Rev l xpert balance enclosures xvs ventilation stations
Technical Reference Summary
Covered Models & Document Scope
Covers Labconco XPert Filtered Balance Enclosures and Protector XVS Ventilation Stations in widths 2', 3', 4', 5', 6' and 8', standard (22.75" high, 8" sash) and tall (32" high, 10" or 12" sash) variants, with depths of 29" and 35.5" (35.5" on 5', 6', 8'). Catalog numbers include 3930300-3930310 (3'), 3930400-3930410 (4'), 3971500-3971641 (5'/6'), 3971800-3971821 (8'), and XVS 4863010-4868012. Accessories covered: Guardian 500 and Guardian Digital 1000 airflow monitors, FilterMate Portable Exhausters, remote blowers, work surfaces, fluorescent light kits, exhaust transition adapters, dampers, storage cabinets, utility shelves, hoses and prefilter kits.
Section-by-Section Technical Breakdown
Defines the exhaust connection geometry, required exhaust volumes and static pressure losses for each enclosure size at face velocities of 60, 80 and 100 fpm, and the number of FilterMate Portable Exhausters required. The exhaust collar is 2" x 10" nominal; 2' through 6' enclosures use one connection, the 8' enclosure uses two. Transition adapters are available for 5" hose or 6" duct.
- Exhaust collar: 2" x 10" (5.1 cm x 25.4 cm) nominal; top or bottom connection (one open, one closed).
- 2' through 6' enclosures require one exhaust connection; 8' enclosure requires two.
- Face velocity range 60-100 fpm; exhaust volume and static pressure loss tabulated per model.
- FilterMate requires 5" Transition Adapter 3912400 (sold with FilterMate).
- Multiple FilterMates require balanced airflow; 5', 6' and 8' enclosures require two or three FilterMates depending on face velocity and size.
Lists Labconco accessory roof-mounted remote blowers with electrical ratings, dimensions, weights and CFM performance at various static pressures. Two frame sizes are offered: 1/4 hp direct drive with 6" ID inlet and 1/4 or 1/3 hp belt drive with 10.87" ID inlet. Explosion-proof versions are available.
- Direct-drive blower: 1/4 hp, 6" ID inlet, 4.25" x 7.38" OD rectangular outlet; 595 CFM at 0.0" S.P., 167 CFM at 0.87" S.P.
- Catalog 4863500: 115 V, 60 Hz, 4.4 A; 4863501: 115/230 V, 50 Hz, 5.6/2.8 A; 7053501 explosion-proof 115 V, 60 Hz, 4.4 A.
- Belt-drive blowers: 1/4 or 1/3 hp, 10.87" ID inlet, 5.5" x 10" OD outlet; adjustable sheave to change RPM.
- Catalog 7068200 (1/4 hp, 115 V, 4.4 A), 7068300 (1/4 hp EP, 4.5 A), 7068400 (1/3 hp, 6.1 A), 7068500 (1/3 hp EP, 6.4 A).
- If connected to house exhaust, an adjustable damper or valve must be installed to control airflow.
Describes FilterMate Portable Exhausters for filtered room exhaust or outside exhaust, with catalog numbers, voltages, filter types, airflow capacities and exhaust connections. HEPA and carbon filter part numbers and efficiencies are listed.
- FilterMate airflow: up to 280 CFM with HEPA filter, up to 220 CFM with HEPA/Carbon combination, when connected with 8' of 5" diameter hose.
- Catalog 3970000-3970004 (115 V/60 Hz) and 3970020-3970024 (230 V/50 Hz).
- 3970002 (115 V) or 3970022 (230 V) recommended when outlet connects to house exhaust; includes 6" nominal thimble connection.
- HEPA filter Part #3707900: 99.99% efficient for particles 0.3 micron.
- HEPA filter with Bag-In/Bag-Out bag Part #3779002.
- Carbon filters: 3923400 organic vapors (12 lb activated), 3923401 formaldehyde (14 lb treated), 3923402 ammonia (16 lb treated).
Specifies nearby duplex receptacles for FilterMate, airflow monitor, light and balance, and describes auxiliary switched outlets on FilterMate and fluorescent light kits. Recommends connecting the airflow monitor to the same switched circuit as the blower.
- FilterMate auxiliary switched outlet rated 8 amps.
- Fluorescent light kit auxiliary outlet can be used for any accessory under 8 amps.
- Airflow monitor should be connected to FilterMate auxiliary switched outlet so monitor is ON when blower is ON.
