Induztr
Engineering Digest Document Viewer Related Models

LH-40 Instruction Manual - Liquid Handler

Technical Reference Summary

Covered Models & Document Scope

LH-40 Liquid Handler for Shimadzu preparative liquid chromatograph systems. Covers standard Fast Injection configuration, optional Accuracy Injection via Analysis Kit (H) (P/N 228-75587-42), optional rinse pump, liquid level detection, multi fraction collector (up to 5) and multi liquid handler (up to 5) configurations, and preparative valve kits for analysis scale (≤5 mL/min, 0.3 mm ID) and preparative scale (5–150 mL/min, 0.8 mm ID).

Section-by-Section Technical Breakdown

pp. 13-16
Sample Injection Methods (Fast and Accuracy Injection)

Describes the standard Fast Injection method where sample is aspirated by the syringe and drawn through needle → needle loop → high-pressure valve → sample loop before injection, and the optional Accuracy Injection method enabled by the Analysis Kit (H) for analytical-scale volumes. The high-pressure valve switches between LOAD and INJ positions and has a pressure capacity of 25 MPa.

  • Fast Injection: sample path needle → needle loop → high-pressure valve → sample loop; valve then rotates to INJ.
  • Accuracy Injection requires optional Analysis Kit (H) P/N 228-75587-42 and uses an injection port and separate analytical high-pressure valve/sample loop.
  • Syringe sizes available: 500 µL, 5 mL, 25 mL; 500 µL syringe is used for Accuracy injection.
  • Sample loop sizes available: 2 mL, 5 mL, 20 mL.
  • High-pressure valve pressure capacity: 25 MPa.
pp. 17-19
Needle Rinsing and Injector Purge

Covers cleaning of the needle outer surface by dipping into the needle rinse port and purging with the syringe, or optionally with a rinse pump for flow rinsing. The optional rinse pump also allows different rinse solutions for internal and external needle rinse.

  • Needle outer surface rinse uses needle rinse port (8); purge is performed by the syringe by default.
  • Optional rinse pump (parts 10 and 11) enables flow rinsing and can shorten purge time.
  • With rinse pump, separate rinse solutions can be selected for internal and external needle rinse.
  • Needle inner surface rinse and injector purge use the needle drain port.
pp. 12, 20
Fractionation and Preparative Functions

Describes fraction collection via the preparative valve, multi fraction collector and multi liquid handler expansion, automatic delay time calculation, fraction overflow prevention, contamination inhibition, and sample rescue on cancellation.

  • Up to 5 fraction collectors can be used; a flow channel selection valve (e.g., FCV-20AH6) is required for multi fraction collector configurations.
  • Up to 5 liquid handlers can be used; flow channel selection valves are required.
  • Automatic delay time calculation optimizes preparative valve timing after flowrate changes.
  • Fraction overflow prevention moves the arm to the next test tube before overflow.
  • Sample rescue function recovers injected samples if preparative work is canceled.
pp. 21-23
Component Parts and Optional Parts

Lists included accessories and optional parts with part numbers, including valve kits, racks, plates, syringes, rinse pump, liquid sensor needle, long needle, leak tray, and communication/valve options.

  • Included: Ethernet cable 2 m (228-61083-41), Drain Tube ASSY (228-75612-41), Remote I/O Cable (228-28253-91) ×2, Liquid Waste Hook (228-62077), Suction Filter ASSY (228-67261-41), Connector (070-51034-02), Signal Cable (228-35047-91).
  • Optional AD Board 228-55519-41 accepts analog signals up to ±10 V; up to four boards for four channels.
  • Valve Kit for Analysis Scale 228-75605-42 (0.3 mm ID, up to 5 mL/min); Valve Kit for Preparative Scale 228-75605-41 (0.8 mm ID, 5–150 mL/min).
  • Rinse Pump 228-75586-41; Liquid Sensor Needle Kit 228-75608-41; Long Needle (H) Kit 228-67203-44.
  • Stainless steel leak tray 228-75609-42 (SUS316L) for THF/DMF.
  • FCV-20AH6 228-65615-58 for flow channel switching; FCV-11AL 228-65611-58 for rinse solution selection.
pp. 24
Preparative Valve Kits and Tray/Rack Combinations

Provides selection criteria for valve kits based on preparative flowrate and for racks/trays based on container type. Notes restrictions such as 1.5 mL vials not usable in fraction racks and stainless steel racks recommended for highly acidic solvents.

  • Analysis scale: up to 5 mL/min, 0.3 mm ID, P/N 228-75605-42.
  • Preparative scale: 5 to 150 mL/min, 0.8 mm ID, P/N 228-75605-41.
  • For refractive index detectors (e.g., RID-20A), use preparative-scale valve kit 228-75605-41 to prevent back pressure on the cell.
  • 1.5 mL vials cannot be used in fraction racks.
  • Use Injection Rack (MTP) when injecting from septum containers with Long Needle (H).
pp. 30
Injection Container and Needle Combinations

Table of compatible containers for short needle, long needle/long needle (H), and liquid sensor needle. LH-40 accepts test tubes up to 165 mm tall.

  • Short needle (default) supports microplates, deep-well plates, micro tubes, 1.5 mL and 4.0 mL vials.
  • Long needle/long needle (H) additionally supports 3.5–32 mL test tubes and 15 mL sample vials.
  • Liquid sensor needle supports micro tubes and 1.5/4.0 mL vials; cannot be used with septum vials.
  • 50 mL test tubes and 250/500/1000 mL sample vials are not supported by any listed needle.
  • Maximum test tube height: 165 mm.
pp. 35
External Input/Output Terminal and Communication

Describes the rear external I/O terminal connectors including PC, INPUT/OUTPUT, INIT button, ETHERNET, RESERVED/OUTPUT, OPTION, FCV, and AD1–AD4 analog inputs. Output functions are configurable via the Utility.

