347-05351
Technical Reference Summary
Covered Models & Document Scope
Covers the Shimadzu DUH series Dynamic Ultra-micro Hardness Tester models DUH-211 (344-04250-58), DUH-211S (344-04251-58), DUH-210 (344-04256-58) and DUH-210S (344-04257-58). DUH-210/DUH-210S additionally support ISO/TS 19278 evaluation for plastics. Optional Vickers and Knoop indenters, x10 objective lens, and DUH-Series/MCT-Series software (installation manual 347-05357) are referenced.
Section-by-Section Technical Breakdown
Specifies the environmental and electrical conditions required for installing the DUH series tester. Room temperature must be 10-35 °C with minimal daily variation, humidity within 80 %, no condensation, no vibration, no direct sunlight, and no corrosive/flammable gases. Electrical requirements include matching the labeled supply voltage, grounding, and mains fluctuation not exceeding 10 % of nominal.
- Room temperature 10-35 °C with minimal variation during a day
- Humidity within 80 %, no condensation
- No vibration, no direct sunlight, no air currents from HVAC directed at instrument
- Installation category II, pollution degree 2, altitude up to 2000 m, indoor
- Mains supply voltage fluctuations not to exceed 10 % of nominal supply voltage
- Ground the instrument; ground PC and optional units with equal potential grounds
- Power switch must be easily accessible for emergency shutdown
Describes the mechanical and electrical installation of the hardness tester, including cable connections between the main body, control unit and PC. Cables include lamp cable, control cable, A/D cable, I/O cable and ground cable. Adjustment steps follow installation.
- Lamp cable (main body-control unit) P/N 344-86719
- Control cable (main body-control unit) P/N 070-01623-72
- A/D cable (control unit-PC) P/N 347-24224
- I/O cable (control unit-PC) P/N 088-50664-56
- Ground cable 1.5 m P/N 344-81775-01
- Power cable 110 V or 230 V P/N 347-20836-01 / 347-20836
- Adjustment procedure follows installation (Section 3.3)
Describes the external view, X-Y sample stage, sample attachment, objective lenses (x10 optional), optical head, triangular pyramid indenter with 115° tip angle, rear panel, control unit, optical system and displacement measuring system.
- Indenter: triangular pyramid with 115° tip angle
- Objective lens x10 is optional
- Sample stage is an X-Y stage
- Displacement measuring system integrated in the loading unit
- Control unit contains main switch and power inlet
Explains dynamic hardness determination from indentation load-depth data, evaluation of plastic and elastic deformation, indentation modulus and work per ISO 14577-1 Annex A, and ISO/TS 19278 for plastics on DUH-210/210S. Also covers Vickers and Knoop hardness by diagonal length measurement with optional indenters, and precautions per ISO 14577-1 and JIS Z2244.
- Dynamic hardness evaluated from both plastic and elastic deformation during indentation
- Load-unload test yields indentation modulus and plastic/elastic indentation work
- ISO 14577-1 Annex A material parameters supported
- ISO/TS 19278 plastics evaluation on DUH-210 and DUH-210S
- Vickers and Knoop hardness require optional indenters and diagonal length reading
- Precautions per ISO 14577-1 and JIS Z2244 included
Lists technical specifications of the loading unit, displacement measuring system, indenter, optical monitor, optical head, sample stage, PC requirements, test modes, data processing, load range, hardness calculation formulae, loading speed, weight and utilities.
- Loading unit and displacement measuring specifications
- Indenter: triangular pyramid 115° tip angle
- Optical monitor and optical head specifications
- Sample stage (X-Y) specifications
- Personal computer requirements
- Test modes, data processing, load range
- Dynamic hardness, Martens hardness and hardness calculation formulae
- Loading speed, weight and utilities
Covers software configuration, starting/stopping the hardness tester, main menu, test execution, analysis functions and Cf-Ap,As correction. Test operation includes stage movement, indenter down, loading and depth monitoring.
- Software configuration and startup/shutdown
- Main menu and test execution
- Analysis functions
- Cf-Ap,As correction
- Test start requires stage moved under indenter
- Loading now indication after depth exceeds 20 µm
Describes daily inspection and periodic inspection/calibration to maintain instrument performance and accurate measurement data. Includes lamp replacement, indentation centering, indenter height adjustment, depth detector adjustment, magnification adjustment, identical focal point adjustment and micrometer adjustment.
