ATP 1603 M — instruction manual

Technical Reference Summary

Covered Models & Document Scope

Operating instructions for the Pfeiffer Vacuum ATP 1603 M magnetically levitated (active magnetic bearing, AMB) turbomolecular pump, part numbers TMDBXXX4XXXX, for light gas (H2/He) pumping on clean applications. M designation = version without Temperature Management System (no TMS). Driven by the OBC electronic drive unit (OBC V4 instructions, part number 129285). Options per ordering guide: copper gasket, purge solenoid valve, air inlet solenoid valve, purge flow reduction device (50 sccm), screws/bolts; inlet flange variants DN200 ISO-F or DN200 CF-F. HHR hand-held remote is NOT included (ordered separately via accessories).

Section-by-Section Technical Breakdown

pp. 17-18
Product description

Magnetically levitated turbopump with active magnetic bearings, no compression stage, designed for light gases (hydrogen/helium) in coating, dry-etching plasma and R&D clean applications. Speed can be set between a minimum and nominal value; M version has no TMS.

  • Scope of delivery: pump, splinter shield in inlet flange, 2 lifting rings, label sheet, instructions; options per ordering guide (copper gasket, solenoid valves, screws, 50 sccm purge reduction device)
  • Stand-by speed is customer-selected between minimum 15000 min-1 and nominal speed
  • Interfaces: splinter shield, inlet flange, electrical connector, water IN/OUT, exhaust flange, purge port, functional grounding
pp. 19-28
Installation

Frame must absorb kinetic energy from sudden rotor blockage (validated per ISO 27892). Defines mandatory inlet flange screw types, torques and clamping forces, permissible pump positions, magnetic field and radiation limits, and water/nitrogen/purge/venting connections. Leak test of the complete pumping line required after installation.

  • Enclosure withstands max static overpressure of 11·10³ hPa for 1 minute when connected per manufacturer recommendations
  • Rotor blockage loads (ATP 1603 M): rotor energy 95 kJ, torque <25 kN·m (1.3 ms, 1 ms delay), bending moment <10 kN·m (1.5 ms), axial force <40 kN (1.4 ms)
  • DN200 ISO-F flange: 12× M10 screws grade 12-9, 30 N·m torque, 161500 N total clamping force, screw length 35 mm
  • DN200 CF-F flange: 24× M8 screws grade 12-9, 20 N·m torque, 266000 N total clamping force, screw length 65 mm
  • Operates in any position; cannot evacuate at atmospheric pressure (transient start at atmosphere allowed); requires backing pump
  • Uniform magnetic field: OK up to 0.5 mT; 0.5–5 mT depends on cooling/gas loads; >5 mT requires shielding; standalone radiation tolerance 10³ Gy
  • Connected inlet-side parts must withstand negative pressure of 1·10⁻³ hPa absolute
  • Cooling water: softened/non-aggressive water; filter on inlet if solid pollution limits cannot be met; manual valve within 3 m
  • Purge: filtered dry nitrogen, max pressure 1–1.5·10³ hPa absolute; 50 sccm flow reducer calibrated for 1.1·10³ hPa; maintain purge whenever rotor is running
  • Purge solenoid valve line connection: max tightening torque 10 N·m; seal with oleo-waterproof glue or PTFE tape
  • Air inlet (venting) solenoid valve: flow 11 slm, reduces braking time; supplied/controlled by the OBC drive unit via Venting connector
  • Twin ferrule fittings: tighten nut 3/4 turn with spanner while holding connector body
pp. 31-34
Operation

Covers preliminary precautions, powering on, pumping start-up, restart after emergency stop, pumping shutdown, powering off, prolonged stoppage and operation monitoring. Explicit lock requirement on the connector while the pump runs.

  • Connector must be locked before use and not disconnected before the pump has stopped
  • Shutdown/prolonged stoppage procedures referenced for later storage (see decommissioning chapter)
pp. 35-39
Maintenance

Maintenance safety instructions, frequency tables, on-site maintenance and exchange of the pump by a replacement product. Includes disconnection, draining of the water circuit and preparation for shipping.

