Rev j filtermate portable exhausters user manual
Technical Reference Summary
Covered Models & Document Scope
Covers Labconco FilterMate Portable Exhausters: 115V/60Hz models 3970000 (HEPA), 3970001 (Carbon), 3970002 (HEPA, canopy to outside), 3970003 (HEPA/Carbon), 3970004 (Carbon/Carbon); 230V/50Hz models 3970020, 3970021, 3970022, 3970023, 3970024; and additional listed models 3970010, 3970011, 3970012, 3970013. Accessory for Protector XVS Ventilation Stations and XPert Balance Enclosures.
Section-by-Section Technical Breakdown
Specifies the electrical supply requirements for 115V and 230V FilterMate models, including circuit rating, motor/blower current, auxiliary outlet capacity, and grounding. Also covers Ground Fault Interrupter (GFIC) requirements.
- Both 115V and 230V models require 10 amp circuits.
- Motor/blower uses 2A; switched auxiliary outlet maximum capacity is 8A.
- 230V model is shipped without a plug; customer must install appropriate plug per local codes.
- Equipment must be grounded (earthed).
- When connecting to a GFIC, ensure the circuit is rated for 15-amp load and functioning correctly.
Defines maximum airflow capabilities for HEPA and combination HEPA/Carbon models, lists catalog numbers with voltage, filter type, maximum airflow, and exhaust connection type. Also covers HEPA and carbon filter part numbers and specifications.
- FilterMate exhausts up to 280 cfm of HEPA filtered air or up to 220 cfm of combination HEPA/Carbon filtered air.
- Airflows achieved when connected to enclosure with an 8' length of 5" hose.
- HEPA Filter Part #3707900 is 99.99% efficient on particles 0.3 micron; normal life expectancy two to five years.
- HEPA Filter Bag-In/Bag-Out Bag Part #3776002 ensures containment during filter changing; Cinching Strap #1968800 ordered separately.
- Carbon filters: Organic Vapors 3923400 (12# activated carbon), Formaldehyde 3923401 (14# treated carbon), Ammonia 3923402 (16# treated carbon).
- Activated carbon adsorbs hydrochloric acid and nitric acid, but not sulfuric acid.
Provides requirements for room air exchanges for carbon-filtered models and location considerations to avoid airflow disruption. Includes an equation to determine minimum room exhaust rate.
- For carbon-filtered models, accumulation of vapor contaminants can be harmful when Rate of Generation exceeds Rate of Room Exhaust.
- Equation: Inlet Vapor Concentration (PPM) x FilterMate Airflow in CFM < TWA (PPM) x Room Exhaust (CFM).
- Enclosure should be located away from drafts, traffic patterns, doors, fans, ventilation registers, fume hoods, and other air handling devices.
- All windows in the room adjacent to the enclosure should be closed.
- FilterMate may be bench mounted, placed on the floor, or suspended on a shelf nearby.
Describes installation procedures for HEPA filters, carbon filters, combination HEPA/Carbon, and dual carbon configurations across specific model numbers. Includes filter types, part numbers, and appropriate uses.
- HEPA filters shipped installed with gasket on downstream side; leak checked at factory; second leak check recommended before use and at least annually.
- Carbon filters installed with gasket side downstream and opposite clamp bolts; tighten clamp bolts to compress gasket 50%.
- Combination HEPA/Carbon: HEPA installed first, carbon filter installed second in perforated plenum exhaust cover; tighten clamps to compress gasket 50%.
- Dual carbon filters do not provide any particulate protection.
- Filter types: HEPA 3707900 (99.99% at 0.3 micron), ULPA 3885700 (99.999% at 0.12 micron), PLEKX 3707904 (99.97% at 0.3 micron with 1.7 lbs carbon).
Instructions for connecting 115V and 230V power to the FilterMate, including use of the switched auxiliary outlet for an airflow monitor.
- 115V: connect supplied power cord to IEC electrical supply plug on back of unit.
- If using 50 Hz operation, blower performance maximum airflow will be reduced by 17%.
- 230V: shipped without a plug; install appropriate plug per local codes.
- Switched auxiliary outlet on back for plugging in airflow monitor.
