HiCube Neo RGA (QMG 250 F1) — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum HiCube Neo RGA compact residual gas analysis systems with integrated turbomolecular pump and QMG 250 mass spectrometer. Six variants: F-series with Faraday detector only (QMG 250 F1/F2/F3, mass ranges 1–100/1–200/1–300 u) and M-series with C-SEM + Faraday detectors (QMG 250 M1/M2/M3, same mass ranges). Options include chamber heating, calibration gas, shut-off valve, transport box, mobile cart, various backing pumps/gauges/filaments, and Pfeiffer MassSpec software.

Section-by-Section Technical Breakdown

pp. 1
Product Overview and Accessories

The HiCube Neo RGA is a compact residual gas analysis system combining an integrated turbomolecular pump unit with a mass spectrometer. Safety is provided by an integrated DigiLine full-range transmitter monitoring total pressure to protect the filament. The system is portable and offers an extended accessory range.

  • Integrated turbomolecular vacuum pump unit and mass spectrometer in one compact unit
  • Integrated DigiLine full-range transmitter monitors total pressure and protects the filament
  • Spare parts: electronics, ion source, filament, analyzer, gauge, pumps/pump service kits, inlet valve
  • Options: mass ranges, filament materials, various backing pumps and gauges, chamber heating, PC with Pfeiffer MassSpec software, shut-off valve, calibration gas option, transport box, mobile cart
pp. 2-3
Specifications — F-Series (QMG 250 F1/F2/F3, Faraday Detector)

The F-series covers three mass ranges (1–100, 1–200, 1–300 u) with Faraday cup detection only. Detection limits range from 4·10⁻¹³ to 7·10⁻¹³ hPa and sensitivity decreases with higher mass range. All variants share identical mechanical, electrical and interface specifications.

  • Detection limit (Faraday): 4·10⁻¹³ (F1), 5·10⁻¹³ (F2), 7·10⁻¹³ hPa/mbar
  • Ar sensitivity (Faraday): 5·10⁻⁴ (F1), 4·10⁻⁴ (F2), 3·10⁻⁴ A/hPa
  • Max. operating pressure (Faraday): 5·10⁻⁴ hPa
  • Rod system: 6 mm diameter, 125 mm length
  • Analyzer operating temperature 200 °C; electronics 5–50 °C; bake-out 300 °C (electronics removed)
  • Dimensions 446 x 583 x 434 mm; weight approx. 22–27 kg (depending on options)
  • Connecting flange DN 16 ISO-KF / DN 40 CF-F (depending on inlet valve)
  • Peak neighbor contribution < 10 / < 20 / < 50 ppm; resolution 0.5–2.5 u at 10% peak height; dwell time 1 ms–16 s/u
  • Supply 100–240 VAC (50/60 Hz); Ethernet interface; power max. 240 W (400 W with heating); peak-ratio reproducibility ± 0.5%
pp. 3
Specifications — M-Series (QMG 250 M1/M2/M3, C-SEM + Faraday)

The M-series adds a channel electron multiplier (C-SEM) to the Faraday detector, extending detection limits to the 10⁻¹⁵ hPa range. The C-SEM limits the analyzer operating temperature to 150 °C and the max. operating pressure to 5·10⁻⁵ hPa. All other specifications match the F-series.

  • Detection limit Faraday/C-SEM: 4·10⁻¹³/3·10⁻¹⁵ (M1), 5·10⁻¹³/4·10⁻¹⁵ (M2), 7·10⁻¹³/5·10⁻¹⁵ hPa
  • Max. operating pressure Faraday/C-SEM: 5·10⁻⁴ / 5·10⁻⁵ hPa
  • Analyzer operating temperature 200 °C, but max. 150 °C when operating with the SEM
  • Same rod system (6 mm x 125 mm), flange options (DN 16 ISO-KF / DN 40 CF-F), bake-out 300 °C (electronics removed)
  • Identical electrical data: 100–240 VAC, Ethernet, 240 W (400 W with heating), ± 0.5% peak-ratio reproducibility

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
M-series (C-SEM) detection limit reaches 3–5·10⁻¹⁵ hPa versus 4–7·10⁻¹³ hPa for Faraday-only F-series p.2 ↗ Choose M-series when trace-gas detection two orders of magnitude lower is required; Faraday suffices for routine RGA work QMG 250 M1/M2/M3 vs F1/F2/F3
Analyzer max. operating temperature with SEM in use is limited to 150 °C instead of 200 °C p.3 ↗ Chamber heating setpoints above 150 °C are incompatible with SEM operation; select F-series or reduce heating QMG 250 M1/M2/M3
Max. operating pressure with C-SEM is 5·10⁻⁵ hPa vs 5·10⁻⁴ hPa for Faraday p.3 ↗ The DigiLine full-range transmitter must interlock filament/SEM protection before 5·10⁻⁵ hPa in SEM mode QMG 250 M1/M2/M3
Bake-out temperature is 300 °C with electronics removed p.2 ↗ Plan bake-out procedures with electronics demounting; do not exceed 200 °C (analyzer) with electronics installed whole series
Electronics operating temperature range is 5–50 °C p.2 ↗ Ensure ambient/enclosure cooling keeps electronics within 5–50 °C, especially with chamber heating active whole series
Power consumption is max. 240 W without heating and max. 400 W with heating p.3 ↗ Size supply circuit and heat dissipation for 400 W per unit when chamber heating option is fitted whole series
Supply voltage 100–240 VAC, 50/60 Hz p.2 ↗ Universal mains input; no voltage variant selection needed whole series
Connecting flange is DN 16 ISO-KF or DN 40 CF-F depending on inlet valve p.2 ↗ Confirm inlet valve option before specifying chamber flange hardware whole series
Mass ranges are 1–100 u (F1/M1), 1–200 u (F2/M2), 1–300 u (F3/M3) p.2 ↗ Select variant by highest gas mass to be analyzed; higher ranges trade detection limit and sensitivity whole series
Ar sensitivity decreases from 5·10⁻⁴ (1–100 u) to 3·10⁻⁴ A/hPa (1–300 u) p.2 ↗ For maximum sensitivity applications prefer the 1–100 u variant whole series
Resolution is 0.5–2.5 u at 10% peak height; dwell time adjustable 1 ms–16 s/u p.2 ↗ Tune resolution/dwell time per application between fast surveys and high-resolution scans whole series
Contribution to neighboring mass < 10 / < 20 / < 50 ppm by mass range variant p.2 ↗ For low-level adjacent-peak separation work, prefer the 1–100 u variant F1/M1 <10 ppm, F2/M2 <20 ppm, F3/M3 <50 ppm
Peak-ratio reproducibility is ± 0.5% p.3 ↗ Suitable for quantitative gas-comparison measurements within ± 0.5% repeatability whole series
Rod system: 6 mm diameter, 125 mm length quadrupole p.2 ↗ Common mechanical design across variants simplifies spare analyzer stocking whole series
Dimensions 446 x 583 x 434 mm; weight approx. 22–27 kg depending on options p.2 ↗ Compact portable unit; verify bench/cart load capacity and transport box compatibility for heavier optioned units whole series
Interface is Ethernet; software is Pfeiffer MassSpec (PC available as option) p.1 ↗ Plan network integration via Ethernet; order the PC option or provide a compatible host whole series
An integrated DigiLine full-range transmitter protects the filament by monitoring total pressure p.1 ↗ No separate pressure-protection gauge needed, but ensure the DigiLine interlock is wired/configured on startup whole series