- For other blower systems, airflow monitor should be switched by the same circuit as the blower exhaust.
Covers unpacking, moving and mounting the enclosure on a work surface. The work surface must be level, rigid, non-porous and chemically resistant, with a dished recessed area. The enclosure front must align within 1/8" of the work surface front and is secured with four screws.
- Enclosure weight: 50 to 200 lbs (22 to 90 kg).
- Do not move the enclosure by tilting it onto a hand truck; use a flat low dolly or floor jack.
- Do not lift by the front Air Foil.
- Work surface should be at least as wide as the enclosure; front alignment within 1/8".
- For stainless steel work surfaces use #10-24 x 3/8" hex machine screws (Labconco P/N 1885806); allow minimum 1/8" clearance under screw head.
- Seal enclosure to work surface with a bead of silicone sealant.
Describes connecting the enclosure to outside exhaust, FilterMate or thimble systems. The exhaust ductwork weight must be supported independently of the enclosure. Transition adapters are available for top or bottom exhaust and for 5" hose or 6" duct.
- Exhaust ductwork weight must be supported independently of the enclosure superstructure.
- Exhaust system should be installed by a qualified HVAC contractor.
- 2' through 6' enclosures accept one 2" x 10" transition adapter; 8' enclosure has two connections.
- Labconco manufactures transition adapters for top or bottom exhaust and for 5" dia. hose or 6" dia. duct.
- Duct material should match the enclosure, procedures and chemical applications.
Explains how to measure average face velocity at the sash opening and states factory ASHRAE 110-1995 test results. Face velocity readings are taken in two rows across the sash opening, 6" from ends and every 12", with rows 3" from upper sash foil and 3" above work surface.
- Recommended average face velocity: 60 to 100 feet per minute.
- Factory ASHRAE 110-1995 rating: less than 0.05 ppm at 4 L/min (AM <0.05).
- Enclosure should be validated whenever relocated, serviced, or at least annually.
- Face velocity profiles and smoke testing should be performed frequently per organization's quality system.
Describes the aerodynamic design elements that provide containment at low flow velocities: Clean-Sweep air foil, upper sash foil, upper dilution air supply, zoned rear perforated baffle, side entry air foils, flush sash, electrical pass-through iris and accessory connections.
- Upper dilution air supply introduces 5 to 7 percent of required air volume.
- Internal depth of 23" supports balances without extending outside enclosure.
- Two enclosure heights: 22.75" or 32"; depths 29" and 35.5" (35.5" on 5', 6', 8').
- Zoned rear perforated baffle has three sections with increasing airflow at bottom; removable or tilt-down for cleaning.
- Flush sash has wiping seal and spring-loaded latch or gas spring; pivots down for operation.
- Electrical pass-through iris allows cords through rear without large opening; solid plugs shipped, iris plugs included.
Provides step-by-step operating procedure: planning, start-up, loading, work techniques, final purging, unloading and shutdown. Key timing requirements include 1 minute purge after start-up and loading, and 2-3 minutes purge before shutdown.
- Allow enclosure to operate unobstructed for 1 minute after start-up.
- After loading, wait one minute to purge airborne contaminants.
- Upon completion, allow 2 to 3 minutes undisturbed operation before shutting down blower.
- Only raise sash for loading.
- Keep all materials inside the lower air foil and work as far to the rear as possible.
- Surface decontaminate objects before removal; cover open trays and containers.
Defines weekly, monthly and annual maintenance tasks including cleaning, airflow verification, face velocity measurement, baffle inspection, lamp replacement and annual certification.
- Weekly: clean interior; operate exhaust and note airflow with visible smoke.
- Monthly: measure face velocity with calibrated thermal anemometer; clean exterior; check rear baffle for blockage.
- Annually: replace fluorescent lamps; have enclosure validated by qualified certification technician.
- Airflow alarms recommended for constant monitoring.
Covers the Guardian 500 Airflow Monitor Kit No. 3944700 or 3944701: component identification, installation, calibration and alarm activation. The monitor provides visual and audible alarms for abnormal airflow.
- Green light indicates normal flow; flashing red light with audible alarm indicates low flow below set point.
- Kit numbers 3944700 or 3944701.
- Permanently installed device for continuous airflow monitoring.
- Calibration procedure and alarm activation described in Chapter 6.
Covers Guardian Digital 1000 Kit No. 3908800 or 3908801: operation, installation, calibration and alternate calibration for constant volume conditions. The monitor displays face velocity, provides RS232 output, night setback mode and auxiliary relay ports.
