321-78382

Technical Reference Summary

Covered Models & Document Scope

This manual (321-78382D, Nov. 2024) covers the Shimadzu Analytical Balance AP Series: AP W Series (AP225W, AP135W, AP225WD, AP125WD, AP324W, AP224W, AP124W), AP X Series (AP324X, AP224X, AP124X), AP Y Series (AP324Y, AP224Y, AP124Y), and AP W-AD Series (AP225W-AD, AP135W-AD, AP225WD-AD, AP125WD-AD, AP324W-AD, AP224W-AD). W-AD variants add automatic doors, touchless sensors, and a standard STABLO-AP ionizer; semi-micro models (AP135W, AP225WD and their -AD variants) have 0.00001 g minimum display and an Adjustable Windbreak Plate.

Section-by-Section Technical Breakdown

pp. 18-20
Installation Environment and Parts Attachment

Specifies the required installation environment and the sequence for attaching glass doors, stopper knobs, the Adjustable Windbreak Plate (semi-micro models), the ionizer, and parts around the pan and protective cover. The W-AD Series glass doors must be fitted with stopper knobs before power-on.

  • Install on a strong, stable, flat table or floor away from air flow, extreme temperature changes, vibration, direct sunlight, dust, electromagnetic waves, magnetic fields, and condensation.
  • Attach glass doors (top, right, left) and verify stopper knobs are attached before turning power ON (W-AD Series).
  • Attach the Adjustable Windbreak Plate on semi-micro models only; removal method is also described.
  • Wire the ionizer (STABLO-AP) and attach parts around the pan and protective cover.
  • Pan area configuration affects weighing performance (W-AD Series).
pp. 37
Basic Operations: Touchless Sensors and Adjustable Windbreak Plate

Describes touchless sensor operation on W-AD Series balances, including LED indication and the two operating modes, and the use of the Adjustable Windbreak Plate on semi-micro models to reduce weighing chamber volume and convection effects.

  • Swipe hand over left or right sensor; blue LED lights while the sensor is actively detecting.
  • A red LED indicates a fault; refer to Troubleshooting (P.153).
  • Touchless sensors are only usable while weight values are displayed, not on menu screens, stand-by mode, or during calibration.
  • Touchless sensor settings are retained after power OFF; position memory settings are reset.
  • Adjustable Windbreak Plate locks only at slide guide scale positions; release handle so the plate top aligns with the scale.
  • The Adjustable Windbreak Plate reduces chamber volume and convection for faster measurement.
pp. 51
Calibration and Perfect Self Calibration (PSC)

Explains how to call the calibration menu, enable or disable PSC, and set enforcement conditions for automatic sensitivity adjustment: temperature fluctuation, precision allowed, and sample mass. Also notes restrictions for verified balances and when using Multi-stand or AP Holder.

  • Press CAL for three or more consecutive seconds to call the calibration menu.
  • PSC can be set ON or OFF in operation mode.
  • Temperature change enforcement condition range: 0.5 to 10 degrees Celsius.
  • Precision Allowed range: 0.0001 to 0.1000 g; lower values cause more frequent adjustment.
  • Sample mass range: 1 g to the weighing capacity.
  • If the weight mark continues to blink, sensitivity calibration has not started; see Troubleshooting (P.153).
  • For EU verified use, operate within the temperature range on the verification label.
  • When using Multi-stand or AP Holder, remove it from the pan or turn PSC off.
pp. 128-152
Connections and Communications with External Equipment

Covers connecting printers, PCs, PLCs and serial equipment, RS232C cable connections, data formats, command list, communication settings, serial communication troubleshooting, output functions, and USB device connection for W-AD/W Series.

  • Serial communication troubleshooting covers cable type, USB serial converter COM number and driver, Balance Keys startup, garbled text, Excel cell movement, Japanese conversion, communication speed, and PC input device interference.
  • Communication settings include standard settings and user settings.
  • Useful output functions include auto-print, interval timer, output timing, GLP output, weighing value GLP output, and print detail settings.
  • USB devices (W-AD/W Series only): save weighing values, calibration records and weighing screens to USB memory; operate with USB keyboard; transfer data with USB barcode readers.
pp. 153-154
Errors and Recovery

Provides troubleshooting tables for display, weighing, calibration, touchless sensor, ionizer, automatic door, and serial communication problems, plus error messages. Each row lists problem, probable cause, recovery action, and reference page.

