MALDImini-1 Instruction Manual

Technical Reference Summary

Covered Models & Document Scope

MALDImini-1 MALDI digital ion trap mass spectrometer (Shimadzu). The manual covers the base instrument, its accessories, optional parts (e.g., Piping Set for Gas Cylinder), sample plates (microscope disposable and SUS), and the MALDImini Server / Console / Maintenance software. No ordering codes or model variants beyond the MALDImini-1 are described.

Section-by-Section Technical Breakdown

pp. 6-10, 144-145
Installation Site and Conditions

Defines the physical, environmental and electrical conditions required to install the MALDImini-1. It specifies bench load, clearances, temperature/humidity windows, power supply ratings and grounding requirements.

  • Bench/floor must support instrument weight of 25 kg plus PC system and table.
  • Minimum clearances: 10 cm rear, 5 cm left/right/top; 70 cm rear/left/right for maintenance; 50 cm table surface to the left for the folding chassis.
  • Room temperature 18 to 26 °C with changes within ±3 °C; humidity 40 to 70 % non-condensing.
  • Indoor location, overvoltage category II, pollution degree 2, elevation ≤ 2000 m.
  • Power supply 100 to 240 V AC (50/60 Hz), 960 VA, voltage fluctuation within ±10 V, dedicated circuit with electrical leakage breaker and earth ground.
pp. 119-129
Connecting the Instrument and Communication Settings

Covers connecting the instrument to power, gas and the controlling PC, and configuring the Ethernet communication settings between the instrument and the MALDImini software.

  • Connect only to the specified power supply; check the nameplate for voltage specification.
  • Ethernet cable connection is required for communication with the instrument.
  • Communication settings are configured in the software (Section 10.1.3).
  • Gas supply uses helium and argon lines with regulator and Swagelok-type connections.
pp. 100-102, 127-129, 145
Gas Used

Describes the gases required for operation (helium and argon), their supply lines, regulator settings and purging procedure.

  • Helium and argon gas supply lines are monitored; low pressure triggers warning WRN??-06-01 / WRN??-06-02.
  • Regulator settings and gas line nut tightness must be checked if pressure is low.
  • Gas line purging procedure is given in Section 8.4.
pp. 14, 21-23, 118
Starting and Stopping the Instrument

Gives the startup, daily startup/shutdown and emergency stop procedures, including the power-off wait time and power outage behavior.

  • After turning power OFF, wait at least 10 seconds before turning it back ON.
  • Emergency shutdown: turn OFF mains switch (rear, above power cable), close gas supply, disconnect power plug.
  • After a power outage the vacuum system restarts automatically when power is restored; long outages require emergency-stop procedure then normal startup and gas purging.
pp. 11-20
Sample Plates and Sample Preparation

Describes the recommended sample plates, their cleaning (basic and chemical), preparation of standard/mixed calibration samples and loading/recording of samples.

  • Use only Shimadzu-recommended sample plates; other plates may damage the instrument.
  • Sample must be dry before opening the door and inserting the plate.
  • Wear powder-free gloves when handling plates and opening the sample door.
  • Cleaning procedures: basic cleaning and chemical cleaning.
pp. 24-39, 142
Basic Operation and Analysis

Covers the analysis workflow: inserting the sample plate, switching on voltages, calibration, MS, MS/MS and MS3 analysis, and switching off/removing the plate.

  • Workflow: set-up, calibration, analysis, settlement.
  • MSn analysis requires appropriate isolation and CID parameters.
  • Wait approximately 30 minutes after applying voltage before measurement for temperature stability; on first switch to [Ready] after mains ON, run an analysis and wait.
pp. 40-93, 130
Advanced Settings and Parameters

Documents the software parameter panels: acquisition settings, raster settings, mass calibration, spectrum processing, display settings and batch acquisition.

  • Acquisition settings include number of measurements and measurement conditions.
  • Mass calibration panel relates time and m/z.
  • Spectrum processing includes baseline subtraction, smoothing and peak detection.
  • Batch acquisition supports batch tables, editing, loading and saving.
pp. 22, 95-108, 147
Maintenance

Lists periodic inspections and maintenance, cleaning, sample door O-ring inspection/replacement, gas line purging, customer QC tests and instrument return/repair procedures.

  • Periodic inspection/calibration should be requested from Shimadzu representative.
  • Customer QC tests: High Mass Test and Mass Resolution Test.
  • Turbomolecular pump overhaul is handled via instrument return.
  • Replacement parts available for 7 years after discontinuation.
pp. 112-118, 140-146
Troubleshooting

Provides problem/cause/corrective-action tables for general and data acquisition problems, error/warning message codes, operating-time notifications, log file checking and power outage handling.

