Shimadzu Analytical Balance Instruction Manual AU Series
Technical Reference Summary
Covered Models & Document Scope
This manual covers the Shimadzu AUW-D, AUW, AUX and AUY series analytical balances. Models: AUW220D and AUW120D (AUW-D dual-range series), AUW320, AUW220, AUW120 (AUW series), AUX320, AUX220, AUX120 (AUX series), AUY220, AUY120 (AUY series). The manual notes that menu settings and calibration display examples are mainly for the AUW220 model; numerical values may differ on other models.
Section-by-Section Technical Breakdown
Describes the main components of the balance, the key panel functions, and the display symbols. It covers the glass doors, pan, anti-draft ring, level screws, connectors, and the function of each key during measurement and menu selection.
- Main components include glass doors (3 places), pan, pan supporter, anti-draft ring, level screws, level indicator, data I/O connector, DC IN connector, RS-232C connector, keyboard connector, below-weigh hook cover, and ground terminal.
- Key panel keys: [POWER], [CAL], [O/T], [UNIT], [PRINT], [1d/10d].
- [POWER] switches between operation and standby; holding for about 3 seconds switches the key notification buzzer on/off.
- [CAL] enters calibration or menu item selection.
- [O/T] tares the balance (displays zero).
- [UNIT] switches units of measurement.
- [PRINT] outputs displayed values to external devices; holding for about 3 seconds outputs date and time (not with AUY).
- [1d/10d] on AUW/AUX/AUY switches display between 1d/10d; on AUW-D switches the weighing range.
- Display symbols include stability mark, tare symbol, zero symbol, weight symbol, bracket, number symbol, menu symbol, add-on symbol, memory symbol, communication symbol, battery symbol, Auto Print symbol, standby mark, and inverse triangle.
Provides capacity, readability, accuracy class, verification scale interval, tare range, calibration weight type, and feature availability for each model series. It distinguishes dual-range AUW-D models from single-range AUW, AUX, and AUY models.
- AUW-D series (AUW220D, AUW120D) are dual-range with capacities 220g/82g and 120g/42g respectively.
- AUW series capacities: AUW320 320g, AUW220 220g, AUW120 120g.
- AUX series capacities: AUX320 320g, AUX220 220g, AUX120 120g.
- AUY series capacities: AUY220 220g, AUY120 120g.
- Accuracy class I for all models.
- Calibration weight: built-in for AUW-D/AUW/AUX series; none for AUY series.
- Number of verification scale intervals: AUW220D 110000, AUW120D 60000, AUW320 160000, AUW220 110000, AUW120 60000, AUX320 160000, AUX220 110000, AUX120 60000.
- Tare range by subtraction equals capacity for each model.
Covers site selection, unpacking and delivery inspection, mechanical installation, leveling, and initial power-on. It specifies power requirements, environmental conditions to avoid, standard packed items, and the automatic self-check and calibration sequence after power-on.
- Install near a power source; verify power voltage conforms to that indicated on the AC adapter.
- Avoid sites with air currents, vibration, direct sunlight, extreme temperature changes, high/low humidity, flammable or corrosive gases, dust, electromagnetic waves, or magnetic fields.
- Install on a sturdy and level tabletop; stone is recommended; edges and corners are generally better than the middle of the room for vibration-free measurement.
- Standard packed items: balance main body, pan, pan supporter, anti-draft ring, AC adapter, adapter cable holder, instruction manual, explanatory operation sheet, inspection certificate, in-use protective cover.
- Level the balance using the two front level screws until the bubble is in the middle of the red circle.
- After power-on, the balance performs a self check and automatically executes calibration (except AUY series); display sequence includes CHE 5, CHE 4, ..., CHE 0, version number, CAL 2, CAL 1, CAL 0, CALEnd, oFF.
- Span calibration immediately after power-up can be aborted by pressing [POWER], but at least one span calibration is necessary before use.
Describes warm-up requirements and span calibration procedures after installation. It covers fully-automatic span calibration by PSC for AUW-D/AUW/AUX series and external-weight span calibration for AUY series.
- Leave power on at standby (warm up) for at least one hour before calibration; for AUW-D series in small range (minimum display 0.01mg), leave power on for at least four hours.
- Perform calibration without people entering/leaving the room and without air flow or vibration.
- For AUW-D/AUW/AUX series, PSC fully-automatic span calibration is ON by default; weight symbol blinks about two minutes before operation.
- For AUY series, span calibration with external weights is the default routine calibration method.
- For accurate measurement, perform span calibration again when the balance is moved or when the installation site temperature changes.
- It is recommended that either PSC or Clock-CAL be set to ON so necessary calibrations operate automatically.