  • Troubleshooting table covers no display, frozen display, flickering display with no stability mark, incorrect results, blinking weight mark, red touchless sensor LED, ionizer POWER/ALARM LEDs, door failure, and 0 g not displayed with NET icon.
  • Serial communication function errors are cross-referenced to P.140.
  • Error messages are listed separately (P.154).
pp. 155-161
Maintenance

Describes maintaining the balance and the ionizer, removing the glass draft shield, precautions when moving or shipping the balance for repair, daily and periodic inspections, and weight types and selection.

  • Ionizer maintenance includes checking and replacing the discharge electrode if bent or missing.
  • Remove the glass draft shield as described.
  • Before transport for repair, remove the pan, pan supporter, shield plate, shield case, stage, Draft shield inner plate (semi-micro), glass doors, and Adjustable Windbreak Plate.
  • Daily inspections and periodic inspections are required; periodic inspections and regular calibrations are requested from Shimadzu service.
  • Weights: types and selection are described.
pp. 162-168
Specifications, Maintenance Parts, and Options

Lists specifications, maintenance parts, and special accessories (options). Also includes technical documentation on electronic balance inspection methods, classes and tolerances, and the menu map.

  • Maintenance parts list and special accessories (options) list are provided.
  • Electronic balance inspection methods include classes and tolerance.
  • Menu map explains how to use the menu map.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
PSC temperature change enforcement condition can be set within a range of 0.5 to 10 degrees Celsius. p.51 ↗ Set the temperature fluctuation threshold for automatic sensitivity adjustment according to the stability of the installation environment. AP Series
PSC Precision Allowed parameter can be set within a range of 0.0001 to 0.1000 g; lower values cause more frequent adjustment. p.51 ↗ Choose a lower value for tighter accuracy at the cost of more frequent automatic adjustment. AP Series
PSC Sample mass can be set within a range of 1 g to the weighing capacity. p.51 ↗ Set the most commonly weighed sample mass so automatic adjustment is optimized for typical use. AP Series
When using a verified balance as a legal measuring instrument in the EU, the balance must be used within the temperature range indicated on the verification label. p.51 ↗ Verify the ambient temperature range on the verification label before legal-for-trade use. EU verified balances
When using the Multi-stand or AP Holder, remove it from the pan or set the PSC function off. p.51 ↗ Avoid automatic sensitivity adjustment errors caused by accessories on the pan. AP Series with Multi-stand or AP Holder
The withstand load of the Draft shield inner plate (semi-micro models only) is 220 g. p.10 ↗ Do not apply more than 220 g load to the plate glass during installation. Semi-micro models
Touchless sensor settings are retained after power OFF, but position memory function settings are reset. p.37 ↗ Reconfigure position memory after power cycling if that function is used. W-AD Series
Touchless sensors cannot be used on the menu screen, in stand-by mode, during calibration, or in any situation other than displaying weight values. p.37 ↗ Return to the weight display before attempting touchless sensor operation. W-AD Series
The Adjustable Windbreak Plate can only be locked at positions where the top of the plate aligns with the slide guide scale; it cannot be locked in the middle of the scale. p.37 ↗ Release the handle at a scale position to lock the plate; intermediate positions are not valid. Semi-micro models
The ionizer AC cord must only be used with a grounding outlet/receptacle. p.6 ↗ Failure to ground the ionizer decreases neutralization characteristics (neutralization time, ion balance). Ionizer STABLO-AP
The ionizer discharge electrode is high voltage and must be replaced if bent or missing. p.8 ↗ Using the ionizer with a bent or missing electrode reduces neutralization performance; contact Shimadzu for replacement. Ionizer STABLO-AP
The design standard period of usage for the product is ten years. p.8 ↗ Using beyond ten years may make it impossible to maintain performance or may cause malfunctions. AP Series
Replacement parts are supplied up until seven years after termination of manufacture. p.12 ↗ Plan spare parts procurement within the supply period. AP Series