  • Indicator states: pink steady = stage initialization error; pink flashing = vacuum operation; light blue = ready; red flashing = error.
  • Error codes format ERR/WRN<system status>-<cause>-<details>.
  • Log files stored in C:\MALDImini-1\Log, separated by date.
  • Operating-time notifications appear in MALDImini Console.
pp. 15, 142-145
Specifications

Lists performance, hardware, software and installation specifications for the MALDImini-1.

  • Laser: Nd:YLF, 349 nm, 60 µJ @ 1 kHz, pulse width < 5 ns, beam divergence 3.2 ± 0.5 mrad.
  • Class 1 laser product per IEC60825-1:2014 and JIS C 6802 (2018).
  • Power: 100 to 240 V AC, 50/60 Hz, 960 VA.
  • Weight: 25 kg.
pp. 147
Maintenance Parts

Lists consumable and repair parts for the instrument.

  • Consumable parts and repair parts are listed in Sections 12.1 and 12.2.
  • Use only specified parts when replacing.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Instrument weight is 25 kg; bench/floor must support instrument plus PC system and table. p.6 ↗ Select a stable bench rated for the full load to prevent tipping or injury. MALDImini-1
Minimum clearance: 10 cm rear, 5 cm left/right/top; 70 cm rear/left/right for maintenance; 50 cm table surface to the left for folding chassis. p.7 ↗ Plan lab layout to allow airflow and service access. MALDImini-1
Operating environment: 18 to 26 °C with changes within ±3 °C; humidity 40 to 70 % non-condensing; indoor, overvoltage category II, pollution degree 2, elevation ≤ 2000 m. p.8 ↗ HVAC must maintain tight temperature stability for reproducible measurements. MALDImini-1
Power supply: 100 to 240 V AC (50/60 Hz), 960 VA; voltage fluctuation within ±10 V; dedicated circuit with electrical leakage breaker and earth ground. p.9 ↗ Do not share the circuit with other instruments; verify capacity and grounding. MALDImini-1
After turning power OFF, wait at least 10 seconds before turning it back ON. p.11 ↗ Prevents instrument failure from rapid power cycling. MALDImini-1
Laser: Nd:YLF, wavelength 349 nm, pulse energy 60 µJ @ 1 kHz, pulse width < 5 ns FWHM, beam divergence 3.2 ± 0.5 mrad. p.15 ↗ Defines the ionization source characteristics relevant to matrix selection and fluence settings. MALDImini-1
Class 1 laser product per IEC60825-1:2014 and JIS C 6802 (2018); internal laser equivalent to Class 4. p.15 ↗ Covers must not be removed; internal service only by Shimadzu personnel. MALDImini-1
Wait approximately 30 minutes after applying voltage before starting measurement; on first switch to [Ready] after mains ON, run an analysis and wait. p.142 ↗ Temperature stabilization is required for mass accuracy and resolution. MALDImini-1
Use only Shimadzu-recommended sample plates; sample must be dry before opening the door. p.11 ↗ Non-recommended plates or wet samples may damage the instrument or impair vacuum. MALDImini-1
Wear powder-free gloves when opening the sample door and attaching/detaching sample plates. p.11 ↗ Prevents contamination that degrades evacuation performance. MALDImini-1
Emergency shutdown: turn OFF mains switch (rear, above power cable), close gas supply, disconnect power plug. p.14 ↗ Defines the safe shutdown sequence for abnormalities. MALDImini-1
After a power outage the vacuum system restarts automatically when power is restored; long outages require emergency-stop procedure then normal startup and gas purging. p.14 ↗ Plan for unattended restarts and purge gas lines after extended outages. MALDImini-1
Error codes format: ERR or WRN<System status code>-<Cause code>-<Details code>, each a 2-digit number. p.144 ↗ Enables interpretation of error messages for corrective action. MALDImini-1
Log files are stored in C:\MALDImini-1\Log, separated by date, readable with a text editor. p.146 ↗ Useful for diagnosing intermittent faults. MALDImini-1
Replacement parts available for 7 years after product discontinuation. p.22 ↗ Long-term maintenance planning should account for this availability window. MALDImini-1
Customer QC tests include High Mass Test and Mass Resolution Test. p.103 ↗ Periodic performance verification can be performed by the customer. MALDImini-1
Do not subject the instrument to impact or vibration; do not move or tilt while operating. p.11 ↗ Prevents misalignment of the optical system and damage to the turbomolecular pump. MALDImini-1
Do not wipe with alcohol or thinner-based solvent; clean with soft cloth dampened with water or diluted neutral detergent, then dry. p.12 ↗ Avoids rust or discoloration of covers. MALDImini-1

