Operation Guide

Technical Reference Summary

Covered Models & Document Scope

Covers the GCMS-QP NX series and GCMS-TQ NX series gas chromatograph mass spectrometers, operated with GCMSsolution Ver. 4.5 or later. Functions specific to the TQ NX series (e.g., CID gas, MRM) are marked and not shown for QP NX models. This Basic Operation Guide covers routine analysis, qualitative and quantitative analysis, shutdown, and maintenance; method development with Smart database is in a separate Method Development Guide.

Section-by-Section Technical Breakdown

pp. 5
Turning ON the Power

Describes the power-on sequence for the GCMS system. Peripheral/accessory equipment and sample pretreatment units must be switched ON before the main GCMS system. For TQ NX series, the CID gas (argon) supply valve must be opened for MRM measurements.

  • Switch ON peripheral/accessory equipment before the main GCMS system.
  • Switch ON sample pretreatment units before the GCMS system.
  • TQ NX series: open CID gas (argon) supply valve for MRM mode.
  • Power ON order: GC, then MS, then PC/printer/display.
  • Start GCMS Real Time Analysis program by double-clicking its icon.
pp. 8-11
System Configuration

Covers checking and setting the modules used for analysis in the [System Configuration] window, including registering modules such as the AOC-30/20i autosampler. For TQ NX series, CID gas usage can be disabled for Q3 Scan/Q3 SIM-only analysis. Auto-tuning result judgment criteria can be configured for EI ion source.

  • Register modules in [Modules Used for Analysis] via [System Configuration] icon.
  • TQ NX: deselect [Use CID Gas] to analyze in Q3 Scan or Q3 SIM only.
  • EI ion source: enable [Check Auto Tuning Result] and set criteria.
  • Auto-tuning criteria defaults: FWHM 0.1, Detector Gain 2, Vacuum Leak (69/28) 2, Maximum Mass Shift 0.1, Relative Intensity Ratio for High m/z (502) 2.
  • Vacuum leak criterion reference: pyrolyzer/OPTIC-4 intensity ratio 1; other pretreatment units intensity ratio 2.
  • Advanced flow technology systems (detector splitting, backflush) may show higher m/z 28; [Vacuum Leak (Intensity Ratio of 69/28)] check not recommended.
pp. 12-14
Vacuum System Startup

Describes opening the carrier gas supply and starting the vacuum system via [Vacuum Control]. Options include vacuum restart mode, automatic vacuum leak check, and stabilization waiting. The leak check compares m/z 28 (nitrogen) and m/z 18 (water) intensities after automatic start.

  • Open carrier gas cylinder valve before starting vacuum system.
  • Vacuum Restart Mode: automatic restart after power recovery; does not start if vacuum system already stopped before power outage.
  • Vacuum Leak Check guideline: AOC-30i/20i, AOC-6000/5000 intensity ratio 2; other accessory/peripheral equipment intensity ratio 5.
  • Leak check available only when ion source is EI and carrier gas is He.
  • Stabilization waiting: about 2 hours for qualitative analysis, 4 hours for quantitative analysis.
  • If leak check fails, search for air leak location.
pp. 14-15
Checking for Vacuum Leakage Manually

Manual leak check procedure using the [Peak Monitor] window after waiting 10 minutes from vacuum system startup. The m/z 28 (nitrogen) peak height must not exceed twice the m/z 18 (water) peak height.

  • Wait 10 minutes after starting vacuum system.
  • Select [Water, Air] in [Monitor Group].
  • Turn ON filament and adjust detector voltage so m/z 18 peak is half the display window height.
  • Acceptable: m/z 28 peak height not more than twice m/z 18 peak height.
  • If m/z 28 > 2 × m/z 18, possible air leak; search for leak location.
pp. 16-21
Autotuning

Covers setting analysis conditions, executing autotuning, and checking results. Autotuning should be performed about 2 hours (qualitative) or 4 hours (quantitative) after vacuum system startup, and periodically even with the vacuum system operating. Calibration curves must be recreated after autotuning.

  • Wait approximately 2 hours (qualitative) or 4 hours (quantitative) after vacuum startup before autotuning.
  • Perform autotuning periodically, e.g., once every two weeks.
  • Tuning mode emission currents: High concentration 20 µA, Normal 60 µA, High sensitivity 150 µA.
  • Estimated autotuning time: QP NX about 3 minutes; TQ NX about 5 min (Q3 Scan/Q3 SIM, no CID gas) or about 7 min (MRM or with CID gas).
  • Acceptance: FWHM 0.5 to 0.7; detector voltage not exceeding 2 kV; relative intensity ratio for m/z 502 at least 2%; m/z 69 peak intensity at least twice m/z 28.
  • Irregular results may indicate vacuum leak, poor column connections, or contaminated ion source.
pp. 23-27
Routine Analysis Flow

Describes routine analysis using existing method and batch files: inspecting consumables, performing autotuning, creating a folder, editing a batch file, and starting batch processing. Batch rows can be edited collectively with [Fill Down] or [Fill Series].

