347-05375
Technical Reference Summary
Covered Models & Document Scope
This manual covers the Shimadzu Micro Compression Tester MCT-211 series. The EMC regulatory section explicitly lists models 344-04252-58 MCT-510, 344-04253-58 MCT-511, 344-04254-58 MCT-210 and 344-04255-58 MCT-211. The main body is P/N 347-24788-41 with optical head 347-25075-44, x10 objective 344-89941-40, x50 objective 344-89964-40, φ50 µm flat upper compression rod 340-47026-01, X-Y stage 347-24205-41, two micrometer heads 339-81876-02, Type 3 sample attachment 344-17737-40 and lower compression plate 347-23123. The control unit is P/N 347-25115; PC, display and printer are not included in the standard configuration.
Section-by-Section Technical Breakdown
Lists the compression tester main body, optical head, objective lenses, compression rod, X-Y stage, micrometer heads, sample attachment and lower compression plate with quantities and part numbers. It also lists the control unit and the cables, AC plug adapter, tool set, manuals, installation CDs and spare fuses supplied with the MCT-211 series.
- Main body P/N 347-24788-41; optical head P/N 347-25075-44.
- Objective lenses: x10 P/N 344-89941-40 and x50 P/N 344-89964-40.
- Flat upper compression rod φ50 µm P/N 340-47026-01.
- X-Y stage P/N 347-24205-41; two micrometer heads P/N 339-81876-02.
- Sample attachment Type 3 P/N 344-17737-40; lower compression plate P/N 347-23123.
- Control unit P/N 347-25115; PC, display and printer are not included.
- Cables: power 347-20836-01 (110 V) or 347-20836 (230 V), lamp 344-86719, control 070-01623-72, A/D 347-24224, I/O 088-50664-56, ground 344-81775-01 (1.5 m).
- Spare part: control-unit fuse 2AT, P/N 347-20424-01, quantity 2.
Defines the environmental and utility conditions for the MCT-211 series, which measures micro displacement of 0.0001 µm under ultra-low test force of 0.2 µN and is highly sensitive to vibration and wind. It gives room temperature, humidity, vibration limits, power supply and grounding requirements, and the outer dimensions of the tester.
- Operating temperature range 10 to 35 °C; room temperature variation 1 °C or lower during operation.
- Humidity 80 % or lower, no condensation; no direct sunlight; no vibration; no air currents from HVAC.
- Installation category II, pollution degree 2, altitude up to 2000 m, indoor use.
- For tests near 10 mN or displacement changes of 0.1 µm or less: temperature fluctuation ±1 °C; horizontal vibration 10 Hz or more ≤0.017 Gal, below 10 Hz ≤0.01 µm; vertical vibration 10 Hz or more ≤0.01 Gal, below 10 Hz ≤0.005 µm.
- Single-phase AC 100–240 V ±10 %, 100 VA, with D-class ground (ground resistance 100 Ω or less).
- Outer dimensions approximately 405 mm × 530 mm × 315 mm / 355 mm; main body weight about 60 kg.
- Mains voltage fluctuations must not exceed 10 % of nominal supply voltage.
Describes placing the main body on a vibration-suppressing table, leveling with the leveling screws, mounting the X-Y stage and micrometers, installing the stage spacer and sample attachment, removing the upper compression rod protection cover, and connecting the cables. It also refers to the separate software installation manual.
- Do not turn ON power until assembly is complete; do not apply impact or external force to the upper compression rod or built-in loading unit.
- Main body weight about 60 kg; carry by three or more people and use the I bolt for crane transport, removing the I bolt afterwards.
- Remove the sliding-stage fixing part and two X-Y stage fixing rubbers before mounting the X-Y stage with four fixing screws.
- Insert micrometers and tighten the micrometer setscrews on both X and Y sides.
- Cable connections: control cable main body I/O–control unit I/O; lamp cable main body Lamp–control unit J31; A/D cable control unit J14–PC AD board; I/O cable control unit J21–PC I/O board; power cable; ground cable.
