AMI 1000 — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum AMI 1000 leak detector for non-destructive leak testing (lab and inline use) of non-porous pharmaceutical packages (vials, syringes, blister, IV bags). Single model; options: customized test chambers, calibration gas supply, venting gas, customer laptop via Ethernet/WiFi, remote data access.
Section-by-Section Technical Breakdown
The AMI 1000 is a scalable leak detector for non-destructive testing in lab and inline applications, delivering deterministic, quantifiable, repeatable results per USP <1207>. It is a turnkey solution with customized test chambers adaptable to vials, syringes, blister, IV bags and other non-porous packages. Accessories and options include calibration gas supply, venting gas, Ethernet/WiFi laptop connection and remote data access.
- Scalable: lab use and inline testing
- Deterministic results per USP <1207> standard
- Customized test chambers for various non-porous package types
- Options: calibration gas supply, venting gas, customer laptop via Ethernet/WiFi
- Remote access to data
Key performance and utility requirements: sensitivity down to 0.4 µm / 2·10⁻⁵ mbar·l/s, CDA supply at 6.3–10 bar rel. per ISO 8573-1 quality 1.3.1, power supply 90–250 V AC with ~1200 W consumption. Operating environment is limited to 20–25 °C and 30–65 % humidity. Software is 21 CFR Part 11 compliant with PDF GMP test/calibration reports and local or LDAP domain authentication.
- Sensitivity: down to 0.4 µm, 2·10⁻⁵ mbar·l/s
- CDA quality: 1.3.1 per ISO 8573-1; pressure 91–145 psig / 6.3–10 bar rel.
- Typical CDA consumption: 2 Nl/cycle
- Calibration gas (optional): 91–145 psig / 6.3–10 bar rel.
- Venting gas (optional): 0–22 psig / 0–1.5 bar rel.
- Power supply: 90–250 V AC; typical consumption 1200 W
- Operating temperature 20–25 °C; humidity 30–65 %
- User interface: 10" multi-touch full HD color screen
- Software: 21 CFR Part 11 compliance, PDF GMP reports, local or LDAP authentication, optional remote data access
The unit runs Windows 10 with a 10" multi-touch display and offers network and peripheral connectivity: 1x LAN (RJ45), WiFi AP, 2x USB 3.0, and 1x HDMI. Physical data: ~130 kg weight, dimensions 913 x 558 x 1071 mm (LxWxH), sound pressure level <53 dB(A).
- Operating system: Windows 10
- Network: 1x LAN (RJ45), WiFi access point
- 2x USB 3.0 (external), 1x HDMI
- Sound pressure level: <53 dB(A)
- Weight: approx. 130 kg
- Dimensions: 913 x 558 x 1071 mm
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Detection sensitivity down to 0.4 µm / 2·10⁻⁵ mbar·l/s p.2 ↗ | Suitable for verifying package integrity at sub-micron defect sizes per USP <1207>; selection must confirm this meets the required leak-rate limit of the application | AMI 1000 |
| CDA (clean dry air) supply is required for operation at 91–145 psig / 6.3–10 bar rel. p.2 ↗ | Plant air infrastructure must deliver regulated compressed air within this range before commissioning | AMI 1000 |
| CDA quality must be 1.3.1 according to ISO 8573-1 p.2 ↗ | Air treatment (particles, water, oil classes 1/3/1) is mandatory; standard shop air will not suffice | AMI 1000 |
| Typical CDA consumption is 2 Nl per cycle p.2 ↗ | Allows sizing of compressed air supply and estimating running cost per test cycle | AMI 1000 |
| Optional calibration gas supply requires 91–145 psig / 6.3–10 bar rel. p.2 ↗ | If calibration gas option is ordered, provide a second regulated gas line at the same pressure range as CDA | AMI 1000 with calibration gas option |
| Optional venting gas pressure range 0–22 psig / 0–1.5 bar rel. p.2 ↗ | Venting line must be regulated to a much lower pressure than CDA; plan a separate low-pressure supply if venting option is used | AMI 1000 with venting gas option |
| Power supply 90–250 V AC, typical consumption 1200 W p.2 ↗ | Single-phase wide-range input; provision ≥1200 W circuit capacity (recommend appropriate breaker/headroom) | AMI 1000 |
| Operating temperature limited to 20–25 °C with 30–65 % humidity p.2 ↗ | Requires climate-controlled room; not usable in standard industrial environments without HVAC — critical for inline installation planning | AMI 1000 |
| Software is 21 CFR Part 11 compliant with PDF GMP test and calibration reports and local or LDAP domain authentication p.2 ↗ | Suitable for regulated pharma environments; plan authentication mode (local vs domain) during IT integration | AMI 1000 |
| Windows 10 operating system with 10" multi-touch full HD screen p.2 ↗ | Consider Windows 10 lifecycle/patching policy in validation and IT planning | AMI 1000 |
| Interfaces: 1x LAN (RJ45), WiFi AP, 2x USB 3.0 (external), 1x HDMI p.3 ↗ | Network data export/remote access via RJ45 or WiFi; USB for peripherals/backup, HDMI for external display | AMI 1000 |
| Weight approx. 130 kg; dimensions 913 x 558 x 1071 mm (LxWxH) p.3 ↗ | Plan handling equipment (two-person lift or hoist) and verify bench/floor loading and footprint in layout | AMI 1000 |
| Sound pressure level <53 dB(A) p.3 ↗ | Acceptable for lab/office-adjacent areas without additional noise mitigation | AMI 1000 |
| Test chambers are customized; adaptable to vials, syringes, blister, IV bags and other non-porous packages p.1 ↗ | Chamber must be specified per package format at ordering time — not a universal off-the-shelf fixture | AMI 1000 |