SplitFlow 80 Neo — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum SplitFlow 80 Neo turbomolecular vacuum pump datasheet/brochure. Two high-vacuum flange variants: DN 63 CF-F and DN 63 ISO-K. Accessories/options listed: electrical adapter, mains cable, connecting cable, CIP adapter, power supply kit and display unit, pressure sensor, mounting kit, safety valve, venting accessories, power failure venting unit, vibration damper.
Section-by-Section Technical Breakdown
Highlights high performance via an extended thermal operating range giving maximum compression and high gas throughput. Compact design and any-orientation mounting make it suitable for portable analytical applications and easy integration into small instruments.
- Maximum compression and high gas throughput due to extended thermal operating range
- Mounting in any orientation
- Compact footprint for integration into small analytical instruments and vacuum systems
- Suitable for portable analytical applications
Lists available accessories and options for the SplitFlow 80 Neo, covering electrical connectivity, integration hardware and vacuum-side components.
- Electrical adapter, mains cable, connecting cable
- CIP adapter
- Power supply kit and display unit
- Pressure sensor
- Mounting kit
- Safety valve
- Venting accessories and power failure venting unit
- Vibration damper
Technical specifications: pumping speeds and compression ratios for N₂, H₂, He, Ar, interstage pumping, gas throughputs, fore-vacuum limits, run-up time, noise level, weight and flange options.
- Max pumping speed: N₂ 67 l/s, H₂ 48 l/s, He 58 l/s, Ar 66 l/s
- Interstage pumping speed: 2.8 l/s (He and N₂)
- Compression ratio: N₂ 1·10¹¹, H₂ 1.4·10⁵, He 1.3·10⁷, Ar 1·10¹¹
- Fore-vacuum/interstage pump compression ratio: 1·10⁵–5·10⁵ (N₂ and He)
- Gas throughput: N₂ 1.3, H₂ 15.3, He 2.7, Ar 0.54 hPa·l/s
- Fore-vacuum max: 22 hPa (N₂, Ar), 14 hPa (H₂)
- Run-up time 75 s; noise 48 dB(A) per EN ISO 2151
- Weight approx. 1.7–3.4 kg depending on HV flange size
- HV flange: DN 63 CF-F or DN 63 ISO-K
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Maximum pumping speed for N₂ is 67 l/s (H₂ 48 l/s, He 58 l/s, Ar 66 l/s) p.2 ↗ | Size the backing pump and conductance accordingly for main gas loads; N₂-base processes get the highest effective speed | SplitFlow 80 Neo (both flange variants) |
| Interstage pumping speed is 2.8 l/s for both He and N₂ p.2 ↗ | Interstage port can be used for differential/tandem applications with limited throughput (~2.8 l/s) | SplitFlow 80 Neo |
| Compression ratio for N₂ is 1·10¹¹ and for H₂ only 1.4·10⁵ p.2 ↗ | Ultimate pressure for hydrogen-bearing loads depends strongly on fore-vacuum quality; hydrogen requires a good backing pump | SplitFlow 80 Neo |
| Fore-vacuum/interstage compression ratio is 1·10⁵–5·10⁵ (N₂ and He) p.2 ↗ | Tandem operation pressure window; select backing stage within this range | SplitFlow 80 Neo |
| Gas throughput N₂ 1.3 hPa·l/s; H₂ 15.3 hPa·l/s; He 2.7 hPa·l/s; Ar 0.54 hPa·l/s p.2 ↗ | Limits continuous gas load sizing; hydrogen throughput is an order of magnitude above nitrogen | SplitFlow 80 Neo |
| Maximum fore-vacuum pressure: 22 hPa for N₂ and Ar, 14 hPa for H₂ p.3 ↗ | Backing pump must maintain ≤22 hPa (≤14 hPa for hydrogen) to avoid overload; tighter limit for H₂ | SplitFlow 80 Neo |
| Run-up time is 75 s p.3 ↗ | Plan process start sequences and interlock delays around ~75 s to full speed | SplitFlow 80 Neo |
| Sound pressure level is 48 dB(A) per EN ISO 2151 p.3 ↗ | Low-noise operation suitable for lab/portable analytical environments | SplitFlow 80 Neo |
| Weight approx. 1.7–3.4 kg depending on HV flange size p.3 ↗ | Light enough for portable/mobile mounting; account for flange choice in mass budgets | SplitFlow 80 Neo |
| HV flange options: DN 63 CF-F and DN 63 ISO-K p.3 ↗ | CF-F variant for UHV/high-leak-tightness requirements; ISO-K for standard vacuum connections | SplitFlow 80 Neo variants |
| Pump can be mounted in any orientation p.1 ↗ | Enables compact, unconventional integration in small analytical instruments without orientation constraints | SplitFlow 80 Neo |
| A power failure venting unit is available as accessory p.1 ↗ | Protects the process/chamber on power loss; include it for unattended operation | SplitFlow 80 Neo |