- Kit numbers 3908800 or 3908801.
- Displays face velocity value.
- Provides RS232 output, night setback mode and several auxiliary relay ports.
- Alternate calibration procedure for constant volume conditions described.
Lists replacement latch kits for the enclosure sash. Specific part numbers and procedures are provided in Chapter 6.
- Replacement latch kits available for sash latch mechanism.
- Refer to Chapter 6 for part numbers and installation.
Lists and describes available accessories: work surfaces, airflow monitors, fluorescent light kits, exhaust transition adapters, dual exhaust connectors, remote blowers, exhaust dampers, FilterMate Portable Exhausters and filters, storage cabinets, utility shelf kits, hose and hose kits, and prefilter kits.
- Work surfaces available in epoxy resin or stainless steel.
- Fluorescent light kit includes auxiliary outlet for accessories under 8 amps.
- Exhaust transition adapters for 5" hose or 6" duct, top or bottom.
- Exhaust dampers for house exhaust systems.
- Storage cabinets, utility shelf kits and holders available.
- Hose, hose clamps and hose kits available.
- Prefilter kits available.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Recommended average face velocity at sash opening is 60 to 100 feet per minute. p.21 ↗ | Sets the airflow target for blower selection and field validation; below 60 fpm containment may be compromised. | All XPert and XVS models |
| Exhaust connection accepts a 2" x 10" (5.1 cm x 25.4 cm) nominal collar; 2' through 6' enclosures have one connection, 8' enclosure has two. p.10 ↗ | Ductwork and transition adapters must match this collar size; 8' units require dual exhaust runs. | All models; 8' uses two connections |
| For a 4' standard XPert/XVS at 60/80/100 fpm, exhaust volume is 150/200/250 CFM with static pressure loss 0.06/0.10/0.16 in. w.g. p.10 ↗ | Blower must deliver at least these CFM at the corresponding system static pressure. | 4' XPert/4' XVS standard height |
| For a Tall 8' XPert at 100 fpm, exhaust volume is 775 CFM with static pressure loss 0.42 in. w.g. and requires 3 FilterMates. p.10 ↗ | Largest airflow demand in the series; house exhaust or multiple FilterMates needed. | Tall 8' XPert |
| FilterMate Portable Exhauster delivers up to 280 CFM with HEPA filter or up to 220 CFM with HEPA/Carbon combination when connected with 8' of 5" diameter hose. p.12 ↗ | For airflows above these limits, multiple FilterMates must be used and balanced. | FilterMate models 3970000-3970024 |
| HEPA filter Part #3707900 is 99.99% efficient for particles 0.3 micron. p.12 ↗ | Suitable for particulate containment; verify compatibility with chemical vapors (use carbon filter). | FilterMate with HEPA filter |
| FilterMate auxiliary switched outlet is rated for 8 amps; fluorescent light kit auxiliary outlet can be used for any accessory under 8 amps. p.13 ↗ | Do not exceed 8 A total load on these convenience outlets. | FilterMate and fluorescent light kit |
| Enclosure models weigh between 50 and 200 lbs (22 to 90 kg). p.14 ↗ | Work surface and supporting structure must safely support this weight plus equipment; use mechanical lift for moving. | All models |
| The front of the enclosure should be aligned within 1/8" of the front of the work surface. p.15 ↗ | Critical for proper airflow entry and sealing; misalignment can affect containment. | All models |
| For stainless steel work surfaces use #10-24 x 3/8" hex machine screws (Labconco P/N 1885806); allow minimum 1/8" clearance under screw head. p.16 ↗ | Specify correct fasteners for stainless steel work surfaces; epoxy surfaces use supplied mounting screws. | Stainless steel work surfaces |
| Upper dilution air supply introduces 5 to 7 percent of the required air volume. p.25 ↗ | Part of total exhaust volume; blower sizing must account for this bypass air. | All XPert and XVS models |
| Internal depth of 23" supports modern balances without extending outside the enclosure. p.26 ↗ | Verify balance dimensions fit within 23" internal depth. | All models |
| Two enclosure heights available: 22.75" or 32"; depths 29" and 35.5" (35.5" on 5', 6', 8'). p.26 ↗ | Tall height typically used for pipet operations, titration or taller equipment. | All models |
| Factory ASHRAE 110-1995 rating: less than 0.05 ppm at 4 L/min (AM <0.05). p.21 ↗ | Baseline containment performance; field certification should confirm similar performance. | All XPert and XVS models |
