E-PDQ EQ — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum E-PDQ compact leak detector series in two variants: E-PDQ EQ (high-pressure version, >14.5–65 psia / 1–4.5 bar abs.) and E-PDQ MQ (low-pressure/vacuum version, 2–14.5 psia / 0.138–1 bar abs.). Brochure also lists accessories and options: LeakTek™, Adaptive Test™, Leak Rx™, fine automatic pressure/flow control, programmable pressure controller, expansion tanks (small/medium/large), two-stage precision pressure regulators, calibration orifices (ECD/ED/ECS, glass orifice), remote exhaust valve with filter, and pre-fill option for large-volume parts.
Section-by-Section Technical Breakdown
Highlights the compact footprint with integrated expansion vessel, fast test setup with LED test status indication, and low operating costs via an automated test circuit. Lists available accessories, software tools and options for calibration and pressure control.
- Compact design with integrated expansion vessel
- Fast test setup with easy-to-read LED test status indication
- Automated test circuit for low operating costs
- LeakTek™ software, Adaptive Test™ utility tools, Leak Rx™
- Fine automatic pressure/flow control device
- Programmable automatic pressure controller for constriction/blockage tests with feedback sensor
- Expansion tank options: small / medium / large
- Two-stage precision pressure regulator (small/medium/large volume)
- Verification orifice ECD, ED, ECS for air calibration; glass orifice calibrated at barometric inlet / vacuum outlet on primary standard
- Remote exhaust valve with filter and pre-fill option for large-volume parts
Comparison table of the two variants covering sensitivity, pressure range, power, environment, interfaces and dimensions. Both variants share the same sensitivity, power supply, user interfaces and mechanical envelope; they differ mainly in operating pressure range.
- Gases: air, nitrogen (both variants)
- Sensitivity: 0.03 cc/min and higher (both variants)
- Pressure range: EQ >14.5–65 psia (1–4.5 bar abs.); MQ 2–14.5 psia (0.138–1 bar abs.)
- Power supply: 24 V DC
- Operating temperature: 59–83 °F / 15–45 °C; humidity 30–80 %
- User interface: RJ45 Ethernet, RS232 serial; software LeakTek with local or domain (LDAP) authentication
- Operating system: Windows 10 remote laptop (no display); network: 1x LAN (RJ45), wired Ethernet
- Weight: approx. 8 lbs / 3 kg; dimensions 10 x 4 x 6.25 in / 254 x 102 x 159 mm
- Pneumatic connections: test port 1/8" Swagelok, air/vacuum supply 1/4" Swagelok
Defines the electrical interface signals for remote control and status reporting. Digital I/O is opto-isolated for 5–30 V DC signals; a single analog output channel provides pressure control feedback.
- Digital inputs (5–30 V DC, opto-isolated): start, stop, test type, remote pressure switch
- Digital outputs (5–30 V DC, opto-isolated): pass, fail, test type, test on, clamp, custom output
- Analog output: single channel, 0–5 V DC pressure control
- Functions marked * depend on configuration
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Minimum detectable leak rate is 0.03 cc/min for both E-PDQ EQ and E-PDQ MQ variants, using air or nitrogen as test gas. p.2 ↗ | Suitable for fine leak testing applications requiring leak rates at or above ~0.03 cc/min; not for ultra-fine (helium-mass-spectrometry-level) leak detection. | whole series |
| E-PDQ EQ operates at >14.5–65 psia (1–4.5 bar abs.); E-PDQ MQ operates at 2–14.5 psia (0.138–1 bar abs.). p.2 ↗ | Selection depends on test part pressure: EQ for pressurized (>1 bar abs.) testing, MQ for vacuum/sub-atmospheric testing; the two are not interchangeable across the full range. | E-PDQ EQ / E-PDQ MQ respectively |
| Power supply requirement is 24 V DC. p.2 ↗ | Integration must provide a 24 V DC supply (no internal mains PSU); factor this into cabinet/power design. | whole series |
| Allowed operating environment: 15–45 °C (59–83 °F) and 30–80 % humidity. p.2 ↗ | Ensure ambient control within test cells; operation outside these limits is not specified and may affect leak-rate accuracy. | whole series |
| Digital inputs/outputs are 5–30 V DC and opto-isolated. p.3 ↗ | PLC I/O can be interfaced directly without level shifting as long as signals stay within 5–30 V DC; isolation provides noise immunity in automated test stands. | whole series |
| Digital output signals include pass, fail, test type, test on, clamp and a custom output. p.3 ↗ | Enables direct wiring of pass/fail verdicts to clamp and station logic in production lines. | whole series (custom output depends on configuration) |
| Analog output is a single 0–5 V DC channel for pressure control. p.3 ↗ | Pressure feedback to external control systems must use this single channel; only one pressure variable can be monitored analog. | whole series |
| Pneumatic connections are 1/8" Swagelok for the test port and 1/4" Swagelok for air/vacuum supply. p.2 ↗ | Specify Swagelok-compatible fittings and tubing in test station plumbing design. | whole series |
| Unit weight is approx. 8 lbs / 3 kg with dimensions 254 x 102 x 159 mm (10 x 4 x 6.25 inch). p.2 ↗ | Compact form factor allows mounting directly on fixtures or within tight machine envelopes; no special lifting required. | whole series |
| Unit has no display; it is operated via a Windows 10 remote laptop over RJ45 Ethernet (1x LAN), with RS232 serial also available. p.2 ↗ | A host PC/laptop with LeakTek software must be provisioned; plan network infrastructure (Ethernet) at each test station. | whole series |
| LeakTek software supports local or domain (LDAP) authentication. p.2 ↗ | Can be integrated into corporate user-management/access-control policies for traceable operation. | whole series |
| Calibration options include verification orifices (ECD, ED, ECS) for air calibration and a glass orifice calibrated at barometric inlet / vacuum outlet against a primary standard. p.1 ↗ | Plan periodic verification against traceable standards to maintain measurement credibility; select the orifice type matching EQ (barometric inlet) or MQ (vacuum outlet) operating mode. | whole series with corresponding options |
| Options include an expansion tank (small/medium/large) and a pre-fill option for large-volume parts. p.1 ↗ | Large test part volumes require the correct expansion tank size and pre-fill option to maintain fill time and pressure stability within test cycle budgets. | whole series, depending on part volume |
| A programmable automatic pressure controller with feedback sensor enables constriction/blockage tests, and a fine automatic pressure/flow control device is available. p.1 ↗ | The same platform can run flow/flow-restriction tests in addition to pressure-decay leak testing, useful for combined part qualification. | whole series with corresponding options |