ASM 390 — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum ASM 390 and ASM 392 helium/hydrogen leak detectors. Both models share the same test methods, power data and dimensions; they differ mainly in helium pumping speed (10 l/s vs 25 l/s) and weight (125 kg vs 130 kg). Document also lists accessories/options: sniffer probe (5 or 10 m hose), helium spray gun, remote control, bottle holder, dust filter.
Section-by-Section Technical Breakdown
Positioning of the ASM 390/392 as high-performance leak detectors for demanding applications in clean environments. Emphasis on fast start-up, high pumping speed, compact ergonomic design, and user-friendly operation.
- Fast start-up, high pumping speed, suited for demanding applications in clean environments
- Compact and ergonomic design with easy access to all testing areas
- Large rotatable color touch display
- Integrated tool box and storage space
Lists available accessories and options for the ASM 390/392 leak detectors.
- Sniffer probe with 5 m or 10 m hose length
- Helium spray gun
- Remote control
- Bottle holder
- Dust filter
Side-by-side specification table for ASM 390 and ASM 392 covering performance, electrical, environmental and interface parameters. Both models are identical in most respects; the key difference is helium pumping speed.
- Test methods: spraying (vacuum) and sniffing tests for both models
- Detectable gases: ⁴He, ³He, H₂
- Min. detectable leak rate (spraying, helium): 1·10⁻¹² mbar l/s / 1·10⁻¹³ Pa m³/s (both models)
- Min. detectable leak rate (sniffing, helium): 1·10⁻⁸ mbar l/s / 1·10⁻⁹ Pa m³/s (both models)
- Helium pumping speed: ASM 390 = 10 l/s, ASM 392 = 25 l/s
- Power supply: 100-240 V, 50/60 Hz; power consumption 800 W, max. 1600 W
- Max. test inlet pressure: 20 mbar
- Start-up time at 20 °C: <2 min; response time <1 s
- Backing pump: multi-stage Roots vacuum pump; roughing capacity 35 m³/h
- Operating temperature: 10-35 °C
- Interfaces: I/O, RS-232, Ethernet (PC, not industrial)
- Sound pressure level: ≤55 dB(A)
Mechanical data including weight, external dimensions and inlet port for both models, plus a dimensional drawing reference.
- Weight approx.: ASM 390 = 125 kg, ASM 392 = 130 kg
- Dimensions (L x W x H): 1072 x 455 x 1025 mm (both models)
- Inlet port: DN 40 ISO-KF (both models)
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Minimum detectable leak rate for helium in spraying (vacuum) test is 1·10⁻¹² mbar l/s (1·10⁻¹³ Pa m³/s) on both ASM 390 and ASM 392. p.2 ↗ | Suitable for ultra-tight, high-integrity leak testing applications; selection between models need not be driven by ultimate sensitivity. | ASM 390 and ASM 392 |
| Minimum detectable leak rate for helium in sniffing test is 1·10⁻⁸ mbar l/s (1·10⁻⁹ Pa m³/s). p.2 ↗ | Sets the practical limit for external sniffer-mode testing; leaks below this rate will not be detected in sniffing applications. | ASM 390 and ASM 392 |
| Helium pumping speed: 10 l/s for ASM 390, 25 l/s for ASM 392. p.2 ↗ | The ASM 392 recovers faster from large helium loads and suits high-throughput spraying tests; ASM 390 is the lighter, lower-speed option. | ASM 390 vs ASM 392 model selection |
| Detectable tracer gases are ⁴He, ³He and H₂. p.2 ↗ | Hydrogen testing is possible (e.g., sensitive method with H₂/H₂ mixtures), not limited to helium only. | ASM 390 and ASM 392 |
| Maximum test inlet pressure is 20 mbar. p.2 ↗ | Test parts must be evacuated below 20 mbar before connecting to the inlet, or connected via appropriate throttling/roughing, to avoid overpressurizing the mass spectrometer stage. | ASM 390 and ASM 392 |
| Start-up time is <2 min at 20 °C and response time is <1 s. p.2 ↗ | Enables rapid cycle times in production leak testing; short warm-up allows late start-up in shift-based workflows. | ASM 390 and ASM 392 |
| Backing is provided by a multi-stage Roots vacuum pump with 35 m³/h roughing capacity. p.2 ↗ | Dry (oil-free) backing simplifies maintenance and avoids oil contamination; 35 m³/h roughing supports gross-leak evacuation of test parts. | ASM 390 and ASM 392 |
| Power supply is 100-240 V AC, 50/60 Hz; power consumption 800 W nominal, max. 1600 W. p.2 ↗ | Runs from standard single-phase mains worldwide; circuit/fuse sizing should account for the 1600 W maximum draw, not just nominal 800 W. | ASM 390 and ASM 392 |
| Operating temperature range is 10-35 °C. p.2 ↗ | Installation environment must be climate-controlled; ambient below 10 °C at start-up may lengthen warm-up beyond the <2 min spec. | ASM 390 and ASM 392 |
| Communication interfaces are I/O, RS-232 and Ethernet (PC-type, not industrial). p.2 ↗ | For PLC integration use the I/O interface; Ethernet is for PC communication only and is not rated for industrial fieldbus use. | ASM 390 and ASM 392 |
| Sound pressure level is ≤55 dB(A). p.2 ↗ | Suitable for cleanroom/laboratory environments without additional acoustic enclosures. | ASM 390 and ASM 392 |
| Inlet port is DN 40 ISO-KF on both models. p.3 ↗ | Test fixtures and connection lines must be specified with DN 40 ISO-KF clamps, centering rings and O-rings for full conductance. | ASM 390 and ASM 392 |
| Weights are approx. 125 kg (ASM 390) and 130 kg (ASM 392); dimensions 1072 x 455 x 1025 mm (L x W x H) for both. p.3 ↗ | Requires rated transport equipment (crane/lift) and a fixed floor position; the narrow 455 mm width suits placement in compact production cells. | ASM 390 and ASM 392 |
| Sniffer probes are available with 5 m or 10 m hose length. p.1 ↗ | Hose length affects sniffing-mode response time; choose 5 m where shorter response is needed, 10 m for larger test volumes. | Optional sniffer probe accessory |
| A dust filter accessory is available for the inlet. p.1 ↗ | Recommended for applications with particle-contaminated test parts to protect the internal pumping system. | ASM 390 and ASM 392 (accessory) |