c142-e024
Technical Reference Summary
Covered Models & Document Scope
Shimadzu MXF-2400 Multi-Channel X-ray Fluorescence Spectrometer; fixed monochromator for simultaneous determination of up to 36 elements, optional scanning monochromator for sequential determination of up to 48 elements; analysis range 4Be, 5B, 6C to 92U; 4 kW thin-window X-ray tube.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Number of elements determined simultaneously | up to 36 elements p.2 ↗ | fixed monochromator |
| Number of elements determined sequentially | up to 48 elements p.2 ↗ | optional scanning monochromator |
| Analysis range | 4Be, 5B, 6C to 92U p.2 ↗ | |
| X-ray tube power | 4 kW p.2 ↗ | Thin Window |
| Analysis time | two minutes p.3 ↗ | about 36 elements |
| Detection limit | a few ppm p.2 ↗ | high sensitivity analysis |
| Power requirements | 3-phase 200 V ±10 %, 50/60 Hz, 5 A p.19 ↗ | T-100 Disk Type Vibration Mill |
| Dimensions | 435 mm dia. × 558 mm high mm p.19 ↗ | T-100 Disk Type Vibration Mill |
| Weight | 120 kg p.19 ↗ | T-100 Disk Type Vibration Mill |
| Power requirements | Single-phase 100 V ±10 %, 50/60 Hz, 2 A p.19 ↗ | TI-100 Vibration Mill |
| Inner volume | 10 mL × 2 mL p.19 ↗ | TI-100 Vibration Mill |
| Dimensions | W580 × D620 × H400 mm p.19 ↗ | TI-100 Vibration Mill |
| Weight | 70 kg p.19 ↗ | TI-100 Vibration Mill |
Operating Limits
- Analysis range: 4Be, 5B, 6C to 92U p.2 ↗
- Maximum simultaneous elements: 36 elements p.2 ↗
- Maximum sequential elements: 48 elements p.2 ↗
- X-ray tube power: 4 kW p.2 ↗