AT-R Series Balance Instruction Manual
Technical Reference Summary
Covered Models & Document Scope
ATX-R and ATY-R series balances. The document covers both series; ATX-R models additionally have Perfect Self Calibration (PSC) and internal weight calibration. Verified balances as legal measuring instruments are distinguished where functions are restricted. Model examples shown include ATX224R and ATY64R.
Section-by-Section Technical Breakdown
Describes unpacking, fitting the pan supporter, pan, pan ring and protective cover, and leveling the balance using the two front level screws. The balance must be leveled before use.
- ATX-R/ATY-R package includes balance main body with windbreak, pan, pan supporter, pan ring, AC adaptor and protective cover.
- Align the two pan notches with the left and right of the main body when placing the pan.
- Turning level screws clockwise as viewed from above raises the balance; counterclockwise lowers it.
- Turn both front level screws counterclockwise until they gently stop, then level.
Covers connecting the AC adapter, the self-check display sequence, and the required warming-up period before calibration or accurate measurement. ATX-R performs an internal weight mechanism check at startup.
- Startup display sequence ends with OFF display; first information shown is the software version.
- ATX-R series automatically checks the internal weight mechanism at power-on, producing a small motor noise.
- Warm up with power ON for at least one hour before calibration or accuracy measurement.
- Warming up is also accomplished in standby mode.
Describes placing samples in the windbreak, reading the display when the stability mark lights, and practices to avoid for stable results. Also covers unit switching and display digit selection.
- Shut the glass door fully before reading the display.
- Do not put hands inside the windbreak, touch containers/samples with bare hands, or weigh samples of different temperatures.
- Use forceps or gloves; allow temperature equilibration inside the glass door.
- Factory default registers only grams; other units must be registered before switching.
- 1d/10d display selection is not applicable to a verified balance as a legal measuring instrument.
Explains menu structure, numeric value entry, menu reset to defaults, menu lock, and output of menu setting information. The default password is 9999.
- Menu reset returns all settings to defaults indicated by asterisks in the menu map.
- Default password is 9999 before shipment.
- Menu setting information can be output to a PC or printer via [TOOLS]-[SYSTEM]-[COND.OUT].
- Menu lock can be released by repeating the lock steps.
Covers span calibration with external weights, Perfect Self Calibration (PSC) for ATX-R, internal weight calibration for ATX-R, and leaving a calibration record. Includes error handling for ERR C and PLS.CAL.
- Span calibration requires placing the calibration weight on the pan and shutting the glass door.
- If no operation is performed within 60 seconds during calibration, ERR C is displayed.
- PSC is available on ATX-R only; when PSC triggers, PLS.CAL is displayed after two minutes of blinking.
- Internal weight calibration is ATX-R only.
- Calibration records can be output to a PC or printer; the balance has no clock function so date/time cannot be output.
- Balance ID is a four-digit management number, default 0000.
Describes the zero tracking function that compensates zero-point drift, and the zero/tare timing change function that selects whether zero/tare waits for stability. Auto zero range is limited to 99 d.
- Zero tracking is ON by default and indicated by the zero tracking symbol.
- Auto zero range upper limit is 99 d, where d is the minimum indication in the displayed unit.
- For a balance with 0.0001 g minimum indication, the zero range upper limit is 0.0099 g in grams and 0.099 ct in carats.
- Zero/tare timing change function default is execution without waiting for the stability mark.
- Zero/tare timing change is not applicable to a verified balance as a legal measuring instrument.
Covers selecting general weighing mode or pouring mode, easy setting of response vs stability, stability detection range, and stability mark lighting timing.
- General weighing mode gives equal emphasis to response and stability.
- Pouring mode is suited to pouring powder or liquid until a target weight is reached; display update is fast.
- Stability detection range options: 0.5d, 1d, 10d, 50d, 100d, 1000d.
- If data output triggered by stability detection is very slow, increase the stability detection range.
