NovoCyte Quanteon Flow Cytometer User Guide

Contents

  1. Overview pp. 6-9
    System overview, fluidics cart connection, compensation and channel configuration
  2. Overview pp. 11-15
    Front matter, glossary of terms and introductory material
  3. Warning pp. 16-28
    General, electrical and laser safety, regulatory compliance and limitations
  4. Overview pp. 29-50
    System requirements, components and Fluidics Station II introduction
  5. Procedure pp. 51-56
    Space requirements, instrument installation, fluidics cart and power connection
  6. Procedure pp. 57-70
    Configuration verification, QC tests and routine instrument operation
  7. Procedure pp. 71-106
    Experiment creation, acquisition settings, compensation, data analysis and file handling
  8. Procedure pp. 107-132
    Preventative and unscheduled maintenance, consumable replacement and calibration
  9. Procedure pp. 133-136
    Troubleshooting tables for common instrument issues
  10. pp. 137-138
    Document revision history
  11. Config pp. 139-142
    Laser and detection channel configuration appendix
  12. Spec pp. 143-146
    Instrument, workstation and monitor technical specifications
  13. p. 139 table, 4 rows
  14. p. 143 table, 15 rows
  15. p. 146 table, 3 rows
  16. p. 125 table, 9 rows
  17. p. 145 table, 19 rows
  18. p. 144 table, 19 rows

Table 13 Laser Wavelength and Detectable Fluorochromes on NovoCyte Quanteon

Laser Wavelength (nm)Typical Fluorescent Fluorochromes
405Pacific Blue, AmCyan, Pacific Orange, Qdot 605, Qdot 655, Qdot 800, BV711
488FITC, EYFP, PI, PerCP, PerCP-Cy5.5, PerCP-eFluor710
561PE, PE-Texas Red, PE-Cy5, PE-Cy5.5, PE-Alexa Fluor 700, PE-Cy7
637APC, Alexa Fluor 680, Alexa Fluor 700, APC-Cy7

Table 15 NovoCyte Quanteon Technical Specifications1

ParameterSpecification
Optics
LaserSolid state laser with on-board thermal-electric cooling and guaranteed thermal stability and beam quality
Laser Power100 mW
Laser Beam Profile10 x 60 µm elliptical beam
Optical Alignment ProcedureFixed, no operator alignment required
Fluorescence DetectionSilicon Photomultiplier (SiPM) with high photon detection efficiency; Individual photodetector for each channel
FSC/SSC SensitivityFSC: 0.4 µm; SSC: 0.1 µm
Fluorescence Resolution<3% CV for CEN
Optical FiltersUser exchangeable, “Smart” filter automatically read by the system
Cell Size0.1 - 50 µm
Fluidics
Flow Cell170 x 290 µm rectangular quartz flow cell
Sample Acquisition Rate100,000 events/second
Sample DeliveryPositive-displacement syringe pump enabling direct absolute count without the need for reference counting beads.
Volumetric Absolute Count Precision< 5%

1 Сноска в тексте страницы не раскрыта.

Table 16 NovoCyte Workstation and Monitor Specifications

Specification
NovoCyte WorkstationSmall form factor (SFF). Intel Core i7 CPU or above, 16G RAM or above, 512G sold state drive or above.
MonitorHigh resolution flat panel with resolution of 1,920 x 1,080 or above
Computer Operating SystemMicrosoft Windows 10 Professional (64 bit) or above with Microsoft Office and NovoExpress software pre-installed

Table 10 Unscheduled Maintenance and When to Perform

Unscheduled MaintenanceWhen to Perform
Debubble• The CV of the data is larger than normal. • QC test failed. • No events showed up after sample collection started for certain period of time. • Instrument was not used for over one week.
Clean the External Surface of the InstrumentKeep the external surface of the instrument clean.
DecontaminationWhen suspecting contamination in the system.
Prime Fluidics System before Long-Term StorageNo plan to run the instrument for more than two weeks.
Prime Fluidics System after Long-Term StorageInstrument has not been used for more than two weeks.
Purge Fluidics System before ShipmentThe instrument needs to be shipped.
Calibrate the NovoCyte Fluidics Station II or NovoCyte Fluidics CartAbnormal warning about the liquid level in the reagent containers is observed.
Clean the Optical FiltersThe optical filters get dirty.
Clean Fluidics System• The CV of the data is larger than normal. • QC test failed. • Excessive debris in display.

