exact replacement part numbers for lamp replacement. whole series
Circuit breaker: 10 Amp for 115V (part #1327201); two 5 Amp breakers for 230V (part #1327204). p.24 ↗ Replacement breakers must match voltage variant; 230V units use two 5 A breakers. whole series
UV ballast is 115V (part #1233400) or 230V (part #1233600). p.24 ↗ Select correct ballast for the unit voltage; do not interchange. whole series
Work surface must be at least as wide and deep as the enclosure and have a dished recessed area for containing primary spills. p.11 ↗ Use Labconco accessory work surfaces with mounting holes for secure installation. whole series
Optional dished epoxy work surfaces: 48" x 26.66" x 1", black #3908402 or gray #3908405. p.20 ↗ Order correct catalog number for color and 4-foot width. whole series
Telescoping base stands: 48" w x 29.25" d, height adjustable 27.5" to 33.5", with fixed feet (#3746702) or 5" casters (#3746712). p.21 ↗ Provides mobile or fixed support; verify height range suits operator and work surface. whole series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Lights do not work. p.22 ↗ Circuit breaker(s) tripped. Reset or replace circuit breaker. Chapter 8
Lights do not work. p.22 ↗ Lamp not installed properly. Inspect lamp installation. Chapter 8
Lights do not work. p.22 ↗ Lamp Wiring disconnected. Inspect lamp wiring. Chapter 8
Lights do not work. p.22 ↗ Defective lamp. Replace lamp. Chapter 8
Lights do not work. p.22 ↗ Light switch is defective. Replace light switch. Chapter 8
Lights do not work. p.22 ↗ UV closure panel not installed. Install UV closure panel (UV lights only). Chapter 8
Lights do not work. p.22 ↗ Magnetic switch for UV closure panel defective. Replace switch (UV lights only). Chapter 8
Lights do not work. p.22 ↗ UV timer circuit board defective. Replace circuit board (UV lights only). Chapter 8
UV light timer does not work. (PCR only) p.22 ↗ No power. Power cable to UV light timer circuit board is disconnected. Verify connection. Chapter 8
UV light timer does not work. (PCR only) p.22 ↗ No lights on. Verify connection. Chapter 8
UV light timer does not work. (PCR only) p.22 ↗ No display. Replace circuit board for UV light timer if it malfunctions. Chapter 8