Covers weighing procedure, changing unit display, switching weighing range on AUW-D series, and stable measurement in the semi-micro range. It includes taring, reading stable values, and notes on temperature equalization.
- Warm up thoroughly (at least one hour) and calibrate before use; for AUW-D series small range (0.01mg), warm up at least four hours.
- Place container on pan, close glass door, wait for stability mark, press [O/T] to tare.
- Place item to be weighed, close glass door, wait for stability, read display.
- Keep glass doors closed except when placing or removing items or calibration weights.
- Equalize temperatures by leaving the item in the chamber before weighing to avoid air convection errors.
- Pressing [UNIT] switches between registered units, piece counting, and specific gravity measurement modes; default registered units are g, pcs, %, ct.
- AUW-D series starts in small range (minimum display 0.01mg); press [1d/10d] to switch to large range (0.1mg).
Explains the WindowsDirect function for transferring measurement results to a personal computer without installing software. It covers settings on the balance and computer, RS-232C cable connection, and troubleshooting for communication problems.
- WindowsDirect allows data transfer to a PC without communication software; the balance acts as a primary device like a keyboard.
- Settings include balance configuration, RS-232C cable connection, and computer setup.
- Troubleshooting covers no operation, issues after PC restart, Windows Vista compatibility, and garbled characters.
- If using a USB serial converter, reassign to COM1 to COM4 if automatically set higher.
- Some notebook PCs disable RS-232C ports for power saving; enable them before use.
- LAN communications may interfere with serial key device settings; try without LAN.
Describes the menu system, menu map, selection procedures, and useful related functions such as settings check display, menu reset, and menu lock.
- Menu selection is used to configure balance functions.
- Settings check display allows reviewing current settings.
- Menu reset returns settings to default.
- Menu lock prevents menu changes; when locked, 'LoCKEd' is displayed briefly if menu entry is attempted.
Covers setting the date, date output style, and time for the built-in clock. This function is available on AUW-D, AUW, and AUX series only.
- Date setting.
- Date output style selection.
- Time setting.
Covers bar graph display, changing the minimum display, and turning the backlight on and off. Minimum display change applies to AUW/AUX/AUY series; backlight applies to AUW series only (not AUW-D).
- Bar graph display can be enabled.
- Minimum display can be changed on AUW/AUX/AUY series only.
- Backlight can be turned on/off on AUW series only (not AUW-D); default is ON.
Details calibration necessity, types, execution with built-in or external weights, calibration check, and calibration settings including PSC, Clock-CAL, PCAL, and external weight value input. Also covers GLP/GMP/ISO compliance settings.
- Types of calibration include span calibration with built-in weight (AUW-D/AUW/AUX), span calibration with external weights, calibration check with built-in weight, and calibration check with external weights.
- PSC fully-automatic span calibration available on AUW-D/AUW/AUX series.
- Clock-CAL fully-automatic span calibration available on AUW-D/AUW series.
- PCAL calibrates the built-in weight (AUW-D/AUW/AUX series only).
- External calibration weight value can be input for E-CAL and PCAL.
- GLP/GMP/ISO compliance settings include calibration report, balance ID, and date printout.
Covers settings for stability and response, stability detection band, zero tracking, and stability mark lighting timing. It includes standard mode, anti-convection mode, high-stability mode, and pouring mode.
- Standard mode for normal conditions.
- Anti-convection mode for drafty environments.
- High-stability mode for unstable readings.
- Pouring mode for fast response.
- Stability detection band can be adjusted.
- Zero tracking can be configured.
- Stability mark lighting timing can be set.
Covers setting units of measurement and percentage conversion. It explains how to register and switch units.
- Units of measurement can be set.
- Percentage (%) conversion function available.
Describes application functions including piece counting, solid specific gravity measurement, liquid density measurement, auto print, interval timer, add-on mode, and formulation mode.
- Piece counting (PCS).
- Solid specific gravity measurement.
- Liquid density measurement.
- Auto Print function.
- Interval Timer (AUW-D/AUW/AUX series only).
- Add-on mode.
- Formulation mode.
Covers connection to electronic printer EP-80 and personal computer via RS-232C, including cable connection, data format, command codes, communication settings, and decimal point symbol in output data.
- Electronic printer EP-80 connection.
- RS-232C connection to personal computer.
- Data format details.
- Command codes for programming.
- Communication settings include standard and user settings.
- Decimal point symbol in output data can be set.
Provides maintenance procedures and transport instructions. It covers cleaning, care, and safe moving of the balance.
- Maintenance procedures for the balance.
- Transport instructions including removing pan and pan supporter, grasping firmly with both hands.
Lists error code displays and their countermeasures, and provides a troubleshooting table for symptoms before, during, and after measurement, during calibration, and during menu settings.
- Error codes: CAL E2, CAL E3, CAL E4, CHE X, Err 0X, Err 20, Err 24.