- Daily inspection and periodic inspection/calibration required
- Periodic inspection/calibration requested from Shimadzu representative
- Lamp replacement procedure (Clause 9.2.1)
- Indentation centering (Clause 9.2.4)
- Indentation depth detector adjustment (Clause 9.2.5)
- Indenter height adjustment (Clause 9.2.6)
- Identical focal point adjustment (Clause 9.2.8)
- Magnification adjustment (Clause 9.2.9)
Provides a phenomenon table with reference clauses and detailed repair procedures for abnormal indentation shape, off-center indentation, improper indenter height, illumination lamp failure, stage position errors, depth display faults, immediate test termination, hardness value mismatch, abnormal length reading, focal point differences, indentation position differences, smoke and abnormal smell.
- Phenomenon table maps symptoms to clauses 10.2.1-10.2.14
- Abnormal indentation shape indicates indenter mounting defect or unlevel sample stage
- Off-center indentation corrected by Clause 9.2.4 adjustment
- Lamp failure causes: lamp not turned on, cable disconnected, lamp burned out
- Depth value not changing after Indenter down: depth detector adjustment or indenter contact
- Loading now not displayed above 20 µm: indenter height improper
- Test ends immediately: depth detector positioned too high
- Hardness mismatch: small test force, vibration, or length measurement system failure
- Smoke or abnormal smell: turn OFF power immediately and contact Shimadzu
Lists maintenance parts and spare parts including fuses for the control unit. Accessories include power cable, lamp cable, control cable, A/D cable, I/O cable, ground cable, AC plug adaptor, tool set, instruction manual and installation CD-ROMs.
- Fuse for control unit (2A) P/N 347-20424-01, quantity 2
- Power cable 110 V P/N 347-20836-01 or 230 V P/N 347-20836
- Lamp cable P/N 344-86719
- Control cable P/N 070-01623-72
- A/D cable P/N 347-24224
- I/O cable P/N 088-50664-56
- Ground cable 1.5 m P/N 344-81775-01
- AC plug adaptor P/N 071-60813-01
- DUH soft installation CD P/N 347-29119
- Machine configuration disk P/N 347-29138 (contains calibration information)
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Room temperature must be maintained between 10 and 35 °C with minimal temperature variation during a day. p.5 ↗ | Install in a temperature-controlled room; daily swings affect measurement stability. | whole series |
| Humidity must be maintained within 80 % with no condensation. p.5 ↗ | Avoid humid environments; condensation may prevent normal operation. | whole series |
| Mains supply voltage fluctuations are not to exceed 10 % of the nominal supply voltage. p.6 ↗ | Verify supply stability and current capacity sufficient for all system components. | whole series |
| Installation environment: IEC installation category II, pollution degree 2, altitude up to 2000 m, indoor. p.5 ↗ | Site must conform to these overvoltage and pollution ratings. | whole series |
| When turning off the power and turning it on again, wait at least 5 seconds before turning it on. p.7 ↗ | Prevents equipment failure from rapid power cycling. | whole series |
| Indenter is a triangular pyramid with 115° tip angle. p.23 ↗ | Vickers and Knoop hardness require optional Vickers or Knoop indenters. | whole series |
| DUH-210 and DUH-210S can evaluate plastics per ISO/TS 19278. p.26 ↗ | Select DUH-210/210S for instrumented micro-indentation testing of plastics/resins. | DUH-210, DUH-210S |
| ISO 14577-1 Annex A material parameters (indentation modulus, plastic and elastic indentation work) are evaluated via load-unload test. p.26 ↗ | Use load-unload test mode for instrumented indentation hardness and modulus. | whole series |
| When test force is 146 mN or less, vibration strongly affects measurement results. p.222 ↗ | Ensure no vibration source is present around the tester before low-load measurement. | whole series |
| When diagonal length is smaller than 20 µm, maximum permissible error corresponds to 1 µm diagonal length (JIS B 7725 Clause 5.3). p.222 ↗ | Small indentations have larger relative length-reading error affecting hardness value. | whole series |
| Reference hardness of the optional Vickers reference sample is the hardness value obtained under 1961 mN test force. p.222 ↗ | Hardness obtained under small test force differs from that under large test force. | whole series |
| Fuse for the control unit is 2 A, P/N 347-20424-01, quantity 2. p.29 ↗ | Use specified fuse rating and part number for replacement. | whole series |