- Stability mark lighting timing can be set to faster or standard; standard gives stricter stability judgment.
Lists displayable units and conversion factors, and describes selecting units and setting user-specified units with custom conversion factor and minimum indication.
- Available units include g, mg, ct, mom and user-specified units.
- Gram conversion: g=1, mg=0.001, ct=0.2, mom=3.75.
- Conversion factor: g=1, mg=1000, ct=5, mom=0.2666667.
- User-specified units allow setting conversion factor and minimum indication.
- User-specified units are not applicable to a verified balance as a legal measuring instrument.
Describes piece counting, percentage weighing, specific gravity measurement (solids and liquids), and formulation modes. Pressing a key for 3 seconds toggles between weighing mode and application function mode.
- Piece counting sets a unit weight and counts pieces.
- Percentage weighing measures weight as a percentage of a reference weight.
- Specific gravity measurement covers solid density and liquid density.
- Formulation mode is for mixing multiple samples according to a formula.
- Application function mode can be used in combination with the comparator function.
Covers connecting external equipment via RS232C and USB device connectors, output timing, auto print, and communication with PCs and printers such as EP-100 or EP-110.
- RS232C connector connects to a printer such as EP-100 or EP-110.
- USB device connector connects to a PC.
- DC IN connector is for the AC adapter.
- Auto print function is indicated by the auto print symbol.
- Output timing change function is available.
Covers maintaining the balance and handling weights, including types of weights and their selection.
- Maintenance procedures are described for keeping the balance in working order.
- Types of weights and their selection are described for calibration.
- Maintenance parts are listed for ATX-R/ATY-R series.
Provides guidance on what to do if problems occur and how to respond to displayed messages such as ERR H, ERR C, WAIT, BUSY and PLS.CAL.
- ERR H at startup: see Responding to Messages.
- ERR C during calibration: calibration not completed; press key and redo, or calibrate internal weight.
- WAIT: calibration record is being output; span calibration starts automatically when output finishes.
- BUSY: something is on the pan; remove it.
- PLS.CAL: unload the pan and execute span adjustment.
Lists specifications for ATX-R/ATY-R series balances, including weight value ranges usable for calibration.
- Specifications are given for ATX-R/ATY-R series.
- The range of weight values that can be used for calibration is referenced in the specifications.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Warm up the balance with power ON for at least one hour before calibration or accuracy measurement. p.29 ↗ | Allow sufficient warm-up time after power-on to achieve stable and accurate weighing results. | ATX-R/ATY-R series |
| If no operation is performed within 60 seconds during span calibration, ERR C (calibration error) is displayed. p.31 ↗ | Complete each calibration step within 60 seconds to avoid calibration failure. | ATY-R series (span calibration procedure shown) |
| The auto zero range upper limit is 99 d, where d is the minimum indication in the displayed unit. p.67 ↗ | Zero range setting is unit-dependent; for 0.0001 g minimum indication the limit is 0.0099 g in grams and 0.099 ct in carats. | ATX-R/ATY-R series |
| Stability detection range can be selected from 0.5d, 1d, 10d, 50d, 100d, 1000d. p.75 ↗ | Increase the stability detection range if data output triggered by stability detection is very slow. | ATX-R/ATY-R series |
| Default password is 9999 before shipment. p.47 ↗ | Use 9999 to perform menu reset or other password-protected operations unless the password has been changed. | ATX-R/ATY-R series |
| Balance ID is a four-digit management number; default ID is 0000. p.63 ↗ | Set a unique four-digit ID for each balance when managing multiple balances; it is included in calibration records. | ATX-R/ATY-R series |
| Gram conversion factors: g=1, mg=0.001, ct=0.2, mom=3.75. p.77 ↗ | Use these factors to convert displayed values between units. | ATX-R/ATY-R series |
