Induztr
Engineering Digest Document Viewer Related Models

QMG 250 F1 — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum PrismaPro F quadrupole mass spectrometer (RGA) brochure covering three Faraday-detector analyzer variants: QMG 250 F1 (1-100 u), QMG 250 F2 (1-200 u), QMG 250 F3 (1-300 u), with options for straight or 90° connection, filament material, standard/extended I/Os, inlet valve, and spare parts (electronics, ion source, filament, analyzer).

Section-by-Section Technical Breakdown

pp. 1
Product Overview and Accessories / Options

The PrismaPro F is a flexible quadrupole mass spectrometer for industrial and analytical applications with modular design and Ethernet connectivity via Pfeiffer MassSpec software. Its lowest detectable partial pressure is 4·10⁻¹³ hPa (mbar). Accessories and configurable options include spare electronics, ion source, filament, analyzer, vacuum gauge/chamber, inlet valve, communication interfaces, mass ranges, filament material, and straight or right-angled connection.

  • Modular design suitable for industrial and analytical environments
  • Data transfer via new generation of Pfeiffer MassSpec software
  • Lowest detectable partial pressure 4·10⁻¹³ hPa (mbar)
  • Options: mass ranges, filament material, straight or right-angled connection, standard or extended I/Os
  • Accessories/spares: electronics, ion source, filament, analyzer, vacuum gauge and chamber, inlet valve, communication interface and test equipment
pp. 2
Technical Data, Specifications and Dimensional Drawings

Technical specifications for QMG 250 F1/F2/F3, all sharing DN 40 CF-F flange, 6 mm x 125 mm rod system, Faraday detector, 200 °C analyzer operating temperature, 300 °C bake-out, and 100-240 VAC supply. Differences lie in mass range, detection limit, argon sensitivity, and neighboring-mass contribution. Dimensions are given for straight and 90°-angle configurations.

  • Mass ranges: F1 1-100 u, F2 1-200 u, F3 1-300 u
  • Detection limit (Faraday): F1 4·10⁻¹³, F2 5·10⁻¹³, F3 7·10⁻¹³ hPa (mbar)
  • Ar sensitivity (Faraday): F1 5·10⁻⁴, F2 4·10⁻⁴, F3 3·10⁻⁴ A/hPa (mbar)
  • Max. operating pressure: 5·10⁻⁴ hPa (mbar) for all variants
  • Rod system: 6 mm diameter, 125 mm length (all variants)
  • Analyzer operating temperature 200 °C; bake-out temperature 300 °C; electronics 5-50 °C
  • Connecting flange DN 40 CF-F; detector: Faraday
  • Resolution at 10% peak height: 0.5-2.5 u; dwell time 1 ms - 16 s/u
  • Dimensions straight: 371.6 x 142 x 128 mm; 90° angle: 271.4 x 142 x 300 mm; weight approx. 2.5 kg
  • Supply voltage: 100 - 240 VAC (50/60 Hz)
pp. 3
Interface and Measurement Reproducibility

All three QMG 250 F variants feature an Ethernet interface for communication and data transfer. Peak-ratio reproducibility is specified as ±0.5% for all models.

  • Interface: Ethernet (all variants)
  • Reproducibility of peak ratio: ±0.5% (all variants)

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
QMG 250 F1 mass range is 1-100 u; F2 is 1-200 u; F3 is 1-300 u. p.2 ↗ Select the variant according to the highest mass/gas species that must be resolved in the process gas spectrum. whole series (per-variant mass range)
Detection limit (Faraday): 4·10⁻¹³ hPa (F1), 5·10⁻¹³ hPa (F2), 7·10⁻¹³ hPa (F3). p.2 ↗ F1 offers the best trace-gas detection performance; longer rod systems trade sensitivity for mass range. whole series
Maximum operating pressure is 5·10⁻⁴ hPa (mbar) for all variants. p.2 ↗ A vacuum system must pump below 5·10⁻⁴ hPa before operating the analyzer to avoid filament damage and scattering. whole series
Analyzer may be operated at 200 °C and baked out at 300 °C. p.2 ↗ Suitable for integration into hot/bakeable UHV systems; electronics must be kept outside the heated zone. whole series
Electronics operating temperature is 5-50 °C. p.2 ↗ Ensure cooling/environment control for the electronics unit; it is not bakeable. whole series
Connecting flange is DN 40 CF-F (ConFlat) on all variants. p.2 ↗ Plan chamber port and CF hardware accordingly; CF allows bake-out compatibility. whole series
Rod system: 6 mm diameter, 125 mm length. p.2 ↗ Defines the physical analyzer length and resolution characteristics; relevant for spare analyzer ordering. whole series
Detector type is Faraday on all three variants. p.2 ↗ No electron multiplier option listed in this document; detection limits correspond to Faraday cup performance. whole series
Resolution at 10% peak height is 0.5-2.5 u; dwell time adjustable 1 ms to 16 s/u. p.2 ↗ Dwell time can be tuned to balance scan speed vs. sensitivity per channel in measurement scripts. whole series
Contribution to neighboring mass: <10 ppm (F1), <20 ppm (F2), <50 ppm (F3). p.2 ↗ For trace analysis adjacent to large peaks, F1 provides the best abundance-sensitivity performance. whole series
Dimensions: straight 371.6 x 142 x 128 mm; 90° angle 271.4 x 142 x 300 mm; weight approx. 2.5 kg. p.2 ↗ Choose straight vs. right-angled connection based on available clearance in the vacuum chamber envelope. whole series
Supply voltage 100-240 VAC at 50/60 Hz. p.2 ↗ Operable on standard worldwide mains without additional transformers. whole series
Interface is Ethernet on all variants. p.3 ↗ Integrate via standard network infrastructure and Pfeiffer MassSpec software; plan IP/networking for commissioning. whole series
Reproducibility of peak ratio is ±0.5% on all variants. p.3 ↗ Supports quantitative trend monitoring and process control based on partial-pressure ratios. whole series
Ar sensitivity (Faraday): 5·10⁻⁴, 4·10⁻⁴, 3·10⁻⁴ A/hPa for F1/F2/F3 respectively. p.2 ↗ Use to convert measured ion currents into partial pressures for argon-containing processes. whole series
Available options include standard or extended I/Os, inlet valve, and choice of filament material. p.1 ↗ Extended I/O enables process interlocking; filament material choice should match process gas chemistry (e.g. oxygen vs. hydrocarbon-rich environments). whole series (ordering options)