Table 15 NovoCyte Quanteon Technical Specifications1

CharacteristicSpecification
User ManagementAdministrative creation of individual user accounts and user groups with privilege control. Configurable roles for individual users. User operation time tracking.
Quality Control
Instrument Daily QCAutomatic daily quality control with Agilent Flow Cytometer QC Particles. Levi-Jennings Plot.
Production QCEach instrument is verified for assembly integrity and performance to specifications.
Sampling
Manual Sample Loading12 X 75 mm tube, 1.5 mL or 2.0 mL tube with Agilent tube holder.
Automatic Sample Loading with NovoSampler QAgilent Flow Cytometer 40 tube rack for 12 x 75 mm tube, 24-well, 48-well, 96-well (flat, V-, U-bottom), 384-well microtiter plates, Agilent Flow Cytometer 24 tube cooling rack for 12 x 75 mm tube, Agilent Flow Cytometer 96-well (flat, U-, V-bottom) plate cooling box.
Operating Conditions
Instrument Dimension (W × D × H)24.4 × 18.1 × 18.8 in (62.0 × 46 ×48.0 cm) 33.5 × 18.1 × 18.8 in (85.0 × 46.0 ×48.0 cm) with NovoSampler Q
Instrument Weight104.0 lb (47.0 kg) 133.0 lb (60.0 kg) with NovoSampler Q
Operating EnvironmentTemperature:15-32℃, Relative Humidity: 80% maximum, Atmosphere Range: 76-106 kPa
Storage EnvironmentTemperature:1-40℃, Relative Humidity: 80% maximum
Power Requirements100-240 VAC, 50-60 Hz
Regulatory Compliance
Indoor and Outdoor UseIndoor use only
Overvoltage CategoryClass II
Equipment ClassClass I Laboratory Equipment
Pollution degree of the intended environment2
Degree of Ingress protectionIPXO

1Specification and performance claims would be based on certain specific experiment conditions.

Table 15 NovoCyte Quanteon Technical Specifications1

SpecificationValue
Sample Flow Rate5-120 µL/min, continuously adjustable
Average Sheath Flow Rate6.5 mL/min
Sample Aspiration Volume5 µL -5 mL
Sample RecoveryAvailable
Compatibility to AutosamplerNo fluidic tubing disassembly or re-connection required.
Fluid Level SensingActive sensing using weight sensors with automated warnings when any fluid level is out of specified range.
Fluid Container Capacity3 L sheath, 3 L waste, 500 mL cleaning, 500 mL rinse. Optional large container for sheath (15 L) and waste (15 L) with Fluidics Cart.
Carryover< 0.1%
Sample Injection Probe (SIP) RinseAutomated flying collar wash of inner and outer SIP surface after each sampling.
Fluidics System MonitoringIn-line pressure sensor monitors the pressure in real time. Automated system recovery when pressure is out of range due to clogging.
Fluidics MaintenanceAutomated startup and shutdown with fluidic system cleaning. Automated user executable cleaning, debubble, rinse, unclog, priming and decontamination. Automated execution of user definable maintenance sequence.
Data Acquisition SoftwareNovoExpress
ParametersHeight and Area for FSC, SSC and all Fluorescent Channels, Width off FSC and Time.
Dynamic Range24 bits; 7.2 decades logarithmic scale
Fluorescence Photodetector Gain Control0-1000 User adjustable, optimized default gain setting for each individual channel.
CompensationFull inter-beam matrix, during or post acquisition.
File ExportFCS 3.0, FCS 3.1; CSV; Batch PDF report
Data ReportAutomatic report. Customizable report. Batch PDF report.
Usage MonitorComprehensive Transaction Log and System Log.

