Technical Reference Summary

Covered Models & Document Scope

This document (P/N 228-92947, Oct. 2018) is the 'Precautions for Use' booklet for the FRC-40 Fraction Collector for Shimadzu High Performance Liquid Chromatograph. It covers installation, operation, maintenance, static-electricity precautions, mobile-phase solvent compatibility, warning labels, and after-sales service for the FRC-40. It references but does not contain the System Guide (228-92935) or the full Instruction Manual (228-92381).

Section-by-Section Technical Breakdown

pp. 7-9
Installation Site and Environment

Specifies environmental and physical installation requirements for the FRC-40, including temperature, humidity, ventilation, spacing, and table stability. It prohibits installation near flammable/explosive gases, strong magnetism, corrosive atmospheres, and spark sources.

  • Room temperature 4 to 35 °C with small daily variation; humidity 20 to 85 % with no condensation.
  • Provide at least 100 mm clearance between the rear face and the wall for heat dissipation.
  • Install on a flat, stable table at least 800 mm deep capable of supporting the entire LC system.
  • Fume hood or exhaust system with about 20 m3/min capacity recommended; nitrogen gas use requires adequate ventilation.
  • Installation category II, pollution degree 2, max altitude 2,000 m, indoor use.
  • Do not install near strong magnetic fields; add a noise filter if the power line has high noise.
  • Earthquake Kit P/N S228-47144-93 (with hook-and-loop fastener) available to prevent falling.
pp. 10-11
Power Supply and Grounding

Defines power supply and grounding requirements to prevent electric shock and ensure stable operation. Voltage must match the label on the rear of the instrument, fluctuations must not exceed 10 % of nominal, and the instrument must be grounded via a three-prong outlet.

  • Connect only to the voltage indicated on the rear label.
  • Power supply voltage fluctuations must not exceed 10 % of the nominal value.
  • Use a circuit breaker that shuts off at a current equal to or below the supply capacity.
  • Ground the instrument; plugging into a grounded three-prong outlet grounds the product.
  • Ground the computer and optional units with equal potential grounds to avoid communication interruption.
  • Provide adequate space for accessing the power switch and power cord for emergency disconnection.
  • Confirm the main power switch on the side face is OFF before plugging in the power cord.
pp. 12
Operation Precautions

Covers safe operating practices including pressure relief before disconnecting high-pressure components, maximum pressure limits for column switching valves, prohibition of flammable aerosols, and protection against static electricity.

  • Before disconnecting tubing or autosampler high-pressure components, shut OFF the pump and confirm pressure has dropped to zero.
  • If pressure does not drop, a blockage may exist; relieve pressure by slowly opening the drain valve with the drain tube in the waste bottle.
  • FCV-12AH, 14AH, or 20AH series column switching valves have a maximum pressure capacity of about 35 MPa; do not use them in 70 MPa systems.
  • Do not use flammable aerosol sprays near the instrument.
  • Never use a cracked reservoir bottle, especially when a helium degassing unit pressurizes it.
  • Keep hands away from the movement range of moving parts.
  • Wear protective goggles and gloves when handling solvents or samples.
pp. 16-20
Static Electricity Prevention

Explains how static electricity accidents occur in LC systems and prescribes five preventive measures, including grounded metal waste containers, reduced container openings, body grounding, drain tube sizing, and humidity control.

  • Use a metal (conductive) container for liquid waste and connect it to ground; optional grounding wire with clip P/N S228-21353-91, 18-L metal container P/N S038-00044, 4-L metal container P/N S038-00043-01.
  • Verify waste container grounding with an electrical tester; oxide coatings or laminates may prevent conductivity.
  • Caps for 18-L and 4-L containers (P/N S228-21354-91) include three 3 mm dia. openings to reduce spark ingress.
  • Ground the body with an antistatic wrist strap using about 1 MΩ resistance; wear anti-static clothing and shoes; use conductive floor mats.
  • Use drain tubes with internal diameter of at least 2 mm for draining large amounts of liquid.
  • Prevent air leaks into tubing; air bubbles can multiply electrostatic charge by a factor of several tens.
  • For low-conductivity liquids (10^-10 S/m max.), add a safe conductive liquid, use as small a waste container as possible, and keep humidity at 65 % or higher.
pp. 21-22
Mobile Phase Solvent Compatibility

Lists solvents that must not be used with PEEK tubing or the degassing unit, pH limits, filtration requirements, and precautions for stainless steel flow paths and autosampler rinse solutions.

  • Do not use concentrated sulfuric acid, concentrated nitric acid, dichloroacetic acid, acetone, THF, dichloromethane, chloroform, DMSO, HFIP, or other fluorine-based organic solvents with PEEK tubing; up to 0.5 % aqueous acetone temporarily is acceptable.
  • Do not use HFIP, hydrogen fluoride, 1,1,2-trichloro-1,2,2-trifluoroethane, Fluorinert FC-40/FC-72/FC-75, perfluoro benzene/octane/decalin/1-methyldecalin/dimethyldecalin/methyl-cyclohexane/dimethyl-cyclohexane, AK-225, ≥30 % nitric acid, ≥40 % sulfuric acid, or hydrogen peroxide with the degassing unit.
  • Use HPLC grade or equivalent solvents filtered at 0.45 µm max.
  • Avoid halogen-ion-containing solvents with stainless steel (SUS316L) flow paths; if used, rinse thoroughly with distilled water immediately after analysis.
  • Avoid aqueous ammonia solutions over 0.1 % concentration.
  • Do not use solutions with pH outside 1 to 13; pH 10 or higher for long periods can damage quartz flow cell.
  • Degas the mobile phase to prevent air bubbles and detector noise.
  • Do not regularly use concentrated (10–50 %) acetic acid or 1 % TFA as autosampler rinse solution; if used, rinse flow lines with distilled water and remove sample rack to vent vapors.
pp. 14
Inspection and Maintenance

Specifies maintenance safety rules: never remove the main cover, disconnect power before service, use specified replacement parts, and keep the instrument dry.

