Shimadzu i-Series High Performance Liquid Chromatograph (HPLC) system specifications leaflet C196-E101C. Covers LC-2070C and LC-2080C systems, each in standard, 3D, UV-model and PDA-model variants, with sample cooler options and conventional/high-speed flow cells; autosampler and column oven features.

Technical Data

Parameter Value Condition
Maximum pressure 44 MPa p.4 ↗ LC-2070C / LC-2080C, 3.0001 to 5 mL/min
System delay volume 650 µL p.4 ↗ LC-2070C
System delay volume 460 µL p.4 ↗ LC-2080C
System delay volume option 460, 1100 µL p.4 ↗ LC-2070C
System delay volume option 650, 1100 µL p.4 ↗ LC-2080C
Sample cooler temperature range 4 to 45 °C p.4 ↗ LC-2080C (included); not included on LC-2070C
Flow cell volume 12 µL p.5 ↗ Conventional cell, 10 mm, TC
Flow cell volume 8 µL p.5 ↗ High-Speed cell, 10 mm, TC
Flow cell pressure rating 12 MPa p.5 ↗ Flow cell
Weight 58 kg p.5 ↗ LC-2070C
Weight 63 kg p.5 ↗ LC-2080C / LC-2080C 3D
Detector wavelength range 200 to 650 nm p.6 ↗ UV detector
Detector wavelength range 200 to 750 nm p.6 ↗ PDA detector
Detector dimensions (W × D × H) 260 × 420 × 210 mm p.6 ↗ UV detector
Detector weight 16 kg p.6 ↗ UV detector
Detector weight 18 kg p.6 ↗ PDA detector
Injection cycle time 14 s p.3 ↗ Autosampler
Maximum sample capacity 1536 samples p.3 ↗ Autosampler, using 384-well MTPs
Carryover < 0.0004 % p.3 ↗ Autosampler
Minimum injection volume 1 µL p.3 ↗ Autosampler, reproducible/linear
Column oven maximum temperature 90 °C p.3 ↗ Forced-air circulation
Column capacity up to six 100 mm columns p.3 ↗ Column oven

Operating Limits

  • Maximum system pressure: 44 MPa (3.0001 to 5 mL/min) p.4 ↗
  • Sample cooler temperature range: 4 to 45 °C p.4 ↗
  • Column oven maximum temperature: 90 °C p.3 ↗
  • Flow cell pressure rating: 12 MPa p.5 ↗
  • Carryover: < 0.0004 % p.3 ↗

Type Code Decoder

Position Token Meaning
Suffix p.4 ↗ 3D 3D variant

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
LC-2070C (UV model, without sample cooler) p.4 ↗ 228-65880-58 UV model, without sample cooler
LC-2070C 3D (PDA model, without sample cooler) p.4 ↗ 228-65881-58 PDA model, without sample cooler
LC-2080C (UV model, with sample cooler) p.4 ↗ 228-65882-58 UV model, with sample cooler
LC-2080C 3D (UV model, with sample cooler) p.4 ↗ 228-65883-58 UV model, with sample cooler
LC-2080C 3D (PDA model, with sample cooler) p.4 ↗ 228-65884-58 PDA model, with sample cooler