ATH 2804 M — instruction manual

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum ATH 2804 and ATH 3204 magnetically levitated turbomolecular pumps (M = version without temperature management system; MT = version with TMS). Applies to part numbers TMAxxxx4xx00 / TMAxxxx4xx5A (light-duty) and TMAxxxx8xx01 (heavy-duty); part numbers not ending in '00'/'01' are customer-specific variants documented separately. Control interface options: Remote, EtherCAT, Profibus, DeviceNet, LonWorks; options include purge solenoid valve and air inlet (venting) solenoid valve. HHR hand-held remote is an accessory ordered separately.

Section-by-Section Technical Breakdown

pp. 17-20
Product description and electronic drive unit

Magnetically levitated (AMB) turbopumps with integrated electronic drive unit for coating, dry-etching plasma, glass coating and R&D. Rotation speed is selectable between minimum and nominal (stand-by speed) to optimize e.g. high-pressure pumping. M version for clean applications; MT version with TMS for applications where condensation must be minimized.

  • Scope of delivery: pump with integrated drive unit, splinter shield in inlet flange, 2 lifting rings, label sheet, manual; optionally copper gasket, flow reduction device, screw set, purge solenoid valve, air inlet solenoid valve
  • Drive unit as standard: RS-232/RS-485 serial link, SERVICE connector for HHR, mains switch, status lights; optional REMOTE / EtherCAT / Profibus / DeviceNet / LonWorks
  • Drive unit powers and controls the water solenoid valve and heating (MT) and can power purge/air-inlet valves (options)
  • Mains connection rating on unit: 200-240 VAC, 50/60 Hz, 10 A
pp. 21-34
Installation (mechanical, water, nitrogen, valves, electrical)

Frame must absorb kinetic energy of a sudden rotor blockage (ISO 27892-verified against crash and bursting); maximum loads perTbl. 1. Mandatory screw types/torques for the inlet flange perTbl. 2. The pump operates in any position but must be connected to a backing pump; magnetic fields up to 0.5 mT guaranteed, up to 5 mT conditionally, >5 mT requires shielding; radiation tolerance 10³ Gy.

  • Rotor-blockage loads (ATH 2804/3204, 25000 min⁻¹, rotor split in two): energy 160 kN·m; torque <50 kN·m (3 ms, 1 ms delay); bending moment <13 kN·m (1.3 ms); axial force <147 kN (<1 ms, only if inlet blocked by rigid parts)
  • No significant axial load observed with standard pendulum valve or free space >100 mm above inlet
  • Inlet flange fastening (mandatory): DN250 CF-F: 32× M8 gr.12-9, 20 N·m each, ≥35 mm long, total clamping 388000 N; DN250 ISO-F: 12× M10, 30 N·m, 197000 N; DN300 CF-F: 32× M10, 30 N·m, 491000 N; DN320 ISO-F: 12× M12, 50 N·m, ≥40 mm, 283000 N; VG 250/300/350: 12× M12, 50 N·m, ≥45/≥40 mm, 283000 N
  • Pump enclosure withstands 11·10³ hPa static overpressure for 1 minute (when connected per manufacturer's recommendations)
  • Connected chamber/parts must withstand −1·10⁻³ hPa relative to atmospheric pressure
  • Install NC isolation valve between turbopump and backing pump
  • Use softened/non-aggressive water; install inlet filter if solid pollution specs unmet; unregulated mains water causes limescale clogging; label 'WATER MAX Pr 7 bar/101 psi'
  • Install LEL detection at 25% LEL in extraction system when pumping flammable gases
  • Mechanical installation always before electrical connection; leak-test the whole line after installation
  • External vibrations/shocks can overload magnetic bearings and stress landing bearings
pp. 35-38
Operation (start-up, shutdown, monitoring)

Start/stop procedures including power-on, pumping start-up, restart after emergency stop, pumping shutdown, power-off, prolonged stoppage, and status monitoring via LED indicators (Tbl. 3).

  • Do not remove the pump connector while the pump is running; lock the connector before use
  • Before working on the inlet flange, wait until the rotor is fully immobilized
  • Pump cannot evacuate at atmospheric pressure; transient start at atmospheric pressure is permissible
pp. 39
Advanced settings

Pump temperature management system (MT) and pump braking functions configured via the drive unit.

