Induztr
Engineering Digest Document Viewer Related Models

ATP 1603 M — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum ATP 1603 M magnetically levitated turbomolecular vacuum pump; dimensional/technical data sheet (3 pages, dated 01.07.2025). Notes available accessories/options: installation kit, cooling water valve, purge gas valve, handheld remote controller, inlet protection screen, heater unit, valve cable and coil, various inlet flange types, various user interfaces; on-board control options Profibus or Remote.

Section-by-Section Technical Breakdown

pp. 1
Product Features, Benefits and Accessories

Positions the ATP 1603 M as a quiet, magnetically bearing turbomolecular pump suited to vibration-sensitive semiconductor and analytical applications. It highlights compatibility with various backing pumps, high compression for light gases, and no wearing parts. A list of accessories and options is provided.

  • Low noise and vibration levels for vibration-sensitive applications in semiconductor manufacturing and analytical instrumentation
  • Magnetic bearing technology — no wearing parts, maintenance-free, long maintenance intervals
  • Compatible with a variety of backing pumps; excellent compression ratio for light gases
  • Accessories/options: installation kit, cooling water valve, purge gas valve, handheld remote controller, inlet protection screen, heater unit, valve cable and coil, various inlet flange types, various user interfaces
pp. 2
Pumping Speed, Compression Ratio and Vacuum Performance

Core vacuum performance data: pumping speeds for N₂, Ar, He and H₂, ISO 5302 compression ratios, ultimate pressure, forevacuum and inlet pressure limits, gas throughput and rotational speed. The rotor speed is variable from 15000 rpm up to nominal speed.

  • Pumping speed: N₂ 1250 l/s, Ar 1180 l/s, He 1220 l/s, H₂ 990 l/s
  • Compression ratio (ISO 5302): N₂ >1·10⁷, Ar >1·10⁸, He 2·10⁴, H₂ 3.5·10³
  • Ultimate pressure < 1·10⁻⁸ mbar
  • Max. forevacuum: N₂ 0.8 mbar, Ar 1.5 mbar; max. inlet vacuum: N₂ 0.4 mbar, Ar 0.027 mbar
  • Gas throughput at final rotational speed (Ar): 600 sccm
  • Rotational speed 37000 rpm, variable 15000 rpm to nominal speed
pp. 2-3
Electrical, Mechanical, Cooling/Baking and Interface Data

Electrical supply requirements, run-up time, power consumption, baking limit, noise/vibration levels, dimensions and flange connections, magnetic field limit, and on-board control options. Recommended backing pumps are ADP/ADS series (dry, magnetically levitated).

  • Controller power supply 200–240 V, 50/60 Hz single-phase; max start-up power 800 W; nominal (stand-by) power 120 W
  • Run-up time to 90% of final rotational speed (pressure exhaust < 0.1 mbar): < 5.3 min
  • Max. baking temperature 120 °C
  • Recommended backing pump: ADP/ADS
  • Sound pressure level (ISO 2151, 1 m, nominal speed) < 48 dB(A); vibration level at nominal speed < 0.01 μm
  • Dimensions Ø 259 x 386 mm (height * depending on flange variant); gas inlet DN 200 ISO-K, outlet DN 40 ISO-KF; N₂ purge flange DN 16 ISO-KF
  • Stainless steel inlet screen as factory setting
  • Permissible radial magnetic field at 10 cm: < 12 mT
  • On-board control options: Profibus, Remote; dimensional drawing and pumping speed curves shown

