ATH 4506 M — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum ATH 4506 M and ATH 4506 MT turbomolecular vacuum pumps (high-temperature/high-flow variants with wide process range and corrosion resistance). Document is a 3-page datasheet/brochure covering specifications, dimensions, pumping speed curves, and available accessories/options.
Section-by-Section Technical Breakdown
Highlights the ATH 4506 M/MT positioning as a high temperature and high flow turbomolecular pump with wide process range and high corrosion resistance. Built-in sensor provides performance diagnostics, maintenance optimization and external communications, with self-diagnostics and optimized temperature management for low cost of ownership.
- Self-diagnostics and built-in sensor for performance diagnostics, maintenance optimization and external communication
- Economical operation via optimized temperature management, low power and cooling water consumption
- Accessories/options: protective seal gas purge valve, cooling water valve, handheld remote control, protective inlet filter, mounting kit
- Inlet flange options: VG350 or VG400; exhaust flanges DN40 or DN50
- Heating options, additional middle purge for harsh processes, rotor temperature sensor, purge presence sensor
- Input power options: 200, 440 or 480 VAC
Provides pumping speeds for N₂, Ar, He and H₂, ultimate pressure, and maximum fore/inlet vacuum limits, identical for M and MT variants. Gas throughput at final rotational speed differs slightly between the two variants, and the MT specifies a fixed recommended purge flow while the M allows 0–100 sccm.
- Pumping speed N₂ 4500 l/s; Ar 4370 l/s; He 3900 l/s; H₂ 3700 l/s (both variants)
- Ultimate pressure < 1·10⁻⁸ mbar
- Max. forevacuum: 3.5 mbar (N₂ and Ar); max. inlet vacuum: 0.5 mbar (N₂), 0.22 mbar (Ar)
- Gas throughput at final rotational speed N₂: 5500 sccm (MT) / 6550 sccm (M); Ar: 2800 sccm (MT) / 3500 sccm (M)
- Recommended purge flow: 100 sccm (MT) vs 0–100 sccm (M)
- Rotational speed 23500 rpm, variable 15000 rpm to nominal speed
- Run-up time to 90% of final speed with exhaust pressure < 0.1 mbar: < 9 min
Electrical power data, heating and cooling water requirements, control interfaces and mechanical dimensions/weight. Both variants share identical electrical, cooling, mounting and interface data; heating temperature applies only to the MT.
- Max. power consumption for start-up: 2000 W; nominal (stand-by) power: 220 W; max. current consumption: 14 A
- Controller power supply: 230/440/480 V AC
- Max. heating temperature 50–95 °C (MT only)
- Cooling water: 60–180 l/h recommended flow, 15–35 °C water temperature
- Pump-mounted controller; on-board control options: Remote, EtherCAT
- Dimensions Ø x H: 450 x 422 mm; weight approx. 140 kg
- Gas inlet: VG350, VG400, VF350; gas outlet: DN 40 ISO-K or DN 50 ISO
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Pumping speed is 4500 l/s for N₂, 4370 l/s for Ar, 3900 l/s for He and 3700 l/s for H₂, identical for ATH 4506 M and MT. p.2 ↗ | Size backing pumps and process throughput calculations on gas-specific speeds; He/H₂ speeds are ~82% of the N₂ rating. | ATH 4506 M and MT |
| Ultimate pressure < 1·10⁻⁸ mbar. p.2 ↗ | Suitable for high/ultra-high vacuum process and analytical applications; system base pressure will be limited by chamber outgassing and backing pump. | Both variants |
| Maximum forevacuum (backing pressure) is 3.5 mbar for N₂ and Ar. p.2 ↗ | Backing pump must maintain ≤3.5 mbar at full gas throughput; oversizing the backing line is unnecessary but undersizing will cause rotor instability/shutdown. | Both variants |
| Maximum inlet vacuum (inlet pressure) 0.5 mbar (N₂) / 0.22 mbar (Ar). p.2 ↗ | Defines the upper inlet pressure for safe continuous operation at full speed when pumping through gases; stays within process windows during gas flooding. | Both variants |
| Gas throughput at final rotational speed: N₂ 5500 sccm (MT) / 6550 sccm (M); Ar 2800 sccm (MT) / 3500 sccm (M). p.2 ↗ | The M variant tolerates higher continuous gas load; select M for high gas-flow processes, MT for high-temperature applications. | ATH 4506 M vs MT |
| Recommended purge flow rate is 100 sccm fixed for MT, adjustable 0–100 sccm for M. p.2 ↗ | Process purge/lean gas shielding must be supplied at these rates; MT requires continuous purge, M allows purge-free operation. | ATH 4506 MT (fixed purge), ATH 4506 M (0–100 sccm) |
| Rotational speed 23500 rpm nominal, variable down to 15000 rpm. p.2 ↗ | Speed can be reduced to ~64% for throughput/temperature control without changing hardware; plan control setpoints accordingly. | Both variants |
| Run-up time to 90% of final rotational speed is < 9 min with exhaust pressure < 0.1 mbar. p.2 ↗ | Sequence pump start with backing pressure < 0.1 mbar to achieve the quoted <9 min start time; important for cycle-time planning. | Both variants |
| Controller power supply 230/440/480 V AC; input power options 200, 440 or 480 VAC. p.1 ↗ | Verify site supply matches one of these voltages before ordering; no 400 V-only version required to be handled specially. | Both variants |
| Max. start-up power consumption 2000 W, nominal/stand-by power 220 W, max. current 14 A. p.3 ↗ | Size supply cable and breaker for 2000 W / 14 A inrush at start-up; steady-state load is only 220 W. | Both variants |
| Maximum heating temperature 50–95 °C, applicable to MT only. p.3 ↗ | MT supports heated operation up to 95 °C for condensable gas handling; the M variant has no heating option in this datasheet. | ATH 4506 MT |
| Cooling water flow 60–180 l/h at 15–35 °C. p.3 ↗ | Provide a water circuit within this flow/temperature window; insufficient flow or >35 °C water degrades temperature management and cost-of-ownership claims. | Both variants |
| Dimensions Ø450 x 422 mm (height), weight approx. 140 kg. p.3 ↗ | Plan mounting structure and handling equipment for 140 kg; inlet flange face is a 450 mm diameter envelope. | Both variants |
| Gas inlet options VG350, VG400, VF350; gas outlet DN 40 ISO-K or DN 50 ISO. p.1 ↗ | Match chamber flange (VG350/VG400/VF350) and backing line (DN40 ISO-K or DN50 ISO) at ordering time; flange choice is an ordering option, not a retrofit. | Both variants |
| On-board control options: Remote and EtherCAT; remote controller is pump-mounted. p.3 ↗ | For automated tools, integrate via EtherCAT; handheld remote control is available as an accessory for manual operation. | Both variants |
| Available options include protective seal gas purge valve, cooling water valve, rotor temperature sensor, purge presence sensor, additional middle purge for harsh processes, and protective inlet filter. p.1 ↗ | For corrosive/harsh processes, order the middle purge and sensors; the purge presence sensor enables interlocking pump start on confirmed purge flow. | Both variants (options) |