  • Temperature management (MT version) controlled by the integrated electronic drive unit
  • Pump braking function available via the HEATING/BRAKING cable
pp. 40-53
Interfaces for control (HHR, REMOTE, RS-232/RS-485, fieldbus)

Control via hand-held remote (DISPLAY/SETUP/SER NUM menus), REMOTE connector with logic inputs (START/STOP/STAND-BY) and dry-contact logic outputs, or RS-232/RS-485 serial link (command list; RS-485 allows multi-pump installations). Fieldbus control is covered in separate manuals.

  • REMOTE connector supports control by direct voltage (Fig. 12) or dry contacts (Fig. 13) plus logic outputs (Fig. 14)
  • RS-485 enables simultaneous installation and control of several pumps
  • Fieldbus interfaces (Profibus, EtherCAT, DeviceNet, LonWorks) documented in separate operating instructions
pp. 54-58
Maintenance and exchange

Maintenance safety instructions, maintenance frequency, on-site maintenance, and the exchange procedure: disconnection from the installation, draining the water circuit, and preparing the pump for shipping.

  • Only trained personnel; nitrogen purge before maintenance on inlet, pump and exhaust lines
  • Exchange: disconnect pump, drain water circuit, prepare for shipping (reuse blanking plates/screws)
pp. 59
Decommissioning and disposal

Shutting down for longer periods, recommissioning procedure, and disposal of the product.

  • Storage after use: see 'Shutting down for longer periods' section
  • RoHS: product components contain no lead (label 'Pb')
pp. 60-63
Malfunctions and fault indication

Malfunction and fault indication (LED behavior) with a malfunction table for diagnosis and remedies.

  • Faults indicated via pump status indicator lights on the drive unit and via the control interfaces
pp. 64-68
Accessories, spare parts and service

Lists accessories (incl. HHR hand-held remote, solenoid valves, screw sets), spare parts, and Pfeiffer Vacuum service solutions.

  • Only Pfeiffer-approved original parts permitted for connection to the equipment
  • Recommended to use the manufacturer's screw set for inlet flange securing
pp. 69-74
Technical data and dimensions

General data, technical/environmental/cooling water/nitrogen/electrical characteristics, pressure and throughput unit conversion tables, and dimension drawings for M and MT versions plus purge and air-inlet solenoid valves.

  • Nominal rotor speed: 25000 min⁻¹ (rating plate: 99 kg, 25000 rpm, M:69 Vrms, 833 Hz, 15 A for ATH2804/3204M example)
  • Includes conversion tables for pressure units and gas throughput units
  • Separate dimension figures for M (Fig. 15) and MT (Fig. 16) versions and solenoid valves (Fig. 17/18)