The balance complies with EN61326-1 class B for emissions and industrial electromagnetic environment for immunity. p.11 ↗ In industrial locations, locate away from strong electromagnetic noise sources and use a separate power source. AP Series
For serial communication with a USB serial converter, the correct COM number must be set in Windows settings and the supplied driver must start normally. p.140 ↗ Check COM port assignment and driver installation if serial communication does not work. AP Series
If communication operations seem strange, set the communication speed on Balance Keys in alignment with the PC performance and avoid the continuous output function. p.140 ↗ Slow communication speed or too-short transmission intervals can cause malfunctions depending on PC processing performance. AP Series
When transporting the balance for repair, remove the pan, pan supporter, shield plate, shield case, stage, Draft shield inner plate (semi-micro), glass doors, and Adjustable Windbreak Plate. p.11 ↗ Failure to remove these parts may damage the draft shield glass or plate glass. AP Series
When moving the balance, do not hold it by the top frame; grip the bottom of the main unit firmly with both hands. p.10 ↗ Holding by the ceiling frame can cause the frame or front glass to loosen and the balance to drop. AP Series
If semi-micro models are moved while connected to the power supply unit, the load on the cable could damage it; always disconnect the cable before transporting. p.10 ↗ Disconnect the power supply unit cable before moving semi-micro models. Semi-micro models
The protective cover must be securely attached with double-sided tape and kept clean to prevent touchless sensor malfunctions. p.10 ↗ A rising or dirty protective cover can cause unintended touchless sensor operations and sample spills. W-AD Series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Nothing is displayed on the screen p.153 ↗ Is the power cord unplugged? Is the main power breaker switched off? Does the voltage match? Check the power supply and voltage, and make sure the connections are correct. P.26
The display stays the same even when a sample is placed on the pan p.153 ↗ Is the pan positioned correctly? Make sure the pan is positioned correctly. P.21
The display flickers, and (Stability Mark) is not displayed p.153 ↗ Has the balance been installed in an unstable location? Remove the effects of vibrations and air flows. Install the balance on a firm base. P.18
The display flickers, and (Stability Mark) is not displayed p.153 ↗ Is the sample overlapping the edge of the pan? Make sure the sample does not overlap the edge of the pan. —
The display flickers, and (Stability Mark) is not displayed p.153 ↗ Is something other than the sample coming into contact with the pan? Make sure that nothing other than the sample comes into contact with the pan. —
The display flickers, and (Stability Mark) is not displayed p.153 ↗ Are the glass draft shields open? Close all glass draft shields and read the display again. —
The measurement results are incorrect p.153 ↗ Has adjustment been performed? Perform adjustment. P.44
The weight mark continues to blink p.153 ↗ There is a chance that changes in the installation environment have drastically misaligned the zero point. Or, a weighing sample has been placed on the pan. Sensitivity calibration won’t start. Check to make sure that nothing has been placed on the pan, and then perform sensitivity calibration manually by pressing [CAL]. Or, press [POWER] and unplug the AC adapter with [STAND-BY] displayed, and then plug in the adapter and press [POWER] again. Sensitivity calibration will be performed automatically. P.51
The weight mark continues to blink p.153 ↗ Does the product use any solid specific gravity measurement or other applications? The product is designed so that automatic sensitivity adjustment is not carried out during specific application operation. If you wish to carry out span calibration, switch to standard measurement mode. P.88
The touchless sensor LED remains red (W-AD Series only) p.153 ↗ The touchless sensor may be malfunctioning or damaged. Turn the balance main body and ionizer power OFF, and then ON again. If the problem is not resolved, contact a Shimadzu Corporation sales location, distributor, or authorized service shop. P.26
The ionizer POWER LED is not lit up. (When the ionizer is installed on the balance main body) p.153 ↗ Is the ionizer power cord unplugged? Check that the ionizer AC adapter and AC cord are connected correctly. P.124
The ionizer POWER LED is not lit up. (When the ionizer is installed on the balance main body) p.153 ↗ Is the ionizer power switch on the back turned ON? Check the ionizer power switch. P.124
