3898400 Rev L Protector Evidence Drying Cabinets

Technical Reference Summary

Covered Models & Document Scope

This manual covers Labconco Protector® Evidence Drying Cabinets (EDC), 2023—Present, in three sizes: 3' single-door (34000xx series), 4' double-door (34040xx series), and 5' double-door (34050xx series). Variants include UV Lamp models (catalog suffixes -00 thru -04) and Wash Down models with drain pump (suffixes -10 thru -14), plus 115 VAC 60 Hz (-00, -10) and 230 VAC 50/60 Hz (-01 thru -04, -11 thru -14) electrical configurations.

Section-by-Section Technical Breakdown

pp. 9
Location and Clearance Requirements

The installation site must be level and of solid construction. The 3' cabinet uses locking casters; the 4' and 5' cabinets use levelling feet. Minimum service clearances are specified for top, sides, and front.

  • Minimum clearance of 6 inches (150 mm) on top and both sides for service.
  • 36 inches (914 mm) clearance at the front for door swing.
  • 3' cabinet equipped with locking casters; 4' and 5' cabinets with levelling feet.
pp. 10
Electrical Requirements

Electrical supply requirements are listed per catalog number. A dedicated outlet with appropriate circuit breaker must be located within 6 feet (1.8 m) of the product and must provide a protective earthing ground.

  • 115 VAC, 60 Hz, 6 A, 1 phase for models 34000-00 & -10, 34040-00 & -10, 34050-00 & -10.
  • 230 VAC, 50/60 Hz, 3 A, 1 phase for models 34000-01 thru -04 & -11 thru -14, 34040-01 thru -04 & -11 thru -14, 34050-01 thru -04 & -11 thru -14.
  • Dedicated outlet within 6 feet (1.8 m) maximum; consult local electrical codes for breaker rating.
pp. 10
Exhaust Requirements

If the optional Exhaust Connection Kit is used, the area above the cabinet exhaust must be clear of obstructions and provide clearance for a 5-inch diameter duct.

  • Clearance for a 5-inch (127 mm) diameter duct required.
  • Area directly above exhaust must be free of structural elements, water and utility lines, or other fixed obstructions.
pp. 10, 25, 26
Drain Requirements

Wash Down models (-10 thru -14) and units with optional drain kit require a floor drain or standpipe within 6 feet (1.8 m). The drain must be below the level of the drain hose exiting the cabinet, even for drain-pump-equipped models.

  • Floor or standpipe drain must be located below the level of the drain hose exiting the cabinet for all models.
  • Cabinet and drain should be in an area not subject to freezing temperatures.
  • Drain line must be routed to a floor drain with an air gap only; never directly connect to a sewer or drain system.
pp. 11, 26
Water Requirements

Wash Down models (-10 thru -14) require a domestic water line within 6 feet (1.8 m) equipped with a shut-off valve ending in a 3/4" male garden hose connector. Water pressure must be 10–40 PSIG.

  • Water system pressure: 10–40 PSIG (0.7–2.8 Bar).
  • Connection: 3/4" male garden hose connector with water shut-off valve.
  • Cabinet water shut-off valve should be in the 'off' position whenever the unit is not being washed down.
pp. 16, 18, 19, 20
Unpacking and Shipping Brace Removal

The cabinet is shipped on a pallet with shipping braces and doors attached at the back. Procedures cover removing shrink wrap, braces, and restraints, and installing doors. Cabinet weights are approximately 300 lbs (3'), 400 lbs (4'), and 510 lbs (5').

  • 3' EDC weighs approximately 300 lbs. (136 kg); 4' EDC 400 lbs. (182 kg); 5' EDC 510 lbs. (232 kg).
  • Manual lifting requires at least 6 persons; mechanical lift recommended.
  • Shipping braces attached by 4 screws per corner; center 5/16" bolt with spacer and nut.
  • 4 ft. and 5 ft. models have two lower door restraints.
pp. 23
Wall Attachment

The EDC must be secured to a wall stud or suitable wall structure using included straps and brackets to minimize overturning risk. Bracket mounting height and pilot hole specifications are given.

  • Top of strap slot positioned 79 inches (200 cm) above the floor.
  • Pilot holes drilled with 1/8" (3 mm) bit to a depth of at least 1-1/2 inches (38 mm).
  • 4 ft. and 5 ft. models include 2 wall brackets; at least one must be used.
  • Do NOT overtighten strap(s).
pp. 25, 26
Drain and Water Connection

UV Lamp models can drain to a bucket (working capacity ~1.5 gallons) or use optional Drain Kit #3928200. Wash Down models have a drain hose routed to a floor drain and a water supply connection with a cabinet shut-off valve.

  • Drain bucket working capacity approximately 1.5 gallons (5.6 L).
  • Optional Drain Kit #3928200 includes hose barb adaptor and 10 ft. (3 m) of 1/2" (13 mm) O.D. polyurethane tubing.
  • Drain line must be routed with an air gap only; never directly connect to sewer or drain system.
  • Wash Down models: connect stainless-steel braided hose to 3/4" male garden hose connector.
pp. 28
Initial Operation Tests

Before use with evidence, perform airflow smoke pattern test, UV light and interlock tests (UV models), and spray gun and drain pump tests (Wash Down models).

