Induztr
Engineering Digest Document Viewer Related Models

ASM 306S — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum ASM 306S compact helium/hydrogen sniffer leak detector. Document covers the single ASM 306S model and its accessories/options: rigid or flexible sniffer probes with various hose lengths, hybrid cables in 2/5/10 m, probe protection filters, dedicated calibrated leaks, trolley, and industrial communication interface options.

Section-by-Section Technical Breakdown

pp. 1
Overview, Features and Accessories

The ASM 306S is a compact leak detector with a small footprint suitable for desktop use or integration into production lines. It offers high sensitivity with repeatable measurements and visual test result management on a large high-resolution color display and on the dedicated sniffer probe. A range of accessories and options is listed.

  • Compact design: suitable for desktop use or integration into production lines
  • High sensitivity and repeatable measurements
  • Test result visualization on large high-resolution color display and on the sniffer probe itself
  • Accessories/options: rigid or flexible sniffer probe with different hose lengths; hybrid cable in 2, 5 or 10 m lengths; probe protection filter; dedicated calibrated leaks; trolley; communication interface compatible with industrial standards
pp. 2
Technical Data and Dimensional Drawing

Key specifications for the ASM 306S sniffer leak detector: test method, detectable gases, minimum detectable leak rates, power supply, timing, environmental limits, interfaces, noise, weight, and dimensions. A dimensional drawing is provided.

  • Test method: sniffing test
  • Detectable gases: ⁴He and H₂
  • Min. detectable leak rate (He, sniffing): 1·10⁻⁷ mbar·l/s (1·10⁻⁸ Pa·m³/s)
  • Min. detectable leak rate (H₂): 5·10⁻⁷ mbar·l/s (5·10⁻⁸ Pa·m³/s)
  • Power supply: 100–240 V, 50/60 Hz; max power consumption 300 W
  • Start-up time at 20 °C: 2 min; response time <1 s
  • Operating temperature: 10–40 °C
  • Interfaces: I/O, RS-232, Ethernet (PC), Profibus, Profinet
  • Sound pressure level: 55 dB(A); weight approx. 22 kg
  • Dimensions (L×W×H): 350 × 305 × 421 mm

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
The ASM 306S performs sniffing tests only (no vacuum test method listed), detecting ⁴He and H₂ tracer gases. p.2 ↗ For vacuum/vessel leak testing another model is required; this unit is suited to atmospheric pressure-line sniffing applications. ASM 306S
Minimum detectable helium leak rate (sniffing): 1·10⁻⁷ mbar·l/s (1·10⁻⁸ Pa·m³/s). p.2 ↗ Sets the test specification limit for helium leak test procedures; tighter leaks cannot be detected with this instrument. ASM 306S
Minimum detectable hydrogen leak rate is 5·10⁻⁷ mbar·l/s, i.e. 5× worse than the helium limit. p.2 ↗ H₂ can be used as tracer gas where helium is unavailable, but the pass/fail threshold must be relaxed by a factor of 5. ASM 306S
Response time <1 s. p.2 ↗ Sniffer scan speed along the test part must be matched to the <1 s response to avoid missing small leaks. ASM 306S
Start-up time is 2 minutes at 20 °C ambient. p.2 ↗ Allow a 2 min warm-up before starting production leak tests or calibrating the instrument. ASM 306S
Power supply 100–240 V, 50/60 Hz, max power consumption 300 W. p.2 ↗ Works on standard single-phase mains worldwide; size supply/fusing for 300 W peak load. ASM 306S
Operating temperature range 10–40 °C. p.2 ↗ Unit must not be used outdoors or in unconditioned production areas outside this range; note calibration was specified at 20 °C. ASM 306S
Communication interfaces: I/O, RS-232, Ethernet (PC), Profibus, Profinet. p.2 ↗ Direct PLC/line integration is possible via Profibus/Profinet or digital I/O without additional converters; Ethernet for PC connection. ASM 306S
Weight approx. 22 kg, dimensions 350 × 305 × 421 mm (L×W×H). p.2 ↗ Footprint fits desktop or line integration; 22 kg allows bench placement but check mounting surface load; a trolley option exists for mobility. ASM 306S
Sound pressure level 55 dB(A). p.2 ↗ Low noise permits use in laboratory or operator-adjacent workstations without extra acoustic measures. ASM 306S
Hybrid cable offered in 2, 5, or 10 m lengths, with rigid or flexible sniffer probes and different hose lengths. p.1 ↗ Select cable/probe length based on test-part reach; longer cables may affect handling in line layouts. ASM 306S with sniffer probe options
Dedicated calibrated leaks are available as accessories, plus a probe protection filter. p.1 ↗ Plan routine calibration with the matching calibrated leak; use the filter to protect the sniffer probe from particulates in production environments. ASM 306S