Covered Models & Document Scope

Pfeiffer Vacuum SpeedAir 3050 air-based leak detector for liquid products; single model, no ordering codes or variants given; sensitivity down to 1.0 µm, pilot-operated valve air supply, configurable options noted only as 'depending on configuration'.

Section-by-Section Technical Breakdown

pp. 1
Overview and Technology

Positioning of the SpeedAir 3050 as a flexible leak detector for a wide range of applications and liquid products, using proven Non-Destructive Testing (NDT) techniques. Claimed to be the most sensitive air-based technology for liquid products compared to helium leak detectors, with repeatable, reliable results and fewer false negatives/positives.

  • Flexible: suitable for a wide range of applications and products
  • Proven: based on global NDT analysis techniques (no damage to materials/components)
  • Accurate: most sensitive air-based technology for liquid products vs helium leak detectors; repeatable, reliable results, fewer false results
pp. 2
Technical Specifications

Key operating data: sensitivity, clean dry air (CDA) supply requirements, calibration and venting gas conditions, electrical supply, environmental limits, and software/user interface features. Includes 10" touch HMI with Leak Rx software compliant with 21 CFR Part 11.

  • Sensitivity: down to 1.0 µm
  • CDA supply: air (pilot operated valves), quality 1.3.1 per ISO 8573-1
  • CDA pressure: 65–145 psig / 4.5–10 bar rel.; typical consumption 2 Nl/cycle
  • Calibration gas: CDA air, 1–20 Torr (pressure listing as printed: 1.3–27 bar abs.)
  • Venting gas: N₂, 2–10 psig (as printed: 140–690 bar rel. — likely misprint)
  • Power supply: 90–250 V AC, typical consumption 1200 W
  • Operating temperature 59–83 °F / 15–45 °C, humidity 30–80 %
  • 10" multi-touch full HD color screen; Leak Rx software (21 CFR Part 11, PDF GMP test/calibration reports, local or LDAP domain authentication)
  • Operating system: Windows 10
pp. 3
Mechanical Data and Connectivity

Physical and interface data of the unit: network and USB connectivity, noise level, weight, and overall dimensions. Note that some values depend on configuration.

  • Network: 1x LAN (RJ45); interfaces: USB, wired Ethernet
  • Sound pressure level: <53 dB(A)
  • Weight: approx. 280 lbs / 127 kg
  • Dimensions (L x W x H): 36 x 24 x 55 inch / 922 x 601 x 1397 mm
  • Values marked * depend on configuration

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Leak detection sensitivity down to 1.0 µm p.2 ↗ Sets the minimum detectable defect size for test-part qualification; defects smaller than 1 µm will not be detected. SpeedAir 3050
Requires CDA per ISO 8573-1 quality class 1.3.1, 65–145 psig (4.5–10 bar rel.), pilot operated valves p.2 ↗ Facility compressed air must meet particle/water/oil class 1.3.1 and a pressure regulator within 4.5–10 bar rel. is mandatory before installation. SpeedAir 3050
Typical CDA consumption is 2 Nl per cycle p.2 ↗ Allows sizing of plant air supply and estimating cost per test cycle. SpeedAir 3050
Calibration gas supply is CDA air at 1–20 Torr p.2 ↗ No external calibration gas bottles needed; supply line must be regulated to the specified Torr range. SpeedAir 3050
Venting gas is N₂ at 2–10 psig p.2 ↗ A regulated nitrogen line must be plumbed to the unit; pressure shown as '140–690 bar rel.' in the document appears to be a misprint and should be verified with the manufacturer. SpeedAir 3050
Power supply 90–250 V AC with typical consumption 1200 W p.2 ↗ Wide-range AC input simplifies global deployment; provision ≥1200 W circuit capacity per unit. SpeedAir 3050
Operating temperature 15–45 °C (59–83 °F) and humidity 30–80 % p.2 ↗ Installation environment (production floor, testing cell) must stay within these limits; not suitable for unconditioned hot/cold areas. SpeedAir 3050
Weight approx. 127 kg (280 lbs) and dimensions 922 x 601 x 1397 mm p.3 ↗ Requires appropriate handling equipment or casters/foundation planning and floor space of roughly 0.92 x 0.60 m footprint with 1.4 m height clearance. SpeedAir 3050
Sound pressure level below 53 dB(A) p.3 ↗ Can be installed near operators without additional acoustic enclosure. SpeedAir 3050
Connectivity: 1x LAN (RJ45), USB, wired Ethernet; runs Windows 10 p.2 ↗ Plan plant network integration and IT security (Windows 10 endpoint) for data exchange and remote reporting. SpeedAir 3050
Leak Rx software with 21 CFR Part 11 compliance, PDF GMP test and calibration reports, local or LDAP domain authentication p.2 ↗ Suitable for GMP-regulated (pharma) production environments; user management can be tied to corporate LDAP/Active Directory. SpeedAir 3050
HMI is a 10" multi-touch full HD color screen p.2 ↗ Operator interface is touch-based; plan enclosure/access layout accordingly, no external monitor required. SpeedAir 3050