  • Inspect consumables and maintenance parts before analysis.
  • Perform autotuning periodically (e.g., once every two weeks); recreate calibration curves after autotuning for quantitative analysis.
  • Use [Fill Down] or [Fill Series] to edit specified batch rows collectively.
  • Start analysis with [Start] on the [Batch] assistant bar.
  • To abort batch processing, click [Stop] on the [Batch] assistant bar.
  • To execute only specified rows, select rows before starting analysis.
pp. 31-36
Creating a Method File (Qualitative Analysis)

Covers setting autosampler, GC, MS, and similarity search parameters for qualitative analysis. Includes injection port temperature, injection mode selection, carrier gas pressure guidance, column oven temperature, interface temperature, scan range, and similarity search library settings.

  • Injection port temperature: 200 °C to 300 °C, considering target component boiling point.
  • Split mode for target component amount 10 ng to 100 ng min.; splitless mode for less than 10 ng.
  • Split ratio: enter -1.0 for splitless mode.
  • Typical carrier gas pressure: middle bore 0.25 mm ID, 30 m: 75–150 kPa; 60 m: 100–250 kPa; semi-wide bore 0.32 mm ID, 30 m: 30–50 kPa; 60 m: 50–100 kPa.
  • Column oven initial temperature: 40 °C to 100 °C.
  • Interface temperature: 200 to 300 °C.
  • Detector voltage adjustment: +0.1 to +0.3 relative to tuning result; increasing by 0.1 kV increases peak intensity by 2 to 4 times.
  • Start Time: set to 0 min initially, then after solvent peak to after solvent elution; End Time = GC Program Time.
  • Filament ON time should be 0.5 minutes less than Start Time (minimum 0).
  • Scan range: end m/z = molecular weight of target compound plus about 100.
pp. 37-39
Sequential Analysis (Qualitative)

Describes creating a batch file with the Batch Table Wizard for qualitative sequential analysis, including unknown-only samples, method file selection, vial number, sample count, injection volume, and data file naming.

  • Use [Batch Table Wizard] and select [Unknown Only].
  • Deselect both [Data Processing] items for qualitative acquisition.
  • If data file name ends with a number, files are named sequentially.
  • Recommended to measure a blank (solvent, etc.) before starting analysis.
pp. 43-46
Analyzing Data (Qualitative)

Covers similarity search, registering target spectra to the spectrum process table, and editing the spectrum process table. Overlapped peaks may lower similarity indices; mass chromatograms can be used to confirm peak purity.

  • Right double-clicking on mass spectra executes a similarity search.
  • Overlapped peaks may cause similarity indices lower than actual values.
  • Register target spectrum to spectrum process table to re-check results or output in reports.
  • Sort spectrum process table chronologically via [Sort Table] on [Edit] menu.
pp. 53-74
Quantitative Analysis

Covers creating a compound table and SIM table, sequential analysis, checking and correcting calibration curves, re-quantifying after correction, checking quantitation results, and printing quantitative reports.

  • Create compound table for quantitative analysis in Scan or SIM mode.
  • Create SIM table when using SIM mode.
  • Check and correct calibration curves before calculating unknown samples.
  • Re-quantify after correcting a calibration curve.
  • Quantitative Browser can be used for multiple samples.
pp. 77-78
Shutting Down GCMS

Describes vacuum system shutdown and turning OFF the power. Proper shutdown procedure must be followed to avoid device failures and data/software damage.

  • Shut down vacuum system before turning OFF power.
  • Follow proper shutdown procedure; improper shutdown is excluded from warranty.
pp. 115-120
Maintenance

Covers maintenance procedures, Easy sTop, and changing replacement guidelines for septa and glass inserts. Consumable status is shown in the instrument monitor; icon background turns black when approaching recommended maximum usage frequency.

  • Hover over consumable icon in instrument monitor to see current state and recommended replacement point.
  • Icon background turns black when consumable approaches recommended maximum usage frequency.
  • Example: septum used 13 times out of maximum 100 times.
  • Replacement frequency may need to be greater than recommended depending on analysis content.
  • Easy sTop maintenance function available.
  • Replacement guidelines for septa and glass inserts can be changed.
pp. 91-92
Ecology Mode (Reducing Carrier Gas Flow During Standby)

Describes reducing carrier gas flow rate during standby using Ecology Mode, set manually or via batch processing.