Explains how to adjust the speed control to achieve correct face velocity when connected to an enclosure. Provides face velocity ranges and adjustment method.
- Face velocity should be adjusted from 60 to 100 fpm.
- Weighing procedures usually require face velocities between 60-80 fpm.
- Face velocity increased by turning speed control counter clockwise; clockwise to decrease.
- Speed control adjustment located on back of FilterMate; temporarily remove hole plug.
- Working at lowest face velocity appropriate for application gives quietest operation.
Recommends monthly maintenance schedule including testing carbon filter condition with gas detector tubes and checking face velocity on HEPA filtered models.
- Test filter condition on carbon filtered models using gas detector tube at intervals of 20% of total estimated time.
- For formaldehyde or carcinogens, check at least every 10% of total estimated time.
- Check face velocity on HEPA filtered models; increase speed control or change HEPA filter when face velocity drops below recommended speed.
- Replace carbon filters when chemical breakthrough is indicated by odor, time, detector tube, or analytical instrumentation.
Defines conditions requiring carbon filter replacement and describes four means of determining when to change filters: odor, detection tubes, time, and analytical instrumentation.
- Carbon filters MUST be replaced when outlet concentration approaches inlet concentration (saturation) or odor in work area becomes intolerable or concentration exceeds TWA.
- Detection tubes: color change indicators measure concentration at exit side; kit with syringe pump and tubing available as Catalog #6924900.
- Each stroke of syringe represents a 100-ml sample.
- When a color change is observed in detector tube, replace filter immediately.
- Time can be used to anticipate saturation for consistent inlet concentrations, but does not replace sampling.
Covers conversion procedures between FilterMate configurations, including converting room exhaust to outside exhaust, combination to dual carbon, HEPA to combination, and converting to 6" hose.
- Converting a Room Exhaust HEPA FilterMate to Outside Exhaust.
- Converting a Combination HEPA/Carbon FilterMate to a Dual Carbon FilterMate.
- Converting a HEPA FilterMate to a Combination HEPA/Carbon FilterMate.
- Converting to 6" Hose and 6" Exhaust Transition increases airflow 15-20%.
Provides a table of problems, probable causes, and corrective actions for common FilterMate issues including blower failure and low face velocity.
- Blower won't turn on: check outlet, circuit breaker/GFIC, blower wiring, blower switch, motorized impeller.
- Low face velocity or poor containment: close sash, replace clogged HEPA filter or increase speed, replace carbon filter loaded with chemicals.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Both 115V and 230V FilterMate models require 10 amp circuits; motor/blower uses 2A and switched auxiliary outlet has maximum capacity of 8A. p.10 ↗ | Electrical supply must be sized for at least 10A; auxiliary outlet can power an airflow monitor up to 8A. | All models |
| When connecting to a Ground Fault Interrupter Circuit (GFIC), ensure the circuit is rated for 15-amp load and is functioning correctly. p.10 ↗ | Lower capacity or defective GFIC may cause excessive activation and cut power during operation. | All models |
| FilterMate exhausts up to 280 cfm of HEPA filtered air or up to 220 cfm of combination HEPA/Carbon filtered air when connected with an 8' length of 5" hose. p.11 ↗ | Maximum airflow depends on filter configuration; use 5" hose of 8' length for rated performance. | HEPA models (280 cfm), HEPA/Carbon and Carbon/Carbon models (220 cfm) |
| HEPA Filter Part #3707900 is 99.99% efficient on particles 0.3 micron; normal life expectancy is two to five years. p.11 ↗ | Plan HEPA filter replacement every 2-5 years depending on usage and room air cleanliness. | HEPA-filtered models |
| Carbon filters: Organic Vapors 3923400 (12# activated carbon), Formaldehyde 3923401 (14# treated carbon), Ammonia 3923402 (16# treated carbon). Activated carbon adsorbs hydrochloric acid and nitric acid, but not sulfuric acid. p.11 ↗ | Select carbon filter type based on chemical being used; do not use activated carbon for sulfuric acid. | Carbon-filtered models |