- Ground the system at one point and keep the ground cable as short as possible.
- Software installation is covered by Installation Manual 347-05357 (DUH-211/MCT-211 series Software Installation Manual).
Gives the X-Y stage travel, area, micrometer resolution and attachment screw pattern, the objective lens arrangement and switching, and the optical head index ranges and readout resolution. It also notes the LED illumination source and its service life.
- Stage elevating distance approx. 60 mm; stage area approx. 125 mm × 125 mm.
- Stage sliding range 25 mm in X and Y; minimum scale division 0.01 mm.
- Attachment set screw M5, pitch 60 mm × 60 mm; HMV-1/2 and DUH series attachments can be mounted.
- Objective lenses: x10 on the right, x50 on the left; switch with the revolver knob until it clicks.
- Optical head index moving range 0 to 250 µm; effective measurement length 200 µm; minimum displayed value 0.01 µm; two-side simultaneous collimation.
- LED illumination source service life approx. 20,000 hours; reduced brightness is recommended to extend life.
States that daily inspection and periodic inspection/calibration are necessary to maintain performance and obtain accurate measurement data. Daily maintenance, inspection and replacement parts are covered in Chapter 9; periodic inspection/calibration is to be requested from the Shimadzu representative.
- Daily maintenance, inspection and replacement parts: see Chapter 9 Maintenance.
- Periodic inspection/calibration: request from Shimadzu representative.
- Replacement cycles for periodic replacement parts are rough estimates and may be shorter depending on environment and frequency.
- Unplug the power cord before inspection, maintenance or parts replacement.
- Never remove the main cover; repairs requiring main-cover removal must be done by a Shimadzu representative.
- Use only the part described in the instruction manual when replacing a part.
Chapter 10 covers phenomena and repair. The manual directs the user to perform inspection and corrective action described in the troubleshooting section and to contact the Shimadzu representative if the problem persists or is not covered. Clause 10.2.3 is referenced for the case where the illumination lamp does not turn ON after brightness adjustment.
- 10.1 Phenomenon and 10.2 Repair are the troubleshooting sections.
- If the illumination lamp does not turn ON, adjust brightness with the control-unit front-panel knob; if still OFF, follow Clause 10.2.3.
- If a problem persists or is not covered, contact the Shimadzu representative.
- Emergency stop: turn OFF the control-unit power switch and disconnect the power cable from the rear of the control unit.
Chapter 11 provides the maintenance parts list. The extracted document text does not include the individual table rows of this list.
- Chapter 11 Maintenance Parts List, Section 11.1 Maintenance Parts List.
- Spare fuse for the control unit is 2AT, P/N 347-20424-01 (from the accessories/spare parts list).
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| The MCT-211 series measures/controls micro displacement of 0.0001 µm under ultra-low test force of 0.2 µN. p.37 ↗ | Installation must minimize vibration and wind; a vibration-suppressing table is the principle requirement. | MCT-211 series |
| Operating temperature range is 10 to 35 °C with room temperature variation of 1 °C or lower during operation; humidity 80 % or lower with no condensation. p.7 ↗ | HVAC and room stability must be verified before acceptance testing; condensation must be avoided. | MCT-211 series |
| For tests at about 10 mN or for detecting displacement changes of 0.1 µm or less: temperature fluctuation ±1 °C; horizontal vibration ≥10 Hz ≤0.017 Gal and <10 Hz ≤0.01 µm; vertical vibration ≥10 Hz ≤0.01 Gal and <10 Hz ≤0.005 µm. p.37 ↗ | These are the acceptance limits for the installation site when high-sensitivity tests are required. | MCT-211 series |
| Power requirement: single-phase AC 100–240 V ±10 %, 100 VA, with D-class ground and ground resistance 100 Ω or less; mains fluctuation not to exceed 10 % of nominal. p.39 ↗ | Verify supply voltage, capacity and grounding before installation; use the Shimadzu-specified power cord. | MCT-211 series |