  • Ecology Mode reduces carrier gas flow during standby.
  • Can be set manually or using batch processing.
  • When ecology mode is set, click [Cancel] before preparing for analysis.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
GCMSsolution Ver. 4.5 or later is required for the GCMS-QP NX and GCMS-TQ NX series. p.1 ↗ Verify software version before commissioning or method transfer. GCMS-QP NX series and GCMS-TQ NX series
Injection port temperature setting range: 200 °C to 300 °C. p.32 ↗ Set according to target component boiling point to ensure vaporization without thermal degradation. whole series
Split mode recommended for target component amount 10 ng to 100 ng min.; splitless mode for less than 10 ng. p.32 ↗ Select injection mode based on expected analyte mass on column. whole series
For splitless mode, enter split ratio -1.0. p.33 ↗ Software convention for splitless injection. whole series
Typical carrier gas pressure: 0.25 mm ID middle bore, 30 m column: 75–150 kPa; 60 m: 100–250 kPa. 0.32 mm ID semi-wide bore, 30 m: 30–50 kPa; 60 m: 50–100 kPa. p.33 ↗ Use as initial pressure setting when no reference method is available; linear velocity is set automatically. whole series
Column oven initial temperature range: 40 °C to 100 °C. p.33 ↗ Set initial oven temperature within this range for method creation. whole series
Interface temperature setting range: 200 to 300 °C. p.34 ↗ Set interface temperature to prevent analyte condensation between GC and MS. whole series
Increasing detector voltage by 0.1 kV increases peak intensity by 2 to 4 times. p.34 ↗ Adjust detector voltage within +0.1 to +0.3 relative to tuning result if peak intensity is too low. whole series
Filament ON time should be set 0.5 minutes less than Start Time (minimum 0). p.34 ↗ Prevents filament from being ON during solvent elution. whole series
Scan range end m/z should be molecular weight of target compound plus about 100. p.34 ↗ Ensures adequate mass range margin for target ions. whole series
Autotuning emission currents: High concentration 20 µA, Normal 60 µA, High sensitivity 150 µA. p.18 ↗ Select tuning mode appropriate to target concentration level for suitable dynamic range. whole series
Estimated autotuning time: QP NX about 3 minutes; TQ NX about 5 min (Q3 Scan/Q3 SIM, no CID gas) or about 7 min (MRM or with CID gas). p.19 ↗ Plan autotuning time in analysis schedules. QP NX and TQ NX series
Autotuning acceptance criteria: FWHM 0.5 to 0.7; detector voltage not exceeding 2 kV; relative intensity ratio for m/z 502 at least 2%; m/z 69 peak intensity at least twice m/z 28. p.21 ↗ Use these limits to judge autotuning results and detect vacuum leak, poor column connection, or contaminated ion source. whole series
Manual vacuum leak check: m/z 28 (nitrogen) peak height must not exceed twice m/z 18 (water) peak height. p.15 ↗ If exceeded, search for air leak location. whole series
Automatic vacuum leak check guideline intensity ratio: AOC-30i/20i and AOC-6000/5000: 2; other accessory/peripheral equipment: 5. p.13 ↗ Set leak check criterion according to connected pretreatment/autosampler equipment. whole series
Stabilization waiting after vacuum system start: about 2 hours for qualitative analysis, 4 hours for quantitative analysis. p.13 ↗ Allow sufficient stabilization time before autotuning and analysis. whole series
Perform autotuning periodically, e.g., once every two weeks, even with vacuum system operating; recreate calibration curves after autotuning. p.23 ↗ Maintain quantitative accuracy by re-running calibration after tuning changes. whole series
Consumable replacement alert: icon background turns black when approaching recommended maximum usage frequency; example septum used 13 times out of maximum 100 times. p.7 ↗ Monitor consumable usage in instrument monitor and replace before exceeding recommended frequency. whole series
High-pressure gas cylinder must not be placed where temperature can exceed 40 °C; verify pressure gauges functional at least once every three months. p.8 ↗ Installation and daily inspection requirement for carrier gas supply. whole series
Do not smoke or use open flames within 5 m of the instrument when using highly combustible gases such as acetylene and hydrogen, or potentially combustible gases such as oxygen and nitrous oxide. p.8 ↗ Site safety requirement for gas cylinder installation. whole series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Vacuum leakage check fails p.14 ↗ Possible air leak Search for the location of the leak, referring to the instruction manual. 2.5 Vacuum System Startup
m/z 28 (nitrogen) peak height more than twice m/z 18 (water) peak height p.15 ↗ Possible air leak Search for the location of the leak; refer to System User's Guide for details on how to check for vacuum leaks. 2.6 Checking for Vacuum Leakage Manually
Autotuning result irregularities (FWHM outside 0.5–0.7, detector voltage exceeding 2 kV, m/z 502 relative intensity below 2%, m/z 69 peak intensity less than tw p.21 ↗ Vacuum leak, poor column connections, or contaminated ion source See Appendix K Maintenance P.115 to implement corrective measures. 2.7.3 Checking Autotuning Results
The hardware configuration for this method is different from the current instrument configuration. The measurement condition in the method file is modified acco p.17 ↗ Method file hardware configuration differs from current instrument configuration Click [OK] and click (Save) on the toolbar. 2.7.1 Setting Analysis Conditions