| For carbon-filtered models, minimum room exhaust rate determined by: Inlet Vapor Concentration (PPM) x FilterMate Airflow in CFM < TWA (PPM) x Room Exhaust (CFM). p.12 ↗ | Ensure room ventilation is sufficient to dilute exhaust below TWA for the chemical. | Carbon-filtered models |
| Carbon filters installed with gasket side downstream and opposite clamp bolts; tighten clamp bolts to compress gasket 50%. p.17 ↗ | Proper gasket compression ensures seal; incorrect installation may cause leakage. | Carbon-filtered models |
| Combination HEPA/Carbon: HEPA installed first with gasket on downstream side; carbon filter installed second in perforated plenum exhaust cover; tighten clamps to compress gasket 50%. p.17 ↗ | Order of filter installation is critical for proper filtration. | Models 3970003, 3970023 |
| Dual carbon filters do not provide any particulate protection. p.18 ↗ | Dual carbon configuration is only for vapor adsorption; not suitable for particulate or biological work. | Models 3970004, 3970024, and 3970003/3970023 when converted |
| If using 50 Hz operation, blower performance maximum airflow will be reduced by 17%. p.24 ↗ | Account for reduced airflow when sizing for 50 Hz applications. | All models on 50 Hz |
| Face velocity should be adjusted from 60 to 100 fpm; weighing procedures usually require 60-80 fpm. p.24 ↗ | Set speed control to achieve face velocity within this range; lower velocities give quieter operation. | All models when connected to enclosure |
| Face velocity increased by turning speed control counter clockwise; clockwise to decrease. p.24 ↗ | Adjustment direction for setting correct face velocity. | All models |
| Test filter condition on carbon filtered models using gas detector tube at intervals of 20% of total estimated time; for formaldehyde or carcinogens, check at least every 10% of total estimated time. p.32 ↗ | Frequent monitoring required for hazardous chemicals to prevent breakthrough. | Carbon-filtered models |
| Detection tube kit including syringe pump and tubing available as Catalog #6924900; each stroke of syringe represents a 100-ml sample. p.33 ↗ | Use specified kit for quantitative breakthrough testing. | Carbon-filtered models |
| Converting to 6" hose and 6" exhaust transition increases airflow 15-20%. p.14 ↗ | Larger hose diameter can improve airflow for demanding applications. | All models (modification) |
| FilterMate may weigh up to 90 lbs. (41 kg); always use more than one person and follow safe-lifting guidelines. p.13 ↗ | Plan for two-person lift during installation or relocation. | All models |
| Locate the FilterMate within approximately 3-feet (1 meter) of the enclosure. p.14 ↗ | Proximity ensures proper hose routing and airflow. | All models when connected to enclosure |
| HEPA filters are leak checked at factory; a second leak check is recommended before using the FilterMate and at least annually thereafter. p.17 ↗ | Regular leak testing ensures containment integrity. | HEPA-filtered models |
| ULPA filter 3885700 is 99.999% efficient for particles 0.12 micron; PLEKX filter 3707904 is 99.97% efficient for 0.3 micron and contains 1.7 lbs carbon. p.18 ↗ | Alternative filter options for higher efficiency or combined particulate/vapor removal. | Models accepting these filter types |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Blower won’t turn on p.42 ↗ | Unit not plugged into outlet | Plug the FilterMate into appropriate electrical service | Chapter 7 |
| Blower won’t turn on p.42 ↗ | Circuit breaker(s) or Ground Fault Interrupter | Reset circuit breaker | Chapter 7 |
| Blower won’t turn on p.42 ↗ | Blower wiring is disconnected | Inspect blower wiring | Chapter 7 |
| Blower won’t turn on p.42 ↗ | Blower switch is defective | Replace switch | Chapter 7 |
| Blower won’t turn on p.42 ↗ | Motorized impeller is defective | Replace motorized impeller | Chapter 7 |
| Low face velocity or poor containment of contaminants p.42 ↗ | Enclosure sash not closed. | Close sash to the lowest position | Chapter 7 |
| Low face velocity or poor containment of contaminants p.42 ↗ | HEPA Filter clogged | Replace HEPA filter or increase speed | Chapter 7 |
| Low face velocity or poor containment of contaminants p.42 ↗ | Carbon filter loaded with chemicals | Replace carbon filter | Chapter 7 |