Induztr
Engineering Digest Document Viewer Related Models

ATH 1600 MT — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum ATH 1600 MT and ATH 1603 M magnetically levitated (magnetic bearing) turbomolecular vacuum pumps. Datasheet/brochure (4 pages) covering both variants' performance data, dimensions, cooling, power, purge gas and interface specifications, plus a list of available accessories and options (installation kit, cooling water valve, purge gas valve, handheld remote controller, inlet protection screen, heater unit, valve cable/coil, various inlet flanges, air cooling system, temperature management system for corrosive processes). The MT version adds an integrated heating system for demanding chemical conditions.

Section-by-Section Technical Breakdown

pp. 1
Product Highlights and Features

ATH 1600 MT / 1603 M are maintenance-free turbomolecular pumps using magnetic bearing technology, eliminating wearing parts and reducing particle contamination/residue build-up (MT version). MT version includes an integrated heating system for demanding chemical conditions. Frequency-controlled DC motors give low noise and vibration, suitable for semiconductor and analytical instrumentation applications.

  • Maintenance-free, no wearing parts (magnetic bearing technology)
  • MT version: reduced particle contamination/residue build-up and integrated heating for chemical conditions
  • Frequency-controlled direct current (DC) motors
  • Compatible with a variety of backing pumps
  • Low noise and vibration, suited to vibration-sensitive applications
  • Accessories/options: installation kit, cooling water valve, purge gas valve, handheld remote controller, inlet protection screen, heater unit, valve cable and coil, various inlet flange types, various user interfaces, temperature management system for corrosive processes, air cooling system
pp. 2-3
Pumping Speed, Compression Ratio and Pressure Limits

Both models offer identical pumping speeds (N₂ 1360–1400 l/s, H₂ 540 l/s) and compression ratios (N₂/Ar > 1·10⁸), with ultimate pressure < 6·10⁻⁹ mbar. The main differences are gas throughput at final speed (1000 vs 4000 sccm N₂) and maximum inlet vacuum / forevacuum limits, where the 1603 M tolerates higher inlet pressures.

  • Pumping speed: N₂ 1360–1400 l/s; Ar 1280–1350 l/s; He 940–970 l/s; H₂ 540 l/s (both models)
  • Compression ratio (ISO 5302): N₂/Ar > 1·10⁸; He 4·10³; H₂ 5·10² (both models)
  • Ultimate pressure < 6·10⁻⁹ mbar (both models)
  • Max. forevacuum: N₂ 2.2 mbar (MT) / 1.7 mbar (1603 M); Ar 3.2/2.5; He 1.6/1.2; H₂ 0.3/0.3 mbar
  • Max. inlet vacuum: N₂ 0.015 mbar (MT) / 0.6 mbar (1603 M); Ar 0.008/0.026 mbar
  • Gas throughput at final speed: N₂ 1000 sccm (MT) / 4000 sccm (1603 M); Ar 500/1200 sccm
  • Recommended purge flow rate: 50 sccm (both models)
pp. 3
Operation, Power, Cooling and Rotational Speed

Both pumps run at 39000 rpm nominal (variable down to 15000 rpm) with run-up to 90% final speed in under 6 minutes at exhaust pressure < 0.1 mbar. Power supply is 200–240 V single-phase 50/60 Hz via the Magpower remote controller; recommended backing pumps are ADP/ADS series. Water cooling at 60 l/h, 15–40 °C is recommended; MT allows max heating temperature 75 °C, both allow baking to 120 °C.

  • Rotational speed 39000 rpm, variable from 15000 rpm up to nominal (both models)
  • Run-up time to 90% of final speed, exhaust < 0.1 mbar: < 6 min
  • Controller power supply: 200–240 V, 50/60 Hz single-phase
  • Max. start-up power consumption 650 W; nominal (stand-by) power 200 W
  • Max. heating temperature 75 °C (MT only); max. baking temperature 120 °C (both)
  • Recommended cooling water flow 60 l/h at 15–40 °C
  • Remote controller type: Magpower; recommended backing pumps: ADP/ADS
  • Sound pressure level (ISO 2151, 1 m): < 48 dB(A); vibration at nominal speed: < 0.01 μm
pp. 3-4
Dimensions, Connections and Interfaces

Both pumps measure Ø305 × 317.2–327.5 mm (height depends on flange variant) and weigh approx. 34 kg, with the controller at approx. 8.7 kg. Gas inlet is DN 200 or DN 250 with DN 40 ISO-KF outlet, N₂ purge flange DN 16 ISO-KF; a stainless steel inlet screen is factory standard. Radial magnetic field at 10 cm must stay below 12 mT; on-board control options are Profibus and Remote.