| Main body weight is about 60 kg; carry by three or more people and use the I bolt for crane transport, removing the I bolt afterwards. p.39 ↗ | Plan rigging and floor/table load capacity; the vibration-suppressing table must be steady. | MCT-211 series |
| X-Y stage: elevating distance approx. 60 mm, area approx. 125 mm × 125 mm, sliding range 25 mm in X and Y, minimum scale division 0.01 mm. p.50 ↗ | Sample size and positioning range are limited to these values; micrometer resolution is 0.01 mm. | MCT-211 series |
| Attachment set screw is M5 with 60 mm × 60 mm pitch; all attachments for Shimadzu HMV-1/2 series and DUH series can be mounted. p.50 ↗ | Existing HMV-1/2 and DUH fixtures are mechanically compatible with the MCT-211 stage. | MCT-211 series |
| Optical head: index moving range 0 to 250 µm, effective measurement length 200 µm, minimum displayed value 0.01 µm, two-side simultaneous collimation. p.52 ↗ | Dimension measurement is limited to 200 µm effective length with 0.01 µm display resolution. | MCT-211 series |
| Objective lenses are x10 (right) and x50 (left); switch with the revolver knob until it clicks. p.51 ↗ | Lenses are spring-mounted and retract on contact but may be damaged; handle with care when mounting or dismounting. | MCT-211 series |
| LED illumination source service life is approx. 20,000 hours; maximum brightness reduces service life. p.45 ↗ | Use reduced brightness where possible to extend lamp life. | MCT-211 series |
| When turning power off and on again, wait at least 5 seconds before turning it on. p.10 ↗ | Prevents equipment failure from rapid power cycling. | MCT-211 series |
| Ground the PC and optional units with equal-potential grounds; ground the system at one point and keep the ground cable as short as possible. p.9 ↗ | Prevents communication interruption and equipment failure between units. | MCT-211 series |
| Control-unit fuse is 2AT, P/N 347-20424-01, supplied as two spare fuses. p.33 ↗ | Replacement fuse rating and part number for maintenance. | MCT-211 series control unit |
| Cable part numbers: power 347-20836-01 (110 V) or 347-20836 (230 V); lamp 344-86719; control 070-01623-72; A/D 347-24224; I/O 088-50664-56; ground 344-81775-01 (1.5 m). p.33 ↗ | Order the correct power cable variant for the site voltage and use only Shimadzu-specified cables. | MCT-211 series |
| EMC compliance (EN61326 class B emissions, minimum immunity) applies to models 344-04252-58 MCT-510, 344-04253-58 MCT-511, 344-04254-58 MCT-210 and 344-04255-58 MCT-211. p.26 ↗ | Locate away from strong electromagnetic noise sources and use a separate power source for such devices. | MCT-510, MCT-511, MCT-210, MCT-211 |
| Replacement parts are available for seven (7) years after product discontinuation; third-party parts availability is determined by their manufacturers. p.22 ↗ | Plan long-term spares and consumables procurement within this window. | MCT-211 series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Illumination lamp is not turned ON p.45 ↗ | Brightness setting too low or lamp not activated | Adjust brightness using the adjustment knob on the control unit front panel and confirm the lamp turns ON. | Clause 3.3.3 |
| Illumination lamp is not turned ON after the above brightness adjustment p.45 ↗ | Fault requiring troubleshooting procedure | Follow the procedure described in Clause 10.2.3 and confirm the lamp turns ON. | Clause 10.2.3 |
| Abnormal condition such as smoke or smell of something burning p.12 ↗ | Equipment abnormality | Turn OFF the power switch of the control unit and disconnect the power cable from the rear of the control unit; check the equipment and contact the service representative before restarting. | In an Emergency |
| Power outage p.12 ↗ | Loss of mains power | Turn OFF the control-unit power switch; lower the Z stage with the Z stage handle to clear the indenter from the sample; slide the sample stage left to the observation position; after power is restored, turn ON the control unit and PC. | During a Power Outage |