  • [PC] connector not used for this instrument.
  • [INPUT] and [OUTPUT] terminals for external devices; output functions configurable in Utility.
  • [INIT] button initializes the instrument.
  • [ETHERNET] connector for computer connection.
  • [FCV] connector controls rinse solution selection valve.
  • [AD1]–[AD4] connectors for analog signal input.
pp. 29-30
Preparation: Tubing and Drain Connections

Covers checking tubing for disconnections/kinks, connecting detector OUT to preparative valve IN with PEEK male nut, and drain tube installation requirements to ensure proper liquid waste flow.

  • Connect detector OUT tube to preparative valve IN tube using PEEK male nut included in valve kit.
  • Use tube internal diameter appropriate for the flowrate.
  • Liquid waste container must be placed lower than the liquid handler (e.g., on floor) and drain tube oriented downward.
  • Drain tube flow channels must not be higher than the drain port.
  • Place liquid waste container in front of the liquid handler; draining to the back is not possible.
  • Use liquid waste hook to secure drain tube end and prevent immersion in waste.
pp. 127-150
Maintenance

Periodic inspection and replacement procedures for high-pressure valve rotor/stator, syringe low-pressure valve rotor/stator, syringe, sample loop, needle seal, plus cleaning and rinsing tasks.

  • Replacing high-pressure valve rotor and stator (removal, installation, reset usage frequency).
  • Replacing syringe unit low-pressure valve rotor and stator.
  • Replacing syringe and checking after replacement.
  • Replacing sample loop and checking after replacement.
  • Replacing needle seal.
  • Cleaning leak tray, exterior, rinsing flow channels, rinsing/inspecting suction filter, lubricating syringe units, cleaning leakage on stage.
pp. 151-162
Technical Information

Tubing and cable connection details, external I/O signal cable connections, detector and A/D board connections, Ethernet connection, product specifications, and maintenance parts lists.

  • Tubing connections and cable connections.
  • External input/output signal cable connections.
  • Detector and A/D board connections.
  • Connecting to [ETHERNET] connector.
  • Specifications: shared, injection functions, fraction collection functions, preparative valve, analog input.
  • Maintenance parts: consumables and replacement parts.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
High-pressure valve pressure capacity is 25 MPa. p.15 ↗ System operating pressure must not exceed 25 MPa at the injection valve. LH-40 standard configuration
Injection volume range from 1 µL to 20 mL by replacing tubing and/or syringes. p.11 ↗ Select syringe and sample loop sizes to cover the required injection volume. LH-40
Available syringe sizes: 500 µL, 5 mL, 25 mL; 500 µL syringe used for Accuracy injection. p.15 ↗ Accuracy injection requires the 500 µL syringe; preparative work typically uses 5 mL or 25 mL. LH-40
Available sample loop sizes: 2 mL, 5 mL, 20 mL. p.15 ↗ Choose loop volume consistent with injection volume and valve kit scale. LH-40
Analysis scale valve kit 228-75605-42 uses 0.3 mm ID tubing for flowrates up to 5 mL/min. p.24 ↗ Use analysis-scale kit for flowrates ≤5 mL/min to avoid excessive delay volume. LH-40 with analysis-scale valve kit
Preparative scale valve kit 228-75605-41 uses 0.8 mm ID tubing for flowrates from 5 to 150 mL/min. p.24 ↗ Use preparative-scale kit for flowrates 5–150 mL/min. LH-40 with preparative-scale valve kit
For refractive index detectors (e.g., RID-20A), use preparative-scale valve kit 228-75605-41 to prevent back pressure on the cell. p.24 ↗ Do not use the analysis-scale kit with RI detectors. LH-40 with RID detector
Optional AD Board 228-55519-41 accepts analog signals up to ±10 V; one channel per board, up to four boards. p.22 ↗ Plan one AD board per detector signal input channel, maximum four. LH-40
Up to 5 fraction collectors and up to 5 liquid handlers can be used; flow channel selection valves required. p.12 ↗ Multi-unit configurations require FCV-20AH6 or equivalent flow channel selection valves. LH-40 multi-unit configurations
LH-40 can be used with test tubes up to 165 mm tall. p.30 ↗ Verify container height does not exceed 165 mm for rack/tray selection. LH-40
1.5 mL vials cannot be used in fraction racks. p.24 ↗ Use larger containers or different racks for fraction collection. LH-40 fraction collection
Stainless steel rack recommended when using highly acidic solvents that may dissolve resins. p.24 ↗ Select stainless steel racks for acidic mobile phases. LH-40 rack selection
Stainless steel (SUS316L) leak tray 228-75609-42 should be purchased if solvents such as THF or DMF are used. p.23 ↗ Use SUS316L leak tray for solvent compatibility with THF/DMF. LH-40
Liquid waste container must be placed lower than the liquid handler and drain tube oriented downward; drain tube flow channels must not be higher than the drain port. p.30 ↗ Improper drain routing causes backflow and poor drainage. LH-40 installation
Place liquid waste container in front of the liquid handler; liquid waste cannot be drained to the back. p.30 ↗ Plan lab bench layout accordingly. LH-40 installation
Use liquid waste hook to secure drain tube end and prevent immersion in liquid waste. p.30 ↗ Prevents static electricity accidents and siphon effects. LH-40 installation
Connect detector OUT tube to preparative valve IN tube using PEEK male nut included in valve kit. p.29 ↗ Use correct fitting to avoid leaks and dead volume. LH-40 installation
Static electricity countermeasures kit grounds static charges in mobile phases at stainless steel couplings. p.33 ↗ Prevents malfunctions from static buildup with organic solvents at high flow rates. LH-40

Related models