| Elevate large bulky objects 3/4" on blocks to allow airflow under the object and into rear baffle exhaust slots. p.28 ↗ | Prevents airflow blockage and turbulence; blocks must be level and secured. | All models |
| Allow enclosure to operate unobstructed for 1 minute after start-up and after loading; allow 2 to 3 minutes before shutdown. p.29 ↗ | Purge times ensure contaminants are removed before work and before blower shutdown. | All models |
| Perchloric acid use and high-level radioisotope materials are prohibited in this enclosure. p.28 ↗ | Do not use for these applications; select appropriate specialized enclosure. | All XPert and XVS models |
| If connected to house exhaust, an adjustable damper (or valve) must be installed to control airflow properly. p.11 ↗ | Required for balancing and maintaining face velocity, especially with multiple enclosures on one system. | House exhaust installations |
| Remote blower 4863500: 115 V, 60 Hz, 4.4 A, 1/4 hp, 6" ID inlet, 595 CFM at 0.0" S.P., 167 CFM at 0.87" S.P. p.11 ↗ | Select blower based on required CFM at system static pressure; performance drops significantly at higher static pressure. | Remote blower 4863500 |
| Maintain humidity level between 45 and 60% to control static electricity in balance enclosures. p.8 ↗ | Static charges can cause weighing errors; ionizers may be used if humidity control is insufficient. | XPert Balance Enclosures |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Contaminants outside of enclosure p.53 ↗ | Improper user techniques for the enclosure. | See "Certifying the Hood" Chapter 3 and "Safety Precautions" Chapter 4 sections in the manual. (Ref. Appendix D) | Chapter 8 Troubleshooting |
| Contaminants outside of enclosure p.53 ↗ | Restriction of the baffle air slots or blockage of the exhaust outlet. | Remove obstruction to ensure that all air slots and the exhaust outlet are unobstructed. | Chapter 8 Troubleshooting |
| Contaminants outside of enclosure p.53 ↗ | External factors are disrupting the enclosure airflow patterns or acting as a source of contamination. | See "Location Requirements" Chapter 2, "Certifying the Enclosure" Chapter 3, and "Safety Precautions" Chapter 4 sections of this manual. (Ref. Appendix D) | Chapter 8 Troubleshooting |
| Contaminants outside of enclosure p.53 ↗ | Enclosure has improper face velocity. | Have enclosure certified and check exhaust system. Check FilterMate filters for loading. Adjust FilterMate speed control. Enclosure should have average face velocity of 60-100 fpm. | Chapter 8 Troubleshooting |
| Airflow monitor malfunction. No power. No lights. No display. p.53 ↗ | Power supply is not plugged into proper voltage; plug in power supply. Verify that airflow monitor interface cables are connected. Check fuses on FilterMate or accessory light. | Plug in power supply to proper voltage; verify interface cables; check fuses. | Chapter 8 Troubleshooting |
| No audible alarm. p.53 ↗ | Alarm has been temporarily silenced using "test/reset" or "enter" buttons. | Check alarm status; alarm may be silenced. | Chapter 8 Troubleshooting |
| Wrong alarm set point. p.53 ↗ | Airflow monitor was not properly adjusted. | Repeat calibration steps outlined in the airflow monitor manual. | Chapter 8 Troubleshooting |
| Constant audible alarm. p.53 ↗ | Check airflow and calibration of airflow monitor. | Check airflow and calibration of airflow monitor. | Chapter 8 Troubleshooting |
| Continuous alarm. p.53 ↗ | Check the face velocity of the enclosure as the airflow of the system may have changed. | If face velocity is correct, calibrate the airflow monitor outlined in the Users Manual. | Chapter 8 Troubleshooting |
| Monitor alarms; air way to airflow monitor sensor is blocked by insects, dust or debris. p.53 ↗ | Airway to airflow monitor sensor is blocked by insects, dust or debris. | Lightly clean the airway with clean air. Be careful not to touch sensitive electrical components. | Chapter 8 Troubleshooting |
| Audible disable will not stay ON p.53 ↗ | An alarm condition must be continuously present before the audible alarm can be silenced. If flow conditions fluctuate near the alarm set point, the airflow monitor will automatically reset itself. | Action should be taken to bring the enclosure airflow into proper operating parameters or adjust the alarm set point lower. | Chapter 8 Troubleshooting |