1 Сноска в тексте страницы не приведена.

Engineering Facts

14 facts · every claim linked to a page
Safety

The instrument must be operated only by personnel trained by Agilent or an authorized representative, so plan operator training before commissioning.

Determines who may legally and safely run the system after installation.

Source
“Agilent flow cytometers and associated accessories are limited to be operated by professionals trained by Agilent Technologies or an authorized representative.”
Installation

The unit weighs over 23 kg, so plan a two-person lift or mechanical aid during site installation.

Affects rigging, manpower and floor/site planning for installation.

Source
“This product is large and the weight is more than 51 lb (23 kg).”
Installation

Moving the instrument after installation is discouraged and requires Agilent professionals to re-install it, so fix the final location before commissioning.

Drives layout decisions and avoids costly re-installation and data-quality risk.

Source
“moving the instrument is not recommended. If you need to move the instrument, be sure to contact Agilent Technologies professionals.”
Safety

The instrument is a Class I laser product with lasers fully enclosed, so no special laser work area is needed except during service.

Simplifies lab laser-safety classification and room requirements at selection.

Source
“Agilent flow cytometer is a Class I laser product according to IEC/EN 60825-1:2014.”
Safety

Optical covers and laser shielding must never be removed by the user, so service access must be planned through Agilent only.

Defines user-permitted service boundaries and laser safety rules on site.

Source
“Modification or removal of the optical covers or laser shielding could result in exposure to hazardous laser radiation.”
Installation

Reagent containers are color-coded and position-coded; wrong placement risks damaging pressure sensors, so verify labeling during setup.

Prevents fluidics station damage and misconnection during installation.

Source
“Please ensure that the container position in the NovoCyte Fluidics Station II is correct”
Installation

Fluidics tubing is color-coded by function (blue rinse, green sheath, yellow clean, red waste), so connect ports accordingly.

Ensures correct plumbing hookup and avoids cross-contamination at install.

Source
“NovoRinse is the rinsing solution and is connected by the blue tubing.”
Operation

Optical configuration should be verified before each experiment and can be changed by swapping filters, requiring a QC test to take effect.

Affects daily workflow and panel-to-configuration matching decisions.

Source
“It is recommended for users to view and verify the current optical configuration before each experiment.”
Operation

If a QC channel is Acceptable or Failed, at least 400 µL must remain in the tube and the test repeated with automatic gain re-adjustment.

Defines minimum QC sample volume and repeat procedure for valid QC.

Source
“Ensure there is at least 400 μL sample remaining in the sample tube.”
Operation

Threshold values are sample-type dependent, with recommended FSC-H ranges given per cell type and microsphere size.

Guides initial threshold setup to avoid debris and signal loss.

Source
“setting the FSC-H threshold as 100,000 ensures the correct collection of the leukocyte signals”
Operation

For 0.1 µm microspheres the SSC gain must be increased to achieve resolution, so low-size particle work needs gain adjustment.

Sets expectations for small-particle detection performance and setup.

Source
“To detect microsphere with 0.1 µm diameter, the photodetector gain of SSC needs to be increased.”
Operation

Default auto-compensation acquisition uses 10,000 events or 20 µL, slow flow, and FSC-H threshold above 100,000.

Defines default compensation run settings engineers must know for setup.

Source
“Stop condition: total 10,000 events or 20μL sample volume.”
Operation

Compensation samples should be prepared at 1–2 × 10^6 /mL for correct automatic compensation.

Affects sample prep protocol and validity of compensation matrix.

Source
“it is recommended to prepare the compensation sample concentration to be 1 x 10^6 ~ 2 x 10^6 /mL.”
Maintenance

The sample injection probe moves downward after confirming, so fingers must be kept clear during probe cleaning.

Pinch/injury hazard during routine probe maintenance.

Source
“Sample injection probe will move downward after clicking the OK button.”