  • Preventive maintenance must be done by trained personnel only (safety label requirement)
  • Exchange procedure: disconnect pump, drain water circuit, prepare pump for shipping
pp. 40
Decommissioning

Shutting down for longer periods, recommissioning after storage, and disposal of the product.

  • Used pumps to be stored per the 'Shutting down for longer periods' procedure
pp. 41-44
Service solutions, Accessories, Spare parts

References Pfeiffer Vacuum service offerings, accessory catalog (including HHR hand-held remote, which is not part of the standard order) and spare parts lists.

  • HHR hand-held remote must be ordered separately via accessories
pp. 45-49
Technical data and dimensions

Contains conversion tables (pressure, gas throughput units), environmental, cooling water, nitrogen and electrical characteristics tables, plus dimension drawings (pump with OBC, purge and air inlet solenoid valves) and weight.

  • Dimensions given for ATP 1603 M with OBC, purge solenoid valve and air inlet solenoid valve (Figs. 7-9)
  • Electrical tables cover both pump electrical characteristics and the customer network requirements
pp. 50-51
EC and UK Declarations of Conformity

EC Declaration of Conformity and UK Declaration of Conformity included at the end of the manual.

  • Declaration of conformity is included with these operating instructions

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Nominal rotor speed is 37000 min-1; stand-by speed is customer selectable between 15000 min-1 and nominal speed. p.1603 ↗ Speed can be tuned per application (e.g. high-pressure pumping) via the OBC drive; selection range starts at 15000 rpm. ATP 1603 M (TMDBXXX4XXXX)
Rotor kinetic energy at nominal speed is 95 kJ; sudden rotor blockage generates transient loads: torque <25 kN·m (1.3 ms duration, 1 ms delay), bending moment <10 kN·m (1.5 ms, 0 ms delay), axial force <40 kN (1.4 ms, 0 ms delay). p.20 ↗ Frame, anchoring screws and adjacent components (e.g. valve plates) must be designed to absorb these peaks; free space >100 mm or a pendulum valve avoids high axial loading. ATP 1603 M
Inlet flange securing is mandatory: DN200 ISO-F = 12× M10 screws grade 12-9 at 30 N·m (length 35 mm, total clamping force 161500 N); DN200 CF-F = 24× M8 screws grade 12-9 at 20 N·m (length 65 mm, total clamping force 266000 N). p.21 ↗ Using wrong screw type, count, grade or torque voids warranty/liability and risks pump detachment at rotor blockage; use torque wrench and original Pfeiffer screw sets. ATP 1603 M, DN200 inlet variants
The pump enclosure withstands a maximum static overpressure of 11·10³ hPa for 1 minute when connected per manufacturer recommendations. p.19 ↗ Defines the burst containment envelope for incident overpressure; connected chambers must withstand 1·10⁻³ hPa absolute negative pressure at the inlet. ATP 1603 M
Magnetic field limits: proper operation guaranteed up to 0.5 mT uniform field; 0.5–5 mT depends on cooling and gas loads; >5 mT requires shielding. Standalone radiation tolerance up to 10³ Gy. p.21 ↗ Site survey of stray magnetic fields (e.g. from ion sources, E-beam tools) is required before installation; shielding mandatory above 5 mT. ATP 1603 M
Inert gas purge must be maintained whenever the rotor is running; purge max pressure 1–1.5·10³ hPa absolute; standard flow reduction device delivers 50 sccm at 1.1·10³ hPa supply pressure. p.26 ↗ Stopping purge while running allows process gases to migrate from fore-vacuum to high-vacuum side and damage bearing housings; set N2 pressure per the purge flow diagram. ATP 1603 M with purge option
Air inlet (venting) solenoid valve has a flow of 11 slm and is supplied/controlled by the OBC electronic drive unit. p.28 ↗ Connect to the OBC 'Venting' connector and trigger via remote control, serial link or fieldbus to shorten braking time; most effective with isolation valves on inlet and exhaust. ATP 1603 M with venting valve option