  • Smoke should move from door to intake prefilter in approximately 5 seconds or less.
  • UV light should turn on with door closed and keylock ON; opening door turns UV off.
  • Drain pump should remove majority of water; some residual may remain due to surface tension or unlevel floor.
pp. 29, 31, 32
Performance Features

The EDC operates using directional airflow, prefiltration, HEPA filtration, and granular activated carbon adsorption. Key components include prefilters, blower, HEPA filter, and carbon filter.

  • Prefilters are MERV 11 rated, removing 65–79% of particles 1.0–3.0 microns and more than 85% of particles 3.0–10 microns.
  • HEPA filters are at least 99.99% efficient in removing particles 0.3 micron.
  • UV lamp generates primary wavelength of 254 nm; replace after 6,000 hours of operation or less.
  • Blower is a motorized, backward curved fan.
pp. 38, 39, 41, 42
Operating Procedures

Procedures for start-up, wipe-down, loading and unloading evidence, wash-down (Wash Down models), and shutdown. Includes weight limits for shelves and hanging rod.

  • Fold-down shelf weight limit: 15 lbs (6.8 kg) each.
  • Cabinet floor and lower storage area weight limit: 20 lbs (9.1 kg).
  • Hanging rod weight limit: 20 lbs (9.1 kg).
  • Wipe down interior with 70% ethanol or suitable disinfectant; do not use bleach on stainless steel liner.
  • UV disinfection time typically 1–8 hours.
pp. 44, 45, 49
Maintenance

Routine maintenance schedule and service operations including prefilter replacement, UV lamp replacement, circuit breaker reset, speed control adjustment, and filter light pressure switch adjustment.

  • Weekly: disinfect interior surfaces, clean UV lamp, operate blower, ensure water valve off.
  • Monthly: wipe exterior with 70% alcohol, check prefilters.
  • Annually: replace HEPA and carbon filters if needed, recertify cabinet.
  • Speed control: turn counterclockwise to increase air velocity, clockwise to decrease.
  • Circuit breakers located on top rear of cabinet; depress white button to reset.
pp. 55
Parts List

Table A-1 lists common replacement parts with part numbers and quantities per unit for 3', 4', and 5' cabinets.

  • Door Prefilter: 3FT #3872600, 4FT #3872601, 5FT #3872602 (package of 6).
  • Door Evidence Prefilter: 3FT #3951802, 4FT #3951822, 5FT #3951832 (package of 10).
  • Cabinet (top) Prefilter #3872700 (package of 6).
  • HEPA Filter #6938100; Granular Activated Carbon Filter #6938200.
  • UV Lamp #1270100 (UV models only).
pp. 56
Dimensions

Table B-1 provides overall dimensions for 3', 4', and 5' models. Interior and lower storage dimensions are also given.

  • 3' model: 36.0" (914 mm) wide, 32.3" (820 mm) deep, 78.9" (2004 mm) high, casters.
  • 4' model: 48.0" (1219 mm) wide, 44.3" (1125 mm) deep, 75.5" (1918 mm) high, legs.
  • 5' model: 60.0" (1524 mm) wide, 56.3" (1430 mm) deep, 75.5" (1918 mm) high, legs.
  • Interior dimensions: 56.0" (1422 mm) height x 26.5" (673 mm) deep x width C.
  • Lower storage dimensions: 12.4" (315 mm) height x 26.5" (673 mm) deep x width D.
pp. 57
Specifications

Power data and environmental conditions for all models. All models use 100 W energy.

  • Power: 115 VAC, 60 Hz, 6 A or 230 VAC, 50/60 Hz, 3 A; energy use 100 W.
  • Indoor use only; maximum altitude 6562 feet (2000 m).
  • Ambient temperature range: 41° to 104°F (5° to 40°C).
  • Maximum relative humidity: 80% for temperatures up to 88°F (31°C), decreasing linearly to 50% at 104°F (40°C).
  • Main supply voltage fluctuations not to exceed ±10% of nominal voltage.
pp. 58, 59
Accessories

Optional accessories include evidence prefilters, paper liners, security tags, hanging clips, exhaust system connection kit, floor drain connection kit, and germicidal cleaners.

