Induztr
Engineering Digest Document Viewer Related Models

ATH 2804 M — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum ATH 2804 M, ATH 2804 MT, ATH 3204 M, ATH 3204 MT magnetically levitated turbomolecular vacuum pumps. M = standard; MT = with integrated heating system (max 85 °C) for corrosive/chemical conditions and reduced particle contamination. Accessories/options: installation kit, cooling water valve, purge gas valve, handheld remote controller, inlet protection screen, heater unit, valve cable and coil, various inlet flange types, various user interfaces, temperature management system for corrosive processes.

Section-by-Section Technical Breakdown

pp. 1
Features, Benefits and Accessories

Overview of the ATH 2804/3204 M/MT series: maintenance-free magnetic bearing technology with no wearing parts, low noise and vibration for vibration-sensitive semiconductor and analytical applications, and compatibility with various backing pumps. The MT version adds an integrated heating system for demanding chemical conditions. Accessories and options include installation kits, valves, remote controllers, inlet screens, heater units and various inlet flanges and user interfaces.

  • Maintenance-free, no wearing parts thanks to magnetic bearing technology
  • MT version: reduced particle contamination/build-up of residue, integrated heating system for chemical/corrosive conditions
  • Low noise (<45 dB(A)) and vibration (<0.01 µm) for vibration-sensitive applications
  • Best-in-class cooling water consumption; compatible with a variety of backing pumps
  • Options: installation kit, cooling water valve, purge gas valve, handheld remote controller, inlet protection screen, heater unit, valve cable and coil, various inlet flange types and user interfaces, temperature management system for corrosive process
pp. 2
Vacuum Performance Data (Pumping Speed, Compression Ratio, Pressures)

Pumping speeds for N₂, Ar, He, H₂ and compression ratios per ISO 5302 for all four models, together with ultimate pressure, maximum forevacuum and maximum inlet vacuum limits. ATH 3204 variants offer higher N₂ (3050 l/s) vs 2350 l/s for ATH 2804; H₂ compression ratio is 4.9·10³ for all models. MT variants have tighter max inlet vacuum limits due to heating.

  • Pumping speed N₂: 2350 l/s (2804 M/MT), 3050 l/s (3204 M/MT)
  • Pumping speed Ar: 2350 l/s (2804), 2900 l/s (3204); He: 2300 l/s (2804), 2500 l/s (3204); H₂: 1650 l/s (2804), 1700 l/s (3204)
  • Compression ratio (ISO 5302): N₂ and Ar 1·10⁹, He 9·10⁴, H₂ 4.9·10³ (all models)
  • Ultimate pressure < 1·10⁻⁸ mbar (all models)
  • Max. forevacuum N₂: 3.4 mbar (M), 3 mbar (MT); Ar: 4.4 mbar (M), 3.7 mbar (MT)
  • Max. inlet vacuum N₂: 0.4/0.38 mbar (M), 0.04 mbar (MT); Ar: 0.03 mbar (M), 0.01 mbar (MT); He/H₂: >1.0 mbar (M only)
pp. 3
Operating, Electrical, Thermal and Gas Throughput Data

Gas throughput at final rotational speed, purge flow, rotational speed and run-up time, plus electrical supply, power consumption, heating/baking temperature limits and cooling water requirements for all models. All variants run at 25 000 rpm (variable 15 000 rpm to nominal) with <8 min run-up. MT variants draw higher stand-by power (590 W) due to heating and are limited to 15–25 °C cooling water.

  • Gas throughput at final speed N₂: >5000 (cold) sccm (M), 2000–2800 sccm (MT); Ar: >2500 (cold) sccm (M), 700–1500 sccm (MT); He/H₂: >5000 (cold) sccm (M only)
  • Recommended purge flow rate: 50 sccm (all models)
  • Rotational speed 25 000 rpm; variable 15 000 rpm to nominal
  • Run-up to 90% final speed < 8 min (with pressure exhaust < 0.1 mbar)
  • Controller power supply: 200–240 V, 50/60 Hz, single-phase; max start-up power 1200 W
  • Nominal (stand-by) power: 160 W (M), 590 W (MT); max heating temperature MT: 85 °C; max baking temperature 120 °C (all)
  • Recommended cooling water flow rate 60 l/h; water temperature 15–40 °C (M), 15–25 °C (MT)
pp. 3-4
Mechanical, Connection and Control Interface Data

Physical dimensions, weights, flange connections, purge flange, magnetic field limits and control options. Pumps use on-board controllers (OBC) with Profibus and Remote options. The ATH 3204 height varies with flange variant (399.3–481.3 mm).