Electrical rating on plate: 44 Vrms, 1300 Hz, 7 A (M version), 37000 rpm, 41 kg. p.8 ↗ Drive-side power supply and cabling sizing must match these values; pump runs only with the Pfeiffer OBC drive unit. ATP 1603 M
Purge solenoid valve line connection: maximum tightening torque 10 N·m; seal with oleo-waterproof glue or PTFE tape; valve has a dust filter that must be removed when connecting to an inert gas line. p.27 ↗ Prevents thread damage and leaks on the N2 line; solenoid is powered and controlled via the OBC (REMOTE connector or serial link). ATP 1603 M with purge solenoid valve option
Twin ferrule fitting assembly: insert pipe until seated, hand-tighten nut, then tighten 3/4 turn with an open-ended spanner while holding the body. p.25 ↗ Ensures leak-tight water circuit connections; leak check of water circuit required after connection. Water circuit connections, ATP 1603 M
Manual lock-out valves for water and nitrogen circuits must be installed within 3 m of the product. p.11 ↗ Design the facility plumbing so LO/TO isolation of both pressurized circuits is possible at the pump during service. ATP 1603 M installations
For flammable gas processes: LEL detection device at 25% LEL must stop chemical supply; oxidizing agent concentration must not exceed 60% of MOC; nitrogen purge required before maintenance on inlet, pump and exhaust lines (NFPA 69-2019 refs; safety systems per EN 50495). p.12 ↗ Integrator must include interlocks, LEL monitoring and purge sequencing in the installation risk assessment for semiconductor/photovoltaic/display processes. ATP 1603 M in flammable/pyrophoric service
Hot surfaces >65 °C possible depending on conditions; cables and pipes must be kept away from surfaces >70 °C; protective gloves per EN ISO 21420. p.12 ↗ Thermal shielding/PPE planning and cable routing needed around the pump during operation and maintenance. ATP 1603 M
Pump storage (new) max 1 year in clean, dry, non-contaminated area within permitted storage temperatures, wrapped in protective envelope with blanking plates in place; always transport upright. p.15 ↗ Spares/stock rotation planning; do not unpack until installation location; reused pumps stored per prolonged shutdown procedure. ATP 1603 M
Only operate with inlet and exhaust connected to pumping lines and a compatible backing pump with isolation valve (NC) between turbo and backing pump; turbo exhaust must route to the dangerous-gas extraction system via the backing pump. p.11 ↗ Never run at atmosphere as primary evacuation; fore-vacuum configuration and gas abatement are mandatory design items. ATP 1603 M
Lifting requirements: lifting rings or minimum 500 mm multi-arm slings; horizontal handling with 2 EN-1492-1 straps; pump weight 41 kg. p.15 ↗ Specify hoist capacity and rigging plan before installation; keep pump suspended until the high vacuum flange is secured. ATP 1603 M
Water cooling requires softened or non-aggressive water; unregulated mains water causes limescale clogging; install a filter if solid pollution characteristics cannot be met; controlled by a tap for flow adjustment (WATER IN / WATER OUT ports, max pressure marked 7 bar / 101 psi on label). p.8 ↗ Use a regulated chilled-water circuit; consider external takeover cooling control if process cannot tolerate water interruptions. ATP 1603 M (and OBC cooling loop where fitted)
M version (no TMS) must not be used with media posing condensables risk; misuse list also bans pumping liquids, solids, dust, corrosive/explosive media and use in explosive atmospheres or ionizing radiation areas. p.14 ↗ Process compatibility screening required; choose a TMS-equipped variant or different pump family for condensable/reactive processes. ATP 1603 M specifically (M = no Temperature Management System)