  • Exhaust System Connection Kit #3869600: 14" long, 5" diameter flexible duct; exhaust system must be capable of 150 CFM (4.2 m3/min) at 0.5" H2O (125 Pa) vacuum.
  • Floor Drain Connection Kit #3928200: includes hose barb adaptor and 10 ft. (3 m) of 1/2" (13 mm) O.D. polyurethane tubing.
  • Security Tags #3902400: package of 100 tamper-evident tags.
  • Hanging Clips #3869500: package of 10.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Electrical supply: 115 VAC, 60 Hz, 6 A, 1 phase for models ending -00 and -10; 230 VAC, 50/60 Hz, 3 A, 1 phase for models ending -01 thru -04 and -11 thru -14. p.10 ↗ Verify site voltage and breaker rating match the model variant before installation. All models
Dedicated outlet must be within 6 feet (1.8 m) maximum of the product and provide a protective earthing ground. p.10 ↗ Plan electrical layout accordingly; improper grounding may create a safety hazard. All models
Minimum clearance: 6 inches (150 mm) on top and both sides; 36 inches (914 mm) at front for door swing. p.9 ↗ Ensure installation site provides these clearances for service and operation. All models
Water supply pressure for Wash Down models: 10–40 PSIG (0.7–2.8 Bar); connection via 3/4" male garden hose connector. p.11 ↗ Install a pressure regulator if line pressure exceeds 40 PSIG. Wash Down models (-10 thru -14)
Drain must be located below the level of the drain hose exiting the cabinet for all models, even drain-pump-equipped models. p.10 ↗ Gravity drainage is required; improper drain elevation may cause incomplete drainage or backflow. All models
Drain bucket working capacity approximately 1.5 gallons (5.6 L) for UV Lamp models. p.25 ↗ Monitor liquid level during decontamination or washdown; empty before overflow. UV Lamp models (-00 thru -04)
Wall bracket strap slot top must be 79 inches (200 cm) above the floor; pilot holes 1/8" (3 mm) diameter, at least 1-1/2 inches (38 mm) deep. p.23 ↗ Ensures proper strap tension and secure wall attachment. All models
Cabinet weights: 3' EDC ~300 lbs. (136 kg), 4' EDC ~400 lbs. (182 kg), 5' EDC ~510 lbs. (232 kg). p.16 ↗ Use mechanical lift or at least 6 persons for manual lifting; ensure floor and path can support weight. All models
Prefilters are MERV 11 rated: remove 65–79% of particles 1.0–3.0 microns and >85% of particles 3.0–10 microns. p.31 ↗ Prefilters capture larger particulates; replace when loaded to maintain airflow. All models
HEPA filters are at least 99.99% efficient in removing particles 0.3 micron. p.31 ↗ HEPA filters provide high-efficiency particulate removal; do not touch media. All models
UV lamp primary wavelength 254 nm; replace after 6,000 hours of operation or less. p.32 ↗ Track UV lamp operating hours and replace proactively to maintain biocidal effectiveness. UV Lamp models (-00 thru -04)
Smoke should traverse from door intake to exhaust prefilter in approximately 5 seconds or less. p.28 ↗ Use smoke test to verify airflow; adjust speed control if slower. All models
Speed control: turn counterclockwise to increase air velocity, clockwise to decrease. p.49 ↗ Adjust to achieve proper face velocity and smoke test result. All models
Weight limits: fold-down shelves 15 lbs (6.8 kg) each; cabinet floor and lower storage area 20 lbs (9.1 kg); hanging rod 20 lbs (9.1 kg). p.34 ↗ Do not exceed these limits to prevent damage or injury. All models
Drain pump should not be operated dry for more than 5 minutes. p.33 ↗ Avoid running pump without liquid to prevent damage. Wash Down models (-10 thru -14)
Exhaust system must be capable of exhausting 150 CFM (4.2 m3/min) at a vacuum of 0.5" of H2O (125 Pa). p.59 ↗ Verify exhaust system capacity when using optional Exhaust Connection Kit #3869600. All models with optional exhaust kit
Environmental conditions: indoor use, max altitude 6562 ft (2000 m), ambient 41–104°F (5–40°C), max RH 80% up to 88°F decreasing to 50% at 104°F. p.57 ↗ Install in a controlled environment within these limits. All models
Main supply voltage fluctuations not to exceed ±10% of nominal voltage; transient overvoltages up to 1500 V (115 V models) or 2500 V (230 V models) allowed. p.57 ↗ Ensure electrical supply quality meets these tolerances. All models

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
UV Light is dim or flickering p.51 ↗ UV lamp near end of life or dirty lamp surface Replace UV lamp (Part #1270100) or clean lamp exterior; replace after 6,000 hours of operation or less. Section 8
Cabinet Blower and/or lights will not turn on p.52 ↗ Circuit breaker tripped or power not connected Reset circuit breaker on top rear of cabinet by depressing white button; check power cord connection. Section 8
UV Light not working p.53 ↗ UV Security Lock Key not in 'ON' position, door not closed, or interlock fault Ensure UV Security Lock Key is turned to 'ON' and door is closed; check interlock function. Section 8
Filter System light is on/flickering p.54 ↗ Prefilters or HEPA filter loaded, or pressure switch needs adjustment Check and replace prefilters if needed; adjust filter light pressure switch per Section 7. Section 8
Pump Unit not coming on p.54 ↗ Switch not in 'Pump' position, circuit breaker tripped, or pump fault Set control switch to 'Pump'; reset circuit breaker; check drain line for obstructions. Section 8