- Troubleshooting table covers display issues, fluctuation, stability mark, repeatability, drift, 'CAL d', 'oL'/' -oL', automatic span calibration frequency, incorrect mass display, unit selection, communication failure, error messages, and menu lock.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Warm-up time before calibration: at least one hour; for AUW-D series in small range (minimum display 0.01mg), at least four hours. p.25 ↗ | Ensure power is left on at standby for the required warm-up period before performing span calibration or critical measurements. | AUW-D/AUW/AUX/AUY series (four hours specifically for AUW-D small range) |
| PSC fully-automatic span calibration is ON by default for AUW-D/AUW/AUX series; weight symbol blinks about two minutes before operation. p.25 ↗ | Allow the balance to perform automatic calibration; avoid vibrations and air flow during PSC operation. | AUW-D/AUW/AUX series |
| AUY series does not perform automatic calibration on power-up; span calibration with external weights is the default routine calibration method. p.24 ↗ | For AUY series, perform span calibration with external weights as described in section 4.5. | AUY series |
| AUW-D series dual-range: small range minimum display 0.01mg, large range minimum display 0.1mg; small range capacities 82g (AUW220D) and 42g (AUW120D). p.29 ↗ | Use [1d/10d] key to switch ranges; measurement exceeding small range capacity automatically switches to large range. | AUW-D series (AUW220D, AUW120D) |
| Default registered units are g, pcs, %, ct; other units must be registered beforehand. p.29 ↗ | If a desired unit cannot be called by [UNIT] key, register it in the unit settings (section 12.1). | All series |
| For a verified balance in the EU, zero-setting function operates within ±0.25e (e = verification scale interval). p.16 ↗ | Understand the zero symbol indication and zero-setting tolerance for legal metrology use. | Verified balances in EU |
| Taring occurs at a weight exceeding 2.0% of capacity; zero-setting occurs within 2.0% of capacity for a verified balance in the EU. p.15 ↗ | Be aware of the distinction between tare and zero-setting when using a verified balance. | Verified balances in EU |
| Level the balance using two front level screws until the bubble is in the middle of the red circle. p.22 ↗ | Proper leveling is required for accurate measurement. | All series |
| Install on a sturdy and level tabletop; stone is recommended; edges and corners are generally better than the middle of the room for vibration-free measurement. p.20 ↗ | Choose an installation site away from vibration sources and air currents. | All series |
| If using a USB serial converter for WindowsDirect, reassign to COM1 to COM4 if automatically set higher. p.37 ↗ | Ensure the COM port number is within the range usable by serial key devices. | WindowsDirect function |
| Some notebook PCs disable RS-232C ports as a power-saving measure; enable RS-232C ports before using WindowsDirect. p.37 ↗ | Check PC power-saving settings if WindowsDirect communication does not operate. | WindowsDirect function |
| LAN communications may interfere with serial key device settings; try using WindowsDirect without LAN. p.37 ↗ | If WindowsDirect communication fails, disconnect LAN or disable LAN to test. | WindowsDirect function |
| For AUW series, backlight default setting is ON; use of backlight requires warm up in mass display such as 'g', not in standby state. p.25 ↗ | When using backlight, ensure the balance is in mass display during warm-up. | AUW series |
| After power-on, the balance performs a self check and automatically executes calibration (except AUY series); display sequence includes CHE 5, CHE 4, ..., CHE 0, version number, CAL 2, CAL 1, CAL 0, CALEnd, oFF. p.24 ↗ | Allow the automatic calibration to complete before use; it can be aborted by pressing [POWER], but at least one span calibration is necessary before use. | AUW-D/AUW/AUX series |
| For accurate measurement, perform span calibration again when the balance is moved or when the installation site temperature changes. p.25 ↗ | Recalibrate after relocation or significant temperature variation. | All series |
| It is recommended that either PSC or Clock-CAL be set to ON so necessary calibrations operate automatically. p.25 ↗ | Enable automatic calibration functions to maintain accuracy. | AUW-D/AUW/AUX series (Clock-CAL for AUW-D/AUW) |
| For AUY series span calibration with external weights: press [CAL] once, 'E-CAL' displayed; press [O/T]; zero display blinks; stability confirmed after about 30 seconds; load weights as indicated; remove weights; 'CAL End' appears. p.26 ↗ | Follow the external weight calibration procedure for AUY series. | AUY series |
| The balance must be used within the temperature range indicated on the verification label when used as a verified balance in the EU. p.20 ↗ | Check the verification label for the allowed temperature range. | Verified balances in EU |