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Nominal rotation speed is 25000 min⁻¹; the operating speed is selectable between a minimum speed and nominal speed (stand-by speed) for application optimization. p.17 ↗ Allows tuning pumping performance for high-pressure operation without hardware change; set stand-by speed via HHR/serial commands. whole series
Rotor sudden-blockage design loads: kinetic energy 160 kN·m, torque <50 kN·m (3 ms duration), bending moment <13 kN·m (1.3 ms), axial force <147 kN (<1 ms, only if inlet blocked by rigid parts). p.23 ↗ Equipment frame and flange fasteners must be dimensioned to absorb these transient loads to prevent pump separation. ATH 2804 and ATH 3204 at nominal speed
DN250 CF-F inlet: 32× M8 screws grade 12-9, 20 N·m each, length ≥35 mm, total clamping force 388000 N; DN300 CF-F: 32× M10, 30 N·m, 491000 N; DN320 ISO-F: 12× M12, 50 N·m, ≥40 mm, 283000 N; VG250/VG300/VG350: 12× M12, 50 N·m, ≥45/≥40 mm, 283000 N. p.23 ↗ Screw type, number, grade and torque are mandatory; under-torque risks loosening, over-torque damages screws. whole series, per inlet flange type
Pump enclosure withstands maximum static overpressure of 11·10³ hPa for 1 minute in case of incident (when connected per manufacturer's recommendations). p.21 ↗ Defines the housing burst/overpressure design limit for safety case calculations of the integrated installation. whole series
Magnetic field: guaranteed operation in uniform fields up to 0.5 mT; 0.5–5 mT operation depends on cooling and gas loads; >5 mT causes excessive rotor heating and requires shielding. p.24 ↗ Verify ambient magnetic fields near magnets/induction equipment; add magnetic shielding otherwise. whole series
Standalone pump withstands radiation levels up to 10³ Gy; operation in areas with ionizing radiation is otherwise misuse. p.24 ↗ Acceptable for most industrial environments but excluded from high-radiation (e.g. nuclear/accelerator) installations unless verified. whole series
The pump cannot evacuate at atmospheric pressure and must be connected to a backing pump; transient start at atmospheric pressure is permissible. p.24 ↗ Always spec a compatible backing pump with an NC isolation valve between turbopump and backing pump. whole series
Connected chamber/parts at the inlet must withstand a negative pressure of 1·10⁻³ hPa relative to atmospheric pressure. p.26 ↗ Chambers, valves and upstream components must be rated for near-full vacuum before connection. whole series
Water circuit maximum pressure labeled 7 bar / 101 psi ('WATER MAX Pr 7 bar/101 psi'); softened or non-aggressive water required, with inlet filter if solid pollution limits cannot be met. p.27 ↗ Use a regulated water supply; unregulated mains water causes limescale clogging requiring full circuit overhaul. whole series (water-cooled)
Manual lockout valves must be installed within 3 m of the product on both the cooling water and nitrogen supply circuits. p.11 ↗ Integrators must include these valves and apply LO/TO before any pump work. whole series
When pumping flammable gases, an electromechanical LEL detection device (alarm at 25% LEL) must stop chemical supply to the pump; oxidant concentration must not exceed 60% of the MOC (NFPA 69-2019); safety systems per EN 50495. p.12 ↗ Mandatory for semiconductor/PV/display installations using flammable or pyrophoric gases; risk assessment is the user/integrator's responsibility. applications with flammable/pyrophoric gases
Drive unit mains supply: 200–240 VAC, 50/60 Hz, 10 A; example pump rating: 69 Vrms, 833 Hz, 15 A at 25000 rpm; pump weight example 99 kg (rating plate ATH2804/3204M). p.18 ↗ Size mains cabling/breakers at ≥10 A and use lifting gear rated for ~100 kg per arm loads. whole series (per rating plate)
Handling requires multi-strand straps minimum 500 mm long with load per arm >100 kg; pump must always be transported upright. p.16 ↗ Crane/hoist capacity and rigging must be specified accordingly; horizontal transport is prohibited. whole series
Axial impact load is negligible with a standard pendulum valve above the inlet or with >100 mm free space; the <147 kN maximum occurs only when the inlet is blocked by rigid parts. p.22 ↗ Install a pendulum valve or maintain ≥100 mm clearance above the inlet to reduce frame loading requirements. whole series
Free storage of a new pump is limited to a maximum of 1 year in a clean, dry, non-contaminated area, in protective envelope with blanking plates fitted. p.16 ↗ Plan stock rotation/recommissioning checks for pumps stored longer than one year. whole series
Hot surfaces >65 °C are present depending on operating conditions; gloves per EN ISO 21420 required, and pipes/cables must be kept away from surfaces >70 °C. p.12 ↗ Route cables/pipework with thermal clearance and define PPE requirements for operators. whole series
Standard control interfaces: RS-232/RS-485 serial link on all units (RS-485 supports multiple pumps); REMOTE connector (logic inputs/outputs) optional; fieldbus options EtherCAT, Profibus, DeviceNet, LonWorks. p.19 ↗ Select the drive-unit interface variant at ordering; fieldbus configuration requires the corresponding separate manual. whole series, per ordering code
MT version includes a temperature management system (heating band, water solenoid valve) to minimize condensation when pumping condensable fluids, provided condensation does not occur inside the pump; M version has no TMS and pumping condensables is classified as misuse. p.17 ↗ Choose MT variant for processes with condensable gas risk; M variant is restricted to clean, non-condensing applications. M vs MT variants
Safety and installation conformity verified per ISO 27892 (rotor crash/bursting tests) and Machinery Directive 2006/42/EC Annex I / EN ISO 12100 risk assessment. p.21 ↗ Integrator can reference these standards for the machinery safety file of the pumping system. whole series