| Ground cable is 1.5 m, P/N 344-81775-01. p.29 ↗ | Ground the instrument and equalize potential grounds of PC and optional units. | whole series |
| Machine configuration disk (CD-ROM) P/N 347-29138 contains calibration information on an individual device. p.29 ↗ | Retain the machine configuration disk for calibration data reference. | whole series |
| Loading now is expected to be displayed after depth value exceeds 20 µm. p.221 ↗ | If not displayed above 20 µm, indenter height is improper; adjust per Clause 9.2.6. | whole series |
| Replacement parts are available for seven years after product discontinuation. p.20 ↗ | Plan long-term maintenance and spare-part stocking within this period. | whole series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Indentation shape is abnormal p.219 ↗ | Indenter mounting part may be defective; or sample surface is not horizontally mounted. | Replace defective part via Shimadzu; make sure sample stage is leveled, then indent a reference sample and check indentation shape. | 10.2.1 |
| Indentation is not at the center of the visual field p.219 ↗ | Indentation position off-center. | Perform adjustment procedure described in Clause 9.2.4. | 10.2.2 |
| Indenter height is improper p.220 ↗ | Indenter touches the sample or does not reach the sample surface when focusing and moving stage to start test. | Refer to Clause 9.2.6. | 10.2.3 |
| Illumination lamp does not turn ON p.220 ↗ | Lamp not turned on; cable not connected between hardness tester and measurement control unit; lamp burned out. | Select [Tool]-[Lamp] menu; verify cable connections per Clause 3.2.5; open top cover and replace lamp per Clause 9.2.1 if burned out. | 10.2.4 |
| “Slide sample stage to load side” is displayed at the start of test p.220 ↗ | Stage has not been moved under the indenter; or switch failure. | Move stage under indenter before starting test; if switch failure, contact Shimadzu for repair. | 10.2.5 |
| Displayed test force value increases but depth value is not displayed after “Indenter down” is displayed at the start of test p.221 ↗ | Indentation depth detector adjustment improper; or indenter is in contact with the sample. | Click <Stop test (F12)>, lower stage, click <Start test (F1)> again. If no change, gently lower indenter fixing attachment with finger momentarily. If depth value appears, adjust detector per Clause 9.2.5. If depth value appears during test without sample, adjust indenter height per Clause 9.2.6. | 10.2.6 |
| Displayed test force and depth values are changed but “Loading now” is not displayed even if the depth value exceeds 20 µm after “Indenter down” is displayed at p.221 ↗ | Indenter height is improper; focal point on sample positioned too low. | Adjust indenter height according to Clause 9.2.6. | 10.2.7 |
| Test ends immediately after “Indenter down” is displayed at the start of test p.221 ↗ | Indentation depth detector adjustment is improper; detector positioned too high. | Move indentation depth detector downward according to Clause 9.2.5. | 10.2.8 |
| Hardness value obtained from indentation length reading result does not conform with the reference hardness sample’s reference value p.222 ↗ | Test force too small; influence of vibration; failure of length measurement system. | Use adequate test force; ensure no vibration source when test force is 146 mN or less; refer to Clause 10.2.14 for length measurement system failure. | 10.2.9 |
| Indentation length leading result is abnormal p.223 ↗ | Magnification adjustment improper; initial condition settings improper; micrometer adjustment improper. | Perform magnification adjustment per Clause 9.2.9; turn power ON again and verify lens magnification setting; if unresolved, micrometer must be repaired by Shimadzu. | 10.2.10 |
| Focal point of the optional objective lens is different from that of the x 50 objective lens p.223 ↗ | Focal point mismatch between optional objective lens and x50 objective lens. | Perform identical focal point adjustment described in Clause 9.2.8. | 10.2.11 |
| Indentation position observed with the optional objective lens is different from that for the x 50 objective lens p.223 ↗ | Indentation centering mismatch. | Perform indentation centering procedure per Clause 9.2.4; if beyond adjustable range, contact Shimadzu. | 10.2.12 |
| Hardness tester emits smoke p.223 ↗ | Equipment fault. | Turn OFF the power supply immediately and contact Shimadzu or authorized dealer/service company. | 10.2.13 |
| Hardness tester emits abnormal smell p.223 ↗ | Equipment fault. | Turn OFF the power supply immediately and contact Shimadzu or authorized dealer/service company. | 10.2.14 |