  • Dimensions Ø x H: 371 x 471.3 mm (2804 M/MT); 371 x 399.3–481.3 mm (3204 M/MT, depending on flange variant)
  • Weight approx.: 98 kg (2804 M), 101 kg (2804 MT), 80–100 kg (3204 M), 103 kg (3204 MT)
  • Gas inlet/outlet: DN 250 / DN 40 ISO-KF or DN 50 ISO-KF (2804); DN 320 / DN 40 ISO-KF or DN 50 ISO-KF (3204)
  • N₂ purge flange: DN 16 ISO-KF (all models)
  • Stainless steel inlet screen factory-fitted on all models
  • Permissible radial magnetic field at 10 cm: 5 mT
  • Sound pressure level (ISO 2151, 1 m, nominal speed): <45 dB(A); vibration <0.01 µm
  • On-board control (OBC): Profibus, Remote; recommended backing pump ADP/ADS (2804 only)

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Recommended purge gas flow rate is 50 sccm through a DN 16 ISO-KF purge flange for all ATH 2804/3204 M/MT models. p.3 ↗ Size purge gas supply/mass-flow control to 50 sccm and plumb a DN 16 ISO-KF line when planning purge for corrosive or particulate-sensitive processes. All models
Controller power supply is 200–240 V, 50/60 Hz single-phase with max start-up power of 1200 W. p.3 ↗ Provide a single-phase 200–240 V circuit rated for ≥1200 W inrush/start-up per pump. All models
Nominal (stand-by) power is 160 W for M variants and 590 W for MT variants due to the integrated heater. p.3 ↗ MT versions roughly quadruple idle power draw; account for this in facility load and thermal budgeting. MT variants vs M variants
Maximum heating temperature for MT versions is 85 °C; maximum baking temperature for all models is 120 °C. p.3 ↗ Keep system bake-out at or below 120 °C; MT heater operation must not exceed 85 °C. All models (baking), MT variants (heating)
Cooling water flow rate is 60 l/h; permissible water temperature is 15–40 °C for M and only 15–25 °C for MT variants. p.3 ↗ MT installations require chilled water (≤25 °C) because of integrated heating; standard chillers setpoints must be verified. M: 15–40 °C; MT: 15–25 °C
Run-up time to 90% of final rotational speed is <8 min at exhaust pressure <0.1 mbar, at 25 000 rpm nominal speed (variable down to 15 000 rpm). p.3 ↗ Ensure backing pump maintains <0.1 mbar exhaust during start-up to achieve the quoted run-up; speed can be throttled to 15 000 rpm for process tuning. All models
Ultimate pressure is <1·10⁻⁸ mbar for all four models. p.2 ↗ Suitable for high/UHV applications; base pressure depends on proper backing and bake-out. All models
Maximum forevacuum (N₂) is 3.4 mbar (M) / 3 mbar (MT); for Ar 4.4 mbar (M) / 3.7 mbar (MT). p.2 ↗ Backing pump sizing must keep forevacuum below these limits for the process gas used; MT versions tolerate slightly lower forevacuum. All models (MT limits stricter)
Maximum inlet vacuum for N₂ is 0.4/0.38 mbar (M) but only 0.04 mbar for MT variants; He/H₂ >1.0 mbar applies only to M versions (not specified for MT). p.2 ↗ At high inlet pressures (soft start / high throughput), MT heating restricts operation; use M variants where high inlet pressure with He/H₂ is required. MT vs M variants
Gas throughput at final speed for N₂ is >5000 sccm (cold) for M variants but only 2000–2800 sccm for MT variants; Ar: >2500 vs 700–1500 sccm. p.3 ↗ MT heating halves or more the high-pressure gas throughput capacity; check process gas loads against MT limits before selecting MT for corrosive processes. MT vs M variants
Pumping speed N₂ is 2350 l/s for ATH 2804 M/MT and 3050 l/s for ATH 3204 M/MT; H₂ speed is 1650 l/s (2804) and 1700 l/s (3204). p.2 ↗ Select ATH 3204 when ~30% higher N₂/Ar speed is needed; H₂ performance is nearly identical between sizes. 2804 vs 3204 variants
Compression ratios (ISO 5302): N₂ and Ar 1·10⁹, He 9·10⁴, H₂ 4.9·10³, identical for all models. p.2 ↗ H₂ compression is relatively low (4.9·10³); for hydrogen-heavy processes verify backing pump base pressure can hold the required inlet H₂ pressure. All models
Gas inlet/outlet connections: DN 250 (ATH 2804) or DN 320 (ATH 3204) inlet with DN 40 ISO-KF or DN 50 ISO-KF outlet; ATH 3204 height varies 399.3–481.3 mm depending on flange variant. p.3 ↗ Verify inlet flange variant ordering (affects height clearance and weight: 80–100 kg for 3204 M) during system layout. ATH 3204 (dimension variation); all models (connections)
Permissible radial magnetic field at 10 cm distance is 5 mT; stainless steel inlet screen is the factory setting on all models. p.4 ↗ Keep strong magnetic sources (magnets, high-current busbars) at least 10 cm away or shield them; remove inlet screen only if particulate generation outweighs protection. All models
Sound pressure level is <45 dB(A) at 1 m (ISO 2151) at nominal speed and vibration is <0.01 µm. p.3 ↗ Pumps can be located close to vibration-sensitive metrology/semiconductor tools without dedicated acoustic or vibration isolation in most cases. All models
Control is via pump-mounted on-board controller (OBC) with Profibus and Remote options; a handheld remote controller is available as an accessory. p.3 ↗ Plan Profibus integration or Remote interface cabling in the tool layout; no external controller rack is needed. All models
Recommended backing pump is ADP/ADS (diaphragm/scroll) for ATH 2804 M/MT; not specified for ATH 3204. p.3 ↗ Pair ATH 2804 with Pfeiffer ADP/ADS backing pumps for a validated combination; for ATH 3204 backing selection must be verified separately against forevacuum limits. ATH 2804 M/MT specifically