  • Never remove the main cover; routine maintenance does not require its removal.
  • Switch OFF power on the back of the main unit and unplug the power cord before inspection, maintenance, or parts replacement.
  • Do not leave the product wet with water; do not wipe with alcohol or paint thinner type solvents.
  • Unplug the power cord and use a dry cloth to wipe dust from plug prongs and mounting surface.
  • Replace parts only with parts specified in the Parts List or Maintenance Parts section.
  • Dispose of liquid waste in accordance with regulations.
pp. 15
Emergency Shutdown and Power Outage

Provides the emergency shutdown procedure and measures to take during a power outage.

  • Emergency shutdown: turn OFF the power switch on the main unit, then disconnect the power cable at the rear.
  • Before reuse after emergency shutdown, inspect the instrument and contact Shimadzu if necessary.
  • During a power outage: turn OFF the power switch on the main unit.
  • After power is restored, confirm installation and operation precautions, then use the standard startup procedure.
pp. 23
Warning Labels

Identifies the warning label on the FRC-40 and its meaning.

  • Warning label P/N S037-72999-02 warns of finger-pinching risk while the drive unit is moving.
  • Do not bring hands or fingers close to the drive unit while it is moving.
  • Obtain a replacement label from Shimadzu if lost or damaged.
pp. 25
After-Sales Service and Replacement Parts

Describes after-sales service, replacement parts availability, periodic inspection/calibration, and disposal precautions.

  • Replacement parts are available for seven years after product discontinuation.
  • Daily maintenance and inspection are described in the Maintenance Instruction Manual.
  • Periodic inspection/calibration should be requested from Shimadzu.
  • Replacement cycles are rough estimates and may require earlier replacement depending on usage.
  • Contact Shimadzu for disposal; dispose in accordance with legal standards.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Room temperature must be between 4 and 35 °C with small variation throughout the day. p.8 ↗ Install in a temperature-controlled environment to maintain performance. FRC-40
Humidity must be kept between 20 and 85 % with no condensation. p.8 ↗ Avoid humid or condensing locations to prevent corrosion and electrical hazards. FRC-40
Provide at least 100 mm space between the rear face of the instrument and the wall. p.8 ↗ Ensures adequate heat dissipation and prevents overheating. FRC-40
Installation table must be flat, stable, and at least 800 mm deep. p.8 ↗ Prevents tipping or falling of the LC system. FRC-40
Power supply voltage fluctuations must not exceed 10 % of the nominal value. p.10 ↗ Use a stable power supply and appropriate circuit breaker to avoid malfunction. FRC-40
FCV-12AH, 14AH, or 20AH series column switching valves have a maximum pressure capacity of about 35 MPa. p.12 ↗ Do not use these valves in 70 MPa systems; ensure system pressure does not exceed valve capacity. FRC-40 with FCV-12AH/14AH/20AH valves
Before disconnecting high-pressure components, confirm pressure has dropped to zero. p.12 ↗ Prevents forceful ejection of liquid or foreign matter. FRC-40
Use drain tubes with an internal diameter of at least 2 mm for draining large amounts of liquid. p.19 ↗ Reduces static charge generation and prevents clogging. FRC-40
Antistatic wrist strap should ground the body with a resistance of about 1 MΩ. p.18 ↗ Protects personnel while discharging static electricity. FRC-40
For low-conductivity liquids (10^-10 S/m max.), maintain humidity at 65 % or higher. p.20 ↗ Helps prevent static electricity formation. FRC-40
Do not use PEEK tubing with concentrated sulfuric acid, concentrated nitric acid, dichloroacetic acid, acetone, THF, dichloromethane, chloroform, DMSO, HFIP, or other fluorine-based organic solvents. p.22 ↗ Prevents tubing rupture and solvent spray. FRC-40 with PEEK tubing
Do not use solutions with pH outside 1 to 13. p.21 ↗ Prevents damage to flow cell and other components. FRC-40
Filter mobile phase solvents at 0.45 µm max. p.22 ↗ Removes microparticles that could cause blockages or damage. FRC-40
Avoid aqueous ammonia solutions with ammonia concentration over 0.1 %. p.22 ↗ Prevents damage to pump plunger and shortened plunger seal life. FRC-40
Replacement parts are available for seven years after product discontinuation. p.25 ↗ Plan long-term maintenance and spare parts inventory accordingly. FRC-40
Earthquake Kit P/N S228-47144-93 (with hook-and-loop fastener) is available. p.9 ↗ Use to secure the instrument against falling during earthquakes. FRC-40
Grounding wire with clip P/N S228-21353-91, 18-L metal container P/N S038-00044, 4-L metal container P/N S038-00043-01, and caps P/N S228-21354-91 are optional products for static prevention. p.17 ↗ Use these parts to implement grounded waste collection. FRC-40
Warning label P/N S037-72999-02 indicates finger-pinching risk while the drive unit is moving. p.23 ↗ Keep hands away from the drive unit during operation. FRC-40