QMG 250 M1 — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum PrismaPro M quadrupole mass spectrometer family, covering three analyzer variants: QMG 250 M1 (mass range 1–100 u), QMG 250 M2 (1–200 u), and QMG 250 M3 (1–300 u). All variants share DN 40 CF-F flange, dual detector (C-SEM / Faraday), Ethernet interface, and options including spare electronics, ion source, filament, analyzer, inlet valve, filament material, straight or 90° connection, and standard or extended I/Os.
Section-by-Section Technical Breakdown
The PrismaPro M is positioned as a flexible quadrupole mass spectrometer for industrial and analytical environments with a modular design and new-generation Pfeiffer MassSpec software for measured-data transfer. The document lists available accessories, spare parts and configurable options.
- Lowest detectable partial pressure: 3 · 10⁻¹⁵ hPa (mbar)
- Accessories/spares: spare electronics, spare ion source, spare filament, spare analyzer
- Vacuum gauge and vacuum chamber, inlet valve, communication interface and test equipment available
- Configurable options: mass ranges, filament material, straight or right-angled connection, standard or extended I/Os
A comparative specification table for QMG 250 M1/M2/M3 covering mass range, detection limits, sensitivity, pressure/temperature limits, flange, detector, resolution, dwell time, dimensions, weight and supply voltage. Dimensional drawings show straight and 90° angle configurations with L/W/H values.
- Mass ranges: M1 = 1–100 u, M2 = 1–200 u, M3 = 1–300 u
- Detection limit Faraday/C-SEM: M1 4·10⁻¹³/3·10⁻¹⁵, M2 5·10⁻¹³/4·10⁻¹⁵, M3 7·10⁻¹³/5·10⁻¹⁵ hPa (mbar)
- Ar sensitivity (Faraday): M1 5·10⁻⁴, M2 4·10⁻⁴, M3 3·10⁻⁴ A/hPa (mbar)
- Max. operating pressure (Faraday/C-SEM): 5·10⁻⁴ / 5·10⁻⁵ hPa (mbar) for all models
- Analyzer operating temperature 200 °C (max. 150 °C with SEM); bake-out temperature 300 °C; electronics 5–50 °C
- Connecting flange DN 40 CF-F; detectors C-SEM and Faraday
- Contribution to neighboring mass: <10 ppm (M1), <20 ppm (M2), <50 ppm (M3); resolution at 10% peak height 0.5–2.5 u; dwell time 1 ms–16 s/u
- Dimensions straight 427.6 x 142 x 128 mm; 90° angle 271.4 x 142 x 356 mm; weight approx. 3.2 kg
- Supply voltage 100–240 VAC (50/60 Hz); interface Ethernet; reproducibility of peak ratio ±0.5 %
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Lowest detectable partial pressure of the series is 3 · 10⁻¹⁵ hPa (mbar), achieved with the C-SEM detector (M1 best case). p.1 ↗ | Ultra-trace gas analysis down to the 10⁻¹⁵ hPa range is possible only when the C-SEM channel is used, not the bare Faraday cup. | whole series, best on QMG 250 M1 |
| Mass ranges are 1–100 u (M1), 1–200 u (M2) and 1–300 u (M3). p.2 ↗ | Variant selection must match the heaviest gas species to be monitored; higher mass range trades off sensitivity and peak separation. | model-specific: QMG 250 M1/M2/M3 |
| Faraday detection limits: 4·10⁻¹³ (M1), 5·10⁻¹³ (M2), 7·10⁻¹³ hPa (mbar). p.2 ↗ | For routine measurements without SEM, the M1 variant gives the best detection limit. | model-specific |
| Argon sensitivity in Faraday mode decreases with mass range: 5·10⁻⁴ (M1), 4·10⁻⁴ (M2), 3·10⁻⁴ A/hPa (mbar). p.2 ↗ | Signal calibration constants must be set per variant; M3 gives ~40% lower sensitivity than M1. | model-specific |
| Maximum operating pressure is 5·10⁻⁴ hPa (mbar) in Faraday mode and 5·10⁻⁵ hPa (mbar) with C-SEM. p.2 ↗ | Process pressures above these limits require pressure reduction or will damage/contaminate the SEM; SEM operation imposes a 10× lower pressure ceiling. | whole series |
| Analyzer can operate up to 200 °C, but only 150 °C when the SEM is installed; electronics rated 5–50 °C. p.2 ↗ | SEM-equipped units must not be used in >150 °C process environments; ambient electronics cooling must be ensured below 50 °C. | whole series |
| Bake-out temperature of the analyzer is 300 °C. p.2 ↗ | Chamber bake-out at up to 300 °C is permissible only if the analyzer electronics are removed/not operated; plan bake-out cycles accordingly. | whole series |
| Connection flange is DN 40 CF-F (ConFlat) on all models. p.2 ↗ | Vacuum chamber port must be prepared as DN 40 CF with copper gasket; enables UHV-tight metal-seal mounting. | whole series |
| Dimensions: straight version 427.6 x 142 x 128 mm; 90° angle version 271.4 x 142 x 356 mm; weight approx. 3.2 kg. p.2 ↗ | Installation clearances and mounting hardware must be chosen per geometry (straight vs. right-angled option); ~3.2 kg is easily supported by the CF flange. | whole series |
| Contribution to neighboring mass is <10 ppm (M1), <20 ppm (M2), <50 ppm (M3). p.2 ↗ | Peak-tail interference rises with mass range; M1 is preferable when resolving adjacent peaks with large dynamic ratio is required. | model-specific |
| Resolution at 10% peak height is adjustable 0.5–2.5 u; dwell time adjustable 1 ms–16 s/u. p.2 ↗ | Scan speed vs. resolution and signal-to-noise can be tuned via parameter settings in software for each application. | whole series |
| Reproducibility of peak ratio is ±0.5 % for all models. p.3 ↗ | Quantitative trend monitoring and ratio-based methods can rely on ~0.5 % repeatability. | whole series |
| Supply voltage is 100–240 VAC at 50/60 Hz; data interface is Ethernet on all models. p.2 ↗ | No dedicated line-voltage variants are needed; integration into networks/PC-based data logging is via standard Ethernet. | whole series |
| Options include filament material choice, straight or right-angled analyzer connection, and standard or extended I/Os. p.1 ↗ | Ordering configuration must specify filament chemistry (application-dependent, e.g. oxygen compatibility), probe geometry, and I/O set at time of order. | whole series |
| Spare parts offered: electronics, ion source, filament, and analyzer; plus vacuum gauge, vacuum chamber, inlet valve and test equipment. p.1 ↗ | A spare filament and ion source should be stocked for uptime-critical installations, as these are consumable/wear components. | whole series |