ASI 35 — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum ASI 35 helium/hydrogen leak detector (spraying and sniffing tests). Accessories/options listed: cables, sniffer module, sniffer probe with different hose lengths, communication interfaces. Inlet port options DN 25 ISO-KF or DN 16 ISO-KF.

Section-by-Section Technical Breakdown

pp. 1
Key Features

The ASI 35 is presented as flexible, accurate, and robust. Its modular design has a small footprint for easy integration into leak detection systems, with high sensitivity for short cycle times and high production throughput, plus long maintenance intervals.

  • Flexible: modular design, small footprint, easy integration into any leak detection system
  • Accurate: high sensitivity, repeatable measurements, suitable for short cycle time and high production throughput
  • Robust: high reliability, long maintenance intervals
pp. 1
Accessories, Spare Parts and Options

Lists available accessories and options for the ASI 35. Includes sniffer hardware and industrial-standard communication interfaces.

  • Cable
  • Sniffer module
  • Sniffer probe with different hose length
  • Communication interface compatible with industrial standards
pp. 2
Technical Data

Core specifications of the ASI 35 covering test methods, sensitivity, gas compatibility, interfaces, power, dimensions and weights. The unit consists of three physical modules: vacuum module, electronic module, and control panel.

  • Test methods: spraying and sniffing tests
  • Detectable gases: ⁴He; ³He; H₂
  • Min. detectable leak rate (Helium, spraying): 5·10⁻¹² mbar l/s / 5·10⁻¹³ Pa m³/s
  • Min. detectable leak rate (Helium, sniffing): 1.5·10⁻⁸ mbar l/s / 1.5·10⁻⁹ Pa m³/s
  • Helium pumping speed: 6 l/s
  • Max. test inlet pressure: 18 mbar
  • Start-up time (20 °C): <3 min
  • Operating temperature: 10–45 °C
  • Communication interfaces: I/O, RS-232, Profibus, Profinet
  • Inlet port: DN 25 ISO-KF, DN 16 ISO-KF
  • Dimensions (L×W×H): vacuum module 279×264×235 mm; electronic module 216×317×111 mm; control panel 241×54×133 mm
  • Weight approx.: vacuum module 15 kg / electronic module 5 kg / control panel 1.3 kg
  • Power supply: 90–240 V (50/60 Hz); max. power consumption 300 W

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Minimum detectable leak rate for helium in spraying (vacuum) test: 5·10⁻¹² mbar l/s (5·10⁻¹³ Pa m³/s) p.2 ↗ Suitable for high-sensitivity vacuum leak testing; spec must not be confused with the much higher sniffing detection limit ASI 35
Minimum detectable leak rate for helium in sniffing test: 1.5·10⁻⁸ mbar l/s (1.5·10⁻⁹ Pa m³/s) p.2 ↗ For atmospheric sniffing applications, design test criteria around this ~4 orders of magnitude worse limit than vacuum mode ASI 35 (sniffing operation)
Detectable tracer gases: ⁴He, ³He and H₂ p.2 ↗ Hydrogen test possible where helium cost/supply is an issue; verify gas-specific sensitivities before certifying ASI 35
Helium pumping speed is 6 l/s p.2 ↗ Determines response/recovery time of vacuum test cycles and pump-down of small test volumes ASI 35
Maximum test inlet pressure is 18 mbar p.2 ↗ Test parts must be evacuated below 18 mbar before opening the leak detector inlet; use a backing/roughing pump above this pressure ASI 35
Start-up time at 20 °C is less than 3 minutes p.2 ↗ Enables short shift startup in production environments; allow ~3 min before first valid measurement ASI 35
Operating temperature range is 10–45 °C p.2 ↗ Installation site must be climate-controlled; outside this range accuracy/reliability are not guaranteed ASI 35
Communication interfaces available: I/O, RS-232, Profibus, Profinet p.2 ↗ Direct PLC integration possible in automated production lines without additional gateways for Profibus/Profinet environments ASI 35
Inlet ports available: DN 25 ISO-KF and DN 16 ISO-KF p.2 ↗ Select KF flange hardware/adapters accordingly; DN 16 reduces conductance versus DN 25 ASI 35
Power supply is 90–240 V, 50/60 Hz with max. 300 W power consumption p.2 ↗ Single-phase worldwide supply compatibility; size supply/breaker for at least 300 W per unit ASI 35
System splits into vacuum module (279×264×235 mm, 15 kg), electronic module (216×317×111 mm, 5 kg) and control panel (241×54×133 mm, 1.3 kg) p.2 ↗ Modular layout allows distributed installation; mounting plates and enclosures must be dimensioned per module, total mass ~21.3 kg ASI 35
Optional sniffer probe is available with different hose lengths p.1 ↗ Hose length selection affects sniffing response time and the stated 1.5·10⁻⁸ mbar l/s sniffing limit; choose shortest practical length ASI 35 with sniffer module/probe option