The ionizer ALARM LED is lit. (When the ionizer is installed on the balance main body) p.153 ↗ If the ALARM LED remains lit even after restarting several times, there may be a malfunction. Turn the power switch OFF and ON again. If the ALARM LED remains lit even after restarting several times, there may be a malfunction. Contact a Shimadzu Corporation sales location or distributor for details. P.124
The ionizer ALARM LED is lit. (When the ionizer is installed on the balance main body) p.153 ↗ Is the discharge electrode dirty? Check if the discharge electrode is dirty. If it is dirty, refer to “11. Maintenance”. P.156
The doors do not open or close even if the door open and close keys are pressed or manual trigger operation is carried out. (W-AD Series only) p.153 ↗ Are the glass doors loose or disconnected? Are any operations being carried out which forcibly stops glass door opening and closing operations? Are the stopper knobs loose or missing? Check that the glass doors and stopper knobs are properly attached, and turn the balance main body power OFF then ON again, or carry out automatic door initialization. If the glass doors or stopper knobs are loose or missing, reattach them before cycling the power. P.18, P.36
0 g is not displayed even though power was turned ON with nothing on the pan. The NET icon will be displayed. p.153 ↗ Has an ionizer, automatic door, or USB memory error occurred during span calibration? Refer to “Error Messages”, and turn the power back ON after resolving the error. P.26, P.154
I set up the serial communication function, but nothing happens. p.140 ↗ Are you using the correct communication cable? Check the type and connection of the connected communication cable. Serial Communication Function Troubleshooting
I am using an USB serial converter, but the serial communication function doesn’t work. p.140 ↗ Has the correct COM number been set? Check the setting for the COM number is correct in the Windows settings. For details on checking and changing the COM number, refer to the Shimadzu website. Serial Communication Function Troubleshooting
I am using an USB serial converter, but the serial communication function doesn’t work. p.140 ↗ Is the driver that is supplied with the USB serial converting starting up normally? There is a chance that the driver provided with the USB serial converter was not set up properly. For details on the setup method, refer to the Shimadzu website. Serial Communication Function Troubleshooting
The serial communication function does not work when the PC is rebooted. p.140 ↗ Have you started up the Balance Keys software? The serial communication function cannot be used if Balance Keys has not been started up. For details on Balance Keys, refer to the Shimadzu website. Serial Communication Function Troubleshooting
Garbled text is loaded onto the PC. p.140 ↗ Are the serial communication function settings correct? The settings for the serial communication function on the balance or PC are incorrect. Reset the settings. ^ [Communication settings] (P.137) Serial Communication Function Troubleshooting
The cells are not transferred when I enter the data on Excel. p.140 ↗ Has Excel been set so that the cell does not move when the [Enter] key is pressed? Click on the [File] tab on Excel, and then click on [Options]. Click on [Advanced Settings], and add a tick to [Move cell after pressing the [Enter] key]. Serial Communication Function Troubleshooting
The cells are not transferred when I enter the data on Excel. p.140 ↗ Has the Japanese conversion function on Windows been set at ON? Set the Japanese conversion function on Windows at OFF. Serial Communication Function Troubleshooting
The operations sometimes seem strange. p.140 ↗ Has the communication speed been correctly set on [Balance Keys]? There are cases in which operations malfunction if the speed of communications is set too slow, depending on the processing performance of the PC. Set this in alignment with the communication speed. There are also cases in which malfunctions occur if the interval set for transmitting data from the balance is set too short. Set this so that subsequent items of data are transmitted after they have been displayed on the screen. Also, avoid using the continuous output function in this event. Serial Communication Function Troubleshooting
The operations sometimes seem strange. p.140 ↗ Is anything touching the PC input device? Make sure nothing is touching the PC keyboard or mouse when data is being transmitted from the balance. Serial Communication Function Troubleshooting