  • Dimensions Ø 305 × 317.2–327.5 mm (height depending on flange variant); weight ≈ 34 kg; controller ≈ 8.7 kg
  • Gas inlet DN 200 or DN 250; outlet DN 40 ISO-KF; N₂ purge flange DN 16 ISO-KF
  • Stainless steel inlet screen as factory setting
  • Permissible radial magnetic field at 10 cm: < 12 mT
  • On-board control options: Profibus, Remote

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Pumping speed for N₂ is 1360–1400 l/s on both ATH 1600 MT and ATH 1603 M (H₂ 540 l/s, He 940–970 l/s, Ar 1280–1350 l/s). p.2 ↗ Identical pumping-speed sizing for both variants; selection between them is driven by gas throughput/inlet pressure limits, not speed. ATH 1600 MT and ATH 1603 M
Ultimate pressure is < 6·10⁻⁹ mbar for both models. p.2 ↗ Suitable for UHV-class applications; backing pump quality and leak-tightness dominate ultimate pressure in practice. whole series
Gas throughput at final rotational speed is 1000 sccm (N₂) for ATH 1600 MT vs 4000 sccm for ATH 1603 M; Ar: 500 vs 1200 sccm. p.3 ↗ Choose ATH 1603 M for high gas-load processes; ATH 1600 MT suits low-throughput chemical/corrosive duty with heating. ATH 1600 MT vs ATH 1603 M
Max. forevacuum for N₂: 2.2 mbar (ATH 1600 MT) / 1.7 mbar (ATH 1603 M); for H₂ only 0.3 mbar on both. p.2 ↗ Backing pump and backing-line conductance must keep fore-vacuum below these values, especially for hydrogen processes. whole series
Max. inlet vacuum (inlet pressure) N₂: 0.015 mbar for ATH 1600 MT but 0.6 mbar for ATH 1603 M; Ar: 0.008 vs 0.026 mbar. p.2 ↗ ATH 1603 M tolerates roughly 40× higher N₂ inlet pressure — key criterion for processes with high transient gas loads. ATH 1600 MT vs ATH 1603 M
Rotational speed is 39000 rpm, variable from 15000 rpm to nominal speed. p.3 ↗ Speed can be reduced to ~38% for process control, reduced vibration or fast cool-down while still pumping. whole series
Run-up time to 90% of final speed is < 6 min with exhaust pressure < 0.1 mbar. p.3 ↗ System ready-time planning; venting conditions during start-up affect this figure. whole series
Controller power supply 200–240 V, 50/60 Hz single-phase; max. start-up power 650 W; nominal (stand-by) power 200 W. p.3 ↗ Single-phase supply suffices; size cabling/fusing for 650 W start-up inrush, not steady-state. whole series
Max. heating temperature 75 °C applies only to ATH 1600 MT; both allow max. baking temperature 120 °C. p.1 ↗ Bakeout up to 120 °C is possible on both; MT's integrated heating (75 °C limit) supports corrosive/condensable process conditions. MT: heating; both: baking
Recommended cooling water flow 60 l/h at 15–40 °C. p.3 ↗ Water cooling circuit (or the optional air cooling system) must be specified at 60 l/h; inlet temperature must remain within 15–40 °C to avoid overheating or condensation. whole series
Recommended purge flow rate is 50 sccm via DN 16 ISO-KF purge flange. p.3 ↗ Set purge gas (N₂) mass-flow to 50 sccm to protect bearings from process gas; the purge flange connection is DN 16 ISO-KF. whole series
Recommended backing pumps are Pfeiffer ADP/ADS series. p.3 ↗ Use diaphragm/screw dry backing pumps (ADP/ADS) for oil-free operation consistent with magnetic-bearing, contamination-free design. whole series
Sound pressure level < 48 dB(A) at 1 m (ISO 2151) and vibration < 0.01 μm at nominal speed. p.3 ↗ Installable near operators and in vibration-sensitive semiconductor/metrology setups without heavy isolation. whole series
Dimensions Ø 305 × 317.2–327.5 mm (height flange-dependent), weight ≈ 34 kg, controller ≈ 8.7 kg. p.3 ↗ Flange variant choice changes envelope height by ~10 mm; support flange and lifting equipment must handle ~34 kg pump plus ~8.7 kg controller. whole series
Gas inlet DN 200 or DN 250; outlet DN 40 ISO-KF; N₂ purge flange DN 16 ISO-KF; stainless steel inlet screen as factory setting. p.3 ↗ Match chamber flange to DN 200/250 variant; keep the factory inlet screen unless particle-ingress risk assessment allows removal. whole series
Permissible radial magnetic field at 10 cm distance is < 12 mT. p.4 ↗ Do not place strong magnets or high-current equipment within 10 cm of the pump or bearing levitation may be disturbed. whole series
On-board control options are Profibus and Remote; remote controller type is Magpower. p.3 ↗ Integrate via Profibus into process control systems, or operate via the Magpower handheld remote controller (listed as an accessory). whole series
Magnetic bearing technology means no wearing parts and no lubricant; MT version reduces particle contamination and residue build-up. p.1 ↗ No oil management or bearing maintenance schedule required; MT is preferable for particle-sensitive semiconductor and corrosive applications. whole series (MT for corrosive/particle-sensitive)