| Menu lock prevents menu entry; 'LoCKEd' is displayed briefly if menu entry is attempted. p.106 ↗ | Remove the menu lock (section 7.4.3) to change menu settings. | All series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| CAL E2 p.105 ↗ | The zero point shift is large during calibration. | Remove items from the pan. In order to postpone calibration, press [POWER] key. | 16.1 Error Code Displays |
| CAL E3 p.105 ↗ | Large span error in PCAL | Use the correct calibration weight. | 16.1 Error Code Displays |
| CAL E4 p.105 ↗ | Large span error in span calibration | Use the correct calibration weight. | 16.1 Error Code Displays |
| CHE X (X is a numeral) (when the display stops here) p.105 ↗ | Internal malfunction | Please contact a service representative. | 16.1 Error Code Displays |
| Err 0X (X is a numeral) p.105 ↗ | Internal malfunction | Please contact a service representative. | 16.1 Error Code Displays |
| Err 20 p.105 ↗ | An improper value setting was attempted. | Enter the correct numbers or decimal points. | 16.1 Error Code Displays |
| Err 24 p.105 ↗ | Power voltage error | Check the power voltage. | 16.1 Error Code Displays |
| Nothing appears in the display. p.106 ↗ | The AC adapter is not securely connected. The power switchboard of the room is turned off. The power voltage is incorrect. | Check power supply and connect correctly. | 16.2 Troubleshooting (Before measurement) |
| The display fluctuates. The stability mark does not readily appear. The measured results have poor repeatability. The display diverges in the same direction. “C p.106 ↗ | Vibration or wind effects | Change the installation site. Change the stability and response settings or change the stability detection band. | 16.2 Troubleshooting (During measurement) |
| The display fluctuates. The stability mark does not readily appear. The measured results have poor repeatability. The display diverges in the same direction. “C p.106 ↗ | Attempting to measure volatile substances | Measure with a lid on it. | 16.2 Troubleshooting (During measurement) |
| The display fluctuates. The stability mark does not readily appear. The measured results have poor repeatability. The display diverges in the same direction. “C p.106 ↗ | The weighed item is electrically charged. | Measure in a metal container. Measure with a metal object larger than the items. | 16.2 Troubleshooting (During measurement) |
| The display fluctuates. The stability mark does not readily appear. The measured results have poor repeatability. The display diverges in the same direction. “C p.106 ↗ | The sample temperature and the temperature inside the weighing chamber differ. | Measure at the same temperature. Leave the item in the chamber before measurement. Change to high-stability mode. | 16.2 Troubleshooting (During measurement) |
| The display fluctuates. The stability mark does not readily appear. The measured results have poor repeatability. The display diverges in the same direction. “C p.106 ↗ | There are air currents entering the weighing chamber. | Leave the glass doors of the weighing chamber open 1 to 2 cm wide when not in use. | 16.2 Troubleshooting (During measurement) |
| The display fluctuates. The stability mark does not readily appear. The measured results have poor repeatability. The display diverges in the same direction. “C p.106 ↗ | Effects of electronic noise or strong electromagnetic waves | Move away from the noise source. | 16.2 Troubleshooting (During measurement) |
| The display fluctuates. The stability mark does not readily appear. The measured results have poor repeatability. The display diverges in the same direction. “C p.106 ↗ | Internal trouble with the balance | Contact a service representative. | 16.2 Troubleshooting (During measurement) |
| “oL” or “-oL” is displayed. p.106 ↗ | The load on the pan is too large. The pan is detached. | Use within the weighing capacity. Place the pan on correctly. | 16.2 Troubleshooting (During measurement) |
| Automatic span calibration executes frequently. p.106 ↗ | Severe temperature variations in the room or the instrument | Move to a location with less temperature fluctuation. | 16.2 Troubleshooting (During measurement) |
| The mass display is incorrect. p.106 ↗ | Span calibration has not been done. | Perform correct span calibration. | 16.2 Troubleshooting (During measurement) |
| The mass display is incorrect. p.106 ↗ | No taring to zero before weighing. | Press the [O/T] key to return mass to zero before weighing. | 16.2 Troubleshooting (During measurement) |
| The desired weighing unit cannot be called by [UNIT] key. p.106 ↗ | The unit has not been registered beforehand. | Set unit beforehand. (See 12.1) | 16.2 Troubleshooting (During measurement) |
| Cannot transmit or receive data to or from computer or device. p.106 ↗ | Communication settings are wrong. | Make the proper communication settings. | 16.2 Troubleshooting (During measurement) |
| Error message appears. p.106 ↗ | Refer to the error code table. | Refer to the error code table. | 16.2 Troubleshooting (During measurement and During calibration) |
| The menu cannot be entered. “LoCKEd” is displayed briefly. p.106 ↗ | The menu is locked | Remove the menu lock. (See 7.4.3) | 16.2 Troubleshooting (Attempt of menu settings) |