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
The indicator does not light up when the power is turned ON. p.140 ↗ Power cable not properly connected or no 100 to 240 V power at distribution panel. Check that the power cable is properly connected and 100 to 240 V power is being supplied to the distribution panel. 9.1.1 General
Unable to connect to instrument. p.140 ↗ Ethernet cable connection problem. Check the connection of the Ethernet cable. 9.1.1 General
The initialization operation does not end. (indicator lights up in pink) p.140 ↗ Stage initialization error. Contact your Shimadzu representative. 9.1.1 General
The initialization operation does not end. (indicator flashes in pink) p.140 ↗ Vacuum operation being performed; sample door may not be closed; instrument may have been inactive for a long time. Check the gauge to see whether the vacuum is changing. Check that the sample door is closed. If the vacuum value has completely stopped changing, contact your Shimadzu representative. 9.1.1 General
The instrument does not enter [Ready] mode. p.140 ↗ Initialization operation or vacuum operation may not be complete. Check that the indicator lights up in light blue. 9.1.1 General
The gas cylinder becomes depleted quickly. p.140 ↗ Gas leak in the connecting parts. Check the connecting parts. 9.1.1 General
The stage vacuum operation does not end. p.140 ↗ Sample door may not be closed, or seal surface of sample door may be contaminated. Clean the seal surface and O-ring. 9.1.1 General
The indicator flashes in red. p.140 ↗ An error was detected. Turn OFF the power to the instrument and restart it. 9.1.1 General
The camera image is not displayed. p.140 ↗ Not specified. Contact your Shimadzu representative. 9.1.1 General
The ion intensity is unstable or low. p.142 ↗ Method file or batch file not appropriate. Check the method file or batch file. 9.1.2 Data Acquisition
The mass accuracy and mass resolution cannot be obtained. p.142 ↗ Laser intensity not appropriate, resulting in insufficient or excessive ions; instrument temperature not stable. Adjust the laser intensity. Wait approximately 30 minutes after applying voltage before starting measurement. On first switch to [Ready] after mains ON, click [Start] to run an analysis and wait. 9.1.2 Data Acquisition
The MSn spectrum cannot be obtained. p.142 ↗ Isolation parameter or CID parameter not appropriate. Check the isolation parameter. Check the CID parameter. 9.1.2 Data Acquisition
ERR02-05-01 or ERR02-05-02 p.144 ↗ Timeout occurred in vacuum operation (rough pumping). Instrument power turned ON while sample door open; or foreign object caught in sample door / door dirty. Close the sample door and turn ON the power again. If error persists, contact Shimadzu. If foreign object/dirt: use ethanol to clean the area where the sample door touches the O-ring, then close door and turn ON power again. 9.2 Error/Warning Messages
ERR02-05-08 or ERR02-05-09 p.144 ↗ Timeout occurred in vacuum operation (TMP). Vacuum performance decreased due to condensation in vacuum piping or other area; or leak due to dirty O-ring contact area. Turn ON the power again and check for the error. If leak suspected: use ethanol to clean the area where the sample door touches the O-ring, close door and turn ON power again. If error persists, contact Shimadzu. 9.2 Error/Warning Messages
ERR04-05-01 p.144 ↗ Decrease in vacuum during measurement was detected. Leakage of vacuum piping and deterioration of vacuum equipment suspected. Make a note of the error code that occurred, and contact your Shimadzu representative. 9.2 Error/Warning Messages
ERR06-05-01 p.144 ↗ Timeout occurred in vacuum operation (rough pumping) after replacing the sample. Foreign object caught in sample door or door dirty. Use ethanol to clean the area where the sample door touches the O-ring, then close the sample door and turn ON the power again. If error persists, contact Shimadzu. 9.2 Error/Warning Messages
ERR06-05-04 p.145 ↗ Timeout occurred in vacuum operation (TMP) after replacing the sample. Vacuum performance decreased due to condensation; or leak due to dirty O-ring contact area. Turn ON the power again and check for the error. If leak suspected: turn ON power again, open door after instrument enters standby mode, then use ethanol to clean the area where the sample door touches the O-ring. If error persists, contact Shimadzu. 9.2 Error/Warning Messages
WRN??-06-01 p.145 ↗ The gas pressure of the helium gas supply line is low. Leak or other abnormality in gas supply line suspected. Check that the regulator settings are correct and that the gas supply line nuts are not loose. 9.2 Error/Warning Messages; ?? varies with instrument state
WRN??-06-02 p.145 ↗ The gas pressure of the argon gas supply line is low. Check that the regulator settings are correct and that the gas supply line nuts are not loose. 9.2 Error/Warning Messages; ?? varies with instrument state
Other error codes p.145 ↗ Not specified. Make a note of the error code that occurred, and contact your Shimadzu representative. 9.2 Error/Warning Messages