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Pumping speed for N₂ is 1250 l/s; He 1220 l/s; Ar 1180 l/s; H₂ 990 l/s p.1603 ↗ Select pump sizing based on the process gas; He and H₂ throughput sizing must account for lower H₂ speed despite high compression capability ATP 1603 M
Compression ratio (ISO 5302) for H₂ is 3.5·10³ and He 2·10⁴, vs >1·10⁷ for N₂ and >1·10⁸ for Ar p.1603 ↗ For hydrogen processes, the forevacuum pressure dominates the ultimate pressure; choose a high-quality backing pump (e.g. ADP/ADS) to reach low ultimate pressures with light gases ATP 1603 M
Ultimate pressure < 1·10⁻⁸ mbar p.1603 ↗ Suitable for high/ultra-high vacuum applications; system base pressure will be limited by foreline pressure for light gases ATP 1603 M
Max. forevacuum pressure: 0.8 mbar (N₂), 1.5 mbar (Ar); max. inlet vacuum 0.4 mbar (N₂), 0.027 mbar (Ar) p.1603 ↗ Backing pump must maintain foreline pressure below these limits or the pump will not sustain vacuum performance ATP 1603 M
Gas throughput at final rotational speed is 600 sccm for Ar p.1603 ↗ Process gas flows above ~600 sccm Ar would exceed the pump's capacity; relevant for sizing gas delivery in sputter/deposition processes ATP 1603 M
Rotational speed is 37000 rpm nominal and is variable from 15000 rpm to nominal speed p.1603 ↗ Speed can be reduced (e.g. to reduce vibration, heat, or to tune compression) via control interface ATP 1603 M
Run-up time to 90% of final rotational speed (exhaust pressure < 0.1 mbar) is < 5.3 min p.1603 ↗ Plan process start-up delays accordingly; ensure backing pressure < 0.1 mbar at start for the quoted run-up time ATP 1603 M
Controller power supply: 200–240 V, single-phase, 50/60 Hz; max start-up power 800 W; nominal (stand-by) power 120 W p.1603 ↗ Specify single-phase mains and a controller sized for the 800 W start-up surge, not the 120 W steady-state power ATP 1603 M
Maximum baking temperature is 120 °C p.1603 ↗ Bakeout of the vacuum system including this pump must not exceed 120 °C (lower than many all-metal chamber bakes; check flange variant) ATP 1603 M
Sound pressure level < 48 dB(A) (ISO 2151, 1 m, nominal speed) and vibration < 0.01 μm p.1603 ↗ Can be installed close to vibration-sensitive instrumentation (electron microscopes, metrology) without heavy vibration isolation ATP 1603 M
Dimensions Ø 259 x 386 mm (height depends on flange variant); inlet DN 200 ISO-K, outlet DN 40 ISO-KF, N₂ purge flange DN 16 ISO-KF p.1603 ↗ Chamber flange design must match DN 200 ISO-K; outlet allows a standard DN 40 ISO-KF foreline; height varies with chosen inlet flange option ATP 1603 M
Permissible radial magnetic field at 10 cm distance: < 12 mT p.1603 ↗ Avoid installation near strong magnetic sources (magnets, MRI, magnetic stirring) or shield the pump; the magnetic bearing system is sensitive to external fields ATP 1603 M
Stainless steel inlet screen is the factory setting p.1603 ↗ Particles protection is provided by default; for particle-free UHV applications the screen may need removal — check application ATP 1603 M
Magnetic bearing technology — no wearing parts, no lubricant, long maintenance intervals p.1603 ↗ Eliminates oil contamination risk and oil-change maintenance; suitable for oil-free/clean processes and for integration with oil-free ADP/ADS backing pumps ATP 1603 M
Recommended backing pumps are the ADP/ADS series (magnetically levitated dry pumps) p.1603 ↗ For oil-free, vibration-sensitive applications, pair with Pfeiffer ADP/ADS; other backing pumps may also be compatible but ADP/ADS is the documented recommendation ATP 1603 M
On-board control options: Profibus or Remote; accessories include handheld remote controller, valve cable and coil, cooling water valve, purge gas valve, heater unit p.1603 ↗ Integration into PLC/ fab control is via Profibus; purge/cooling valve accessories enable automated gas purge and cooling control via valve cables ATP 1603 M