| Conversion factors for selected units: g=1, mg=1000, ct=5, mom=0.2666667. p.77 ↗ | Multiply gram value by the conversion factor to obtain the value in the selected unit. | ATX-R/ATY-R series |
| User-specified units allow setting the conversion factor and minimum indication as required. p.77 ↗ | Custom units can be configured for special applications, but not on a verified balance as a legal measuring instrument. | ATX-R/ATY-R series (not verified balances) |
| Zero tracking function is ON by default. p.65 ↗ | Zero-point fluctuations due to temperature changes and other factors are automatically compensated when the indication is zero. | ATX-R/ATY-R series |
| Zero/tare timing change function default is execution without waiting for the stability mark. p.69 ↗ | Zero/tare can be performed immediately without waiting for stability; this function is not applicable to a verified balance as a legal measuring instrument. | ATX-R/ATY-R series (not verified balances) |
| PSC (Perfect Self Calibration) is available on ATX-R only. p.56 ↗ | ATX-R models can perform self-calibration; ATY-R models do not have PSC. | ATX-R series |
| Internal weight calibration is ATX-R only. p.58 ↗ | Only ATX-R models can calibrate the internal weight. | ATX-R series |
| When PSC triggers, PLS.CAL is displayed after two minutes of blinking; unload the pan and execute span adjustment. p.57 ↗ | If PLS.CAL appears, remove any load from the pan and perform span calibration. | ATX-R series |
| The balance has no clock function; date and time cannot be output. p.61 ↗ | Calibration records cannot include date/time stamps from the balance itself. | ATX-R/ATY-R series |
| RS232C connector connects to a printer such as EP-100 or EP-110; USB device connector connects to a PC. p.21 ↗ | Use RS232C for printer connection and USB device for PC connection. | ATX-R/ATY-R series |
| 1d/10d display selection is not applicable to a verified balance as a legal measuring instrument. p.37 ↗ | Verified balances cannot reduce the minimum displayed digit by one digit. | Verified balances |
| Stability mark lighting timing can be set to faster or standard; standard gives stricter stability judgment and more reliable data. p.75 ↗ | Choose faster timing for higher throughput, standard timing for higher accuracy. | ATX-R/ATY-R series (not verified balances) |
| Do not disconnect the AC adapter while [WAIT] or [SET] is displayed. p.41 ↗ | Disconnecting during these states risks corrupting data in the scale. | ATX-R/ATY-R series |
| If calibration does not end normally and the balance stops, do not move it; turn the power on and start it up correctly so the internal weight is correctly held. p.31 ↗ | Moving the balance after a failed calibration may cause failure because the internal weight is not held correctly. | ATX-R/ATY-R series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| ERR H p.29 ↗ | Displayed at startup; see Responding to Messages. | See Responding to Messages (page 139). | Responding to Messages, page 139 |
| WAIT p.31 ↗ | The calibration record is being output. | When output has finished, span calibration will start automatically. | |
| BUSY p.31 ↗ | There is something placed on the pan. | Take the item off the pan and follow the procedure. | |
| ERR C p.31 ↗ | No operation was performed within 60 seconds during span calibration. | Press the key and repeat the operation from the beginning. | |
| ERR C p.53 ↗ | Span calibration was not completed for one of the following reasons: too large a discrepancy between the zero point of the balance and the sensitivity; a container has been placed on the pan; the pan is not on the balance; too large a discrepancy in the value of the internal weight. | Press the key and redo the operation from the beginning. If the same display reappears, calibrate the internal weight (page 58). | |
| ERR C p.59 ↗ | The internal weight has not been calibrated for one of the following reasons: the wrong weight has been placed on the pan; no operation has been performed within 60 seconds of the flashing weight value or zero display. | Press the key and repeat the operation from the beginning. | |
| PLS.CAL p.57 ↗ | PSC triggered and two minutes passed since mass display blinking started. | Unload the thing on the pan and execute span adjustment. | Push the key when you must keep measuring after PLS.CAL is displayed; return to blinking mass display. |